K190054 · Becton, Dickinson and Company · FMF · Oct 21, 2019 · General Hospital
Device Facts
Record ID
K190054
Device Name
BD Insulin Syringe
Applicant
Becton, Dickinson and Company
Product Code
FMF · General Hospital
Decision Date
Oct 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Becton Dickinson insulin syringes are intended for subcutaneous injection of U-100 insulins
Device Story
BD Insulin Syringe is a sterile, single-use, non-pyrogenic piston syringe designed for subcutaneous injection of U-100 insulin. Device consists of graduated barrel, plunger rod, and permanently attached needle/hub assembly. Available in 0.3mL, 0.5mL, and 1mL capacities with various needle gauges (27G-31G) and lengths (6mm-16mm). Operated manually by patient or healthcare provider to draw and inject insulin. Output is the delivery of a precise insulin dose. Benefits include accurate dosing and sterile delivery of medication. Used in home or clinical settings.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench functional performance testing (ISO 8537, ISO 9626) and biocompatibility testing (ISO 10993). Bench testing included air/liquid leakage, pull forces, needle break-off force, penetration force, scale permanency, filling/injection force, volumetric accuracy, dead space, and corrosion resistance. Biocompatibility testing included cytotoxicity, sensitization, reactivity, irritation, toxicity, pyrogenicity, and hemocompatibility.
Indicated for subcutaneous injection of U-100 insulin in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
BD Ultra-Fine™ Insulin Syringe, BD Ultra-Fine™ II Insulin Syringe, BD Insulin Syringes with the BD Ultra-Fine™ needle and BD Insulin Syringes with Micro-Fine™ IV Needle (K170386)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 21, 2019 Becton Dickinson and Company Meriam Youssef Senior Manager Regulatory Affairs 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K190054
Trade/Device Name: BD Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: September 19, 2019 Received: September 23, 2019
Dear Meriam Youssef:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190054
Device Name BD Insulin Syringe
| Indications for Use (Describe) | Becton Dickinson insulin syringes are intended for subcutaneous injection of U-100 insulins |
|--------------------------------|---------------------------------------------------------------------------------------------|
|--------------------------------|---------------------------------------------------------------------------------------------|
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K190054 510(k) Summary
| Submitted By: | Meriam Youssef<br>Sr. Manager Regulatory Affairs, BD Medical- Diabetes Care<br>1 Becton Drive<br>Franklin Lakes, NJ 07417<br>Tel: 201 847 6557<br>Fax: 201 847 5307 | | | |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|--|--|
| Date Prepared: | September 19, 2019 | | | |
| Device Name: | Trade Name: | BD Insulin Syringe™ | | |
| | Common Name: | Syringe, Piston | | |
| | Regulation Name: | Piston syringe | | |
| | Regulation: | 21 CFR 880.5860 | | |
| | Product Code: | FMF | | |
#### Legally marketed predicate device to which substantial equivalence is being claimed:
- BD Ultra-Fine™ Insulin Syringe, BD Ultra-Fine™ II Insulin Syringe, BD Insulin Syringes with the . BD Ultra-Fine™ needle and BD Insulin Syringes with Micro-Fine™ IV Needle (K170386)
#### Device Description:
The subject BD Insulin Syringes are a plastic syringe designed for subcutaneous injection of a desired dose of U 100 insulin. The BD Insulin Syringe consists of a graduated barrel, plunger rod and needle/hub assembly. The BD Insulin Syringes are sterile, single use, and non-pyrogenic. These devices operate on the principles of a piston syringe. The BD Insulin Syringe with 6mm length cannula have a sub-brand name of BD Veo™ Insulin Syringe.
