K212764 · Medwell Technology, Ltd. · FMF · Nov 4, 2022 · General Hospital
Device Facts
Record ID
K212764
Device Name
Medwell Technology Ltd Syringes, various sizes
Applicant
Medwell Technology, Ltd.
Product Code
FMF · General Hospital
Decision Date
Nov 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medwell Technology Polycarbonate Syringes are intended to inject fluids into or withdraw fluids from the body.
Device Story
Medwell Technology Polycarbonate Syringes are manual, single-use piston syringes used to inject or withdraw fluids. The device consists of a rigid polycarbonate barrel, a plunger made of ABS polymer, and a synthetic rubber stopper (tip) lubricated with medical-grade silicone oil. The barrel features graduated markings (1-30 mL) for volume measurement. The device is available with either luer slip or luer lock tips. It is intended for use by healthcare professionals in clinical settings. The device is provided sterile (via ethylene oxide) or non-sterile for further processing. The user manually operates the plunger to draw or expel fluids. The device facilitates fluid delivery or aspiration, aiding in medication administration or sample collection.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included compliance with ISO 7886-1 (syringe performance), ISO 80369-7 (luer connectors), and ISO 10993-1 (biocompatibility). Sterile barrier integrity was validated via ASTM F1886/F1886M-16, ASTM F88/F88-15, and ASTM F1929-15. Sterilization validated per ISO 11135.
Technological Characteristics
Materials: Polycarbonate barrel, ABS polymer plunger, synthetic rubber elastomer tip, medical-grade silicone oil. Dimensions: 1cc - 30cc volume. Connectivity: None. Energy: Manual. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 7886-1, ISO 80369-7, ISO 10993-1, ASTM F1886/F1886M-16, ASTM F88/F88-15, ASTM F1929-15.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from the body. For prescription use.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food & Drug Administration (FDA). The logo features a stylized caduceus symbol on the left, followed by the letters "FDA" in a bold, sans-serif font. To the right of "FDA" is the text "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION", also in a sans-serif font. The color scheme is primarily blue and white.
November 4, 2022
Medwell Technology Ltd Jonathan Gilbert Regulatory/Clinical Affairs Consultant to Medwell Technology Ltd 1641 Jeurissen Lane Chanhassen, Minnesota 55317
Re: K212764
Trade/Device Name: Medwell Technology Polycarbonate Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: October 11, 2022 Received: October 12, 2022
#### Dear Jonathan Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens
-S3
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212764
#### Device Name
Medwell Technology Polycarbonate Syringes
Indications for Use (Describe)
The Medwell Technology Polycarbonate Syringes are intended to inject thuids into or withdraw fluids from the body.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212764 510(k) Summary
| 1. Sponsor: | Medwell Technology Ltd<br>8th Building, Pujing, Fumin Industry Zone, DaLang Town<br>Dongguan City, Guandong 523770 China |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Jon Gilbert, Consultant<br>906.361.3237<br>jgilb.raca@gmail.com |
| 2. Preparation Date: | November 1, 2022 |
| 3. Subject Device<br>Common Name: | Piston Syringes |
| Trade Name: | Medwell Technology Polycarbonate Syringes |
| Regulation Number: | 21 CFR 880.5860 |
| Regulation Name: | Piston syringe |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | FMF |
| 4. Predicate Device | MHC Standard and NRFit Tip Syringes<br>K171131, cleared November 8, 2017 |
| 5. Device Description: | The Medwell Technology PC Syringes are provided<br>sterile or in bulk non-sterile for further processing (e.g.<br>sterilization). They are single use devices consisting of<br>rigid polycarbonate barrels with luer slip or luer lock tips<br>and various colored plungers with a synthetic rubber tip<br>(stopper). The syringe barrels are printed with graduated<br>markings in cc (1- 30 milliliters) indicating the volume of<br>liquid inside the various size syringe barrels. |
