K210464 · Anhui Tiankang Medical Technology Co., Ltd. · FMF · Aug 24, 2021 · General Hospital
Device Facts
Record ID
K210464
Device Name
Auto Disable Syringe
Applicant
Anhui Tiankang Medical Technology Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Aug 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AUTO DISABLE SYRINGE is intended for use in the suction of vaccine for medical purposes. Additionally , after injection to the body, the plunger can be automatically locked by the triggered mechanism to prevent the re-use of this syringe.
Device Story
Auto Disable Syringe is a sterile, single-use hypodermic syringe with an integrated fixed needle. Device consists of a calibrated polypropylene barrel, movable plunger, polyisoprene rubber stopper, and steel clip mechanism. Input is manual aspiration of vaccine; device transforms input via plunger movement to deliver fixed dose; output is vaccine injection. Upon completion of injection, internal steel clip triggers to lock plunger, preventing reuse. Used in clinical settings by healthcare professionals for immunization. Benefits include prevention of needle reuse and associated blood-borne pathogen transmission. Device is provided sterile via ethylene oxide.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including functional performance (ISO 7886-1, ISO 7886-3, ISO 7864, ISO 9626), biocompatibility (ISO 10993-1), and sterility/shelf-life validation (ISO 11135, ASTM D4169, ASTM F1980).
Technological Characteristics
Materials: Polypropylene (barrel/plunger), polyisoprene rubber (piston), stainless steel (needle). Energy: Manual. Dimensions: 0.05ml-1.0ml capacity; 23G-30G needles. Sterilization: Ethylene Oxide (10^-6 SAL). Standards: ISO 7886-1/3, ISO 7864, ISO 9626, ISO 10993, ISO 11135, ASTM F88, ASTM F1886, ASTM F1929, ASTM F1980.
Indications for Use
Indicated for vaccine administration in patients requiring hypodermic injection. Intended for single-use only; features automatic locking mechanism to prevent reuse.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
BD SoloShot Mini Syringe/ BD Auto Disable Syringe (K201234)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 24, 2021
Anhui Tiankang Medical Technology Co.,Ltd. Bai Baodong General Manager No.228 Weiyi Road, Economic Development Zone Tianchang, Anhui 239300 China
Re: K210464
Trade/Device Name: Auto Disable Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: July 23, 2021 Received: July 26, 2021
Dear Bai Baodong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT. Alan M. Stevens Assistant Director Injection Devices Team DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210464
Device Name AUTO DISABLE SYRINGE
#### Indications for Use (Describe)
The AUTO DISABLE SYRINGE is intended for use in the suction of vaccine for medical purposes. Additionally , after injection to the body, the plunger can be automatically locked by the triggered mechanism to prevent the re-use of this syringe.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K210464 510(K) Summary
#### Date prepared: August 24, 2021
#### Submitter Name and Address:
| Submitter Name: | Anhui Tiankang Medical Technology Co., Ltd. |
|-----------------|-------------------------------------------------------------------------------|
| Address: | No.228 Weiyi Road, Economic Development Zone, Tianchang City.<br>Anhui, China |
| Contact Person: | Bai Baodong, General Manager |
| TEL: | +86-550-7309187 |
| FAX: | +86-550-7308901 |
| E-mail: | tkquality@126.com |
### Submission Devices Information:
Trade/Proprietary Name: AUTO DISABLE SYRINGE Classification name: Syringe, piston Class: II. Product codes: FMF Regulation number: 21CFR880.5860,
Submission Type: 510(K)
#### Predicate Device Information:
Company Name: Becton, Dickinson, and Company Device name: BD SoloShot Mini Syringe/ BD Auto Disable Syringe 510(K) Number: K201234
#### Device Description
1) Indication for use:
The AUTO DISABLE SYRINGE is intended for use in the suction and injection of vaccine for medical purposes. Additionally, after injection to the plunger can be automatically locked by the triggered mechanism to prevent the re-use of this syringe.
2) Design of syringe:
The AUTO DISABLE SYRINGE is designed for the hypodermic injection of the intended micro
{5}------------------------------------------------
scale solution. It has the automatic function in the structure of the syringe which after injection of an intended fixed dose locks the syringe to prevent re-use of the syringe. It can be widely used for injection of vaccine for the immunization.
3) Structure of syringe (with needle):
The syringe consists of a calibrated hollow barrel, a movable plunger, a rubber stopper (piston) assembled at end of the plunger, and a steel clip installed between the barrel and the plunger that functions to prevent re-use of the syringe.
