CONVERTORS DRAPE MATERIAL
Device Facts
| Record ID | K964125 |
|---|---|
| Device Name | CONVERTORS DRAPE MATERIAL |
| Applicant | Allegiance Healthcare Corp. |
| Product Code | KKX · General, Plastic Surgery |
| Decision Date | Dec 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4370 |
| Device Class | Class 2 |
Intended Use
The Convertors® Drapes and Leggings are devices intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.
Device Story
Convertors® Drapes and Leggings function as sterile barriers; composed of nonwoven polyester fibers adhesively laminated to 1.25 mil embossed polyethylene film. Used in clinical/surgical settings to isolate incision sites from contaminants. Protects patients by preventing microbial transfer; aids clinicians in maintaining sterile fields. No active components; passive barrier technology.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993-1 (cytotoxicity, sensitization, irritation/intracutaneous reactivity). Physical performance testing included hydrostatic head, impact penetration, synthetic blood penetration (ASTM 1670-1995), Phi X174 viral penetration (ASTM 1671-1995), Trapezoid Tear, Grab Tensile, Mullen Burst, Basis Weight, and flammability (Class 2).
Technological Characteristics
Nonwoven adhesively bonded polyester fibers laminated to 1.25 mil embossed blended polyethylene film. Passive barrier device. Flammability Class 2. Tested per ASTM 1670-1995 and ASTM 1671-1995.
Indications for Use
Indicated for use as a protective patient covering to isolate surgical incision sites from microbial and other contamination during surgical procedures.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
- Convertors® Drape Materials
Related Devices
- K023419 — TRILAMINATE DRAPES/TIBURON · Allegiance Healthcare Corp. · Feb 11, 2003
- K023425 — DRAPES WITH ABSORBENT REINFORCEMENT · Allegiance Healthcare Corp. · Nov 6, 2002
- K981818 — CONVERTORS TRILAMINATE DRAPES · Allegiance Healthcare Corp. · Jul 10, 1998
- K993795 — CONVERTORS SURGICAL DRAPES · Allegiance Healthcare Corp. · Jan 12, 2000
- K970554 — CONVERTORS SMS DRAPES · Allegiance Healthcare Corp. · May 27, 1997