Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs

K182948 · Ansell Healthcare Products, LLC · KGO · Mar 22, 2019 · General, Plastic Surgery

Device Facts

Record IDK182948
Device NameGammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs
ApplicantAnsell Healthcare Products, LLC
Product CodeKGO · General, Plastic Surgery
Decision DateMar 22, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Intended Use

Gammex PI Hybrid Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Sterile, disposable, powder-free surgical gloves; synthetic rubber blend of polyisoprene and polychloroprene; polyurethane inner coating for donning. Used by operating room personnel to protect surgical wounds from contamination. Subject device is identical to predicate (K151694) with added labeling for chemotherapy drug permeation resistance. Tested per ASTM D6978-05 for breakthrough detection times against 29 chemotherapy agents. Provides barrier protection during surgical procedures and handling of chemotherapy drugs.

Clinical Evidence

No clinical data; bench testing only. Device performance validated via ASTM D6978-05 chemotherapy permeation testing, ASTM D3577-09 physical properties, and ISO 10993 biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity).

Technological Characteristics

Synthetic rubber blend (polyisoprene/polychloroprene); polyurethane inner coating. Sterile (radiation, 10^-6 SAL). Meets ASTM D3577-09 (dimensions, physical properties), ASTM D5151-06 (freedom from holes), ASTM D6124-06 (powder-free), and ASTM D6978-05 (chemotherapy permeation).

