← Product Code [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO) · K231973

# PolyIsoprene Powder Free Black Colour Radiation Attenuating Surgical Glove (K231973)

_Ineo Tech Sdn Bhd · KGO · Sep 29, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K231973

## Device Facts

- **Applicant:** Ineo Tech Sdn Bhd
- **Product Code:** [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO.md)
- **Decision Date:** Sep 29, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Intended Use

The polyisoprene powder free, black colour surgical gloves are radiation attenuating and intended to be worn by operating room personal to protect a surgical wound from contamination

## Device Story

Disposable, sterile, powder-free polyisoprene surgical glove; contains tungsten-based particles for radiation attenuation. Worn by operating room personnel during fluoroscopic/radiographic procedures. Protects surgical wounds from contamination; reduces operator hand exposure to ionizing scattered/secondary radiation. Available in two thickness models. Used in surgical settings; manual operation by clinicians. Benefits include barrier protection and reduced radiation dose to hands.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM standards for physical dimensions, tensile strength, elongation, residual powder, and radiation attenuation (ASTM D7866). Biocompatibility confirmed via ISO 10993 (irritation, sensitization, systemic toxicity, cytotoxicity). Endotoxin and pyrogenicity testing performed per ASTM D7102 and USP standards.

## Technological Characteristics

Material: Polyisoprene with tungsten-based radiation-attenuating particles. Manufacturing: Latex dipping and sulphur vulcanization. Standards: ASTM D3577, ASTM D5151, ASTM D6124, ASTM D7866. Sterilization: Radiation (SAL 10^-6). Form factor: Disposable, powder-free, black, sizes 6-9. Two thickness models (Model 1: min 0.21mm; Model 2: min 0.29mm).

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Emerson & Co. S.r.l ([K192933](/device/K192933.md))

## Reference Devices

- [K022873](/device/K022873.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health and Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".

September 29, 2023

Ineo Tech Sdn Bhd Pierre Hoerner Company Director PT5825, Jalan Cassia Selatan 3/11 Taman Perindustrian Batu Kawan Bandar Kassia, Penang 14110 Malaysia

Re: K231973

Trade/Device Name: PolyIsoprene Powder Free Black Colour Radiation Attenuating Surgical Glove Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I. reserved Product Code: KGO, OPH Dated: June 25, 2023 Received: July 3, 2023

Dear Pierre Hoerner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231973

## Device Name

Polyisoprene Powder Free Black Colour Radiation Attenuating Surgical Glove Model 1: Min. 0.21mm (thickness) Model 2: Min. 0.29mm (thickness)

## Indications for Use (Describe)

The polyisoprene powder free, black colour surgical gloves are radiation attenuating and intended to be worn by operating room personal to protect a surgical wound from contamination

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K231973: 510(k) SUMMARY

[per 21 CFR 807.92]

#### 1. Submitter Information

| Name:<br>Address: | Ineo Tech Sdn. Bhd.<br>PT5825, Jalan Cassia Selatan 3/11 Taman Perindustrian<br>Batu Kawan, 14110, Bandar Cassia, Penang, Malaysia |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Phone:            | +33 (6) 70 21 95 45                                                                                                                |
| E-Mail:           | pierre@ineotech-my.com                                                                                                             |
| Contact Name:     | Dr Pierre Hoerner, Ph.D                                                                                                            |
| Date of summary:  | 27 Sept, 2023                                                                                                                      |

#### 2. Device Identification

| Trade name:          | Polyisoprene Powder Free Black Color Radiation Attenuating<br>Surgical Gloves |
|----------------------|-------------------------------------------------------------------------------|
| Classification name: | Surgeon's glove                                                               |
| Product Code:        | KGO (primary code); OPH                                                       |
| Regulation number:   | 21 CFR 878.4460                                                               |
| Device Class:        | Class I (general controls)                                                    |
| Assigned 510k number | K231973                                                                       |

#### 3. Identification of Predicate Devices

- . 510(k) number K192933 by Emerson & Co. S.r.1 is identified as the primary predicate device
- 510(k) number K022873 is considered as reference device to support specific testing or . device features.

#### 4. Device Description

The Polyisoprene Powder Free Black Color Radiation Attenuating Surgical Glove is a disposable sterile powder-free medical device made of polyisoprene that is intended to be

{4}------------------------------------------------

worn by operating room personnel to protect a surgical wound from contamination. The glove contains radiation-attenuating particles to reduce the exposure from harmful ionizing scattered or secondary radiation on the operator's hand during fluoroscopic procedures. The glove is proposed in two (2) options. The Model 2 are made with exactly the same material and composition and differ only by their thicknesses and therefore their attenuation properties.

#### ട. Indications for Use

The Polyisoprene Powder Free Black Color Surgical Gloves are radiation attenuating and intended to be worn by operating room personal to protect a surgical wound from contamination.

