K211728 · Jiangsu Kangyou Medical Instrument Co., Ltd. · FMF · Oct 5, 2021 · General Hospital
Device Facts
Record ID
K211728
Device Name
Sterile hypodermic syringes for single use
Applicant
Jiangsu Kangyou Medical Instrument Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Oct 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile hypodermic syringes for single use are intended to inject fluids into or withdraw fluids from the body.
Device Story
Sterile hypodermic syringe for single use; consists of barrel, plunger, piston, and optional hypodermic needle (tube, hub, cap). Manual operation by clinician; used to aspirate or inject fluids. Available in various syringe volumes (1ml-60ml) and needle gauges (16G-30G). Device facilitates fluid delivery or withdrawal; supports clinical decision-making by enabling precise dosing and sampling. Benefits patient through sterile, single-use delivery mechanism. Used in clinical settings.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via biocompatibility (ISO 10993), sterility (ISO 11135), and mechanical/functional testing (ISO 7886-1, ISO 7864, ISO 80369-7, ISO 9626).
Technological Characteristics
Materials: Polypropylene (barrel, plunger, hub, cap), Polyisoprene (piston), Stainless steel (needle tube). Manual operation. Luer Slip and Luer Lock connectors. EO sterilization (SAL 10^-6). Standards: ISO 7886-1, ISO 7864, ISO 80369-7, ISO 9626, ISO 10993, ASTM F1980-16, ASTM F88/F88M-15, ASTM F1929-2015.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from the body in patients requiring such procedures.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 5, 2021
Jiangsu Kangyou Medical Instrument Co., Ltd. Xiang Yao Product Manager Tangzhuang Yaotang Town Jintan, Jiangsu 213223 China
Re: K211728
Trade/Device Name: Sterile hypodermic syringes for single use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: August 20, 2021 Received: August 30, 2021
Dear Xiang Yao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211728
#### Device Name
Sterile hypodermic syringes for single use
Indications for Use (Describe)
The Sterile hypodermic syringes for single use are intended to inject fluids into or withdraw fluids from the body.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K211728 510(k) summary
Preparation Date: October 5, 2021
# l Submitter
Device submitter: Jiangsu Kangyou Medical Instrument Co., Ltd. Tangzhuang, Yaotang Town 213223 Jintan, Jiangsu, People's Republic of China
Contact person: Xiang Yao Product Manager Phone: 13813652889 Fax: 86 519 82582461 Email: export@kym.cn
# II Device
Trade Name of Device: Sterile hypodermic syringes for single use Common Name: Piston Syringe with Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe with Needle Regulatory Class: II Product code: FMF Review Panel: General Hospital
# III Predicate Devices
| Trade name: | Sterile Single-use Syringe with Needle |
|-------------------------|--------------------------------------------|
| Common name: | Piston Syringe with Needle |
| Classification: | Class II, 21 CFR 880.5860 |
| Product Code: | FMF |
| Premarket Notification: | K163161 |
| Manufacturer: | JiangXi HongDa Medical Equipment Group Ltd |
# IV Device description
The Sterile hypodermic syringes for single use are intended for manual and single use only to aspirate and inject of fluids for medical purpose, and consist of syringe (barrel, plunger, piston) and hypodermic needle (needle tube, hub, needle cap). The proposed device is available in a variety of syringe and needle sizes.
| Syringe Size | Needle Gauge | Needle Length |
|---------------------------|------------------------|---------------------------|
| Available in 1ml、2ml、3ml、 | Available in 16G, 18G, | Available in 8mm (5/16"), |
{4}------------------------------------------------
| 5ml、10ml、20ml、30ml、 | 20G, 21G, 22G, 23G, | 13mm (1/2"), 15mm (3/5"), |
|---------------------|---------------------|------------------------------------------------------------------|
| 50ml, 60ml | 24G, 25G, 26G, 27G, | 20mm (4/5"), 25mm (1"), |
| | 29G, 30G | 30mm (1 1/6"), 32mm (1 1/4"),<br>33mm(1 3/10"), 38mm (1<br>1/2") |
The different gauge of needles could fit every specification of syringes. The needle is optional.
| Gauge<br>Syringe | 30G | 29G | 27G | 26G | 25G | 24G | 23G | 22G | 21G | 20G | 18G | 16G |
|------------------|-----|-----|-----|-----|-----|-----|-----|-----|-----|-----|-----|-----|
| 1ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 2ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 3ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 5ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 10ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 20ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 30ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 50ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
| 60ml | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
# V Indications for use
The Sterile hypodermic syringes for single use are intended to inject fluids into or withdraw fluids from the body.
