← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K212764

# Medwell Technology Ltd Syringes, various sizes (K212764)

_Medwell Technology, Ltd. · FMF · Nov 4, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K212764

## Device Facts

- **Applicant:** Medwell Technology, Ltd.
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** Nov 4, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Medwell Technology Polycarbonate Syringes are intended to inject fluids into or withdraw fluids from the body.

## Device Story

Medwell Technology Polycarbonate Syringes are manual, single-use piston syringes used to inject or withdraw fluids. The device consists of a rigid polycarbonate barrel, a plunger made of ABS polymer, and a synthetic rubber stopper (tip) lubricated with medical-grade silicone oil. The barrel features graduated markings (1-30 mL) for volume measurement. The device is available with either luer slip or luer lock tips. It is intended for use by healthcare professionals in clinical settings. The device is provided sterile (via ethylene oxide) or non-sterile for further processing. The user manually operates the plunger to draw or expel fluids. The device facilitates fluid delivery or aspiration, aiding in medication administration or sample collection.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing included compliance with ISO 7886-1 (syringe performance), ISO 80369-7 (luer connectors), and ISO 10993-1 (biocompatibility). Sterile barrier integrity was validated via ASTM F1886/F1886M-16, ASTM F88/F88-15, and ASTM F1929-15. Sterilization validated per ISO 11135.

## Technological Characteristics

Materials: Polycarbonate barrel, ABS polymer plunger, synthetic rubber elastomer tip, medical-grade silicone oil. Dimensions: 1cc - 30cc volume. Connectivity: None. Energy: Manual. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 7886-1, ISO 80369-7, ISO 10993-1, ASTM F1886/F1886M-16, ASTM F88/F88-15, ASTM F1929-15.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- MHC Standard and NRFit Tip Syringes ([K171131](/device/K171131.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food & Drug Administration (FDA). The logo features a stylized caduceus symbol on the left, followed by the letters "FDA" in a bold, sans-serif font. To the right of "FDA" is the text "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION", also in a sans-serif font. The color scheme is primarily blue and white.

November 4, 2022

Medwell Technology Ltd Jonathan Gilbert Regulatory/Clinical Affairs Consultant to Medwell Technology Ltd 1641 Jeurissen Lane Chanhassen, Minnesota 55317

Re: K212764

Trade/Device Name: Medwell Technology Polycarbonate Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: October 11, 2022 Received: October 12, 2022

#### Dear Jonathan Gilbert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan M.
Stevens
-S3

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212764

#### Device Name

Medwell Technology Polycarbonate Syringes

Indications for Use (Describe)

The Medwell Technology Polycarbonate Syringes are intended to inject thuids into or withdraw fluids from the body.

Type of Use (Select one or both, as applicable)

| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black"></path> </svg>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16"></rect> </svg>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div>                                                           |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K212764 510(k) Summary

| 1. Sponsor:                       | Medwell Technology Ltd<br>8th Building, Pujing, Fumin Industry Zone, DaLang Town<br>Dongguan City, Guandong 523770 China                                                                                                                                                                                                                                                                                                                                                               |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact                           | Jon Gilbert, Consultant<br>906.361.3237<br>jgilb.raca@gmail.com                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 2. Preparation Date:              | November 1, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| 3. Subject Device<br>Common Name: | Piston Syringes                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name:                       | Medwell Technology Polycarbonate Syringes                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Regulation Number:                | 21 CFR 880.5860                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Regulation Name:                  | Piston syringe                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Panel:             | General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulatory Class:                 | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code:                     | FMF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| 4. Predicate Device               | MHC Standard and NRFit Tip Syringes<br>K171131, cleared November 8, 2017                                                                                                                                                                                                                                                                                                                                                                                                               |
| 5. Device Description:            | The Medwell Technology PC Syringes are provided<br>sterile or in bulk non-sterile for further processing (e.g.<br>sterilization). They are single use devices consisting of<br>rigid polycarbonate barrels with luer slip or luer lock tips<br>and various colored plungers with a synthetic rubber tip<br>(stopper). The syringe barrels are printed with graduated<br>markings in cc (1- 30 milliliters) indicating the volume of<br>liquid inside the various size syringe barrels. |
| 6. Indication for Use:            | The Medwell Technology Polycarbonate Syringes are<br>intended to inject fluids into or withdraw fluids from the body.                                                                                                                                                                                                                                                                                                                                                                  |