| Syringe size (mL) | Bevel | Cannula length (mm) | Gauge (G) |
|-------------------|-------|---------------------|-----------|
| 0.3mL | 3 | 12.7mm | 29G |
| 0.3mL | 3 | 12.7mm | 30G |
| 0.3mL | 3 | 8mm | 29G |
| 0.3mL | 3 | 8mm | 31G |
| 0.3mL | 3 | 6mm | 31G |
| 0.5mL | 3 | 12.7mm | 28G |
| 0.5mL | 3 | 12.7mm | 29G |
| 0.5mL | 3 | 12.7mm | 30G |
| 0.5mL | 3 | 8mm | 31G |
| 0.5mL | 3 | 6mm | 31G |
| 1mL | 3 | 16mm | 27G |
| 1mL | 3 | 12.7mm | 28G |
| 1mL | 3 | 12.7mm | 29G |
| 1mL | 3 | 12.7mm | 30G |
| 1mL | 3 | 8mm | 30G |
| 1mL | 3 | 8mm | 31G |
| 1mL | 3 | 6mm | 31G |
| 1mL | 3 | 16mm | 27G |
The BD Insulin Syringes are offered in the following sizes:
{5}------------------------------------------------
# Indications for Use:
Becton Dickinson insulin syringes are intended for subcutaneous injection of U-100 insulins.
Differences in Indications for use: The indications for use are the same as the predicate.
# Substantial Equivalence Discussion:
| | | Predicate Device | Comparison |
|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| General Information<br>Feature | Subject Device:<br>BD Insulin Syringe | BD Ultra-Fine™ Insulin<br>Syringe, BD Ultra-Fine™<br>II Insulin Syringe, BD<br>Insulin Syringes with the<br>BD Ultra-Fine™ needle<br>and BD Insulin Syringes<br>with Micro-Fine™ IV<br>Needle | |
| 510(k) Number | K190054 | K170386 | N/A |
| Device Classification | Syringe Piston | Same | Same |
| Regulation Number | 880.586 | Same | Same |
| Product Code | FMF | Same | Same |
| Indications of Use | Becton Dickinson insulin<br>syringes are intended for<br>subcutaneous injection of<br>U 100 insulins | Becton Dickinson insulin<br>syringes are intended for<br>subcutaneous injection of<br>U-100 insulins. | Same |
| Specific Drug Use | U-100 Insulin | Same | Same |
| Capacity | 1mL,<br>0.5mL,<br>0.3mL | 1mL | The subject device<br>includes a 0.5mL and<br>0.3ml model. |
| Bevel | 3 | 3 | Dimensional<br>modifications to<br>primary bevel of<br>cannula |
| Cannula Gauge Size(s) | 27G, 28G, 29G, 30G, 31G | 27G, 28G, 29G, 30G, 31G | Same |
| Nozzle Tip Type | Snap fit needle hub for<br>0.3mL and 0.5mL designs<br>for permanently<br>attached needles<br>Integral barrel for 1mL<br>design for<br>permanently attached | Integral barrel for 1mL<br>design for<br>permanently attached<br>needles | Introduction of one glue<br>well design for 1mL<br>barrel configuration |
| Scale Markings | needles<br>1 unit increments and ½<br>unit increments (0.3mL)<br>1 unit increments (0.5mL)<br>2 units increments (1mL) | 2 units increments<br>(1mL) | Addition of ½ unit scale<br>mark increments |
| General Information<br>Feature | Subject Device: | Predicate Device | Comparison |
| General Information<br>Feature | Subject Device:<br>BD Insulin Syringe | BD Ultra-Fine™ Insulin<br>Syringe, BD Ultra-Fine™<br>II Insulin Syringe, BD<br>Insulin Syringes with the<br>BD Ultra-Fine™ needle<br>and BD Insulin Syringes<br>with Micro-Fine™ IV<br>Needle | |
| Packaging Configuration | Polybag for self-contained<br>syringe and blister pack<br>(multiple blister packs per<br>perforated sheet) offerings | Polybag for self-contained<br>syringe and blister pack<br>(individual blister pack)<br>offerings | Addition of retention<br>ring to 1mL<br>self-contained insulin<br>syringe<br>Multiple blister packs<br>per perforated sheet |
| Cannula Length Size(s) | 6mm, 8mm, 12.7mm,<br>16mm | 6mm, 8mm, 12.7mm,<br>16mm | Same |
| Cannula (Needle) material | 304 Stainless Steel | 304 Stainless Steel | Same |
| Cannula (Needle) shield<br>material and Cap (for<br>plunger rod) material | Polyethylene | Polyethylene | Modification to the co-<br>monomer of<br>polyethylene material |
| Cannula cover color | Orange | Orange | Same |
| Lubricant | Medical Grade Silicone | Medical Grade Silicone | Modification to<br>lubricant formulation<br>carrier solvent |
