WELLMIEN SURGICAL MASK

K101000 · Wellmien (Suzhou) Import & Export Trading Co., Ltd. · FXX · Aug 9, 2010 · General, Plastic Surgery

Device Facts

Record IDK101000
Device NameWELLMIEN SURGICAL MASK
ApplicantWellmien (Suzhou) Import & Export Trading Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateAug 9, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The Wellmien Surgical Masks (for single use only) are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism, body fluid, and particulate aerosol transfer.

Device Story

Pleated 3-ply surgical mask; single-use. Construction: outer/inner layers 100% spun-bond polypropylene; middle layer 100% meltblown polypropylene filter media; soft elastic ear-loops; malleable aluminum nose piece. Used in clinical settings by healthcare workers and patients to provide a physical barrier against microorganisms, body fluids, and aerosols. Device functions as a passive protective barrier; no active components or software.

Clinical Evidence

No clinical data. Bench testing only: Bacterial Filtration Efficiency (ASTM F2101-07), Particle Filtration Efficiency (ASTM F2299-03), Differential Pressure, Flammability (16 CFR 1610), Biocompatibility (ISO 10993), and Fluid Resistance (ASTM F1682-07).

Technological Characteristics

3-ply pleated mask. Materials: 100% spun-bond polypropylene (outer/inner), 100% meltblown polypropylene (middle). Malleable aluminum nose piece. Elastic ear-loops. Standards: ASTM F2101-07 (BFE), ASTM F2299-03 (PFE), 16 CFR 1610 (flammability), ISO 10993 (biocompatibility), ASTM F1682-07 (fluid resistance).

