Allograft Delivery Device (OFAC-C)

K233368 · Bioventus, LLC · FMF · Jun 25, 2024 · General Hospital

Device Facts

Record IDK233368
Device NameAllograft Delivery Device (OFAC-C)
ApplicantBioventus, LLC
Product CodeFMF · General Hospital
Decision DateJun 25, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2

Indications for Use

The Allograft Delivery Device is intended to be used for the delivery of hydrated allograft to an orthopedic surgical site.

Device Story

The Allograft Delivery Device is a sterile, single-use, disposable instrument designed for orthopedic surgery. It consists of a cannula barrel and a push rod. The device functions by containing hydrated allograft bone graft material within the cannula and expressing it through a single distal hole into the surgical site via manual depression of the syringe plunger and push rod. It is intended for use by clinicians during orthopedic procedures requiring allograft placement. The device facilitates controlled delivery of graft material to the surgical site, potentially improving the efficiency of bone graft placement. It is provided sterile and is compatible with pre-existing allograft syringes via a male double-threaded lock interface.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included sterilization validation (ISO 11137), shelf-life validation (ASTM F1980-21), sterile barrier testing (ASTM F1886, F2096, F88), biocompatibility (ISO 10993-1), functional verification (cannula buckling, plunger force, connection strength, durability), shipping/transit testing (ASTM D4169-22), and usability testing (IEC 62366-1).

Technological Characteristics

Materials: Biocompatible medical-grade polypropylene. Form factor: Curved cannula barrel with push rod; 5.5 mm distal exit hole. Capacity: Up to 4.6 cc. Connectivity: None (manual). Sterilization: Gamma irradiation (SAL 10^-6). Standards: ISO 11137-1/2/3, ASTM F1980-21, ASTM F1886/F1886M-16, ASTM F2096-11, ASTM F88/F88M-21, ISO 10993-1, ASTM D4169-22, IEC 62366-1.