Indications for Use

Indicated for operating room personnel to protect surgical wounds from contamination. Tested for use with chemotherapy drugs per ASTM D6978-05.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 22, 2019 Ansell Healthcare Products, LLC. Don Cronk Sr Manager, QARA / Technical Services 2301 Robb Drive Reno, Nevada 89523 Re: K182948 Trade/Device Name: Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: February 21, 2019 Received: February 21, 2019 Dear Don Cronk: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Clarence W. Murray lii III -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182948 #### Device Name Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs ## Indications for Use (Describe) Gammex PI Hybrid Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Tested chemotherapy drugs are as follows: | Test Chemotherapy drug & Concentration | Minimum Breakthrough Detection Time in Minutes | |----------------------------------------|------------------------------------------------| | Bleomycin sulfate (15.0 mg/ml) | >240 | | Busulfan (6.0 mg/ml) | >240 | | Carboplatin (10.0 mg/ml) | >240 | | Carmustine (3.3 mg/ml) | 23.6 | | Cisplatin (1.0 mg/ml) | >240 | | Cyclophosphamide (20.0 mg/ml) | >240 | | Cytarabine (100.0 mg/ml) | >240 | | Dacarbazine (10 mg/ml) | >240 | | Daunorubicin (5.0 mg/ml) | >240 | | Docetaxel (10.0 mg/ml) | >240 | | Doxorubicin Hydrochloride (2.0 mg/ml) | >240 | | Epirubicin (2.0 mg/ml) | >240 | | Etoposide (20.0 mg/ml) | >240 | | Fludarabine (25.0 mg/ml) | >240 | | Fluorouracil(50.0 mg/ml) | >240 | | Gemcitabine (Gemzar) (38.0 mg/ml) | >240 | | Idarubicin (1.0 mg/ml) | >240 | | Ifosfamide (50.0 mg/ml) | >240 | | Irinotecan (20.0 mg/ml) | >240 | | Mechlorethamine HCl(1.0 mg/ml) | >240 | | Melphalan (5.0 mg/ml) | >240 | | Methotrexate (25.0 mg/ml) | >240 | | Mitomycin C (0.5 mg/ml) | >240 | | Mitoxantrone (2.0 mg/ml) | >240 | | Oxaliplatin (2.0 mg/ml) | >240 | | Paclitaxel (Taxol) (6.0 mg/ml) | >240 | | Rituximab (10.0 mg/ml) | >240 | | ThioTEPA (10.0 mg/ml) | 36.1 | | Vincristine Sulfate(1.0 mg/ml) | >240 | {3}------------------------------------------------ Prescription Use (Part 21 CFR 801 Subpart D) - X Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 510k Summary K182948 #### Submitter Ansell Healthcare Products LLC. 111 Wood Avenue South, Suite 210 Iselin, NJ 08830 USA #### Contact Person: Don Cronk Associate Director, Regulatory Affairs Americas Phone: 775-470-7106 don.cronk@ansell.com #### Date Prepared March 14th 2019 #### Name of Device | Trade Names: | Gammex PI Hybrid Surgical Gloves Tested for Use with Chemotherapy Drugs | |----------------------------|-------------------------------------------------------------------------| | Common Name: | Surgeon's Gloves | | Classification Name: | Surgeon's Gloves | | Classification Regulation: | 21 CFR 878.4460 | | Device Class: | I | | Product Code: | KGO | | Classification Panel: | General and Plastic Surgery | | 510k: | K182948 | | | | #### Legally Marketed Predicate Device K151694 – Gammex PI Hybrid Surgical Gloves #### Device Description Gammex Pl Hybrid Surgical Gloves are sterile and disposable devices. Gloves are made of synthetic rubber blend of polyisoprene and polychoroprene. A polyurethane polymer coating is applied to the inner surface of the glove to make donning eay. Ansell has performed independent laboratory testing on the already cleared device to support a labeling modification for the subject device that the gloves are safe and effective for handling chemotherapy agents. #### Indication for Use Statement Gammex PI Hybrid Surgical Gloves are intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. {5}------------------------------------------------ # Chemotherapy drug Permeation (Average breakthrough detection time in minutes) (ASTM D6978-05) | Test Chemotherapy drug & Concentration | Minimum Breakthrough Detection Time (Sample<br>1,2,3)(mins) | |----------------------------------------|-------------------------------------------------------------| | Bleomycin sulfate (15.0 mg/ml) | >240 | | Busulfan (6.0 mg/ml) | >240 | | Carboplatin (10.0 mg/ml) | >240 | | Carmustine (3.3 mg/ml) | 23.6(23.8,26.5,23.6) | | Cisplatin (1.0 mg/ml) | >240 | | Cyclophosphamide (20.0 mg/ml) | >240 | | Cytarabine (100.0 mg/ml) | >240 | | Dacarbazine (10 mg/ml) | >240 | | Daunorubicin (5.0 mg/ml) | >240 | | Docetaxel (10.0 mg/ml) | >240 | | Doxorubicin Hydrochloride (2.0 mg/ml) | >240 | | Epirubicin (2.0 mg/ml) | >240 | | Etoposide (20.0 mg/ml) | >240 | | Fludarabine (25.0 mg/ml) | >240 | | Fluorouracil(50.0 