PROTECTOR RETRACTOR & PROTECTOR RETRACTOR WITH DRAPE
Device Facts
| Record ID | K954824 |
|---|---|
| Device Name | PROTECTOR RETRACTOR & PROTECTOR RETRACTOR WITH DRAPE |
| Applicant | Medical Creative Technologies, Inc. |
| Product Code | KKX · General, Plastic Surgery |
| Decision Date | Apr 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4370 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
This device is to be used to protect and retract an incision in the abdominal wall. This is accomplished by placing one o-ring through the incision into the peritoneum cavity while the other o-ring remains outside of the incision. The outside o-ring is then rolled down against the skin. This lines the wound to protect against contamination and holds the incision open.
Device Story
Sterile, disposable abdominal wound retractor; consists of open-ended cylinder with flexible o-ring at each end; optional integral drape. Device functions by inserting one o-ring into peritoneal cavity and rolling the external o-ring against skin to line wound, prevent contamination, and maintain incision opening. Used in surgical settings by physicians. Replaces combined use of separate wound edge protector drapes and stainless steel self-retaining retractors. Simplifies surgical field by eliminating bulky metal retractor assemblies.
Clinical Evidence
Bench testing on simulated abdominal wall and pig studies verified wound protection and retraction capabilities. Clinical evidence includes a pilot study (n=41) and ongoing pivotal clinical study as part of the DEXTERITY® Pneumo Sleeve system. Results indicate safe and effective performance for wound protection and retraction across four available sizes.
Technological Characteristics
Sterile, single-use disposable device. Construction: flexible o-rings and cylinder (material type unspecified). Form factor: four sizes to accommodate varying incision dimensions. No energy source, software, or connectivity.
Indications for Use
Indicated for patients undergoing abdominal surgery requiring incision protection and retraction.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
- Wound edge protector drape
- Self-retaining retractor
Reference Devices
- DEXTERITY® Pneumo Sleeve
Related Devices
- K020435 — WOUND PROTECTOR/RETRACTOR · Applied Medical Resources Corp. · Apr 15, 2002
- K041711 — ALEXIS WOUND RETRACTOR, MODELS C8312, C8301, C8302, C8303, C8304 · Applied Medical Resources Corp. · Aug 26, 2004
- K140064 — SURGISLEEVE WOUND PROTECTOR · Covidien · Mar 12, 2014
- K050256 — APPLE MEDICAL/OB MOBIUS ELASTIC RETRACTOR · Apple Medical Corp. · Apr 26, 2005