# VI Comparison of technological characteristics with the predicate devices
The Sterile hypodermic syringes for single use have intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between Sterile hypodermic syringes for single use and predicate devices are the specification of syringe volume, needle gauge and needle length. However, syringes will be selected by physician per injection requirement and this difference does not affect indication for use. Additionally, the performance of syringe and needle has been evaluated and the test results met the requirements of ISO 7886-1 and ISO 7864. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.
| | Table 6-1 Substantial equivalence discussion | |
|--|----------------------------------------------|--|
| | | |
| Device<br>feature | Subject Device | Predicate Device K163161 | Comment |
|------------------------------|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------|
| Device feature | Subject Device | Predicate Device K163161 | Comment |
| Syringe type | Standard piston syringe | Standard piston syringe | Same |
| Indications for use | The Sterile hypodermic syringes for single use are intended to inject fluids into or withdraw fluids from the body. | Sterile Single-use Syringe with Needle is intended to inject fluids into or withdraw fluids from the body. | Same |
| Product code | FMF | FMF | Same |
| Regulation number | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | II | II | Same |
| Principle of operation | For manual use only | For manual use only | Same |
| Needle gauge | 16G, 18G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 29G, 30G, | 18G, 20G, 21G, 22G, 23G, 25G, 26G, 27G, 28G, 29G, 30G | Difference 1 |
| Needle Length | 5/16", 1/2", 3/5", 4/5", 1", 1 1/6", 1 1/4", 1 3/10", 1 1/2" | 1/2", 5/8", 1", 1 1/4", 1 1/2" | |
| Syringe Volume | 1ml、2ml、3ml、5ml、10ml、20ml、30ml、50ml、60ml | 1ml, 3ml, 5ml, 10ml, 20ml, 60ml | |
| Connector Type | Luer Slip and Luer Lock | Luer Slip and Luer Lock | Same |
| main structure and materials | Barrel<br>Polypropyle ne | Barrel<br>Polypropyle ne (PP) | Same |
| | Plunger<br>Polypropyle ne | Plunger<br>Polypropyle ne (PP) | |
| | Piston<br>Polyisopren e | Piston<br>Polyisopren e | |
| | Needle tube<br>Stainless steel | Needle tube<br>Stainless Steel, SUS304 | |
| | Hub<br>Polypropyle ne | Needle hub<br>Polypropyle ne (PP) | |
| | Needle cap<br>Polypropyle ne | Needle cap<br>Polypropyle ne (PP) | |
| Performance specifications | Complies with ISO 7886-1 | Complies with ISO 7886-1 | Same |
| Single Use | Yes | Yes | Same |
| Device<br>feature | Subject Device | Predicate Device K163161 | Comment |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| SAL | 10-6 | 10-6 | Same |
| Biocompatibility | Cytotoxicity<br>No cytotoxicity | Cytotoxicity<br>No cytotoxicity | Same |
| | Intracutaneous reactivity<br>No intracutaneous reactivity | Irritation<br>No intracutaneous reactivity | Same |
| | Sensitization<br>No skin sensitization | Sensitization<br>No skin sensitization | Same |
| | Systemic Toxicity<br>No systemic toxicity | Systemic Toxicity<br>No systemic toxicity | Same |
| | Hemolysis<br>No Hemolysis | Hemolysis<br>No Hemolysis | Same |
| | Pyrogen<br>No Pyrogen | Pyrogen<br>No Pyrogen | Same |
| Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
{5}------------------------------------------------
{6}------------------------------------------------
# Difference 1
The syringe volume, needle gauge and length of subject devices are different from the predicate device. However, this difference is just in dimension. Different needle specification will be selected by physician per patient's condition. This difference does not affect intended use. In addition, differences were addressed through ISO 7886-1, ISO 7864, ISO 9626 and ISO 80369-7. Therefore, the differences on syringe volume, needle gauge and length do not raise different question of safety and effectiveness.