mg/ml) | >240 | | Gemcitabine (Gemzar) (38.0 mg/ml) | >240 | | Idarubicin (1.0 mg/ml) | >240 | | Ifosfamide (50.0 mg/ml) | >240 | | Irinotecan (20.0 mg/ml) | >240 | | Mechlorethamine HCl(1.0 mg/ml) | >240 | | Melphalan (5.0 mg/ml) | >240 | | Methotrexate (25.0 mg/ml) | >240 | | Mitomycin C (0.5 mg/ml) | >240 | | Mitoxantrone (2.0 mg/ml) | >240 | | Oxaliplatin (2.0 mg/ml) | >240 | | Paclitaxel (Taxol) (6.0 mg/ml) | >240 | | Rituximab (10.0 mg/ml) | >240 | | ThioTEPA (10.0 mg/ml) | 36.1(36.1,37.0,36.7) | | Vincrinstine Sulfate(1.0 mg/ml) | >240 | | Vincrinstine Sulfate(1.0 mg/ml) | >240 | {6}------------------------------------------------ | | Predicate Device | | Subject Device | Comparison | |----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| | Trade Name | Gammex PI Hybrid<br>Surgical Gloves | | Gammex PI Hybrid Surgical Gloves Tested for Use with<br>Chemotherapy Drugs | Different | | 510(k) Number | K151694 | | K182948 | Different | | Submitter | Ansell Healthcare<br>Products LLC | | Ansell Healthcare Products LLC | Same | | Product Code | KGO | | KGO | Same | | Regulation Number | 21 CFR 878.4460 | | 21 CFR 878.4460 | Same | | Regulation Name | Non-powdered<br>Surgeon's Glove | | Non-powdered Surgeon's Glove | Same | | Indications for Use | These gloves are<br>intended to be worn<br>by operating room<br>personnel to protect<br>a surgical wound<br>from<br>contamination. | | Gammex PI Hybrid Surgical Gloves are intended to be worn<br>by operating room personnel to protect a surgical wound<br>from contamination. These gloves were tested for use with<br>Chemotherapy Drugs as per ASTM D6978-05. Standard<br>Practice for Assessment of Medical Gloves to Permeation by<br>Chemotherapy Drugs.<br><br>Tested chemotherapy drugs are as follows: | Similar with the<br>exception of the<br>chemotherapy claim | | | | | Test Chemotherapy drug &<br>Concentration | Minimum<br>Breakthrough<br>Detection<br>Time | | | | | Bleomycin sulfate<br>(15.0 mg/ml) | >240 | | | | | Busulfan<br>(6.0 mg/ml) | >240 | | | | | Carboplatin<br>(10.0 mg/ml) | >240 | | | | | Carmustine<br>(3.3 mg/ml) | 23.6 | | | | | Cisplatin<br>(1.0 mg/ml) | >240 | | | | | Cyclophosphamide<br>(20.0 mg/ml) | >240 | | | | | Cytarabine<br>(100.0 mg/ml) | >240 | | | | | Dacarbazine<br>(10 mg/ml) | >240 | | | | | Daunorubicin<br>(5.0 mg/ml) | >240 | | | | | Docetaxel<br>(10.0 mg/ml) | >240 | | | | | Doxorubicin Hydrochloride<br>(2.0 mg/ml) | >240 | | | | | Epirubicin<br>(2.0 mg/ml) | >240 | | | | | | | | | Fludarabine<br>(20.0 mg/ml) | >240 | | | | | Fluorouracil<br>(25.0 mg/ml) | >240 | | | | | Gemcitabine (Gemzar)<br>(38.0 mg/ml) | >240 | | | | | Idarubicin<br>(1.0 mg/ml) | >240 | | | | | Ifosfamide<br>(50.0 mg/ml) | >240 | | | | | Irinotecan<br>(20.0 mg/ml) | >240 | | | | | Mechlorethamine HCl<br>(1.0 mg/ml) | >240 | | | | | Melphalan<br>(5.0 mg/ml) | >240 | | | | | Methotrexate<br>(25.0 mg/ml) | >240 | | | | | Mitomycin C<br>(0.5 mg/ml) | >240 | | | | | Mitoxantrone<br>(2.0 mg/ml) | >240 | | | | | Oxaliplatin<br>(2.0 mg/ml) | >240 | | | | | Paclitaxel (Taxol)<br>(6.0 mg/ml) | >240 | | | | | Rituximab<br>(10.0 mg/ml) | >240 | | | | | ThioTEPA<br>(10.0 mg/ml) | 36.1 | | | | | Vincrinstine Sulfate<br>(1.0 mg/ml) | >240 | | | | Prescription or<br>Over- The Counter-Use | Over-The-Counter-Use | Over-The-Counter-Use | Same | | | Materials | Synthetic rubber blend of<br>polyisoprene and<br>polychloroprene | Synthetic rubber blend of polyisoprene and<br>polychloroprene | Same | | | Coating | Polyurethane polymer<br>inner coating to aid<br>donning | Polyurethane polymer inner coating to aid donning | Same | | | Design | Single use | Single use | Same | | | | Sterile | Sterile | Same | | | | Powder-free | Powder-free | Same | | | | hand specific | hand specific | Same | | | | Beaded cuff | Beaded cuff | Same | | | Color | White | White | Same | | | Dimensions and<br>physical properties | Meets ASTM D3577 | Meets ASTM D3577 | Same | | | Sterility | Sterile | Sterile | Same | | | Sterilization method | Radiation | Radiation | Same | | | Sterility Assurance Level<br>(SAL) | 10-6 SAL | 10-6 SAL | Same | | | Freedom from holes | Meets ASTM D3577-<br>09(2015)<br>Inspection level/AQL: GI/<br>AQL 0.65 (multiple sampling<br>plan)<br>Lot Size: 35,001 to 150,000<br>pieces<br>Sample size: 100 pieces<br>Accept number: 0<br>Actual found: 0 | Meets ASTM D3577-<br>09(2015)<br>Inspection level/AQL: GI/AQL<br>0.65 (multiple sampling plan)<br>Lot Size: 35,001 to 150,000<br>pieces<br>Sample size: 100 pieces<br>Accept number: 0<br>Actual found: 0 | Same | | | Powder-Free | Meets ASTM D 6124-06 The<br>averaged