K994244 · Allegiance Healthcare Corp. · FYA · Feb 18, 2000 · General, Plastic Surgery
Device Facts
Record ID
K994244
Device Name
CONVERTORS SURGICAL GOWNS, STERILE
Applicant
Allegiance Healthcare Corp.
Product Code
FYA · General, Plastic Surgery
Decision Date
Feb 18, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
Surgical apparel are devices intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids and particulate material.
Device Story
Convertors® Surgical Gowns are protective apparel for operating room personnel. Constructed from degradable spunlaced nonwoven fabric, the gowns are available in unreinforced, fabric-reinforced (additional nonwoven layer in chest/sleeves), and poly-reinforced (additional polyolefin film layer in chest/sleeves) configurations. Worn by surgeons and staff during surgical procedures to prevent cross-contamination of microorganisms, body fluids, and particulates between the patient and the wearer. The device functions as a physical barrier. No electronic or software components.
Clinical Evidence
Bench testing only. Materials evaluated for biocompatibility per ISO 10993 Part-1, including cytotoxicity, sensitization, and primary skin irritation. Materials met industry-recognized test methods for intended use.
Indicated for use by operating room personnel during surgical procedures to provide a barrier against microorganisms, body fluids, and particulate material for both the patient and the wearer.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K102652 — NON REINFORCED SURGICAL GOWNS, FILM REINFORCED SURGICAL GOWNS, FABRIC REINFORCED SURGICAL GOWNS · Gri Medical & Electronic Technology Co., Ltd. · Feb 3, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
## FEB 1 8 2000
Image /page/0/Picture/1 description: The image shows the word "Allegiance" in a bold, sans-serif font. To the left of the word is a stylized graphic that resembles a cluster of dots arranged in a cross-like shape. The word "Allegiance" is the primary focus of the image, and the graphic serves as a visual element associated with the text.
K9942yy
Allegiance Healthcare Corporation 1500 •• Ξ – egan Road McGa.v. Park. Illinois 60085-6787 847.473.1500
FAX: 847.785.2461
## SMDA REQUIREMENTS
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Convertors® Surgical Gowns
| Manufacturer: | Allegiance Healthcare Corporation<br>One Butterfield Trail<br>El Paso, Texas 79906 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Affairs Contact: | Sharon Robbins<br>1500 Waukegan Road MPWM<br>McGaw Park, IL 60085 |
| Telephone: | (847) 785-3311 |
| Date Summary Prepared: | September, 1999 |
| Common Name: | Convertors® Surgical Gowns |
| Classification: | Class II per 21CFR § 878.4040 |
| Predicate Device: | Isolyser Industries Enviroguard Surgeons Gowns |
| Description: | The gowns are comprised of a single base layer<br>of degradable spunlaced nonwoven fabric in<br>gown configurations of unreinforced, fabric<br>reinforced and poly-reinforced. The fabric<br>reinforced gown contains an additional layer of<br>spunlaced nonwoven fabric in the chest and<br>sleeves area. The poly-reinforced gown contains<br>an additional layer of polyolefin film in the chest<br>and sleeves of the gown. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Allegiance" in a bold, sans-serif font. To the left of the word is a graphic of a cross made up of small dots. The word is in black and the background is white.
Allegian se Healthcare Corporation 1500 Wajyegan Road McGaw -ark. Illinois 60085-6787 847.473 ・500
FAX: 8ニー 785.2461
# SMDA REQUIREMENTS (continued)
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Convertors® Surgical Gowns
| Intended Use: | Surgical apparel are devices intended to be worn<br>by operating room personnel during surgical<br>procedures to protect both the surgical patient<br>and the operating room personnel from the<br>transfer of microorganisms, body fluids and<br>particulate material. |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence: | The Convertors® Surgical gowns are<br>substantially equivalent to the Isolyser<br>Enviroguard gowns in that: |
| | - the intended use is the same<br>- the performance attributes are<br>similar |
| Summary of testing: | All materials used in the fabrication of this<br>Convertors® Surgical gowns were evaluated through<br>biological qualification safety tests as outlined in ISO<br>10993 Part-1 "Biological Evaluation of Medical<br>Devices". The biocompatibility tests performed were<br>cytotoxicity, sensitization, and primary skin irritation.<br>These materials also were tested in accordance with<br>industry recognized test methods and were found to<br>be acceptable for the intended use. |
ー | 3,00000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and its head turned to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
AUG 18 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Sharon Robbins Regulatory Affairs Manager Allegiance Healthcare Corporation 1500 Waukegan Road McGaw Park, Illinois 60085-6787
Re: K994244
Trade/Device Name: Convertors® Surgical Gowns Regulation Number: 878.4040 Regulation Name: Surgical apparel Regulatory Class: II Product Code: FYA Dated: December 14, 1999 Received: December 16, 1999
Dear Ms. Robbins:
This letter corrects our substantially equivalent letter of February 18, 2000.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Robbins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Chiu S. Lin, Ph.D.
Chiu S. Lin, Ph. B Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Allegiance
Allegiance Healthcare Corporation 1500 Vi Engegan Road McGar. Fark. Illinois 60085-6787 847.2-3 -500 AX 5-- 785.2461
Page 1 of 1
510(k) Number (if known):
Unknown K99A244
Device Name:
Convertors® Surgical Gowns, Sterile
Indications For Use:
The Convertors® Surgical Gowns are devices intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids and particulate material.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The Counter Use Prescription Use ਾ (Per 21 CFR 801.109) (Division Sign-Off) Division of Dental, Infection C and General Hospital Device 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.