← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K211482

# Disposable Syringe with Needle (K211482)

_Shandong Qinkai Medical Industry Co., Ltd. · FMF · Jan 24, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K211482

## Device Facts

- **Applicant:** Shandong Qinkai Medical Industry Co., Ltd.
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** Jan 24, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Sterile single-use syringe with needle is intended to inject fluids into or withdraw fluids from the body.

## Device Story

Disposable manual syringe with needle; used for fluid injection or withdrawal. Components include polypropylene barrel, plunger, piston (polyisoprene), needle hub, and stainless steel (SUS304) needle. Available in various volumes (1ml-50ml) and needle gauges (18G-30G). Operated manually by clinicians; not for use with syringe pumps. Provided sterile via EO gas. Benefits include reliable fluid delivery/aspiration for clinical procedures.

## Clinical Evidence

Bench testing only. No clinical data. Testing included physical, mechanical, and chemical verification per ISO 7886-1, ISO 9626, ISO 7864, and ISO 80369-7. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogen, hemolysis). Sterile barrier integrity verified via ASTM F88, F1929, and F1886. Sterilization and shelf-life (3 years) validated per ISO 11135 and ISO 10993-7.

## Technological Characteristics

Materials: Polypropylene (barrel, plunger, hub, cap), Polyisoprene (piston), Stainless Steel SUS304 (needle). Manual operation. Luer slip/lock connectors. EO gas sterilization (SAL 10^-6). Complies with ISO 10993 (biocompatibility), ISO 7886-1 (syringe), ISO 7864/9626 (needle), and ISO 80369-7 (connectors).

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Sterile Single-use Syringe with Needle ([K163161](/device/K163161.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 24, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shandong Qinkai Medical Industry Co., Ltd. Alice Gong Manager South Section of Quancheng Road, Medical Equipment Industrial Park, Chengwu County Heze Citv. Shandong Province 274200 China

#### Re: K211482

Trade/Device Name: Disposable Syringe with Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: December 14, 2021 Received: December 27, 2021

#### Dear Alice Gong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211482

Device Name Disposable Syringe with Needle

Indications for Use (Describe)

Sterile single-use syringe with needle is intended to inject fluids into or withdraw fluids from the body.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K211482 510(K) SUMMARY

| Preparation Date:  | January 24, 2022                                                                                                                                                                          |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name:    | Shandong Qinkai Medical Industry Co., Ltd.,<br>South Section of Quancheng Road,<br>Medical Equipment Industrial Park,<br>Chengwu County, Heze City,<br>Shandong Province,<br>China 274200 |
| Contact Person:    | Mr. Ou Kai Jiang<br>General Manager                                                                                                                                                       |
| Telephone Number:  | (86) 0530-6112108                                                                                                                                                                         |
| Fax Number:        | (86) 0530-6112108                                                                                                                                                                         |
| E-mail Address:    | 490566042@qq.com                                                                                                                                                                          |
| Trade Name:        | Disposable Syringe with Needle                                                                                                                                                            |
| Regulation Name:   | Syringe, Piston                                                                                                                                                                           |
| Regulation Number: | 21 CFR 880.5860                                                                                                                                                                           |
| Product Code:      | FMF                                                                                                                                                                                       |
| Device Class:      | Class II                                                                                                                                                                                  |
| Regulation Name:   | Hypodermic Single Lumen Needle                                                                                                                                                            |
| Regulation Number: | 21 CFR 880.5570                                                                                                                                                                           |
| Product Code:      | FMI                                                                                                                                                                                       |
| Device Class:      | Class II                                                                                                                                                                                  |
| Predicate Device:  | K163161 Sterile Single-use Syringe with Needle                                                                                                                                            |

### Device Description

The subject device is for single use only, which is comprised of syringes and needles with various specifications. The syringe consists of barrel, plunger, piston, and the needle consist of needle hub, needle cap. The proposed devices shall be operated manually, and are not intended for use with syringe pumps. The proposed devices are available in a variety combination of needle size and syringe volume. The syringe sizes, needle gauges and lengths are provided in the following table.

| Syringe Size                                            | Needle Gauge                                                       | Needle Length                               |
|---------------------------------------------------------|--------------------------------------------------------------------|---------------------------------------------|
| Available in 1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml | Available in 18G, 20G, 21G, 22G, 23G, 25G, 26G, 27G, 28G, 29G, 30G | Available in 1/2", 5/8", 1", 1 1/4", 1 1/2" |

The subject device is provided sterile and is for single use only. The EO gas sterilization process is validated with a resulting sterility assurance level (SAL) of 106. It meets requirements of ISO 11135, sterilization of

{4}------------------------------------------------

health-care products for ethylene oxide - requirements for the development, validation and routine control of a sterilization process for medical devices.

