K231618 · Original Mattress Factory, Inc. · FXX · Jul 28, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231618
Device Name
3 Layer Surgical Mask (134252531)
Applicant
Original Mattress Factory, Inc.
Product Code
FXX · General, Plastic Surgery
Decision Date
Jul 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
When properly worn, the 3 Layer Surgical Masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate materials. This device is non sterile and for single use only.
Device Story
Single-use, flat-pleated surgical face mask; 3-layer construction; outer/inner layers spun-bond polypropylene; middle layer melt-blown polypropylene filter; polyester/spandex ear loops; moldable nose piece. Used in clinical settings or general use to reduce exposure to blood, body fluids, and particulates. Protects both patient and wearer. No electronic components or software.
Clinical Evidence
No clinical data. Bench testing only. Performance validated against ASTM F2299 (PFE), ASTM F1862 (fluid resistance), ASTM F2101 (BFE), 16 CFR 1610 (flammability), and EN 14683 (differential pressure). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).
Indicated for use by patients and healthcare workers to protect against the transfer of microorganisms, body fluids, and particulate materials in infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical Face Mask (K202595)
Related Devices
K210679 — Medical Face Mask · Wuxi Yushou Medical Appliances Co., Ltd. · Jan 18, 2022
K210767 — Surgical Masks, Model:FE-1 · Hunan Triplex Precision Medical Devices Co., Ltd. · Jun 10, 2021
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
July 28, 2023
Original Mattress Factory Inc. Nathan Elseser Director of Personal Protective Equipment 4930 State Road Cleveland, Ohio 44134
Re: K231618
Trade/Device Name: 3 Layer Surgical Mask (134252531) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 31, 2023 Received: June 2, 2023
Dear Nathan Elseser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231618
Device Name 3 Layer Surgical Mask
Indications for Use (Describe)
When properly worn, the 3 Layer Surgical Masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate materials. This device is non sterile and for single use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### 1. Submitter Information:
Company Name: The Original Mattress Factory, Inc. d.b.a. Original Medical Supply
Address: 4930 State Road, Cleveland, Ohio, USA 44134
Phone Number: 216-661-8388
Contact Person: Nathan Elseser
Email: Nelseser@OriginalMattress.com
#### Subject Device Information
| Type of 510(k): | Traditional |
|--------------------|-----------------------|
| Common Name: | Surgical Face Mask |
| Trade Name: | 3 Layer Surgical Mask |
| Classification: | Mask, Surgical |
| Review Panel: | General Hospital |
| Product Code: | FXX |
| Regulation Number: | 21 CFR 878.4040 |
| Regulation Class: | Class II |
#### Predicate Device Information:
| Sponsor: | Premier Guard USA LLC |
|--------------------|--------------------------------------------------------------------|
| Address: | 460 Briarwood Drive, Suite 400 |
| Common Name: | Surgical Face Mask |
| Trade Name: | Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical Face Mask |
| 510(k) Number: | K202595 |
| Product Code: | FXX |
| Regulation Number: | 21 CFR 878.4040 |
| Regulation Class: | Class II |
#### 2. Device Description
Subject Device (3 Layer Surgical Mask) is single use, 3 layers, flat-pleated style face mask with ear loops and moldable nose piece. The outer layer and inner facing layer of face mask consist of spun-bond polypropylene, and the middle layer consists of melt blown polypropylene filter. Each mask contains polyester / spandex blend ear loops to secure the mask over the user's face and mouth with moldable nose piece to firmly fit over the nose.
{4}------------------------------------------------
### 3. Intended Use
When properly worn, the 3 Layer Surgical Mask are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate materials. This device is non sterile and for single use only.
## 4. Test Summary
Subject device has been evaluated for safety and performance by lab bench testing according to the following standards:
- ASTM F2299 Standard Test Method for Determining the Initial Efficiency of Materials Used in Surgical face masks to Penetration by Particulates Using Latex Spheres.
- ASTM F1862 Standard test method for resistance of Surgical face masks to penetration by synthetic blood (Horizontal projection of fixed volume at a known velocity)
- ASTM F 2101-19 Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Surgical face mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus.
- 16 CFR Part 1610 STANDARD FOR THE FLAMMABILITY OF CLOTHING TEXTILES
- ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks
- . During use, the Non Woven Face Mask will directly contact with user's skin, so we have it tested to demonstrate conformance to the following standards.
