← Product Code [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO) · K222350

# Sterile Nitrile Surgical Gloves Powder Free (K222350)

_Jr Engineering & Medical Technologies (M) Sdn. Bhd. · KGO · Apr 13, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K222350

## Device Facts

- **Applicant:** Jr Engineering & Medical Technologies (M) Sdn. Bhd.
- **Product Code:** [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO.md)
- **Decision Date:** Apr 13, 2023
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Intended Use

A Sterile Nitrile Surgical Gloves Powder Free is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

## Device Story

Sterile, disposable synthetic rubber latex surgical gloves; intended for use by operating room personnel to protect surgical wounds from contamination. Device functions as a physical barrier. Available in sizes 6 through 9. Sterilized via Gamma Radiation to SAL 10^-6. Shelf life 3 years. Used in clinical/surgical settings. Protects patient by preventing contamination during procedures.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D5151 (freedom from holes), ASTM D3577 (tensile strength, elongation, dimensions), ASTM D6124 (residual powder), and ISO 10993 (biocompatibility: irritation, sensitization, cytotoxicity, systemic toxicity, pyrogenicity, bacterial endotoxin).

## Technological Characteristics

Material: Synthetic rubber latex (ASTM D3577-2019). Form factor: Surgical gloves, sizes 6-9. Sterilization: Gamma radiation (SAL 10^-6). Connectivity: None. Software: None. Physical properties: Tensile strength >17MPa (unaged), >12MPa (aged); Ultimate elongation >650% (unaged), >490% (aged).

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Sterile Nitrile Surgical Gloves, Powder Free ([K170515](/device/K170515.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

April 13, 2023

JR Engineering & Medical Technologies (M) SDN. BHD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504

Re: K222350

Trade/Device Name: Sterile Nitrile Surgical Gloves Powder Free Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: March 13, 2023 Received: March 13, 2023

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K22350

#### Device Name

Sterile Nitrile Surgical Gloves Powder Free

#### Indications for Use (Describe)

A Sterile Nitrile Surgical Gloves Powder Free is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label>           |
|------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510K SUMMARY

#### K22350

Date of Preparation: April 10, 2023 As required by: 21CFR § 807.92

#### A. APPLICANT INFORMATION

|    | Submitter Name                       | JR Engineering & Medical Technologies (M) SDN. BHD                                                                                            |
|----|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
|    | Address                              | Lot 8 &10, Jalan Zurah 3 & Lot 1&3, Jalan Zurah 3A/1, Pusat<br>Perindustrian 2, 44200 Rasa, Hulu<br>Selangor, Selangor Darul Ehsan, Malaysia. |
|    | Phone                                | +603-60572081                                                                                                                                 |
|    | Fax                                  | +603-60572181                                                                                                                                 |
|    | E-mail                               | ganeshjrmt@gmail.com                                                                                                                          |
|    | Contact Person                       | Mr. Ganesan Subramaniam                                                                                                                       |
|    | Designation                          | Managing Director                                                                                                                             |
|    | Contact Number                       | +6012 224 6677                                                                                                                                |
|    | Contact Email                        | ganeshjrmt@gmail.com                                                                                                                          |
|    | US Agent                             | Manoj Zacharias                                                                                                                               |
|    | US Agent Address                     | Liberty Management Group Ltd.<br>75 Executive Dr. Ste 114 Aurora, IL 60504 USA                                                                |
|    | US Agent Telephone                   | 1 (630) 270-2921                                                                                                                              |
|    | US Agent Fax                         | 1 (815) 986-2632                                                                                                                              |
|    | US Agent Email                       | manoj@libertymanagement.us                                                                                                                    |
| B. | Device Identification                |                                                                                                                                               |
|    | Common Name                          | Surgical Gloves                                                                                                                               |
|    | Device Name                          | Sterile Nitrile Surgical Gloves Powder Free                                                                                                   |
|    | Product Proprietary or<br>Trade Name | JR MEDIC                                                                                                                                      |
|    | Classification Name                  | Surgeon's Gloves                                                                                                                              |

1 KGO 21 CFR 878.4460

#### C. Predicate Device

Product Code

Device Classification

Regulation Number

510k Number Common Name Device Name Classification name Device Classification Product Code Regulation Number

K170515 Surgical Gloves Sterile Nitrile Surgical Gloves, Powder Free Surgeon's Gloves 1 KGO 21 CFR 878.4460

#### D. Description of the Device

The proposed device, Sterile Nitrile Surgical Gloves Powder Free is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.

{4}------------------------------------------------

The proposed device is made of synthetic rubber latex, as per standard ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves.

The classification is: Type II - gloves compounded from synthetic rubber latex.

The proposed device is Sterile Nitrile Surgical Gloves Powder Free, and is produced in sizes 6, 6 ½, 7, 7 ½, 8, 8 ½, and 9. All sizes share the same White color.

