K210643 · Shandong Zhushi Pharmaceutical Group Co., Ltd. · FXX · Aug 4, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210643
Device Name
Disposable Surgical Face Mask
Applicant
Shandong Zhushi Pharmaceutical Group Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Aug 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
The Disposable Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
Device Story
Disposable surgical face mask; three-layer, flat-pleated design; ear-loop attachment; malleable polyethylene nose piece. Used in clinical settings by healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material. Single-use; non-sterile. Provides physical barrier protection; performance verified via fluid resistance, filtration efficiency, and flammability testing.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM F2101-19 (BFE ≥98%), ASTM F2299-03 (PFE ≥98%), ASTM F1862M-17 (fluid resistance at 160 mmHg), EN 14683:2019 (differential pressure <6 mmH2O/cm²), and 16 CFR Part 1610 (flammability Class 1). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10 (non-cytotoxic, non-irritating, non-sensitizing).
Technological Characteristics
Three-layer flat-pleated mask. Inner/outer layers: spun-bond polypropylene; middle layer: melt-blown polypropylene. Ear-loops: spandex and polyester. Nose piece: malleable polyethylene wire. Dimensions: 17.5cm x 9.5cm, 17.5cm x 9.0cm, 17.0cm x 9.0cm, 14.5cm x 9.5cm. Non-sterile, single-use. Tested per ASTM F2100-19 standards.
Indications for Use
Indicated for healthcare workers and patients as a protective nose and mouth covering during medical and surgical procedures where there is a risk of exposure to microorganisms and body fluids.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K210433 — Surgical Face Mask (Ear loops and Tie-on) · Wuhan Dymex Healthcare Co., Ltd. · Jun 7, 2021
K202580 — Disposable Surgical Mask · Qingdao Likangyuan Medical Device Co., Ltd. · Jun 7, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 4, 2021
Shandong Zhushi Pharmaceutical Group Co., Ltd % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K210643
Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 21, 2021 Received: June 28, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210643
Device Name Disposable Surgical Face Mask
#### Indications for Use (Describe)
The Disposable Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 201 Subpart D) | Over-The-Counter Use (21 CFR 201 Subpart C) |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
510k number: K210643 Date: July 26, 2021 This summary of 510(k) is being submitted in accordance with requirements of 21 CFR 807.92.
#### 1.0 Submitter's information
Name: Shandong Zhushi Pharmaceutical Group Co., Ltd Address: No.6 Shande Road, Shan County, Heze City, Shandong, China Contact: Mr. Junhui Zhu Phone Number: 86-530-7150111 Fax number: 86-530-7150111 Date of Preparation: Jul.26, 2021 Email: 2307426957@gq.com
#### Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Subject Device information
Type of 510(k): Traditional Trade name: Disposable Surgical Face Mask Common name: Surgical mask Classification name: Mask, Surgical Model(s): ZS-B, ear-loop.
#### 3.0 Classification
Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Panel: Surgical apparel
#### 4.0 Predicate device information
Manufacturer: Guangdong Haiou Medical Apparatus Co., Ltd.
{4}------------------------------------------------
Device: Disposable Medical Surgical Face Mask 510(k) number: K203200 Classification name: Mask, Surgical Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Panel: Surgical apparel
# 5.0 Device description
The Disposable Surgical Face Mask is single use, three-layer, flat-pleated style with ear-loop and nose piece. The mask is manufactured with three layers, the inner and outer layers, and the ear-loop is made of spandex and polyester, and the middle layer is made of melt blown fabrics. During use, the ear-loop shall be tied over user ear. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polyethylene wire. The Disposable Surgical Face Mask is provided in blue color. The mask is sold non-sterile and are intended to be single use, disposable devices.
# 6.0 Indication for Use Statement
The Disposable Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
### 7.0 Technological Characteristic Comparison
Provided below is a comparison of the proposed device with the predicate device.
