← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K253068

# Profoject Insulin Syringes; Profoject Safelock Disposable Insulin Syringe (K253068)

_CMT Health PTE., Ltd. · FMF · Feb 23, 2026 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K253068

## Device Facts

- **Applicant:** CMT Health PTE., Ltd.
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** Feb 23, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

Profoject™ Insulin Syringes (Model A): The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. Profoject™ Insulin Syringes (Model B): The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. Profoject™ Safelock Disposable Insulin Syringe The device is intended for the delivery of U-40 or U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.

## Device Story

Profoject™ Insulin Syringes and Safelock Disposable Insulin Syringes are sterile, single-use piston syringes for subcutaneous insulin delivery. Available in 0.3mL, 0.5mL, and 1mL volumes. Model A features integrated hub/barrel; Model B features separate hub/barrel. Safelock version includes manual safety shield to prevent accidental needle-sticks. Operated by patients or clinicians in home or clinical settings. Device functions via manual plunger depression to deliver insulin dose. Safelock mechanism activated by user via fingertip or thumb. Benefits include accurate insulin dosing and reduced risk of sharps injury. Performance verified through ISO standards for sterility, biocompatibility, and mechanical integrity.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by non-clinical bench testing, including ISO 7864 (needles), ISO 8537 (insulin syringes), ISO 23908 (sharps protection), and ISO 10993 (biocompatibility).

## Technological Characteristics

Piston syringe design; materials include polypropylene (barrel/shield) and stainless steel (needle). Lubricant: Polydimethylsiloxane. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Connectivity: None. Software: None. Standards: ISO 7864, ISO 9626, ISO 8537, ISO 23908, ISO 10993-1, ISO 11135.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Insulin Syringe ([K223453](/device/K223453.md))
- Kendall Monoject Magellan Insulin Safety Syringe ([K061492](/device/K061492.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

February 23, 2026

CMT Health Pte. Ltd
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120A
Madison, NJ 07940

Re: K253068
Trade/Device Name: Profoject™ Insulin Syringes; Profoject™ Safelock Disposable Insulin Syringe
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF, MEG
Dated: January 12, 2026
Received: January 12, 2026

Dear Dave Yungvirt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253068 - Dave Yungvirt
Page 2

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K253068 - Dave Yungvirt
Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director, Injection Devices
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K253068  |   |
|  Device Name Profoject™ Insulin Syringes, Profoject™ Safelock Disposable Insulin Syringe  |   |
|  Indications for Use (Describe) Profoject™ Insulin Syringes (Model A): The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. Profoject™ Insulin Syringes (Model B): The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. Profoject™ Safelock Disposable Insulin Syringe The device is intended for the delivery of U-40 or U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

510(k) Summary
[As required by 21 CFR 807.92]

## 1. Submission Information

510(k) Number: K253068
Date: 13 February 2026
Type of 510(k) Submission: Traditional
Submitter: CMT HEALTH PTE. LTD.
150 BEACH ROAD, #28-05, GATEWAY WEST, SINGAPORE, 189720
Contact Person: Monica Ma
E-mail: ra@cmthealth.com
Tel: +65 6846 1379

## 2. Device Description

Proprietary Name: Profoject™ Insulin Syringes, Profoject™ Safelock Disposable Insulin Syringe
Common Name: Syringe, Piston; Syringe, Antistick
Regulation Name: Piston syringe
Product Code: FMF, MEG
Device Class: II
Regulation Number: 21 CFR 880.5860
Review Panel: General Hospital

Indications for use - Rx Only:

Profoject™ Safelock
Disposable Insulin Syringe
The device is intended for the delivery of U-40 or U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks.

Indications for use – Rx and OTC:

Profoject™ Insulin Syringes (Model A)
The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.

Profoject™ Insulin Syringes (Model B)
The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.

Device Description:

Profoject™ Insulin Syringes
The Profoject™ Insulin Syringes are categorized into Model A and Model B, the difference is that Model A utilizes an integrated hub and barrel design, and Model B has a separate hub and barrel. The proposed device are 0.3mL, 0.5mL, and 1mL syringes designed for subcutaneous injection of a desired dose of insulin, consisting of needle cap, needle tube, plunger stopper, plunger, barrel and hub (only Model B). Profoject™ Insulin Syringes are sterile, single-use, non-toxic and work on the principle of a piston syringe.

Profoject™ Safelock
Disposable Insulin Syringe
The Profoject™ Safelock Disposable Insulin Syringe are 0.3mL, 0.5mL, and 1mL syringes designed for subcutaneous injection of a desired dose of insulin, consisting of needle cap, needle tube, plunger stopper, plunger, barrel, collar and protective shield. The Profoject™ Safelock Disposable Insulin Syringe is a sterile, single-use, non-toxic and work on the

{5}

principle of a piston syringe.

