K162180 · Berpu Medical Technology Co., Ltd. · FMF · Dec 29, 2016 · General Hospital
Device Facts
Record ID
K162180
Device Name
Disposable Insulin Syringe
Applicant
Berpu Medical Technology Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Dec 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The disposable insulin syringe is intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of insulin into parts of the body below the surface skin.
Device Story
Disposable insulin syringe with fixed needle; used for manual aspiration and subcutaneous injection of U-100 insulin. Device consists of polypropylene barrel, plunger, and protective caps, polyisoprene rubber piston, and stainless steel (SUS304) needle. Available in 0.3ml, 0.5ml, and 1ml volumes with various needle gauges (27G-31G) and lengths (5mm-12mm). Operated manually by clinician or patient. Pushing plunger rod generates force to inject insulin into subcutaneous tissue. Provided sterile, single-use. Benefits include accurate insulin delivery for diabetes management.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via physical, mechanical, and chemical testing per ISO 9626, ISO 7864, and ISO 8537. Biocompatibility confirmed per ISO 10993. Sterilization validated per ISO 11135.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, caps), Polyisoprene (piston), Stainless Steel SUS304 (needle). Dimensions: 0.3-1.0ml volume; 27G-31G needles; 5-12mm lengths. Principle: Manual piston-driven injection. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 9626, ISO 7864, ISO 8537, ISO 10993, ASTM F88, ASTM F1140, ASTM F1929.
Indications for Use
Indicated for patients requiring manual aspiration and subcutaneous injection of U-100 insulin.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile Insulin Syringe for single use with fixed needle (K072739)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is depicted in a simple, black line drawing, and the text is also in black, creating a clean and professional look.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 29, 2016
Berpu Medical Technology Co., Ltd % Diana Hong Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai 200120 CHINA
Re: K162180
Trade/Device Name: Disposable Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: November 28, 2016 Received: December 2, 2016
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows a signature and the name Tina Kiang-S. The signature is a stylized design with loops and curves. The name is written in a simple, sans-serif font, with the first name above the last name.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162180
Device Name Disposable Insulin Syringe
Indications for Use (Describe)
The disposable insulin syringe is intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of insulin into parts of the body below the surface skin.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exhibit #2 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: __K162180
- 1. Date of Preparation: 12/22/2016
- 2. Sponsor Identification
### Berpu Medical Technology Co., Ltd
No. 14 Xingji Road, Yongxing Street, Longwan District, 325000, Wenzhou, Zhejiang Province
Establishment Registration Number: 3004496829
Contact Person: Jundong Tan Position: Management Representative Tel: +86-577-86651999 Fax: +86-577-86630389 Email: BERPU@BERPU.COM
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### Identification of Proposed Device 4.
Trade Name: Disposable Insulin Syringe Common Name: Insulin syringe with needle
### Regulatory Information
Classification Name: Syringe, Piston; Classification: II; Product Code: FMF Regulation Number: 21CFR 880.5860 Review Panel: General Hospital;
Indication for Use:
The disposable insulin syringe is intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of insulin into parts of the body below the surface skin.
### Device Description
The proposed device is a syringe with fixed needle tube which is intended for injection the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin. The insulin is injected to subcutaneous tissue by pushing force generated through pushing plunger rod of the insulin syringe. The proposed devices are provided sterile, single use.
The proposed device is available in 0.3ml, 0.5ml and 1ml volumes.