Below is the list of different combinations:

| No. | Syringe Size | Needle Gauge & Length                          |
|-----|--------------|------------------------------------------------|
| 1   | 1 ml         | 30G*1/2", 29G*1/2", 28G*5/8"                   |
| 2   | 2 ml         | 27G*1/2", 26G*1/2", 25G*1"                     |
| 3   | 3 ml         | 25G*1", 23G*1 1/4", 22G*1 1/4"                 |
| 4   | 5 ml         | 23G*1 1/4", 22G*1 1/2", 21G*1 1/2"             |
| 5   | 10 ml        | 22G*1 1/2", 21G*1 1/2", 20G*1 1/2"             |
| 6   | 20 ml        | 22G*1 1/2", 21G*1 1/2", 20G*1 1/2", 18G*1 1/2" |
| 7   | 30 ml        | 21G*1 1/2", 20G*1 1/2", 18G*1 1/2"             |
| 8   | 50 ml        | 21G*1 1/2", 20G*1 1/2", 18G*1 1/2"             |

#### Indications for Use

| Characteristic                              | Predicate Device<br>Sterile Single-use Syringe with<br>Needle K163161                                               | Subject Device<br>Disposable Syringe with Needle<br>K211482                                                         |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Indications for Use                         | Sterile Single-use Syringe with<br>Needle is intended to inject fluids<br>into or withdraw fluids from the<br>body. | Sterile single-use syringe with<br>needle is intended to inject fluids<br>into or withdraw fluids from the<br>body. |
| Prescription Only<br>or<br>Over the Counter | Prescription Only                                                                                                   | Prescription Only                                                                                                   |

# Discussions of differences in Indications for Use statement

The indications for use statement for the subject device is identical to the predicate device.

#### Technological Characteristics

The table below includes a comparison of the technological characteristics between the new device and those of the predicate device:

| Technological<br>Characteristic | Predicate Device<br>Sterile Single-use Syringe<br>with Needle<br>K163161 | Subject Device<br>Disposable Syringe with<br>Needle<br>K211482 | Comments |
|---------------------------------|--------------------------------------------------------------------------|----------------------------------------------------------------|----------|
| Product Code                    | FMF and FMI                                                              | FMF and FMI                                                    | Same     |
| Regulation<br>Number            | 21 CFR 880.5860, 21 CFR<br>880.5570                                      | 21 CFR 880.5860, 21 CFR<br>880.5570                            | Same     |
| Class                           | II                                                                       | II                                                             | Same     |