- o ISO 10993-5, Biological Evaluation Of Medical Devices -- Part 5: Tests For InVitro Cytotoxicity
- ISO 10993-10, Biological Evaluation Of Medical Devices Part 10: Tests O For Irritation And Skin Sensitization.
{5}------------------------------------------------
## 5. Summary of Comparison and Technological Characteristics:
## Table 1 - Elements of Comparison
| Elements of Comparison | Subject Device | Predicate Device (K202595) | Comments | |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|----------------------------------------------|
| PRODUCT CODE | FXX | FXX | SAME | |
| Regulation Number | 21 CFR 878.4040 | 21 CFR 878.4040 | SAME | |
| Class | II | II | SAME | |
| Product Name | 3 Layer Surgical Mask | Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical Face Mask | – | |
| GENERAL COMPARISON | | | | |
| Intended Use | When properly worn, the 3 Layer Surgical Masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate materials. This device is non sterile and for single use only. | The Premier Guard USA 3 Layer Ear Loop ASTM Level 3 Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile | SAME | |
| Composite | Flat Pleated, 3 Layers, ear loops | Flat Pleated, 3 Layers, ear loops | SAME | |
| Material<br>S | Outer Facing | Spun-Bond<br>Polypropylene | Spun-Bond<br>Polypropylene | SAME |
| | Middle | Melt Blown<br>Polypropylene<br>fiber | Melt Blown fiber | SAME |
| | Inner Facing | Spun-Bond<br>Polypropylene | Spun-Bond<br>Polypropylene | SAME |
| | Nose Piece | Polyethylene<br>coated annealed<br>carbon steel<br>wire | Polyethylene coated<br>annealed carbon<br>steel wire | SAME |
| | Ear Loops | Polyester,<br>Spandex | Nylon, Spandex | Different¹ |
| | Color | Blue, White | Blue | SIMILAR |
| | Dimensions (width) | 95.0mm (+/-<br>5.0mm) | 95.0mm | SAME |
| Dimensions (length) | | 175mm (+/-<br>5.0mm) | 175mm | SAME |
| Use | | Single Use | Single Use | SAME |
| Sterility | | Non-Sterile | Non-Sterile | SAME |
| ASTM F2100 Level | | Level 3 | Level 3 | SAME |
| Fluid Resistance Performance ASTM F<br>1862 | | Lot 1: 32 out of<br>32 pass at 160<br>mmHg<br>Lot 2: 32 out of<br>32 pass at 160<br>mmHg<br>Lot 3: 31 out of<br>32 pass at 160<br>mmHg | Pass at 160mmHg | Both Predicate<br>and Subject<br>Device PASS |
| Particulate Filtration Efficiency ASTM F<br>2299 | | Lot 1: 99.80%<br>Lot 2: 99.83%<br>Lot 3: 99.87% | ≥ 98% | Both Predicate<br>and Subject<br>Device PASS |
| | | | | |
| Bacterial Filtration Efficiency ASTM F 2101 | Lot 1: 99.8%<br>Lot 2: 99.8%<br>Lot 3: 99.8% | $\ge 98%$ | Both Predicate<br>and Subject<br>Device PASS | |
| Differential Pressure (Delta P) EN<br>14683:2019+AC : 2019 | Lot 1: 4.8 mmHg<br>Lot 2: 4.7 mmHg<br>Lot 3: 4.6 mmHg | < 6mmH20/cm² | Both Predicate<br>and Subject<br>Device PASS | |
| Flammability 16CFR 1610 | Class 1, Non-<br>Flammable | Class 1, Non-<br>Flammable | Both Predicate<br>and Subject<br>Device PASS | |
| Cytotoxicity | Comply with ISO<br>10993-5 non-<br>cytotoxic | Comply with ISO<br>10993-5 non-<br>cytotoxic | SAME | |
| Irritation | Comply with ISO<br>10993-10 non-<br>irritating | Comply with ISO<br>10993-10 non-<br>irritating | SAME | |
| Sensitization | Comply with ISO<br>10993-10 non-<br>sensitizing | Comply with ISO<br>10993-10 non-<br>sensitizing | SAME | |
{6}------------------------------------------------
{7}------------------------------------------------
*Notes 1 - Different material used in the ear loops of the masks. Subject Device was tested for & passed biocompatibility and performance testing. This material difference shows no effect on performance or safety of Subject Device as compared to Predicate Device.