#### Primary Subject Device Reference Device Comparison Characteristic Predicate Device 510(k) number K231973 K192933 K022873 Regulation No. 21 CFR 878.4460 21 CFR 21 CFR 878.4460 Same regulation 878.4460 number KGO Product Code KGO (primary) KGO Same product code OPH Device Class Class I Class I Class I Same device class Device type Powder-free Powder-free Powder-free sterile surgical Same indicated use sterile surgical sterile surgical glove glove glove

#### 6. Comparison to Predicate Devices

{5}------------------------------------------------

| Indications for use statement | The polyisoprene powder free, black color surgical gloves are radiation attenuating and intended to be worn by operating room personnel to protect a surgical wound from contamination. | The polyisoprene powder free, black color surgical gloves are radiation attenuating and intended to be worn by operating room personnel to protect a surgical wound from contamination | A powder-free surgeon's glove is a device made of synthetic latex that may bear a trace amount of glove powder and is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. The radiation attenuating surgical glove is intended to be used during medical procedures where hands are necessarily exposed to radiation in order to offer some degree of protection to the hand from radiation. This includes surgical procedures that require the use of fluoroscopy or radiography | Similar                                                                                             |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Color                         | Black                                                                                                                                                                                   | Black                                                                                                                                                                                  | Black                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same color                                                                                          |
| Size Range                    | 6 to 9                                                                                                                                                                                  | 6 to 9                                                                                                                                                                                 | 6 to 9                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same size range                                                                                     |
| Use limit of the device       | Single use                                                                                                                                                                              | Single use                                                                                                                                                                             | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same use limit                                                                                      |
| Technology                    | Latex Dipping & Sulphur vulcanization                                                                                                                                                   | Latex Dipping & Sulphur vulcanization                                                                                                                                                  | Latex Dipping & Sulphur vulcanization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same dipping and vulcanization technology.                                                          |
| Elastomer                     | Non Natural Rubber Latex (PolyIsoprene)                                                                                                                                                 | Non Natural Rubber Latex (PolyIsoprene))                                                                                                                                               | Non Natural Rubber Latex (PolyChloroprene)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | All gloves are made with Non Natural Rubber Latex elastomer.                                        |
| RadiationAttenuation Material | (Tungsten based composition)                                                                                                                                                            | (Tungsten based composition)                                                                                                                                                           | (Tungsten based composition)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | All gloves are made without Lead and contain tungsten as Radiation-Attenuation Material composition |
| Standards met                 | ASTM D3577<br>ASTM D5151<br>ASTM D6124<br>ASTM D7866                                                                                                                                    | ASTM D3577<br>ASTM D5151<br>ASTM D6124<br>ASTMD7866                                                                                                                                    | ASTM D3577<br>ASTM D5151<br>ASTM D6124<br>ASTM D7866                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same performance standards met                                                                      |

{6}------------------------------------------------

| Sterilization                      | Sterile (SAL 106). Radiation. | Sterile (SAL 106). Ethylene Oxide | Sterile (SAL 10-6). Radiation | Sterilization process is Radiation or Ethylene Oxide, SAL 10-6 |
|------------------------------------|-------------------------------|-----------------------------------|-------------------------------|----------------------------------------------------------------|
| Biocompatibilty<br>See Table below | Meets requirements            | Meets requirements                | Meets requirements            | All met current requirements                                   |

#### 7. Summary of Non-Clinical Performance Testing

The Polyisoprene Powder Free Black Color Radiation Attenuating Surgical Glove possesses the following technological characteristics:

| Test Method | Test Purpose                    | Acceptable criteria                                                                                                                                                            | Results                                                                                                                                  |
|-------------|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D5151  | - Freedom from holes            | Criteria:<br>AQL 1.5                                                                                                                                                           | PASS<br>AQL=0.65                                                                                                                         |
| ASTM D3577  | Physical Dimensions<br>- Length | Criteria:<br>Size 6 to 9 ≥265mm<br>Size 6 to 9 ≥285mm                                                                                                                          | PASS<br>All sizes ≥285mm<br>(range:298-301mm)                                                                                            |
|             | - Width                         | Criteria:<br>size 6.0: 76 $±$ 6mm<br>size 6.5: 83 $±$ 6mm<br>size 7: 89 $±$ 6mm<br>size 7.5: 95 $±$ 6mm<br>size 8: 102 $±$ 6mm<br>size 8.5: 108 $±$ 6mm<br>size 9: 114 $±$ 6mm | PASS<br>83-84mm (size 6.5)<br>90-91mm (size 7)<br>96-97mm (size 7.5)<br>106-107mm (size 8)<br>110-113mm (size 8.5)<br>112-115mm (size 9) |

