← Product Code [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF) · K024052

# MERIT MEDICAL 1-ML SYRINGE (K024052)

_Merit Medical Systems, Inc. · FMF · Feb 21, 2003 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K024052

## Device Facts

- **Applicant:** Merit Medical Systems, Inc.
- **Product Code:** [FMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF.md)
- **Decision Date:** Feb 21, 2003
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Merit Medical 1-mL Syringes are used to inject fluids into, or withdraw fluids from, the body.

## Device Story

Merit Medical 1-mL Syringe is a manual piston syringe consisting of a calibrated hollow barrel and movable plunger. Features a male Luer Lock connector for attachment to hypodermic needles or other female Luer-compatible devices. Operated manually by a clinician to inject or withdraw fluids. Device provides a standard mechanical means for fluid delivery or aspiration.

## Clinical Evidence

Bench testing only; conformance to ISO 7886-1:1993 standard.

## Technological Characteristics

Manual piston syringe; male Luer Lock connector; calibrated barrel; plunger mechanism. Conforms to ISO 7886-1:1993.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Becton Dickinson (BD) single-use Hypodermic and Insulin Syringes ([K980580](/device/K980580.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K024052

### 510(K) Summarv

FEB 2 1 2003

- Dec. 3. 2002 1. Preparation Date
- Merit Medical Systems, Inc. (MMS) 2. Submitted Bv 1600 West Merit Pkwy. South Jordan, UT 84095

#### Contact Person/ Prepared By

Stephanie A. Erskine Director, Regulatory Affairs, MMS Phone (801) 208 4349 Fax (801)253 1684 Email serskine@merit.com

| 3. | Device Identification |                                  |
|----|-----------------------|----------------------------------|
|    | Trade Name            | Merit Medical 1-mL Syringe       |
|    | Common Name           | Syringe, Hypodermic Syringe      |
|    | Classification Name   | Piston Syringe (21 CFR 880.5860) |

- 4. Predicate Device(s) K980580, Becton Dickinson (BD) single-use Hypodermic and Insulin Syringes
#### 5. Device Description

Merit Medical Systems 1-mL Syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male Luer Lock connector (nozzle) for attaching the female Luer connector (hub) of a hypodermic single lumen needle, or for attaching other devices with a female Luer.

#### 6. Intended Use

Merit Medical 1-mL Syringes are used to inject fluids into, or withdraw fluids from, the body.

#### 7. Statement of Intent to Conform to ISO 7886-1;1993

Merit Medical Systems. Inc. intends to establish that its 1-mL Syringes conform to the FDA-recognized Consensus Standard, ISO 7886-1:1993 Sterile hypodermic needles for single use- Part 1: Syringes for manual use, before marketing the devices. Data supporting conformance with the standard, with minor exceptions and deviations identified in the premarket notification submission, will be available before marketing the device.

#### 8. Conclusion

The Merit Medical 1-mL Syringe is safe and effective for its intended use.

{1}------------------------------------------------

# 9. Substantial Equivalence/ Conformity with Standards

#### Similarities/ Differences of the proposed device when compared to 9.1 the predicate:

### 9.1.1 Intended Use

The Becton Dickinson (B-D) General Purpose syringes are intended for general purpose fluid aspiration/ injection. Merit Medical 1-mL Syringes are used to inject fluids into, or withdraw fluids from, the body. As such, the Intended Uses of the MMS and B-D syringes are equivalent.

#### 9.1.2 Materials

Materials used in the manufacture of MMS syringes are typically used in the manufacture of general-purpose syringes, including the predicate device.

#### 9.1.3 Design

The design of the MMS syringe is typical for syringes, including that of the predicate.

# 9.1.4 Operational Principles

The MMS syringe is manually operated by advancing and withdrawing the plunger in the barrel. The operating principles are identical for all manual syringes, including the predicate.

# 9.1.5 Technology

The same fundamental technology is used in the design of the MMS svringes as is employed in the design of all manual syringes, including the predicate.

# 9.1.6 Safety and Performance

MMS has provided a statement that its syringes will conform to the requirements of ISO 7886-1:1993, an FDA- recognized consensus standard, before marketing the devices. This statement and the data that will be collected to support conformance will be used to demonstrate safety and performance in lieu of demonstrating substantial equivalence with the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract image of an eagle.

FEB 2 1 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Stephanie A. Erskine Director, Regulatory Affairs Merit Medical Systems, Incorporated 1600 West Merit Parkway South Jordan, Utah 84095

Re: K024052

Trade/Device Name: Merit Medical 1-mL Syringe Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: December 5, 2002 Received: December 6, 2002

Dear Ms. Erskine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Erskine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Suarez

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### INDICATION(S) FOR USE STATEMENT*

Merit Medical 1-mL Syringes are used to inject fluids into, or withdraw fluids from, the body.

Signature of 510(k) Submitter:

Printed Name of Submitter:

Stephanie A. Erskine Director, Regulatory Affairs Merit Medical Systems, Inc.

Date:

*Suggested language and format to meet the requirements of sections 513(i) of the Federal Food, Drug, and Cosmetic Act, as amended, and sections 807.92(a)(5) and 801.4 of the Code of Federal Regulations, Title 21.

Concurrence of Office of Device Evaluation

510(k) Number

K624052

Division Sign-Off Office of Device Evaluation

Prescription Use

✓

OR Over-The-Counter Use

Patuxo Crescenti

y, General Hospital. n of Anesthesiolor tion Control, Dent

510(k) Number. K

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K024052](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/FMF/K024052)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).