FLORASEAL, MODEL FS350D

K083354 · Adhezion Biomedical · NZP · Jul 20, 2009 · General, Plastic Surgery

Device Facts

Record IDK083354
Device NameFLORASEAL, MODEL FS350D
ApplicantAdhezion Biomedical
Product CodeNZP · General, Plastic Surgery
Decision DateJul 20, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.4370
Device ClassClass 2
AttributesTherapeutic

Intended Use

FLORASEAL™ Microbial Sealant is Intended for use after topical operative skin preparations, with standard surgical draping, and prior to a surgical incision. The product is used to reduce the risk of skin flora contamination throughout a surgical procedure.

Device Story

FloraSeal is a film-forming, cyanoacrylate-based microbial sealant. Provided in a ready-to-use applicator consisting of a thermoformed blister tray with an attached sponge tip. Applied topically to the skin by clinical staff after standard skin preparation and before surgical incision. The cyanoacrylate polymerizes to form a protective film that immobilizes indigenous skin flora, reducing the risk of contamination throughout the surgical procedure. The device is used in clinical/surgical settings. Benefits include sustained reduction of microbial colonization (99.9% reduction maintained for 24 hours).

Clinical Evidence

Clinical study involved volunteer subjects with a 14-day washout period. FloraSeal and Povidone-Iodine were applied to the right inguinal region. Bacterial immobilization was evaluated at 15 minutes, 4 hours, and 24 hours. Results demonstrated a 99.9% reduction in microbial colonization within 15 minutes, maintained for 24 hours.

Technological Characteristics

Cyanoacrylate-based film-forming sealant. Applicator consists of a thermoformed blister tray with a sponge tip, packaged in a Tyvek pouch. Evaluated for MVTR, setting time, surface coverage, hydrostatic pressure, and water impact penetration. Biocompatible.

Indications for Use

Indicated for use on patients undergoing surgical procedures to reduce the risk of skin flora contamination. Applied after topical skin preparation and before incision, in conjunction with standard surgical draping.

Regulatory Classification

Identification

A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.

Special Controls

*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083354 ### 510 (K) SUMMARY . JUL 20 2009 | Submitted by: | Adhezion Biomedical LLC<br>One Meridian Blvd<br>Suite 1802<br>Wyomissing, PA 19610<br>Phone: 484-334-2929 | | | |--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Richard G. Jones<br>Sr. Vice President<br>Phone: 610-431-2398<br>Email: rjones@adhezion.com | | | | Date of Summary: | July 14, 2009 | | | | Device Trade Name: | FloraSeal Microbial Sealant | | | | Common or Used Name: | Microbial Sealant | | | | Classification Name: | Surgical Drape and Drape Accessory<br>Product Code: NZP | | | | Predicate Device: | Integuseal Microbial Sealant | | | | Device Description: | Flora Seal Microbial Sealant is a film forming, cyanoacrylate based microbial sealant provided in a ready to<br>use applicator. Each applicator consists of a thermoformed blister tray with a heat sealed lid with an<br>attached applicator sponge tip. This applicator tray with sponge tip is contained in an outer Tyvek pouch. | | | | Indications for Use | <table><tr><td>FLORASEAL™</td><td>Microbial Sealant is Intended for use after topical operative skin<br/>preparations, with standard surgical draping, and prior to a surgical<br/>incision. The product is used to reduce the risk of skin flora<br/>contamination throughout a surgical procedure.</td></tr></table> | FLORASEAL™ | Microbial Sealant is Intended for use after topical operative skin<br>preparations, with standard surgical draping, and prior to a surgical<br>incision. The product is used to reduce the risk of skin flora<br>contamination throughout a surgical procedure. | | FLORASEAL™ | Microbial Sealant is Intended for use after topical operative skin<br>preparations, with standard surgical draping, and prior to a surgical<br>incision. The product is used to reduce the risk of skin flora<br>contamination throughout a surgical procedure. | | | | Substantial Equivalence: | FloraSeal is substantially equivalent to InteguSeal. This is based on comparable dispensing application<br>design, same indication for use, bench, in vitro bacterial immobilization studies as well as clinical study, plus<br>animal and clinical studies for both efficiency and safety including biocompatibility.<br><br>The clinical test included volunteer subjects who participated in a washout period of 14 days prior to the<br>evaluation for stabilization of skin bacterial flora. The test site was the right inguinal region where FloraSeal<br>and Povidone - lodine were applied and evaluated for indigenous bacterial immobilization at 15 minutes, 4<br>and 24 hours. The treatments reduced microbial colonization by 99.9% within 15 minutes of application and<br>maintained this reduction for 24 hours. | | | | Other Testing: | Bench test were preformed with Integuseal as control and showed FloraSeal to provide results that were<br>equal to or better than Integuseal. These test included effect on wound closure strength, film integrity over<br>time, flexibility properties (FloraSeal significantly better), MVTR, setting time, surface coverage, hydrostatic<br>pressure and water impact penetration. | | | {1}------------------------------------------------ The following criteria have been met for the releasable 510(k) Summary ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ - . The summary includes only the information that is also covered in the body of the 510(k) - The summary does not contain any puffery or unsubstantiated labeling claims. . - . The summary does not contain any raw data, i.e., contains only summary data. - The summary does not contain any trade secret or confidential commercial information. . - The summary does not contain any patient identification information. . {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three talons, representing service to the people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Richard G. Jones Senior Vice President Regulatory Affairs, Ouality Assurance and Clinical Affairs Adhezion Biomedical LLC One Meridian Boulevard, Suite 1B02 Wyomissing, Pennsylvania 19610 ## JUL 2 0 2009 Re: K083354 Trade/Device Name: FloraSeal Microbial Sealant Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: II Product Code: NZP Dated: June 18, 2009 Received: June 18, 2009 Dear Mr. Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Jones Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Anthony D. Weston for. Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K083354 KU8554 Device Name: Flora Seal Microbial Sealant Indications For Use: Flora Seal™ Microbial Sealant is intended for use after topical operative skin preparations, with standard surgical draping, and prior to a surgical incision. The product is used to reduce the risk of the skin flora contamination throughout a surgical procedure. Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shala H. Margolis, MD (Division Sign-Off) (Division Sign-Off) Division of Anesthesiology, General Hospital Division of Anesthesiology, General Pouises Division of Antoch, Dental Devices Page 1 of 510(k) Number. K083354
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...