| 6. Indication for Use: | The Medwell Technology Polycarbonate Syringes are<br>intended to inject fluids into or withdraw fluids from the body. |
{4}------------------------------------------------
| 7. Substantially Equivalent (SE) Comparison: | | | |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------|
| Table 2. Technological Characteristics & Substantial Equivalence Table | | | |
| Item | Propose Device<br>K212764 | Predicate Device<br>K171131 | Comparison* |
| Product name | Sterile and Non-sterile<br>Hypodermic Syringes for<br>Single Use | Sterile and Non-sterile<br>Hypodermic Syringes for<br>Single Use | Same |
| Product Code | FMF | FMF | Same |
| Regulation No. | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | CLASS II | CLASS II | Same |
| Syringe Volume | 1cc - 30cc | 1cc - 30cc | Same |
| Nozzle Type | Luer slip; Luer lock | Luer slip; Luer lock | Same |
| Configuration and<br>material | Barrel Polycarbonate | Barrel Polycarbonate | Same |
| | Plunger ABS polymer | Plunger ABS Polymer | |
| | Plunger Tip Elastomer | Plunger Tip Elastomer | |
| | Silicone Oil Medical Grade | Silicone Oil Medical Grade | Same |
| Operation Mode | Manual Use | Manual Use | Same |
| Syringe<br>Performance | Complies with ISO<br>7886-1:2017 | Complies with ISO 7886-1:2017 | Same |
| Biocompatibility | Complies with ISO<br>10993 series standards<br>for patient contact<br>profile > 24 hrs < 30<br>days | Complies with ISO<br>10993 series standards<br>for patient contact profile<br>> 24 hrs < 30 days | Same |
| Rx Use / OTC Use | Yes / No | Yes / No | Same |
| Single Use | Yes | Yes | Same |
| Shelf-life | 3 years | 1 year | Same** |
| Sterilization | EO | EO | Same |
| SAL | 10-6 | 10-6 | Same |
| Label/Labeling | Complies with 21 CFR<br>801 | Complies with 21 CFR<br>801 | Same |
*Subject device PC syringe components are comprised of same materials and manufactured using same manufacturing methods. K171131 references identical materials and manufacturing methods for standard PC syringes (NRFIT tip configurations excluded).
**Predicate device was cleared with performance testing supported by 1 year shelf life. 3 year shelf life assigned for subject device is consistent with performance data testing (accelerated aging, device performance and packaging verification) and raises no new questions about safety or performance.
The clinical technique, indications for use, technical specifications, materials used, sterility status (validation and sterility assurance level) as well as the biocompatibility status are identical between the subject and predicate syringes.
{5}------------------------------------------------
## 8. Non-Clinical Performance Testing:
The performance of the subject Medwell Technology PC Syringes is identical to the predicate devices as no material, manufacturing or technological changes have occurred.
Testing per ISO 7886-1, ISO 80369-7, and ISO 10993-1 were conducted to demonstrate conformance to the recognized standards and substantial equivalence with the predicate device.
## Biocompatibility Testing:
In accordance with ISO 10993-1, the syringe complies with the standards for patient contact profile > 24 hrs < 30 days.
## Sterile Barrier Packaging Test
Sterile barrier packaging testing was performed on the proposed device, which include visual inspection (ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15) and dye penetration test (ASTM F1929-15). The test result showed that the device package can maintain its integrity.
#### Sterilization and Shelf-Life Test
The EO sterilization method has been validated per ISO 11135 and defines the routine control and monitoring parameters. The shelf life of the Sterile Hypodermic Syringe for Single Use is 3 years, determined based on stability studies which includes accelerated aging. Sterilization and shelf-life testing listed were performed on the proposed device.
#### Animal and Clinical Tests
Neither animal nor clinical testing was required to demonstrate performance of the subject and predicate devices. Product functionality has been adequately assessed by non-clinical tests.
#### 9. Conclusion:
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Medwell Technology PC Syringes are substantially equivalent to the Standard Syringes of K171131 with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.