| Item | Proposed device | Predicated device | Remark |
|----------------|-------------------------------|---------------------------------------|-----------|
| Product Code | FXX | FXX | Same |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Product name | Disposable Surgical Face Mask | Disposable Medical Surgical Face Mask | Same |
| 510(k) No. | K210643 | K203200 | Different |
| Models | ZS-B, ear-loop. | HO-KZ01 | Different |
| Composite | Flat Pleated, 3 layers | Ear loop, flat pleated, 3 layers | Same |
| Specification | 17.5cm×9.5cm<br>17.5cm×9.0cm | 17.5cm×9.5cm | Different |
| | Table 3 - General Comparison |
|--|------------------------------|
| | |
{5}------------------------------------------------
| | | 17.0cm×9.0cm | | | |
|-----------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------|
| | | 14.5cm×9.5cm | | | |
| | Intended Use | The Disposable Surgical<br>Face Mask is indicated as a<br>protective nose and mouth<br>covering for healthcare<br>workers and patients involved<br>in medical and surgical<br>procedures. The masks are<br>indicated in any procedure or<br>situation where there is a risk<br>of exposure to<br>microorganisms and body<br>fluids. | Disposable Medical Surgical<br>Face Mask is indicated as a<br>protective nose and mouth<br>covering for healthcare<br>workers and patients involved<br>in medical and surgical<br>procedures. The masks are<br>indicated in any procedure or<br>situation where there is a risk<br>of exposure to<br>microorganisms and body<br>fluids. | Same | |
| | | OTC use | Yes | Yes | Same |
| | | Internal layer | Spun-bond polypropylene | Spun-bond polypropylene | Same |
| | | Middle layer | Melt blown polypropylene | Melt blown polypropylene | Same |
| | Mat<br>erial | External layer | Spun-bond polypropylene | Spun-bond polypropylene | Same |
| | | Nose piece | Malleable polyethylene<br>wire | Malleable polyethylene<br>with aluminum wire | Same |
| | Ear-loop | | spandex +polyester | polyester | Similar |
| Color | | Blue | Blue | Same | |
| | | 17.5cm×9.5cm, ±5% | | | |
| | | 17.5cm×9.0cm, ±5% | W: 17.5cm | Different | |
| | Dimension (W×L) | 17.0cm×9.0cm, ±5% | L: 9.5cm | | |
| | | 14.5cm×9.5cm, ±5% | | | |
| Sterility | | Non-Sterile | Non-Sterile | Same | |
| | Single Use | Yes | Yes | Same | |
| | Sterile | No | No | Same | |
| | Fluid Resistance<br>Performance | 31/32 Passed at 160mmHg<br>Level 3 ASTM F1862-17 | Level 3: 160 mmHg | Same | |
| | Particulate Filtration<br>Efficiency | 32/32 Passed ≥98%<br>ASTMF2299-03 | ≥98% | Same | |
| | Bacterial Filtration<br>Efficiency | 32/32 Passed at ≥98%<br>ASTM F2101-19 | ≥98% | Same | |
| Differential Pressure | | 32/32 Passed at <6<br>mmH2O/cm2<br>EN 14683: 2019, Annex C | <6.0 mmH2O/cm2 | Similar | |
| Flammability | | 32/32 Passed ≥ Second's<br>burn Time-Class 1 16 CFR<br>Part 1610 | Class 1, Non-Flammable | Same | |
| Cytotoxicity | Under the conditions of the<br>study, the subject device<br>was non-cytotoxic | Under the conditions of the<br>study, the predicate device<br>was non-cytotoxic | Same | | |
| Irritation | Under the conditions of the<br>study, the subject device was<br>non-irritating | Under the conditions of the<br>study, the predicate device<br>was non-irritating | Same | | |
| Sensitization | Under the conditions of the<br>study, the subject device was<br>non-sensitization | Under the conditions of the<br>study, the predicate device<br>was non-sensitization | Same | | |
{6}------------------------------------------------
# 8. Summary of Non-Clinical Performance Testing
#### Performance Testing Summary
| Test Method | Purpose | Pass Criteria | Results |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|-------------------------------------------------------------------------------------------------|
| ASTM F2101-19Standard<br>TestMethod for<br>Evaluating the Bacterial<br>Filtration | The purpose of the test is to<br>evaluate the Bacterial filtration<br>efficiency (BFE) (%) | ≥98% | 3 lots tested with<br>total 96 samples,<br>94/96 Passed at<br>≥98% /Pass |
| EN 14683: 2019, Annex<br>C Medical face masks -<br>Requirements and<br>test methods according to<br>ASTM F2100:2019 | The purpose of the test is to<br>evaluate the Different pressure<br>(Delta-P) | <6.0<br>mmH2O/cm² | 3 lots tested with<br>total 96 samples,<br>92/96 Passed<br><6 mmH2O/cm²<br>/ Pass |
| ASTM F2299-03<br>Standard Test Method<br>for Determining the Initial<br>Efficiency of Materials<br>Used in Medical Face<br>Masks to Penetration by<br>Particulates Using Latex<br>Spheres according to<br>ASTMF2100:2019 | The purpose of the test is to<br>evaluate the Sub-micron<br>particulate filtration efficiency<br>at 0.1 micron, % (PFE) | ≥98% | 3 lots tested with<br>total 96 samples,<br>96/96 Passed at<br>≥98% / Pass |
| ASTM F1862M-17<br>Standard Test Method for<br>Resistance of Medical<br>Face Masks to<br>Penetration by Synthetic<br>Blood (Horizontal<br>Projection of Fixed<br>Volume at a Known<br>Velocity) according to<br>ASTMF2100:2019for<br>pass result | The purpose of the test is to<br>evaluate the Resistance to<br>penetration by synthetic blood,<br>Minimum pressure in mmHg | Fluid<br>resistant<br>claimed at<br>160 mm Hg | 3 lots tested with<br>total 96 samples,<br>95 of 96 test articles<br>passed at 160mmHg<br>/Pass |
{7}------------------------------------------------
| 16 CFR Part 1610 Standard<br>for the Flammability of<br>Clothing according to ASTM<br>F2100:2019 | The purpose of the test is to<br>evaluate the Flame spread | Class 1 | 3 lots tested with<br>total 96 samples,<br>96/96 Passed ≥3<br>seconds burn Time-<br>Class 1 /<br>Pass |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------|
- Biocompatibility Testing According to ISO 10993-1:2009, the nature of body contact ● for the subject device is Surface Device category, Skin Contact and duration of contact is A-Limited (≤24h). The following tests for the subject device were conducted to demonstrate that the subject device is biocompatible and safe for its intended use:
{8}------------------------------------------------
1) In vitro Cytotoxicity Test per ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity,
2) Skin Sensitization Tests per ISO 10993-10:2010 Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitization,
3) Skin Irritation Tests per ISO 10993-10:2010 Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitization.
#### 9. Summary of Clinical Performance Test
No clinical study is included in this submission
#### 10.0 Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K210643, the Disposable Surgical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K203200.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.