Indications for Use

Indicated for health care workers and patients in medical and surgical procedures to prevent transfer of microorganisms, body fluids, and particulate aerosols.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K101000 #### Section 5: 510(k) Summary: This summary of 510k safety and effectiveness information is being submitted in accordance with the requirements of 21CFR 807.92 #### 1. Contact Information AUG - 5 ZUIU Submitter & Foreign Manufacturer Identification Wellmien (Suzhou) Imp. & Exp. Trading Co., Ltd 1-711, No.636 Ganjiang East Road Suzhou, Jiangsu Province, China, Zip code: 215005 Tel: 86-512-81881706 ### Submitter's FDA Registration Number : N/A US Agent and Contact Person Chengyu Shen Manton Business and Technology Services 5 Carey Street, Pennington, NJ 08534 Tel: 608-217-9358, Email: cyshen@aol.com Date of Summary: March 23, 2010 Date of Revision 1: June 5, 2010 #### 2. Regulatory Information Name of Device: Surgical Mask (For Single Use Only) Panel: General Hospital Class: Class 2 Regulation Number: 21 CFR 807.4040 Product Code: FXX Classification Name: Mask, Surgical #### Predicate Device Information: 3. K060776: Surgical Face Mask (Yellow, White, Green, Pink) by Hong Ya Non-Woven Products Co., Ltd #### 4. Device description: WELLMIEN's surgical masks (for single use only) are pleated 3-ply mask. Inner and outer layers are made of 100% spun-bond polypropylene. Middle player is made of 100% meltblown polypropylene filter media。 Ear-loops are made of soft elastic loops. The nose piece for all WELLMIEN's face masks is malleable aluminum wire. All of the materials used in the construction of the WELLMIEN's surgical masks are being used in currently marketed devices. (see Predicate Device information ) {1}------------------------------------------------ #### 5. Intended Use: The Wellmien Surgical Masks (for single use only) are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism, body fluid, and particulate aerosol transfer. #### 6. Comparison to Predicate Devices WELLMIEN Surgical Mask (for single use only) is compared with it Predicate Device: K060776: Surgical Face Mask (Yellow, White, Green, Pink) by Hong Ya Non-Woven Products Co., Ltd. The design, material, and performance of both devices are very similar, and the two products are substantially equivalent in safety and effectiveness. | Description | Our Device | Predicate Device (K060776) | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------| | Indication for<br>Use | Nose and mouth covering for health<br>care workers and patients to prevent<br>microorganism, body fluid, and<br>particulate aerosol transfer. | Same | | Layers | Three | Three | | Materials | Outer layer is made of 100% spun-bond polypropylene. | Outer layer is made of 100% spun-bond polypropylene. | | | Middle player is made of 100% meltblown polypropylene filter media. | Middle player is made of 100% meltblown polypropylene filter media. | | | Inner layer is made of 100% spun-bond polypropylene. | Inner layer is made of 100% spun-bond-polypropylene. | | | Ear-loops are made of soft elastic loops.<br>The nose piece is malleable aluminum wire | Ear-loops are made of soft elastic loops.<br>The nose piece is malleable aluminum wire | | Dimensions | Large: 18x9cm, Medium:<br>17.5x9.5cm, Small: 14.5x9 cm | N/A | | Mask style | Flat Pleated | Flat Pleated | | Design | Ear Loop | Ear Loop | | Color | Blue | Yellow, White, Green, Pink | | NIOSH<br>certification | N/A | N/A | Comparison of Intended Use, Design, Material, and Specifications {2}------------------------------------------------ | Comparison of Device Performance | | | |------------------------------------|------------------------|-----------------------------| | Test | Our Device | Predicate Device (K060776) | | Fluid Resistance<br>Performance | Fluid Resistance | Fluid Resistance | | Particle Filtration<br>Efficiency | 99.8% | 96.8% | | Bacteria Filtration<br>Efficiency | 99.9% | At least 99.9% | | Differential<br>Pressure (Delta-P) | 3.7-4.0 (mm Water/cm²) | Average 2.34 (mm Water/cm²) | | Flammability | Class 1 None Flammable | Class 1 None Flammable | | Biocompatibility | Biocompatible | Biocompatible | #### 7. Discussion of Non-Clinical Tests Performed to Determine Substantial Equivalence The following non-clinical tests were performed to determine substantial equivalence. Tests were conducted following the recommended procedures outlined in the respective consensus standards. Test results met all relevant requirements in the test standards, and are comparable to the predicate device. - Bacterial Filtration Efficiency (BFE) : ASTM F2101-07 (1) - Pressure Differential (Delta P) (2) - (3) Latex Particle Challenge(PFE): ASTM F2299-03 - Flammability: 16 CFR 1610 (4) - Biocompatibility per ISO 10933 (ર) - Fluid Resistant Synthetic Blood Penetration Resistant Test: ASTM (6) F1682-07 More detailed comparison of the design, technical, and performance characteristics to the predicate device are summarized in Section 7: Summary Report. More details of non-clinical tests are summarized in Section 7.5 and 7.6. #### 8. Discussion of Clinical Tests Performed Not applicable #### 9. Conclusions The WELLMIEN's surgical masks (for single use only) have the same intended use and technological characteristics as the predicate devices. {3}------------------------------------------------ Moreover, bench testing contained in this submission demonstrates that the technological characteristics do not raise any new questions of safety or effectiveness。 WELLMIEN's surgical masks (for single use only) are substantially equivalent to the predicate device : {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Wellmien (Suzhou) Import and Export Company, Limited C/O Mr. Chengyu Shen Manton Business and Technology Services 5 Carey Street Pennington, New Jersey 08534 Re: K101000 Trade/Device Name: Wellmien Surgical Mask (for single use only) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: June 2, 2010 Received: June 25, 2010 Dear Mr. Shen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. AUG - 9 2010 {5}------------------------------------------------ ## Page 2- Mr. Shen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting . ___ . (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Susan Ruines Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## 01000 ## Indications for Use 510(k) Number (if known): K101000 AUG - 9 2010 Device Name: Wellmien Surgical Mask (for single use only) Indications for Use: The Wellmien Surgical Masks (for single use only) are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism, body fluid, and particulate aerosol transfer. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Eliz. H. F. Llanetto Wills (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Division of Anesthesiology, General Douises Division Control, Dental Devices 01 000 510(k) Number:
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...