{8}------------------------------------------------
### 6. Non-clinical Tests Performed on the Subject Device
The proposed subject device was tested and conformed to the following standards and recommendations in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission.
### Table 2 - Performance Testing
| Test Methodology | Purpose | Subject Device | Acceptance Criteria | Result |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------|
| Fluid Resistance<br>Performance<br>(ASTM F 1862) | This test method is used to<br>evaluate the resistance of<br>medical face masks by the<br>impact of a small volume<br>(approx 2ml) of high velocity<br>stream of synthetic blood.<br>Pass/Fail determinations are<br>based on visual detection of<br>synthetic blood penetration. | Lot 1: 32 out of 32<br>pass at 160 mmHg<br>Lot 2: 32 out of 32<br>pass at 160 mmHg<br>Lot 3: 31 out of 32<br>pass at 160 mmHg | 29 out of 32 pass at<br>120 mmHg | PASS |
| Particulate<br>Filtration<br>Effeciency (ASTM F<br>2299) | The purpose of this test is to<br>measure the initial particle<br>filtration effeciency of<br>materials using<br>monodispersed aerosols<br>containing suspended latex<br>sphere particulates of 0.1 um<br>diameter. | Lot 1: 99.80%<br>Lot 2: 99.83%<br>Lot 3: 99.87% | ≥ 98% PFE | PASS |
| Bacterial Filtration<br>Effeciency (ASTM F<br>2101) | The purpose of this test is to<br>measure the Bacterial<br>Filtration Effeciency of the<br>mask as specified in ASTM F<br>2101. | Lot 1: 99.8%<br>Lot 2: 99.8%<br>Lot 3: 99.8% | > 98% BFE | PASS |
| Flammability 16<br>CFR 1610 | The purpose of this test<br>method is to determine the<br>flammability characteristics of<br>the mask as specified in 16<br>CFR 1610. Materials in the<br>construction of face masks<br>shall meet the requirements<br>of Class 1, Normal<br>Flammability specified by 16<br>CFR part 1610. | Lot 1: DNI<br>Lot 2: DNI<br>Lot 3: DNI<br>All 3 Lots met Class 1<br>Flammability<br>standards | Class 1 | PASS |
| Differential<br>Pressure EN<br>14683:2019 | The purpose of this test<br>method is to determine the<br>overall breathability through<br>the mask barrier in terms of<br>mmH2O/cm³. The mask<br>surface is subjected to<br>constant air flow of 8.0 L/min<br>and the pressure differential<br>is measured. | Lot 1: 4.8mmHg<br>Lot 2: 4.7mmHg<br>Lot 3: 4.6mmHg | < 6.0 mmH20/cm² | PASS |
| Test Methodology | Purpose | Subject Device | Acceptance<br>Criteria | Result |
| Cytotoxicity<br>(ISO 10993-5) | The purpose of the test is to<br>determine the biological<br>reactivity of a mammalian cell<br>culture in response to the test<br>article. | Proposed Device<br>scored zero (0) for<br>all reactivity. Non-<br>cytotoxic. | Subject Article must<br>score reactivity results<br>not greater than 2, or<br>a mild reactivity. | PASS |
| Irritation<br>(ISO 10993-10) | To evaluate the potential skin<br>irritation caused by the<br>extraction of the test article<br>extract contacting with the skin<br>surface of rabbits. | All Scores were<br>Zero(0). Proposed<br>Device was<br>determined to be<br>Non-irritant. | Test article<br>subjected skin areas<br>are graded based on<br>reactivity after 24,<br>48, & 72 hours. | PASS |
| Sensitization<br>(ISO 10993-10) | To evaluate the potential test<br>article extracts to cause skin<br>sensitization in the guinea pig<br>according to GPMT Method. | All Scores were<br>Zero(0). Proposed<br>Device was<br>determined to be<br>Non-sensitizer. | Test article<br>subjected skin areas<br>are graded based on<br>reactivity after 24 &<br>48 hours. | PASS |
{9}------------------------------------------------
{10}------------------------------------------------
#### Table 3 - Biocompatibility Testing
#### 7. Summary of Clinical Testing
There were no clinical studies performed or included in this submission.
#### 8. Conclusion
The conclusion drawn from the nonclinical tests and substantial equivalence comparison with predicate device demonstrates that the subject device in this 510(k) submission, 3 Layer Surgical Mask, is as safe, effective, and performs as well, or better than the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.