← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K210914

# Sterile Syringe for Single Use (K210914)

_Jiangsu Suyun Medical Material Co., Ltd. · FMF · Dec 17, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K210914

## Device Facts

- **Applicant:** Jiangsu Suyun Medical Material Co., Ltd.
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** Dec 17, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Sterile Syringe for Single Use is intended for use by health care professionals for general purpose aspiration and injection of fluids.

## Device Story

Manual, single-use piston syringe; available with or without needle. Components: polypropylene barrel, plunger, needle hub, protective cap; synthetic rubber piston; stainless steel (SUS304) needle tube. Used by healthcare professionals in clinical settings for fluid aspiration and injection. Device operates via manual plunger displacement. Output is the delivery or withdrawal of fluids. Benefits include sterile, single-use fluid management. Sterilized via Ethylene Oxide (SAL 10^-6).

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis) and performance testing (seal strength, particulate matter, connector standards, needle requirements) confirm compliance with ISO and ASTM standards.

## Technological Characteristics

Materials: Polypropylene (barrel, plunger, hub, cap), synthetic rubber (piston), stainless steel SUS304 (needle). Manual operation. Luer lock or luer slip connectors; central or eccentric locations. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 7886-1, ISO 7864, ISO 9626, ISO 6009, ISO 80369-7/20, ISO 10993 series, ASTM F756, ASTM F88, ASTM F1929, USP <788>.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- BD Emerald™, Single Use, Hypodermic Syringe ([K113241](/device/K113241.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name in a sans-serif font. The logo is simple and professional, conveying the agency's authority and mission.

December 17, 2021

Jiangsu Suyun Medical Material Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai, 200120 China

Re: K210914

Trade/Device Name: Sterile Syringe for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: November 10, 2021 Received: November 17, 2021

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K210914

Device Name Sterile Syringe for Single Use

Indications for Use (Describe)

The Sterile Syringe for Single Use is intended for use by health care professionals for general purpose aspiration and injection of fluids.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   

 Food and Drug Administration
   

 Office of Chief Information Officer
   

 Paperwork Reduction Act (PRA) Staff
   

 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

# Tab #6 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section

807.92. The assigned 510(k) Number: _ _ _ K210914_

- Date of Preparation: 9/3/2021 1.
- 2. Sponsor Identification

Jiangsu Suyun Medical Materials Co., Ltd. No.18 Jinqiao Road, Dapu Industrial Park, Lianyungang City, Jiangsu, 222000, China

Establishment Registration Number: 3003717263

Contact Person: Guangning Xu Position: Management Representative Tel: +86-518-85462073 Fax: +86-518-85608151 Email: quality@suyunmedical.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)

## Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

### 510(k) Summary

#### 4. Identification of Proposed Device

Trade Name: Sterile Syringe for Single Use Common Name: Piston Syringe

### Regulatory Information

Classification Name: Syringe, Piston Classification: II; Product Code: FMF; Regulation Number: 21CFR 880.5860 Review Panel: General Hospital;

Indication for Use Statement:

The Sterile Syringe for Single Use is intended for use by health care professionals for general purpose aspiration and injection of fluids.

Device Description

### Syringe with needle

The Syringe with needle is intended for manual and single use only. It consists of six components 1) Barrel (luer lock or luer slip) 2) Plunger 3) Piston 4) Needle Tube 5) Protective Cap and 6) Needle Hub.

The proposed device is available in various combination of syringe volume, connector type (luer lock or luer slip), connector location (central type or eccentric type) and needle size.

## Syringe without needle

The Syringe without needle is intended for manual and single use only. It consists of three components 1) Barrel (luer lock or luer slip) 2) Plunger and 3) Piston.

The proposed syringe is available in various combination of syringe volume, connector type (luer lock or luer slip), and connector location (central type or eccentric type).

The proposed device is sterilized by Ethylene Oxide to achieve a SAL of 106 and supplied in sterility maintenance package which could maintain the sterility of the device during the shelf life of 5 years.