# 3. Predicate Device Identification

Insulin Syringes

K223453 - Insulin Syringe

Safelock Disposable Insulin Syringe

K061492 - Kendall Monoject Magellan Insulin Safety Syringe

# 4. Non-Clinical Test Conclusion

Non-clinical verification of the Projoject™ Insulin Syringes, Projoject™ Safelock Disposable Insulin Syringe have been conducted to evaluate their safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the performance of the proposed device in accordance with relevant test methods cited below:

# Performance Testing

1. ISO 7864 Fourth edition 2016-08-01 Sterile hypodermic needles for single use - Requirements and test methods
2. ISO 9626 Second edition 2016-08-01 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
3. ISO 8537 Third edition 2016-03-15 Sterile single-use syringes, with or without needle, for insulin
4. ISO 23908: 2011 Sharps Injury Protection – Requirements and Test Methods – Sharps Protection Features for Single Use Hypodermic Needles, Introducers for Catheters and Needles Uses for Blood Sampling
5. ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
6. ASTM F1140/F1140M-13 (2020) Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
7. ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
8. ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
9. USP - NF &lt;71&gt; Sterility Tests
10. ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
11. ISTA 3A: 2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
12. ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
13. USP - NF &lt;788&gt; Particulate Matter in Injections
14. USP - NF &lt;85&gt; Bacterial Endotoxin Testing
15. ISO 11607-1 Second edition 2019-02 [Including AMD1:2023] Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including Amendment 1 (2023)]

{6}

3

Usability Testing

FDA guidance document Medical Devices with Sharps Injury Prevention Features.

Biocompatibility Testing

The biocompatibility evaluation for the Profoject™ Insulin Syringes, Profoject™ Safelock Disposable Insulin Syringe were conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as an externally communicating with prolonged, blood path indirect contact.

1. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2. ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
3. ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
4. USP &lt;151&gt; Rabbit Pyrogen Study
5. ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
6. ISO 10993-4 Third edition 2017-04 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood
7. ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials

5. Substantial Equivalent Based on Assessment of Clinical Performance Data

Clinical data was not including in this submission

{7}

# 6. Substantially Equivalent Comparison Conclusion

Table 1 - Profoject™ Insulin Syringe

|  Parameters |   | Proposed Device | Predicate Device | Remark  |
| --- | --- | --- | --- | --- |
|  1 | 510(k) Number | K253068 | K223453 | --  |
|  2 | 510(k) Holder | CMT HEALTH PTE. LTD. | Promisemed Hangzhou Meditech Co., Ltd. | --  |
|  3 | Trade Name | Profoject™ Insulin Syringes | Insulin Syringe | --  |
|  4 | Common Name | Syringe, Piston, needle, hypodermic, single lumen | Syringe, Piston, needle, hypodermic, single lumen | Same  |
|  5 | Regulation Name | Piston Syringe, Hypodermic single lumen needle | Piston Syringe, Hypodermic single lumen needle | Same  |
|  6 | Product Code | FMF | FMF | Same  |
|  7 | Regulation Number | 21 CFR 880.5860 | 21 CFR 880.5860 | Same  |
|  8 | Review Panel | General Hospital | General Hospital | Same  |
|  9 | Device Class | II | II | Same  |
|  10 | Indications for use | The Profoject™ Insulin Syringes are intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. | Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. | Same  |
|  11 | Type of use | Prescription use and over the-counter use | Prescription use and over the-counter use | Same  |
|  12 | Operating Principle | The insulin is injected to subcutaneous tissue by pushing force generated through pushing plunger of the insulin syringe. | The insulin is injected to subcutaneous tissue by pushing force generated through pushing plunger rod of the insulin syringe. | Same  |
|  13 | Specific drug use | Insulin | Insulin | Same  |
|  14 | Volume | 0.3ml, 0.5ml, 1ml | 0.3ml, 0.5ml, 1ml | Same  |
|  15 | Gauge | 27G, 28G, 29G, 30G, 31G | 32G, 31G, 30G, 29G, 28G | Different Note 1  |
|  16 | Length | 6 mm, 8 mm, 10 mm, 13 mm, 15mm | 6 mm, 8 mm, 12 mm  |   |
|  17 | Needle cover color | Red (U-40) and orange (U-100) | Red (U-40) and orange (U-100) | Same  |
|  18 | Lubricant composition | Needle point lubricant: Polydimethylsiloxane Barrel lubricant: Polydimethylsiloxane | Aminofuntional siloxane | Different Note 2  |
|  19 | Barrel transparency | Transparent | Transparent | Same  |
|  20 | Biocompatibility | No cytotoxicity No irritation reactivity No significant evidence of skin sensitization | No cytotoxicity No irritation reactivity No significant evidence of skin sensitization | Same  |

{8}

Table 2 - Profoject™ Safelock Disposable Insulin Syringe

|  Parameters |   | Proposed device | Predicate Device | Remark  |
| --- | --- | --- | --- | --- |
|  1 | 510(k) Number | K253068 | K061492 | --  |
|  2 | 510(k) Holder | CMT HEALTH PTE. LTD. | TYCO HEALTHCARE | Same  |
|  3 | Trade Name | Profoject™ Safelock Disposable Insulin Syringe | Kendall Monoject Magellan Insulin Safety Syringe | --  |
|  4 | Common Name | syringe, antistick | syringe, antistick | Same  |
|  5 | Regulation Name | Piston syringe | Piston syringe | Same  |
|  6 | Product Code | MEG | MEG | Same  |