{5}------------------------------------------------

| Configuration<br>and material | Barrel, Plunger, Needle hub<br>and Needle cap is made of<br>Polypropylene (PP).<br>Piston is made of<br>Polyisoprene.<br>Needle tube is made of<br>Stainless Steel, SUS304 | Barrel, Plunger, Needle hub<br>and Needle cap is made of<br>Polypropylene (PP).<br>Piston is made of<br>Polyisoprene.<br>Needle tube is made of<br>Stainless Steel, SUS304 | Same                 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Operation Mode                | For manual use only                                                                                                                                                        | For manual use only                                                                                                                                                        | Same                 |
| Syringe Volume                | 1ml, 3ml, 5ml, 10ml, 20ml,<br>60ml                                                                                                                                         | 1ml, 2ml, 3ml, 5ml, 10ml,<br>20ml, 30ml, 50ml                                                                                                                              | Similar<br>Comment 1 |
| Connector Type                | Luer Slip and Luer Lock                                                                                                                                                    | Luer Slip and Luer Lock                                                                                                                                                    | Same                 |
| Needle Gauge                  | 18G, 20G, 21G, 22G, 23G,<br>25G, 26G, 27G, 28G, 29G,<br>30G                                                                                                                | 18G, 20G, 21G, 22G, 23G,<br>25G, 26G, 27G, 28G, 29G,<br>30G                                                                                                                | Same                 |
| Needle Length                 | 1/2", 5/8", 1", 1 1/4", 1 1/2"                                                                                                                                             | 1/2", 5/8", 1", 1 1/4", 1 1/2"                                                                                                                                             | Same                 |
| Biocompatibility              | meet ISO 10993<br>requirements.                                                                                                                                            | meet ISO 10993 requirements.                                                                                                                                               | Same                 |
| Sterilization                 | EO Sterilization                                                                                                                                                           | EO Sterilization                                                                                                                                                           | Same                 |
| SAL                           | 10-6                                                                                                                                                                       | 10-6                                                                                                                                                                       | Same                 |
| Single Use                    | Single Use                                                                                                                                                                 | Single Use                                                                                                                                                                 | Same                 |
| Label/Labeling                | Complied with 21 CFR part<br>801                                                                                                                                           | Complied with 21 CFR part<br>801                                                                                                                                           | Same                 |

#### Discussions of differences in technological characteristics

## Comment 1

The subject device has more syringe volume specification than the predicated device, however the subject device has met the requirements of product performance standards and biocompatibility standards. The volume specifications are also within the overall range included in the predicate. This difference does not affect intended use, so it will not raise new questions of safety or effectiveness.

## Performance Testing

Bench tests were conducted to verify that proposed device meet all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that proposed device complies with the following standards:

| Test Item      | Standard                                                                                                                                                         |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Syringe        | ISO 7886-1: 2017 Sterile hypodermic syringes for single use - Part 1:<br>Syringes for manual use                                                                 |
| Needle         | ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical<br>devices - Requirements and test methods                                           |
| Needle         | ISO 7864: 2016 Sterile hypodermic needles for single use - Requirements and<br>test methods                                                                      |
| Luer connector | ISO 80369-7: 2016 Small-bore connectors for liquids and gases in healthcare<br>applications - Part 7: Connectors for intravascular or hypodermic<br>applications |

A. Physical, Mechanical and Chemical Tests performed on the proposed device:

{6}------------------------------------------------

B. Sterile Barrier Packaging Testing performed on the proposed device:

| Test Item                      | Standard               |
|--------------------------------|------------------------|
| Seal strength                  | ASTM F88/F88M-2015     |
| Detecting seal leaks           | ASTM F1929-2015        |
| Determining integrity of seals | ASTM F1886/F1886M-2016 |

- C. Sterilization and Shelf Life Testing performed on the proposed device:

| Test Item                | Standard                                                                                                                                                        |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EO residue               | ISO 10993-7:2008                                                                                                                                                |
| ECH residue              | ISO 10993-7:2008                                                                                                                                                |
| Bacteria Endotoxin Limit | USP 36-NF 31<85>                                                                                                                                                |
| Shelf Life Evaluation    | Physical, Mechanical, Chemical, Package and<br>Sterility Tests were performed on aging samples<br>to verify the shelf life of 3 years for the subject<br>device |

- D. The hub is made of pigmented material and the hub color for different needle gauge comply with ISO 6009:2016.
- E. Biocompatibility Testing is conducted according to ISO 10993-1:2018:

| Test Item                      | Standard          |
|--------------------------------|-------------------|
| Cytotoxicity Test              | ISO 10993-5:2009  |
| Skin Sensitization Test        | ISO 10993-10:2010 |
| Intracutaneous Reactivity Test | ISO 10993-10:2010 |
| Acute Systemic Toxicity Test   | ISO 10993-11:2017 |
| Pyrogen Test                   | ISO 10993-11:2017 |
| Hemolysis Test                 | ISO 10993-4:2017  |
| Particulate                    | USP <788>         |

#### Conclusions

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The subject Device, Disposable Syringe with Needle, K211482 is substantially equivalent to the predicate Device, Sterile Single-use Syringe with Needle, K163161.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K211482](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K211482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