Indications for Use

Indicated for the delivery of hydrated allograft to an orthopedic surgical site in patients undergoing orthopedic surgery.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size. June 25, 2024 Bioventus LLC Shanna Hu Sr. Manager RA, FES Technology 4721 Emperor Blvd., Suite 100 Durham, North Carolina 27703 Re: K233368 Trade/Device Name: Allograft Delivery Device (OFAC-C) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: May 24, 2024 Received: May 24, 2024 Dear Shanna Hu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long H. Chen =S Digitally signed by Long H. Chen-S Date: 2024.06.25 15:21:19 -04'00' Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use indications for Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. Submission Number (if known) K233368 Device Name Allograft Delivery Device (OFAC-C) Indications for Use (Describe) The Allograft Delivery Device is intended to be used for the delivery of hydrated allograft to an orthopedic surgical site. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Bioventus. The logo consists of a circular, interwoven design in shades of purple and orange on the left. To the right of the design is the word "bioventus" in a sans-serif font, colored in purple. Manufacturer: 510(k): Bioventus 4721 Emperor Blvd., Suite 100 Durham, NC 27703 USA #### K233368 510(k) Summary # 510(k) Summary Allograft Delivery Device | Submitter Information: | | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company: | Bioventus LLC<br>4721 Emperor Boulevard<br>Suite 100<br>Durham, NC 27703 | | Contact: | Shanna Hu<br>Senior Manager, Regulatory Affairs<br>(661) 362-6694<br>shanna.hu@bioventus.com<br><br>Kellie Stefaniak<br>Vice President, Regulatory Affairs<br>(919) 566-1224<br>kellie.stefaniak@bioventus.com | | Date Prepared: | June 24, 2024 | | Device Information: | | | Trade Name: | Allograft Delivery Device | | Common Name: | Graft Delivery Device | | Regulation Name: | Piston Syringe | | Regulation Number: | 21 CFR 880.5860 | | Regulatory Classification: | Class II | | Product Code: | FMF | | Proposed Panel: | General Hospital | | Predicate Device: | | | Name: | Bi-Portal Bone Graft Delivery Device (KGM1 Graft Delivery Canula) | Spinal Surgical Strategies, LLC (Kleiner Device Labs) K142661 {4}------------------------------------------------ # Device Description: The Allograft Delivery Device is a sterile, single-use, disposable allograft delivery device intended for the delivery of hydrated allograft bone graft material to an orthopedic surgical site. The delivery device consists of a cannula used for containing and delivering the allograft material to the surgical site and a push rod required for expressing the allograft material from the cannula. The cannula component is an open bore tube with a male double-threaded lock interface that mates with a pre-existing allograft syringe. The cannula component may be used to dispense allograft to a patient orthopedic surgical site. When used as an ancillary device to orthopedic surgical procedures that require the placement of allograft Delivery Device can deliver a prepared amount of allograft material. # Indications for Use: The Allograft Delivery Device is intended to be used for the delivery of hydrated allograft to an orthopedic surgical site. #### Substantial Equivalence: The Allograft Delivery Device was evaluated to verify substantial equivalence. The indication for use of the subject device is the same as the predicate device. The comparison table provides an overview of the indications, key features, and technological comparison of the device to the predicate device. These comparisons support substantial equivalence for the intended use, technology, function, and mechanism of action to the predicate. The Allograft Delivery Device does not introduce any new issues of safety or efficacy relative to the predicate device. #### Comparison Table: | Attributes | Allograft Delivery Device<br>(Subject Device) | Bi-Portal Bone Graft<br>Delivery Device<br>(Predicate Device) | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant | Bioventus LLC | Spinal Surgical Strategies,<br>LLC (Kleiner Device Labs) | | Regulation No. & | 21 CFR 880.5860, Piston | 21 CFR 880.5860, Piston | | Name | Syringe | Syringe | | Product Code | FMF | FMF | | 510(k) Number | K233368 | K142661 | | Indications for Use | The Allograft Delivery<br>Device is intended to be<br>used for the delivery of<br>hydrated allograft to an<br>orthopedic surgical site. | The Bi-Portal Bone Graft<br>Delivery Device is<br>intended to be used for the<br>delivery of hydrated<br>allograft or autograft to an<br>orthopedic surgical site. | {5}------------------------------------------------ | Prescription / OTC Use | Prescription Use | Prescription Use | |---------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------| | Type of Use | Single-Use, Disposable | Single-Use, Disposable | | Mechanism of<br>Operation | Graft dispersed from device<br>tip by depressing syringe<br>plunger and by the action of<br>a push rod. | Graft dispersed from<br>device tip by depressing<br>plunger | | Graft Injection | Single distal hole delivery<br>into surgical space | Bilateral sidehole delivery<br>into surgical space | | Device Materials | Biocompatible Medical<br>grade polypropylene | Biocompatible Medical<br>Grade Plastics | | Device Components | Cannula Barrel and Push<br>Rod | Barrel, Plunger and Funnel | | Bone Graft Exit Hole<br>Diameter | Single distal hole<br>5.5 mm | Two sideholes:<br>5.6 mm x 14.8mm | | Barrel Shape | Round, Curved | Rectangular, Straight | | Volume Capacity | Up to 4.6 cc in Cannula<br>Barrel | Up to 4.0 cc in Barrel | | Sterility Claim /<br>Sterility Assurance<br>Level (SAL) | Sterile, SAL 1 x 10-6 | Sterile, SAL 1 x 10-6 | | Sterilization Method | Gamma irradiation | Ethylene Oxide gas | # Performance Data: The following performance data were provided in support of the substantial equivalence determination: # Sterilization and Shelf-life Sterilization validation was conducted to confirm that the subject device reached 10℃ sterility assurance level and the performance of the device and packaging are in line with expectations after sterilization requirements per ISO 11137-1:2006, ISO 11137-2:2013, and ISO 11137-3: 2017 standards. The sterilization process has been successfully validated and the sterility tests are within satisfactory limits. The shelf-life validation study was conducted under the accelerated aging condition according to ASTM F1980-21 to verify the packaging integrity as well as the physical properties of the Allograft Delivery Device are within pass limits both before and after the aging process. The results successfully demonstrate an up to 1 year shelf life. {6}------------------------------------------------ Sterile barrier testing was conducted in compliance with the following FDA-recognized consensus standards: - . Visual Inspection, F1886/F1886M-16 - Bubble Leak, ASTM F2096-11 - Seal Strength, ASTM F88/F88M-21 # Biocompatibility Testing The biocompatibility evaluation for the Allograft Delivery Device was conducted in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" issued on September 8, 2023, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following tests were conducted: - Cytotoxicity - Pyrogenicity - Irritation - Acute Systemic Toxicity ● - Sensitization ● All of the tests received passing results, demonstrating that the Allograft Delivery Device meets the applicable requirements outlined in the FDA guidance on ISO 10993-1. #### Functional verification - о Cannula Buckling - o Plunger Force - Cannula Connection Strength o - No gross separation of fluids from allograft - Cannula Durability o - o Push Rod Function/Strength # Shipping/Transit Shipping/Transit testing was conducted as per ASTM D4169-22 #### Usability testing Usability testing was conducted as per IEC 62366-1 edition 1.1 2020-06 The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is substantially equivalent to the predicate device. {7}------------------------------------------------ # Conclusion: The Allograft Delivery Device is substantially equivalent to the Bi-Portal Bone Graft Delivery Device in intended use, mechanism of operation, and patient-contacting materials. Performance testing has been conducted to demonstrate the subject device and predicate device technological characteristics are equivalent and that the subject device is as safe and as effective as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...