{9}

|  7 | Regulation Number | 21 CFR 880.5860 | 21 CFR 880.5860 | Same  |
| --- | --- | --- | --- | --- |
|  8 | Review Panel | General Hospital | General Hospital | Same  |
|  9 | Device Class | II | II | Same  |
|  10 | Indications for use | The device is intended for the delivery of U-40 or U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks. | The device is intended for the delivery of U-100 insulin. The needle stick prevention feature of the device, once activated, guards against accidental needle-sticks. | Different Note 1  |
|  11 | Design Features | -Manually operated safety feature to prevent accidental needle sticks.
-Graduation markings | -Manually operated safety feature to prevent accidental needle sticks.
-Graduation markings | Same  |
|  12 | Mechanism of Action | Mechanical delivery of U-40 and U-100 insulin. Needle-stick prevention feature is manually activated by the user by a finger-tip or thumb operation. | Mechanical delivery of U-100 insulin. Needle-stick prevention feature is manually activated by the user by a finger-tip or thumb operation. | Different Note 2  |
|  13 | Components | Sterile Syringe
Single lumen needle
Safety shield/feature | Sterile Syringe
Single lumen needle
Safety shield/feature | Same  |
|  14 | Prescription vs. OTC | Prescription Use Only | Prescription Use Only | Same  |
|  15 | Sterile vs. Non-Sterile | Sterile | Sterile | Same  |
|  16 | Sterilization Method | EO | Gamma Radiation | Different Note 3  |
|  17 | Shelf life | 5 years | 5 years | Same  |
|  18 | Single Use vs. Reusable | Single Use | Single Use | Same  |
|  19 | Non-pyrogenic | Yes | Yes | Same  |
|  20 | Needle Size | 29G, 30G, 31G | 29G, 30G | Different Note 4  |
|  21 | Needle Length | 1/2", 5/16", 3/8", 1/4" | 1/2", 5/16" |   |
|  22 | Wall Thickness | Regular Wall, Thin Wall | Regular Wall |   |
|  23 | Lubricant | Polydimethylsiloxane | Silicone | Different Note 5  |
|  24 | Needle Cap Color | U-40 Insulin: Red
U-100 Insulin: Orange | Orange | Different Note 6  |
|  25 | Device Materials (barrel and shield) | Polypropylene and Polypropylene | Polypropylene and Polypropylene | Same  |
|  Differences between proposed device and predicate device, reference device:  |   |   |   |   |
|  **Note 1:**
The proposed device and the predicate device have the same indications for use of insulin injection and the proposed device has another syringe type for U-40 insulin injection. The difference between U-40 insulin  |   |   |   |   |

{10}

syringe and U-100 insulin syringe is the insulin concentration delivery, the U-40 insulin syringe has red needle cap for color marking in accordance with ISO 8537 to distinguish the insulin concentration. In addition, the tolerance on graduated capacity has been validated in accordance with ISO 8537. Therefore, the differences between U-40 insulin syringes and U-100 insulin syringes do not raise any new questions of safety and effectiveness.

**Note 2:**
The mechanism of action of the insulin syringes of the proposed device and predicate device is the same. The difference between U-40 insulin syringe and U-100 insulin syringe is the insulin concentration delivery, the U-40 insulin syringe has red needle cap for color marking in accordance with ISO 8537 to distinguish the insulin concentration.

**Note 3:**
Although the proposed device and the predicate device were sterilized differently, the proposed device was sterilized by EO to achieve a SAL rating of $10^{-6}$, which is the same as the predicate device, and therefore does not raise new questions of safety and effectiveness when compared to the predicate device.

**Note 4:**
The proposed device's needle size, needle length and wall thickness are different from the predicate device. This difference does not affect intended use. The differences on wall thickness does not raise new questions of safety and effectiveness when compared to the predicate device. In addition, differences in wall thickness between the predicate and proposed device were addressed through ISO 8537 and ISO 9626 performance testing.

**Note 5:**
The material of lubricant is different between the proposed device and predicate device. This difference does not raise any new questions of safety and effectiveness when compared to the predicate device. The biocompatibility test of the proposed device was conducted to demonstrate that the proposed device met the biocompatibility requirements.

**Note 6:**
The insulin syringe needle cap colors are different among the proposed device and predicate device. The U-40 insulin syringe has red protective cap for color marking in accordance with ISO 8537 to distinguish the insulin concentration. In addition, biocompatibility on the needle caps was conducted and supports that this difference does not impact the safety and effectiveness of the device to claim substantially equivalent to the predicate device.

**The Conclusions:**
The conclusions drawn from the non-clinical tests demonstrate that the devices are as safe, as effective, and performs as well as the legally marketed devices identified in the submission. Thus the subject devices are substantially equivalent to the predicate devices.

7

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K253068](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K253068)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