# VII Performance data
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
Biocompatibility of the Sterile hypodermic syringes for single use were evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device - Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:
| Cytotoxicity | ISO 10993-5: 2009 |
|---------------------------|--------------------|
| Skin sensitization | ISO 10993-10: 2010 |
| Hemolysis | ISO 10993-4: 2017 |
| Intracutaneous reactivity | ISO 10993-10: 2010 |
| Acute systemic toxicity | ISO 10993-11: 2017 |
| Pyrogenicity | ISO 10993-11: 2017 |
{7}------------------------------------------------
Particulate matter
USP <788>
### Sterilization and shelf life testing
The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters. Bacteria Endotoxin Limit is carried out according to USP42-NF37 <85> Bacterial Endotoxins Test.
The testing is performed according to the following standards:
| EO residue | ISO 10993-7:2008 |
|--------------------|------------------|
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin | USP42-NF37 <85> |
The shelf life of 5 year is determined based on stability studies which include ageing test according to FDA recognized standard ASTM F1980-16.
Package integrity testing was conducted on the final, packaged, and sterile devices after environmental conditioning and simulated transportation. All packaging deemed acceptable for protection of product and sterility maintenance.
The testing is performed according to the following standards:
| Seal strength | ASTM F88/F88M-15 |
|----------------|------------------|
| Seal integrity | ASTM F 1929-2015 |
### Performance testing
Performance testing is performed according to the following standards:
| ➤ ISO 7886-1: 2017 | |
|----------------------------------------------------------|----------------------------------------------|
| Appearance | Clause 5 of ISO 7886-1:2017 |
| Limits for acidity or alkalinity | Clause 6.2 of ISO 7886-1:2017 |
| Limits for extractable metals | Clause 6.3 of ISO 7886-1:2017 |
| Lubricant | Clause 7 of ISO 7886-1:2017 |
| Tolerance on Graduated capacity | Clause 8 of ISO 7886-1:2017 |
| Scale | Clause 9.1- 9.4 of ISO 7886-1:2017 |
| Barrel flanges | Clause 10.2 of ISO 7886-1:2017 |
| Plunger stopper / Plunger assembly | Clause 11 of ISO 7886-1:2017 |
| Nozzle | Clause 12 of ISO 7886-1:2017 and ISO 80369-7 |
| Dead Space | Clause 13.1 of ISO 7886-1:2017 |
| Freedom from air and liquid leakage past plunger stopper | Clause 13.2 of ISO 7886-1:2017 |
| Force to operate the piston | Clause 13.3 of ISO 7886-1:2017 |
| Fit of plunger stopper/ plunger in barrel | Clause 13.4 of ISO 7886-1:2017 |
| ➤ ISO 7864: 2016 | |
| Cleanliness | Clause 4.3 of ISO 7864: 2016 |
| Limits for acidity or alkalinity | Clause 4.4 of ISO 7864: 2016 |
{8}------------------------------------------------
| Limits for extractable metals | Clause 4.5 of ISO 7864: 2016 |
|------------------------------------------|----------------------------------------------------------|
| Tubular needle designation | Clause 4.6 of ISO 7864: 2016 |
| Colour coding | Clause 4.7 of ISO 7864: 2016 |
| Needle hub | Clause 4.8 of ISO 7864: 2016, ISO 80369-<br>and ISO 6009 |
| Needle cap | Clause 4.9 of ISO 7864: 2016 |
| Needle tube | Clause 4.10 of ISO 7864: 2016 and ISO<br>9626:2016 |
| Needle Point | Clause 4.11 of ISO 7864: 2016 |
| Bond between Tube and Hub | Clause 4.12 of ISO 7864: 2016 |
| Patency of Lumen | Clause 4.13 of ISO 7864: 2016 |
| > ISO 80369-7:2016 | |
| Leakage by pressure decay | Clause 6.1.2 of ISO 80369-7: 2021 |
| Positive pressure liquid leakage | Clause 6.1.3 of ISO 80369-7: 2021 |
| Sub-atmospheric pressure air leakage | Clause 6.2 of ISO 80369-7: 2021 |
| Stress cracking | Clause 6.3 of ISO 80369-7: 2021 |
| Resistance to separation from axial load | Clause 6.4 of ISO 80369-7: 2021 |
| Resistance to separation from unscrewing | Clause 6.5 of ISO 80369-7: 2021 |
| Resistance to overriding | Clause 6.6 of ISO 80369-7: 2021 |
| > ISO 9626:2016 | |
|------------------------|-----------------------------|
| Stiffness | Clause 5.8 of ISO 9626:2016 |
| Resistance to breakage | Clause 5.9 of ISO 9626:2016 |
# VIII Conclusion
The Sterile hypodermic syringes for single use are substantially equivalent to its predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.