- 5. Identification of Predicate Device
510(k) Number: K072739 Product Name: Sterile Insulin Syringe for single use with fixed needle
Manufacturer: ShanDong Weigao Group Medical Polymer Products Co., Ltd
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test provided in this submission include:
{5}------------------------------------------------
Physical, Mechanical and Chemical Tests performed on the proposed device
| Materials | Clause 3 of ISO 9626:1991/AMD-1:2001 |
|-----------------------------------|---------------------------------------|
| Surface finish | Clause 4 of ISO 9626:1991/AMD-1:2001 |
| Cleanliness | Clause 5 of ISO 9626:1991/AMD-1:2001 |
| Limits for acidity and alkalinity | Clause 6 of ISO 9626:1991/AMD-1:2001 |
| Size designation | Clause 7 of ISO 9626:1991/AMD-1:2001 |
| Dimensions | Clause 8 of ISO 9626:1991/AMD-1:2001 |
| Stiffness | Clause 9 of ISO 9626:1991/AMD-1:2001 |
| Resistance to breakage | Clause 10 of ISO 9626:1991/AMD-1:2001 |
| Resistance to corrosion | Clause 11 of ISO 9626:1991/AMD-1:2001 |
| Cleanliness | Clause 4 of ISO 7864:1993 |
|----------------------------------|----------------------------|
| Limits for acidity or alkalinity | Clause 5 of ISO 7864:1993 |
| Limits for extractable metals | Clause 6 of ISO 7864:1993 |
| Size designation | Clause 7 of ISO 7864:1993 |
| Colour coding | Clause 8 of ISO 7864:1993 |
| Needle hub | Clause 9 of ISO 7864:1993 |
| Sheath | Clause 10 of ISO 7864:1993 |
| Needle tube | Clause 11 of ISO 7864:1993 |
| Needle point | Clause 12 of ISO 7864:1993 |
| Performance | Clause 13 of ISO 7864:1993 |
| Freedom from extraneous matter | Clause 5 of ISO 8537:2007 |
|-------------------------------------|-----------------------------|
| Limits for extraneous matter | Clause 6 of ISO 8537:2007 |
| Lubrication of syringes and needles | Clause 7 of ISO 8537:2007 |
| Range of size | Clause 8 of ISO 8537:2007 |
| Graduated scale | Clause 9 of ISO 8537:2007 |
| Barrel | Clause 10 of ISO 8537:2007 |
| Piston/plunger assembly | Clause 11 of ISO 8537:2007 |
| Needle tubing and needles | Clause 13 of ISO 8537:2007 |
| Performance of assembled syringe | Clause 14 of ISO 8537:2007 |
| Dose Accuracy testing | Clause 9.1 of ISO 8537:2007 |
Sterile Barrier Packaging Testing performed on the proposed device:
| Seal strength | ASTM F88/F88-09 |
|-------------------|----------------------|
| Internal pressure | ASTM F1140/F1140M-13 |
| Dye penetration | ASTM F 1929-12 |
{6}------------------------------------------------
Sterilization and Shelf Life Testing performed on the proposed device:
| SAL | 10-6 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Validation method | ISO 11135:2014 |
| EO residue | ISO 10993-7:2008 |
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin Limit | USP 38-NF 33 <85> |
| LAL Pyrogen Test | ISO 10993-11:2006 |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package and Sterility<br>Tests were performed on accelerated aging samples to<br>verify the claimed shelf life of the device |
Biocompatibility Testing
The patient-contact materials of Disposable Insulin Syringe are identified and biocompatibility testing is performed according to ISO 10993 standards
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
#### Substantially Equivalent (SE) Comparison 8.
| Item | Proposed Device | Predicate Device<br>K072739 | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|------------------------|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Product Code | FMF | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Regulation No. | 21CFR 880.5860 | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Classification | II | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Indication for Use | The disposable insulin syringe is intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of insulin into parts of the body below the surface skin. | The sterile insulin syringe for single use with fixed needle is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Configuration and Material | Configuration Material Protective end cap Polypropylene (PP) Plunger Polypropylene (PP) Barrel Polypropylene (PP) Piston Polyisoprene Rubber Needle Stainless Steel (SUS304) Needle cover Polypropylene (PP) | | | | | | | | | | | | | | | Configuration Material Protective end cap Polypropylene (PP) Plunger Polypropylene (PP) Barrel Polypropylene (PP) Piston Polyisoprene Rubber Needle Stainless Steel (SUS304) Needle cover Polypropylene (PP) | | | | | | | | | | | | | | |
| Syringe Volume | 0.3ml, 0.5ml, 1.0ml | 0.5ml, 1.0ml | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Needle Gauge | 31G, 30G, 29G, 28G, 27G | 30G, 29G, 28G | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Needle Diameter | 0.25mm, 0.30mm, 0.33mm,0.36mm, 0.40mm | 0.30mm, 0.33mm, 0.36mm | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Needle Length | 5mm, 6mm, 8mm, 12mm | Unknown | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Performance Test | Comply with<br>ISO 9626<br>ISO 7864<br>ISO 8536 | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Sterile | EO Sterilized | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Single Use | Single Use | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Biocompatibility | Cytotoxicity No cytotoxicity Irritation No irritation reactivity Sensitization No significant evidence of skin sensitization Systemic Toxicity No significant evidence of systemic toxicity Hemolysis No evidence of hemolysis | | | | | | | | | | | Conform with ISO 10993 | | | | | | | | | | | | | | | | | | |
## Table 1 Comparison of Technology Characteristics
{8}------------------------------------------------
| K162180 | S001 | E2 510(k) Summary |
|---------|------|-------------------|
|---------|------|-------------------|
| Pyrogen | No evidence of |
|---------|----------------|
| | pyrogens |
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.