NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs

K150973 · Terang Nusa Snd Bhd · KGO · Aug 4, 2015 · General, Plastic Surgery

Device Facts

Record IDK150973
Device NameNUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs
ApplicantTerang Nusa Snd Bhd
Product CodeKGO · General, Plastic Surgery
Decision DateAug 4, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Intended Use

This Surgeons glove is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves are tested for use with Chemotherapy Drugs. Testing based on ASTM D6978-05

Device Story

NUZONE X2G is a powder-free, synthetic polychloroprene surgical glove. Designed for use by operating room personnel to provide a barrier against contamination of surgical wounds. Device is also tested for resistance to permeation by specific chemotherapy drugs per ASTM D6978-05. Used in clinical/surgical settings; provides physical protection for both patient and clinician during procedures involving chemotherapy agents.

Clinical Evidence

Bench testing only. Permeation testing performed for ten chemotherapy drugs (Carmustine, Cisplatin, Cyclophosphamide, Dacarbazine, Doxorubicin Hydrochloride, Etoposide, Paclitaxel, Fluorouracil, Thiotepa, Vincristine Sulfate) per ASTM D6978-05. Breakthrough detection times reported for each agent.

Technological Characteristics

Material: Synthetic polychloroprene rubber. Form factor: Powder-free surgical glove. Testing standard: ASTM D6978-05 for chemotherapy drug permeation. Non-sterile/sterile status not explicitly detailed, but intended for surgical use.

Indications for Use

Indicated for use by operating room personnel to protect surgical wounds from contamination. Suitable for use during handling of chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are depicted in a simple, flowing line style. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 22, 2015 Terang Nusa, Sdn. Bhd. Mr. Robert Hill Regulatory Director 1. Jalan 8 Pengkalan Chepa 2 Industrial Zone 16100 Kota Bharu MALAYSIA Re: K150973 Trade/Device Name: NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's glove Regulatory Class: I Product Code: KGO, LZC Dated: July 1, 2015 Received: July 5, 2015 Dear Mr. Hill: This letter corrects our substantially equivalent letter of August 4, 2015. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, *Tejashri Purohit-Sheth, M.D.* Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150973 ### Device Name NUZONE X2G Polychloroprene Synthetic Green Surgical Gloves Powder Free, Tested for Use with Chemotherapy Drugs #### Indications for Use (Describe) This Surgeons glove is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloves are tested for use with Chemotherapy Drugs. Testing based on ASTM D6978-05 | DRUG | Minimum Breakthrough Detection Time<br>(Specimen 1/2/3)<br>Minutes | |--------------------------------------------------|--------------------------------------------------------------------| | Carmustine (BCNU), 3.3 mg/ml (3,300ppm) | 30.6 (30.6, 30.6, 30.8) | | Cisplatin, 1.0 mg/ml (1,000ppm) | Up to 240 | | Cyclophosphamide (Cytoxan), 20 mg/ml (20,000ppm) | Up to 240 | | Dacarbazine (DTIC), 10.0 mg/ml (10,000ppm) | Up to 240 | | Doxorubicin Hydrochloride, 2.0 mg/ml (2,000ppm) | Up to 240 | | Etoposide (Toposar), 20.0 mg/ml (20,000ppm) | Up to 240 | | Paclitaxel (Taxol), 6.0 mg/ml (6,000ppm) | Up to 240 | | Fluorouracil, 50.0 mg/ml (50,000ppm) | Up to 240 | | Thiotepa, 10.0 mg/ml (10,000ppm) | 31.2 (31.2, 60.8, 61.3) | | Vincristine Sulfate, 1.0 mg/ml (1,000ppm) | Up to 240 | Please note that the following drugs have low permeation times: Carmustine (BCNU) 30.6 Thiotepa 31.2 Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...