| Cannula bonding<br>adhesive (UV cured) | UV Cured Adhesive | UV Cured Adhesive | New 0.3mL & 0.5mL<br>UV Cured Adhesive |
| Scale marking ink | Solvent based Markem<br>black colored ink | Solvent based Markem<br>black colored ink | Same |
| Plunger rod material | Polystyrene | Polystyrene | Remove colorant<br>additive from the<br>plunger rod component |
| Plunger tip (stopper)<br>material | Rubber stopper | Rubber stopper | Same |
| Barrel and Hub Material | Polypropylene | Polypropylene | New material<br>formulation of<br>polypropylene |
| Single Use Only | YES | YES | Same |
| Non-Pyrogenic | YES | YES | Same |
| Sterile (10-6) | YES | YES | Same |
{6}------------------------------------------------
{7}------------------------------------------------
## Discussion:
Apart from the above mentioned modifications to the design and materials, there are no technological differences between the subject and the predicate devices. The design modifications were assessed through performance testing per ISO 8537 and ISO 9626. The material changes were evaluated per ISO 10993-1.
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness.
## Non-Clinical Testing:
Following BD's Quality System processes, required testing was conducted to validate the cumulative modifications made to the subject devices.
Substantial Equivalence is being supported with full performance testing the current BD Insulin Syringe devices currently marketed.
## Bench functional performance
- Testing was conducted according to ISO 8537:2016 Sterile single-use syringes, with or without needle, ● for insulin, and ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices. This testing included Air Leakage, Cannula Pull Force, Hub Pull Force, Hub Pull Force, Liquid Leakage, Needle Break-Off Force, Penetration Force, Scale Print Permanency, Shield Pull Force, Syringe Filling Force, Syringe Injection Force (Force to Expel Water), Volumetric Accuracy, Dead Space and Corrosion Resistance.
#### Biocompatibility testing
- Testing was conducted according to relevant sections of ISO 10993:2019 Biological evaluation of ● medical devices. This testing included Cytotoxicity, Sensitization (GP Max), Sensitization, Intracutaneous Reactivity, Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Subacute/Subchronic Toxicity, Hemocompatibility, and Genotoxicity (Bacterial and Mammalian). The medical device is classified as: externally communicating, blood path indirect, with prolonged contact duration.
## Additional Testing:
- ASTM F2148: 2013 Standard Practice for Evaluation of Delayed Contact Hypersensitivity Using the Murine Local Lymph Node Assay (LLNA), ISO 11137-1:2006/Amd 2013(E) Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. ISO 11137-2:2013(E) Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose, ISO 11137-3: 2017(E) Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine control, and ISO 11737-1:2018(E) Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products .
Results of testing demonstrated the BD Insulin Syringes met requirements for its intended use and is as safe and effective as its predicate device.
## Clinical Testing:
Not Applicable
## Conclusion:
The modifications to the design, dimensions, and materials of the subject device met the standards. The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness.
The BD Insulin Syringes are substantially equivalent to the BD Ultra-Fine™ Insulin Syringe, BD Ultra-Fine™ II Insulin Syringe, BD Insulin Syringes with the BD Ultra-Fine™ needle and BD Insulin Syringes with Micro-Fine™ IV Needle cleared under K170386 with respect to the indications for use, target populations, treatment method, use environment and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.