- ડ. Identification of Predicate Device
510(k) Number: K113241 Product Name: BD Emerald™, Single Use, Hypodermic Syringe

{5}------------------------------------------------

#### 6. Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications and is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > ISO 10993-5:2009 Biological evaluation of medical Devices-Part 5: Tests for in Vitro Cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization
- > ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity
- > ISO 10993-4:2017 Biological Evaluation of Medical Devices-Part 4: Selection of Tests for Interactions with Blood
- > ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
- > USP<788> Particulate Matter for Injections (Method 1 Light Obscuration Particle Count Test)
- > ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- > ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods
- > ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices
- A ISO 6009:2016 Hypodermic needles for single use - Colour coding for identification
- > ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-Part 7: Connectors for intravascular or hypodermic applications
- > ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications-Part 20: Common test methods
- > ISO 7886-1:2017 Sterile hypodermic syringes for single use- Part 1: Syringes for manual use

#### 7. Clinical Test Conclusion

No clinical study is included in this submission.

{6}------------------------------------------------

#### Substantially Equivalent (SE) Comparison 8.

| ITEM                        | Proposed Device                                                                                                                                      | Predicate Device<br>K113241                                                                                                                                          | Remark    |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product                     | Sterile Syringe for Single Use                                                                                                                       | BD EmeraldTM, Single Use, Hypodermic Syringe                                                                                                                         | /         |
| Product Code                | FMF                                                                                                                                                  | FMF                                                                                                                                                                  | Same      |
| Regulation<br>Number        | 21 CFR 880.5860                                                                                                                                      | 21 CFR 880.5860                                                                                                                                                      | Same      |
| Class                       | Class II                                                                                                                                             | Class II                                                                                                                                                             | Same      |
| Indication for<br>Use       | The Sterile Syringe for Single Use<br>is intended for use by health care<br>professionals for general purpose<br>aspiration and injection of fluids. | The BD EmeraldTM Single Use,<br>Hypodermic Syringe is intended<br>for use by health care<br>professionals for general purpose<br>aspiration and injection of fluids. | Same      |
| Components and<br>materials | Barrel                                                                                                                                               | Polypropylene (PP)                                                                                                                                                   |           |
|                             | Plunger                                                                                                                                              | Polypropylene (PP)                                                                                                                                                   |           |
|                             | Piston                                                                                                                                               | Synthetic rubber                                                                                                                                                     |           |
|                             | Needle hub                                                                                                                                           | Polypropylene (PP)                                                                                                                                                   | Different |
|                             | Protective<br>cap                                                                                                                                    | Polypropylene (PP)                                                                                                                                                   |           |
|                             | Needle tube                                                                                                                                          | Stainless Steel<br>SUS304                                                                                                                                            |           |
|                             | Barrel                                                                                                                                               | Polypropylene (PP)                                                                                                                                                   |           |
|                             | plunger                                                                                                                                              | Polypropylene (PP)                                                                                                                                                   |           |
|                             | Piston                                                                                                                                               | Unknown                                                                                                                                                              |           |
|                             | Needle hub                                                                                                                                           | Polypropylene (PP)                                                                                                                                                   |           |
|                             | Protective<br>cap                                                                                                                                    | Unknown                                                                                                                                                              |           |
|                             | Needle tube                                                                                                                                          | Stainless Steel                                                                                                                                                      |           |
| Operation Mode              | For manual use only                                                                                                                                  | For manual use only                                                                                                                                                  | Same      |
| Single Use                  | Single Use                                                                                                                                           | Single Use                                                                                                                                                           | Same      |
| Label/Labeling              | Complied with 21 CFR part 801                                                                                                                        | Complied with 21 CFR part 801                                                                                                                                        | Same      |

# Table 1 General Comparison of the Sterile Syringe for Single Use

Different - Components and materials

The components of the proposed device are same with the predicate device but the materials are not exactly the same. However, the proposed device had been tested for biocompatibility and the result does not show any adverse effect. Therefore, the difference on the proposed device doesn't raise new questions of safety and effectiveness.