At the end of the barrel, a fixed an unmovable needle is installed, and a needle cap covers the needle. The zero line and nominal capacity line are printed outside of barrel. The main materials are PP, latex free (polyisoprene rubber) and stainless steel.
The AUTO DISABLE SYRNGE is for Prescription (Rx) use only
### Product models:
| Description | Needle Length |
|--------------------------|----------------------------|
| 1cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 1cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 1cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 1cc(ml), with needle:26G | 3/8" and 1/2" |
| 1cc(ml), with needle:27G | 3/8" and 1/2" |
| 1cc(ml), with needle:28G | 3/8" and 1/2" |
| 1cc(ml), with needle:29G | 3/8" and 1/2" |
| 1cc(ml), with needle:30G | 3/8" and 1/2" |
| Description | Needle Length |
|----------------------------|----------------------------|
| 0.5cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.5cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.5cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.5cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.5cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.5cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.5cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.5cc(ml), with needle:30G | 3/8" and 1/2" |
| Description | Needle Length |
|----------------------------|----------------------------|
| 0.4cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.4cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.4cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.4cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.4cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.4cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.4cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.4cc(ml), with needle:30G | 3/8" and 1/2" |
{6}------------------------------------------------
| Description | Needle Length |
|----------------------------|----------------------------|
| 0.3cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.3cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.3cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.3cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.3cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.3cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.3cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.3cc(ml), with needle:30G | 3/8" and 1/2" |
| Description | Needle Length |
|-----------------------------|----------------------------|
| 0.25cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.25cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.25cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.25cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.25cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.25cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.25cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.25cc(ml), with needle:30G | 3/8" and 1/2" |
| Description | Needle Length |
|----------------------------|----------------------------|
| 0.2cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.2cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.2cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.2cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.2cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.2cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.2cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.2cc(ml), with needle:30G | 3/8" and 1/2" |
| Description | Needle Length |
|----------------------------|----------------------------|
| 0.1cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.1cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.1cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.1cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.1cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.1cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.1cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.1cc(ml), with needle:30G | 3/8" and 1/2" |
| Description | Needle Length |
|-----------------------------|----------------------------|
| 0.05cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.05cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.05cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" |
| 0.05cc(ml), with needle:26G | 3/8" and 1/2" |
| 0.05cc(ml), with needle:27G | 3/8" and 1/2" |
| 0.05cc(ml), with needle:28G | 3/8" and 1/2" |
| 0.05cc(ml), with needle:29G | 3/8" and 1/2" |
| 0.05cc(ml), with needle:30G | 3/8" and 1/2" |
{7}------------------------------------------------
## Comparison of technological characteristics with the predicate:
| Elements of<br>comparison | Subject device | Predicate device: | Comments |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------|
| AUTO DISABLE<br>SYRINGE | BD SoloShot Mini<br>Syringe/ BD Auto Disable<br>Syringe (K201234) | | |
| Indications for<br>Use | The AUTO DISABLE<br>SYRINGE is intended<br>for use in the suction<br>and injection of vaccine<br>for medical purposes.<br>Additionally, after<br>injection to the body,<br>the plunger can be<br>automatically locked by<br>the triggered<br>mechanism to prevent<br>the re-use of this<br>syringe. | BD SoloShot Mini<br>Syringe/ BD Auto<br>Disable Syringe is<br>intended for aspiration<br>and injection of fluids. | Comment 1. |
| Principle of<br>operation | Normal | Normal | Identical |