residual powder<br>content for the glove during<br>process validation is 0.48mg<br>per glove | Meets ASTM D 6124-06 The<br>averaged residual powder<br>content for the glove during<br>process validation is 0.48mg<br>per glove | Same | | | Biocompatibility Primary<br>Skin Irritation<br>ISO 10993-<br>10:2010 | Under the conditions of the study<br>(per ISO 10993-10), the device is not<br>an irritant | Under the conditions of the study<br>(per ISO 10993-10), the device is not<br>an irritant | Same | | | Biocompatibility Dermal<br>Sensitization - ISO 10993-<br>10:2010 | Under the conditions of the study<br>(per ISO 10993-10), not a sensitizer | Under the conditions of the study (per<br>ISO 10993-10), not a sensitizer | Same | | | Biocompatibility Acute Systemic<br>Toxicity - ISO 10993-11: 2006 | Under the conditions of the study,<br>there was no mortality or evidence<br>of systemic toxicity | Under the conditions of the study,<br>there was no mortality or evidence<br>of systemic toxicity | Same | | | Shelf Life | 3 years | 3 years | Same | | {7}------------------------------------------------ {8}------------------------------------------------ The subject device is manufactured from synthetic rubber blend of polyisoprene with polyurethane polymer inner coating to aid donning. The predicate device is manufactured from the same formulation. The subject device's materials are functionally equivalent to those of the subject device is identical to the prior cleared version of the device in all regards with the exception of the chemotherapy claims found within the Indications for Use. The subject device meets the applicable requirements for surgeon's gloves with regard to dimensions and sizes, physical properties, freedom from holes, powder residues, and protein content as found in the following standards: ASTM D5151 and ASTM D6124. The subject device passes biological reactivity testing for dermal sensitization, irritation, and acute systemic toxicity in accord with the ISO 10993-10 and ISO 10993-11. #### Non-Clinical Summary: A clinical study was not conducted on the subject or predicate devices. Technological Characteristics Gammex PI Hybrid Surgical Gloves, tested for use with chemotherapy drugs, have the following technological characteristics as compared to ASTM or equivalent standards: {9}------------------------------------------------ | Technological Characteristics | Standard/Test/FDA Guidance | Result Summary | |--------------------------------------|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Dimensions | ASTM D3577-09 | Meets ASTM D3577-09 requirements for length, width and thickness | | Length | Minimum 265mm | Average 303mm | | Palm Width(size) | (mm) | Average value in mm | | 5.5 | 70±6 | 75 | | 6 | 76±6 | 80 | | 6.5 | 83±6 | 85 | | 7 | 89±6 | 91 | | 7.5 | 95±6 | 97 | | 8 | 102±6 | 103 | | 8.5 | 108±6 | 112 | | 9 | 114±6 | 117 | | Thickness | (mm) | Average value in mm | | Finger | Minimum 0.10 | 0.22 | | Palm | Minimum 0.10 | 0.2 | | Cuff | Minimum 0.10 | 0.15 | | Physical Properties | ASTM D3577-09 | Meets ASTM D3577-09 requirements for tensile strength and elongation at break<br>before and after accelerated aging | | Freedom from holes | ASTM D3577-09<br>ASTM D5151-06 | Meets ASTM D3577-09 and ASTM D5151-06 requirements of AQL 1.5 | | Powder-Free | ASTM D3577-09<br>ASTM D6124-06 | Meets Applicable requirement for Powder Free; ≤ 2 mg per glove | | Sterility | ANSI/AAMI/ISO 11137-1:2006 | Meets ANSI/AAMI/ISO 11137-1:2006 requirement of 10-6 SAL | | Biocompatibility: | | | | ISO in vitro cytotoxicity | ISO 10993-5:2009 | Under the conditions of the study, the device extract was found to be cytotoxic and therefore the device extracts were evaluated by ISO 10993-11 - Test for systemic toxicity. From the Acute Systemic Toxicity device extracts, the device extracts did not elicit a systemic response in the animal model | | ISO Skin Irritation Study | ISO 10993-10:2010 | Under the conditions of the study, not an irritant | | ISO Maximization Sensitization Study | ISO 10993-10:2010 | Under the conditions of the study, not a sensitizer | | ISO Systemic Toxicity Study | ISO 10993-11: 2006 | Under the conditions of the study, there was no mortality or evidence of systemic toxicity | | Chemotherapy Permeation Standard | ASTM D6978 - 05(2013) | Under the conditions of the study, the permeation is acceptable | #### Conclusion The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is as safe, as efferms as well as or better than the legally marketed device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...