BD INSULIN SYRINGE

K024112 · Becton, Dickinson & CO · FMF · Jan 9, 2003 · General Hospital

Device Facts

Record IDK024112
Device NameBD INSULIN SYRINGE
ApplicantBecton, Dickinson & CO
Product CodeFMF · General Hospital
Decision DateJan 9, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

BD Insulin Syringes are intended for the subcutaneous injection of insulins

Device Story

BD Insulin Syringe (Ultra-Fine and Ultra-Fine II) is a manual piston syringe for subcutaneous insulin delivery. Device consists of graduated barrel, plunger rod, and needle/hub assembly. Available in 1cc, 1/2cc, and 1/3cc capacities with 30g or 31g needles. Operated by patients or clinicians; manual injection process. Provides precise insulin dosing; single-use, sterile, non-toxic, non-pyrogenic. Benefits include reliable subcutaneous delivery of insulin.

Clinical Evidence

Bench testing only. Tests included needle pull-out force, hub pull-off forces, needle angularity, needle break-off testing, and dose accuracy. Results demonstrate performance equivalent to predicate devices.

Technological Characteristics

Piston syringe design; graduated barrel; plunger rod; needle/hub assembly. Needle gauges: 30g (12.7mm) and 31g (8mm). Sterilization: Gamma irradiation. Single-use, disposable. No electronic components or software.

Indications for Use

Indicated for subcutaneous injection of insulin in patients requiring insulin therapy.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K024112 # 510K SUMMARY OF SAFETY AND EFFECTIVENESS #### l . Submitted By: John Schalago Manager, Regulatory Affairs Becton Dickinson Consumer Healthcare 1 Becton Drive Franklin Lakes, NJ 07417-1883 201-847-5663 Phone: 201-848-0457 Fax: #### 2. Device Name: BD Insulin Syringe - Ultra-Fine™ and Ultra-Fine™ II Trade Name: Insulin Syringe Common Names: Classification Name: Piston Syringe #### 3. Predicate Device: BD Insulin Syringe - K941657 and K955235 Manufactured by: Becton Dickinson Consumer Healthcare #### 4. Device Description: The BD Insulin Syringes - Ultra-Fine™ and Ultra-Fine™ II are designed for the subcutaneous injection of a desired dose of insulin. The syringe has a graduated barrel, a plunger rod and needle/hub assembly. The needle shield is colored orange. The Ultra-Fine™ Insulin Syringe has a 30g x 1/2" (12.7mm) needle and is available in 1cc (100 units), 1/2 cc (50 units), and 1/3 cc (30 units) sizes. The Ultra-Fine™ II 31g x 1/2" (8mm) needle and is available in 1cc (100 units), ½ cc (50 units), and 1/3 cc (30 units) sizes. These devices operate on the principles of a piston syringe. The syringe fluid path is sterile (gamma irradiation sterilization), non-toxic, non-pyrogenic and single use, disposable. Continued... 00001.6 {1}------------------------------------------------ # 510K Summary of Safety and Effectiveness (Continued) #### 5. Intended Use: BD Insulin Syringes are intended for the subcutaneous injection of insulins #### Technological Characteristics: 6. The BD Ultra-Fine™ and Ultra-Fine™ II insulin syringes and the predicate devices have the identical technological characteristics and perform as piston syringes. The only difference between the principal devices and the predicate device is the needle gauge. #### 7. Performance: Bench tests relating to the performance of the needle length were conducted. The tests performed included needle pull-out force, hub pull-off forces, needle angularity, needle break-off testing and dose accuracy. The results demonstrate that the BD Insulin Syringes performs equivalent to the predicate devices and is safe and effective when used as intended. The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E, under which a device can be marketed without pre-approval or classification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US patent Laws or their application by the courts. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three heads, depicted in black. JAN 0 9 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. John Schalago Manager, Regulatory Affairs Becton Dickinson & Company 1 Becton Drive Franklin Lakes, New Jersey 07417-1880 Re: K024112 Trade/Device Name: BD Insulin Syringe-Ultra-Fine™ and Ultra-Fine™ II Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: December 11, 2002 Received: December 13, 2002 Dear Mr. Schalago: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Schalago Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Attachment XI.1 Page 1 of 1 # 510(k) Number (if known): < 024112 Device Name: Becton Dickinson Insulin Syringe Indications For Use: Becton Dickinson insulin syringes are intended for subcutaneous injection of insulins. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) (Division Sign-Off) v. General Hospital ision of Anest ction Control. De 510(k) Number. K024112 000024
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...