| Needle length | 3/8", 1/2", 1" | 3/8", 5/8", 3/4", 1" | Comment 2. |
| Needle gauge | 23G, 24G, 25G, 26G,<br>27G, 28G, 29G, 30G | 23G, 24G, 25G, 27G | Comment 3 |
| Syringe<br>capacity | 0.05ml, 0.10ml, 0.20ml,<br>0.25ml, 0.30ml, 0.40ml,<br>0.50ml, 1.0ml | 0.05ml, 0.10ml, 0.50ml, | Comment 4 |
| Reuse<br>Prevention<br>(Safety)<br>Feature | Auto-disabled, prevents<br>syringe re-use | Auto-disabled,<br>prevents syringe<br>re-use | Identical |
| Lubricant for<br>barrel | Silicone oil | Silicone oil | Identical |
| Barrel<br>transparency | Transparent and clear | Transparent and clear | Identical |
| Gradations<br>legibility | Legible | Legible | Identical |
| Material | Barrel: PP | Barrel: PP | Comment 5 |
| | Plunger: PP | Plunger: PP | |
| | Piston: Latex Free | Piston: rubber | |
| | (Polyisoprene Rubber) | Needle: stainless steel | |
| | Needle: stainless steel | | |
| Performance | Conform to ISO7886-3<br>and ISO7886-1,<br>ISO9629, ISO7864 | Conform to ISO7886-3<br>and ISO7886-1,<br>ISO9629, ISO7864 | Identical |
| Biocompatibility | Conform to ISO10993 | Conform to ISO10993 | Identical |
| Labeling | Meet the requirements<br>of 21CFR Part 801 | Meet the requirements<br>of 21CFR Part 801 | Identical |
| Sterility | EO sterilization 10-6 | EO sterilization 10-6 | Identical |
{8}------------------------------------------------
Discussions of differences in technological characteristics
## Comment 1:
The indications for use are similar for both devices (to the extent that both devices are intended for aspiration and injection of fluids for medical purposes). The predicate device defines that the device is for general use. The subject device is specifically used for vaccine administration, which is included in the scope of general use. In addition, the subject device includes a statement regarding the locking mechanism to prevent re-use. This mechanism is included in the predicate design, but is not included in the indications for use statement. The differences to the indications for use statement between the predicate and the subject device are acceptable.
## Comment 2, 3 and 4:
The specifications of the proposed device and predicate device are different; however, both of them comply with the same recognized standards. The different sizes of needle are addressed through performance testing.
## Comment 5:
The piston used the new technology material to replace the traditional rubber material for less toxicity (Polyisoprene Rubber). The safety testing report of biocompatibility and function performance testing report demonstrates that the change does not negatively impact safety and performance.
{9}------------------------------------------------
## Non-Clinical Testing
## Functional Performance Testing
The AUTO DISABLE SYRINGE described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards:
ISO 7864:2016, Sterile hypodermic needles for single use - Requirements and test methods
- . Fragmentation test
- . Determination of flow rate
- . Lubricant: Penetration force and drag force
- . Bonding strength
ISO 7886-3, sterile hypodermic syringes for single use - Part 3: Auto-disabled syringes for fixed-dose immunization and ISO7886-1, sterile hypodermic syringes for single use - Part 1: Syringes for manual use
- . Limits for acidity and alkalinity
- . Limits for extractable metals
- . Lubricant: Penetration force and drag force
- . Tolerance on nominal capacity
- Graduated scale ●
- . Barrel- dimensions and flanges
- Plunger stopper and assembly - fitness of stopper /plunger in barrel
- Integrated needle
- . Sharps protection features
- . Dead space
- Freedom from air and liquid leakage ●
- Auto-disable syringe feature .
ISO9626, Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
- Limits for acidity and alkalinity
- designated metric size ●
- Stiffness .
- Resistance to breakage ●
- Resistance to corrosion .
## Sterility, Shipping and Shelf-Life
- . EO sterilization validation and package validation conducted in accordance with ISO11135:2014, ISO11607-1:2019, ISO11607:2019
{10}------------------------------------------------
- . Sterilant residuals were evaluated per ISO10993-7:2008. Average daily dose to patient of EO less than 4mg/device and ECH less than 9mg/device
- . Package integrity testing, after environmental conditioning and simulated transportation in accordance with ASTM D4169-16, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
- . Sterile Barrier Packaging Testing performed on the proposed device:
- o Seal strength ASTM F88/F88-15
- o Visual Inspection ASTM F 1886-16
- o Dye penetration ASTM F1929-15
- . Shelf life of 5 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
## Biocompatibility Testing (ISO10993-1:2018)
In accordance with ISO 10993-1, the syringe is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 hours). The following testing was conducted:
- . Cytotoxicity (per ISO10993-5:2009)
- . Sensitization (per ISO10993-10:2014)
- Irritation (per ISO10993-10:2017) .
- Acute Systemic Toxicity (per ISO10993-11:2017) .
- Pyrogenicity (per ISO10993-4:2017) ●
- Hemocompatibility (per ISO10993-4:2017) .
- . Particulate matter ( per USP<788>)
## Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The AUTO DISABLE SYRINGE is substantially equivalent to BD SoloShot Mini Syringe/ BD Auto Disable Syringe with respect to the indications for use and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.