← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K960049

# BRANNON PORTSYRINGE (K960049)

_James K. Brannon · FMF · May 24, 1996 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K960049

## Device Facts

- **Applicant:** James K. Brannon
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** May 24, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Brannon PortSyringe is intended for fluid aspiration and transfer of the fluid aspirate to vacuum specimen tubes and bottles.

## Device Story

Brannon PortSyringe (BPS) is a 5cc piston syringe integrated with a blood specimen collection device (BSCD) within the plunger unit. Device functions as a conventional syringe for fluid aspiration via a hypodermic needle attached to the distal nozzle. Once aspirated, fluid is transferred to vacuum specimen tubes by inserting tubes into the integrated holder; vacuum pressure induces the plunger to advance, moving fluid from the barrel through a narrow conduit in the plunger to the specimen tube. Alternatively, blood can be transferred to culture bottles via the hypodermic needle. Used in clinical settings (e.g., intensive care) by healthcare providers. Benefits include improved phlebotomy efficiency, safety, and patient comfort by enabling simultaneous aspiration and transfer without additional assembly.

## Clinical Evidence

Clinical trial conducted at University of Iowa Hospitals and Clinics (1989) involving intensive care patients requiring multiple blood samples. Study demonstrated improved phlebotomy safety, efficiency, and patient comfort compared to traditional procedures. Bench testing confirmed successful consecutive aspirations and fluid transfers without failure.

## Technological Characteristics

5cc piston syringe with integrated BSCD in plunger. Materials are identical to B-D disposable piston syringe. Fluid path includes a centrally disposed narrow conduit in the plunger, sealed by a rubber-sleeved needle. Mechanical operation; no energy source or software.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- B-D disposable piston syringe (unspecified 510(k))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

MAY 24 1996
K960049

James K. Brannon, M.D.
(Contact Person)
5729 Canterbury Drive
Culver City, CA 90230-6547
(310) 216-0207 Office
(310) 216-0207 Fax
(310) 898-2834 Pager

February 21, 1996

510(k) SUMMARY
AMENDED
K960049

Name of Device: Piston Syringe 80 FMF
Common/Usual Name: Syringe
Trade/Proprietary Name: Brannon PortSyringe

The piston syringe, made available in a 5cc size, of this 510(k) notification, known as the Brannon PortSyringe (BPS), is a combination of two legally marketed devices; (1) a piston syringe and (2) a blood specimen collection device (BSCD). The BSCD is integral with the plunger unit of the piston syringe with the BSCD communicating with the fluid chamber of the piston syringe via a centrally disposed and substantially narrow conduit. The fluid collection needle is sealed by a rubber sleeve. Appreciate that the needle is not manufactured by the applicant nor a manufacturing firm identified by the Applicant, but rather purchased from an FDA recognized manufacturer/distributor. Further appreciate that the BPS is manufactured with identical material as the predicate piston syringe.

Sealing of the conduit with the fluid collection needle allows the piston-plunger unit and the barrel to function as a conventional syringe. Functional operation simply requires that one compression fits a hypodermic needle of any gauge to the distal nozzle tip of the barrel. One is then able to aspirate body fluids as the predicate piston syringe. After a given amount of fluid is aspirated into the barrel of the syringe, a minimal amount will suffice, a plurality of vacuum specimen tubes can be inserted into the tube-holder. The vacuum specimen tubes are advanced over the sleeved needle while stabilizing the tube-holder. In so doing, the vacuum within the specimen tube induces the piston-plunger unit to advance distally as the aspirated fluid is transferred to the vacuum specimen tube proximally.

{1}

James K. Brannon, M.D.
Brannon PortSyringe
K960049
February 21, 1996
510(k) Summary
Page 2

If one desires blood for culturing, the blood aspirate is transferred to a vacuum blood culture bottle by inserting the hypodermic needle directly into the vacuum bottle as is done using a predicate piston syringe.

Regarding blood collection with simultaneous intravenous catheter insertion, the BPS is securely fitted to the female hub of a percutaneously inserted intravenous catheter. One should briefly aspirate to assure vein patency. A plurality of vacuum specimen tubes are then inserted into the tube-holder as described in this summary.

The technological differences of the BPS include fluid aspiration through the distal nozzle tip of the syringe barrel and subsequent transfer of the fluid aspirate through the piston-plunger unit and into a plurality of vacuum specimen tubes. The predicate piston syringe only allows influx and efflux of a fluid through the distal nozzle tip of its syringe barrel.

Regarding safety and effectiveness, a clinical trial was conducted at the University of Iowa Hospitals and Clinics (UIHC) in January of 1989. The title of the investigation was “Safe, Simple &amp; Efficient Fluid Extraction”, with a report made in the UIHC publication Pacemaker, October 1989. The patient population included intensive care patients requiring multiple blood samples for vacuum specimen tubes and vacuum blood culture bottles. Use of the investigational device, which required some assembly, was as outlined in this summary. Appreciate that the new device of this summary does not require any assembly other than the attachment of a hypodermic needle. It was shown in this investigation that traditionally cumbersome phlebotomy procedures were made safer with improved efficiency, proficiency, and patient comfort.

Further appreciate that the Brannon PortSyringe meets all of the biocompatibility and performance standards (efficacy) of the predicate piston syringe and predicate BSCD. The Brannon PortSyringe is manufactured with identical material as the predicate piston syringe (B-D disposable piston syringe) and there is no deviation from this standard. Further, the Brannon PortSyringe was used during consecutive aspirations and fluid transfers to vacuum specimen tubes at the bench and there were no failures. The Brannon PortSyringe works safely and effectively as outlined in the Premarket Notification 510(k).

{2}

James K. Brannon, M.D.
Brannon PortSyringe
K960049
February 21, 1996
510(k) Summary
Page 3

In conclusion, the above summary elucidates the physical characteristics which constitute the Brannon PortSyringe. The summary further shows by comparison that the indicated use, fluid aspiration and transfer of the fluid aspirate to vacuum specimen tubes and bottles, is identical to the predicate piston syringe and the predicate BSCD. The clinical trial conducted at the UIHC demonstrates that the Brannon PortSyringe can be used safely and effectively in a manner identical to the predicate piston syringe and BSCD. Therefore, the Brannon PortSyringe of this 510(k) summary is claimed to be substantially equivalent to a predicate piston syringe.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K960049](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K960049)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