{7}------------------------------------------------

| Test Method                                          | Test Purpose                                                                                               | Acceptable criteria                                                                                    | Results                                                                                                                                                                                                                  |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                      | - Thickness                                                                                                | Criteria:<br>Min. 0.1mm                                                                                | PASS<br>Model 1:<br>Finger: 0.27-0.29mm<br>Palm: 0.24-0.25mm<br>Cuff: 0.23-0.24mm<br><br>Model 2:<br>Finger: 0.33-0.39mm<br>Palm: 0.33-0.35mm<br>Cuff: 0.31-0.33mm                                                       |
| ASTM D3577                                           | Physical Properties<br>- Tensile Strength                                                                  | Criteria:<br>- Before accelerated aging:<br>17 MPa min<br>- After accelerated aging:<br>12 MPa min     | PASS<br>18.7-19.6MPa<br><br>18.2-18.5MPa                                                                                                                                                                                 |
|                                                      | - Strength at 500%                                                                                         | Criteria<br>7.0 MPa max                                                                                | PASS<br>4.1-4.9MPa                                                                                                                                                                                                       |
|                                                      | - Ultimate Elongation                                                                                      | Criteria:<br>- Before aging: 650 % min<br>- After aging: 490 % min                                     | PASS<br>700-720%<br>620-645%                                                                                                                                                                                             |
| ASTM D6124                                           | - Residual powder                                                                                          | Criteria:<br><2mg / glove                                                                              | PASS<br>0.05-0.18mg/glove                                                                                                                                                                                                |
| Test Method                                          | Test Purpose                                                                                               | Acceptable criteria                                                                                    | Results                                                                                                                                                                                                                  |
| ASTM D7866                                           | Attenuation<br>requirements<br>(Measured in the<br>thinnest portion of the<br>glove, i.e cuff)             | Criteria:<br>60 kVp: 29%<br>80 kVp:22%<br>100 kVp:18%<br>120 kVp:15%                                   | PASS<br>Model 1<br>60kVp: 42.0-43.0%<br>80kVp: 34.4-35.0%<br>100kVp:28.9-29.4%<br>120kVp: 25.1-25.4%<br>Model 2:<br>60kVp:50.5-51.3%<br>80kVp: 42.0-42.7%<br>100kVp:36.2-36.7<br>120kVp: 31.2-31.6                       |
| Biocompatibility<br>ISO 10993-10, Skin<br>Irritation | - Determine if an<br>extract of the product<br>produces skin<br>Irritation                                 | - No skin irritation                                                                                   | PASS<br>Under the condition of the<br>study:<br>- No skin irritation was<br>observed. The skin<br>irritation response<br>category of both polar and<br>non-polar extract of the<br>test item in rabbit is<br>negligible. |
| ISO 10993-10, Dermal<br>Sensitization                | - Determine if an<br>extract of the product<br>produces Dermal<br>Sensitization after<br>repeated exposure | - No skin sensitization                                                                                | - No skin sensitization<br>was produced in guinea<br>pigs.                                                                                                                                                               |
| ISO 10993-11, Acute<br>Systemic Toxicity             | - Determine if an<br>extract of the product<br>causes systemic<br>toxicity in vivo                         | - Meet systemic toxicity<br>test                                                                       | - The tested glove meets<br>the requirement of acute<br>systemic toxicity test.                                                                                                                                          |
| Test Method                                          | Test Purpose                                                                                               | Acceptable criteria                                                                                    | Results                                                                                                                                                                                                                  |
| ISO 10993-5,<br>Cytotoxicity                         | - In vitro cytotoxicity<br>on extract                                                                      | - Comparable cytotoxicity<br>to predicate device                                                       | - Both devices exhibited a<br>slight reactivity at 50x<br>dilution concentration.                                                                                                                                        |
| ASTM D7102-22,<br>Endotoxin                          | - Determination of<br>bacterial endotoxin on<br>gloves                                                     | - Endotoxin level<br>according to<br>ASTMD7102-17 less than<br>20 EU/device on 3<br>production batches | - Endotoxin determination<br>performed according to<br>ASTMD7102-17 showed<br>12.9, 12.8 and 15.4                                                                                                                        |
| ISO 21582-21,<br>Pyrogenicity                        | - Pyrogen testing                                                                                          | - Non pyrogenic                                                                                        | - Extract of test item<br>evaluated for pyrogen test<br>is non pyrogenic and<br>meets the requirements of<br>pyrogenicity test as per<br>US Pharmacopeia                                                                 |

{8}------------------------------------------------

{9}------------------------------------------------

#### 8. Clinical Data

Clinical data is not needed for medical glove 510(k) submissions.

#### 9. Conclusion Statement

The conclusions drawn from the nonclinical tests demonstrate that the Polyisoprene Powder Free Black Color Radiation Attenuating Surgical Glove is as safe, as effective, and performs as well as or better than the legally marketed devices identified.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K231973](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K231973)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