{4}------------------------------------------------

| 7. Substantially Equivalent (SE) Comparison:                           |                                                                                                     |                                                                                                  |             |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------|
| Table 2. Technological Characteristics & Substantial Equivalence Table |                                                                                                     |                                                                                                  |             |
| Item                                                                   | Propose Device<br>K212764                                                                           | Predicate Device<br>K171131                                                                      | Comparison* |
| Product name                                                           | Sterile and Non-sterile<br>Hypodermic Syringes for<br>Single Use                                    | Sterile and Non-sterile<br>Hypodermic Syringes for<br>Single Use                                 | Same        |
| Product Code                                                           | FMF                                                                                                 | FMF                                                                                              | Same        |
| Regulation No.                                                         | 21 CFR 880.5860                                                                                     | 21 CFR 880.5860                                                                                  | Same        |
| Class                                                                  | CLASS II                                                                                            | CLASS II                                                                                         | Same        |
| Syringe Volume                                                         | 1cc - 30cc                                                                                          | 1cc - 30cc                                                                                       | Same        |
| Nozzle Type                                                            | Luer slip; Luer lock                                                                                | Luer slip; Luer lock                                                                             | Same        |
| Configuration and<br>material                                          | Barrel Polycarbonate                                                                                | Barrel Polycarbonate                                                                             | Same        |
|                                                                        | Plunger ABS polymer                                                                                 | Plunger ABS Polymer                                                                              |             |
|                                                                        | Plunger Tip Elastomer                                                                               | Plunger Tip Elastomer                                                                            |             |
|                                                                        | Silicone Oil Medical Grade                                                                          | Silicone Oil Medical Grade                                                                       | Same        |
| Operation Mode                                                         | Manual Use                                                                                          | Manual Use                                                                                       | Same        |
| Syringe<br>Performance                                                 | Complies with ISO<br>7886-1:2017                                                                    | Complies with ISO 7886-1:2017                                                                    | Same        |
| Biocompatibility                                                       | Complies with ISO<br>10993 series standards<br>for patient contact<br>profile > 24 hrs < 30<br>days | Complies with ISO<br>10993 series standards<br>for patient contact profile<br>> 24 hrs < 30 days | Same        |
| Rx Use / OTC Use                                                       | Yes / No                                                                                            | Yes / No                                                                                         | Same        |
| Single Use                                                             | Yes                                                                                                 | Yes                                                                                              | Same        |
| Shelf-life                                                             | 3 years                                                                                             | 1 year                                                                                           | Same**      |
| Sterilization                                                          | EO                                                                                                  | EO                                                                                               | Same        |
| SAL                                                                    | 10-6                                                                                                | 10-6                                                                                             | Same        |
| Label/Labeling                                                         | Complies with 21 CFR<br>801                                                                         | Complies with 21 CFR<br>801                                                                      | Same        |

*Subject device PC syringe components are comprised of same materials and manufactured using same manufacturing methods. K171131 references identical materials and manufacturing methods for standard PC syringes (NRFIT tip configurations excluded).

**Predicate device was cleared with performance testing supported by 1 year shelf life. 3 year shelf life assigned for subject device is consistent with performance data testing (accelerated aging, device performance and packaging verification) and raises no new questions about safety or performance.

The clinical technique, indications for use, technical specifications, materials used, sterility status (validation and sterility assurance level) as well as the biocompatibility status are identical between the subject and predicate syringes.

{5}------------------------------------------------

## 8. Non-Clinical Performance Testing:

The performance of the subject Medwell Technology PC Syringes is identical to the predicate devices as no material, manufacturing or technological changes have occurred.

Testing per ISO 7886-1, ISO 80369-7, and ISO 10993-1 were conducted to demonstrate conformance to the recognized standards and substantial equivalence with the predicate device.

## Biocompatibility Testing:

In accordance with ISO 10993-1, the syringe complies with the standards for patient contact profile > 24 hrs < 30 days.

## Sterile Barrier Packaging Test

Sterile barrier packaging testing was performed on the proposed device, which include visual inspection (ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15) and dye penetration test (ASTM F1929-15). The test result showed that the device package can maintain its integrity.

#### Sterilization and Shelf-Life Test

The EO sterilization method has been validated per ISO 11135 and defines the routine control and monitoring parameters. The shelf life of the Sterile Hypodermic Syringe for Single Use is 3 years, determined based on stability studies which includes accelerated aging. Sterilization and shelf-life testing listed were performed on the proposed device.

#### Animal and Clinical Tests

Neither animal nor clinical testing was required to demonstrate performance of the subject and predicate devices. Product functionality has been adequately assessed by non-clinical tests.

#### 9. Conclusion:

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Medwell Technology PC Syringes are substantially equivalent to the Standard Syringes of K171131 with respect to the indications for use, target populations, treatment method, and technological characteristics.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K212764](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K212764)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