{7}------------------------------------------------

| ITEM               |                     | Proposed Device                        | Predicate Device<br>K113241                  | Remark    |
|--------------------|---------------------|----------------------------------------|----------------------------------------------|-----------|
| Product            |                     | Sterile Syringe for Single Use         | BD EmeraldTM, Single Use, Hypodermic Syringe | /         |
|                    | Volume              | 1ml, 2ml, 5ml, 10ml, 20ml, 30ml, 50ml  | 2ml, 3ml, 5ml, 10ml                          | Different |
| Syringe            | Connector Type      | Luer Lock/ Luer Slip                   | Luer Lock/ Luer Slip                         |           |
|                    | Connector location  | Central/Eccentric                      | Central                                      |           |
| Needle             | Gauge               | 22G                                    | 21G                                          | Different |
|                    | Length              | 30mm, 32mm                             | 40mm                                         |           |
|                    | Syringe Performance | Complied with<br>ISO 7886-1            | Complied with<br>ISO 7886-1                  | Same      |
| Needle Performance |                     | Complied with<br>ISO 7864,<br>ISO 9626 | Complied with<br>ISO 7864,<br>ISO 9626       | Same      |

Table 2 Performance Comparison of the Sterile Syringe for Single Use

#### Different - Syringe

The volume and connector location of proposed device is different from the predicate device. For the volume, the proposed device has more volumes, 1ml, 30ml and 50ml. For the connector location, the proposed device has more connector locations. In addition, all specifications of proposed device had been tested for performance and the test results could meet the requirements according to ISO 7886-1:2017. The differences do not raise new questions of safety and effectiveness.

### Different- Needle Gauge and Length

The needle gauge and length of the proposed device is different from the predicate device. However, the needle with a diameter of 0.7mm (21G) and a length of 30mm or 32mm are covered by the performance standards. The needles have been tested and the test results comply with related standards requirements. The differences do not raise new questions of safety and effectiveness.

{8}------------------------------------------------

| ITEM                     | Proposed Device                | Predicate Device<br>K113241                    | Remark    |
|--------------------------|--------------------------------|------------------------------------------------|-----------|
| Product                  | Sterile Syringe for Single Use | BD Emerald™, Single Use, Hypodermic<br>Syringe | /         |
| Patient-contact Material |                                |                                                |           |
| Barrel                   | Polypropylene (PP)             | Barrel<br>Polypropylene (PP)                   | /         |
| Plunger                  | Polypropylene (PP)             | Plunger<br>Polypropylene (PP)                  | /         |
| Piston                   | Synthetic rubber               | Piston<br>Unknown                              | Different |
| Needle tube              | Stainless Steel SUS304         | Needle tube<br>Stainless Steel                 | /         |
| Needle hub               | Polypropylene (PP)             | Needle hub<br>Polypropylene (PP)               | /         |
| Biocompatibility         |                                |                                                |           |
| Cytotoxicity             | No cytotoxicity                | Conforms to ISO 10993 series standards         | Different |
| Irritation               | No intracutaneous reactivity   |                                                |           |
| Sensitization            | No skin sensitization          |                                                |           |
| Systemic Toxicity        | No systemic toxicity           | /                                              |           |
| Hemolysis                | No Hemolysis                   | /                                              |           |
| Pyrogen                  | No Pyrogen                     | /                                              | /         |
| Sterilization            | EO Sterilization               | EO Sterilization                               | Same      |
| SAL                      | 10-6                           | 10-6                                           | Same      |
| Endotoxin Limit          | 20 EU per device               | 20 EU per device                               | Same      |

Table 3 Safety Comparison of the Sterile Syringe for Single Use

Different- Patient-contact material

The patient-contact material of piston of the predicate device is unknown. However, the biocompatibility tests were conducted on the material consisted of the proposed device and the test result shows that the material are biocompatible.

- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K210914](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K210914)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