The proposed device is sterilized using Gamma Radiation method to achieve the Sterility Assurance Level (SAL) of 10 6 while packaged to maintain sterility. The shelf life is 3 years.

#### E. Technological Characteristics

| Characteristic                                                                                    | Subject device K222350                                                                                                                                                                            | Predicate device K170515                                                                                                                                                                                                                                                          | Remarks   |
|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code                                                                                      | KGO                                                                                                                                                                                               | KGO                                                                                                                                                                                                                                                                               | Same      |
| Regulation No.                                                                                    | 21 CFR 878.4460                                                                                                                                                                                   | 21 CFR 878.4460                                                                                                                                                                                                                                                                   | Same      |
| Class                                                                                             | 1                                                                                                                                                                                                 | 1                                                                                                                                                                                                                                                                                 | Same      |
| Intended Use                                                                                      | A Sterile Nitrile Surgical Gloves<br>Powder Free is a device made of<br>synthetic rubber intended to be<br>worn by operating room<br>personnel to protect a surgical<br>wound from contamination. | The powder-free nitrile surgical<br>Gloves, is a sterile and single use<br>devise made of synthetic rubber<br>intended to be worn by operation<br>room personnel to protect a surgical<br>wound from contamination. The<br>gloves do not contain lubricating or<br>dusting powder | Similar   |
| Powdered or Powder free                                                                           | Powder free                                                                                                                                                                                       | Powder free                                                                                                                                                                                                                                                                       | Same      |
| Material                                                                                          | Synthetic Rubber                                                                                                                                                                                  | Synthetic Rubber                                                                                                                                                                                                                                                                  | Same      |
| Color                                                                                             | White                                                                                                                                                                                             | White                                                                                                                                                                                                                                                                             | Same      |
| Classification as per<br>ASTMD3577-19,<br>Standard Specification<br>for Rubber Surgical<br>Gloves | Type II - gloves compounded<br>from Synthetic rubber latex                                                                                                                                        | Type II - gloves compounded from<br>Synthetic rubber latex                                                                                                                                                                                                                        | Same      |
| Sterilization                                                                                     | Radiation, SAL-10-6                                                                                                                                                                               | ETO/as well as Radiation, SAL- 10-6                                                                                                                                                                                                                                               | Similar   |
| Label and Labeling                                                                                | Meet FDA's label Requirements                                                                                                                                                                     | Meet FDA's label Requirements                                                                                                                                                                                                                                                     | Same      |
| Type of use                                                                                       | Over the counter use                                                                                                                                                                              | Over the counter use                                                                                                                                                                                                                                                              | Same      |
| Dimensions Length:-<br>Min 265mm                                                                  | 307mm                                                                                                                                                                                             | 265mm min                                                                                                                                                                                                                                                                         | Similar   |
| Size: Width                                                                                       |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                   |           |
| 6.0 (76±6mm)                                                                                      | 78mm                                                                                                                                                                                              | 76±6mm                                                                                                                                                                                                                                                                            |           |
| 6.5 (83±6mm)                                                                                      | 85mm                                                                                                                                                                                              | 83±6mm                                                                                                                                                                                                                                                                            | Similar   |
| 7.0 (89±6mm)                                                                                      | 88mm                                                                                                                                                                                              | 89±6mm                                                                                                                                                                                                                                                                            |           |
| 7.5 (95±6mm)                                                                                      | 97mm                                                                                                                                                                                              | 95±6mm                                                                                                                                                                                                                                                                            |           |
| 8.0 (102±6mm)                                                                                     | 103mm                                                                                                                                                                                             | 102±6mm                                                                                                                                                                                                                                                                           |           |
| 8.5 (108±6mm)                                                                                     | 110mm                                                                                                                                                                                             | 108±6mm                                                                                                                                                                                                                                                                           |           |
| 9.0 (114±6mm)                                                                                     | 116mm                                                                                                                                                                                             | 114±6mm                                                                                                                                                                                                                                                                           |           |
| Cuff, Palm, Finger Tip<br>Thickness<br>Min 0.10mm                                                 | Cuff- 0.12mm<br>Palm- 0.18mm<br>Finger Tip- 0.21mm                                                                                                                                                | Cuff, Palm & Finger has a min<br>0.10mm thickness                                                                                                                                                                                                                                 | Similar   |
| Tensile Strength<br>Unaged<br>17MPa minimum                                                       | 24.20MPa                                                                                                                                                                                          | 17MPa min                                                                                                                                                                                                                                                                         | Similar   |
| Ultimate Elongation<br>Unaged<br>650% minimum                                                     | 871%                                                                                                                                                                                              | 650% min                                                                                                                                                                                                                                                                          | Similar   |
| Stress at 500%                                                                                    |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                   |           |
| Unaged<br>7.0 MPa Max                                                                             | 6.1MPa                                                                                                                                                                                            | 7.0MPa max                                                                                                                                                                                                                                                                        | Similar   |
| Tensile Strength                                                                                  |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                   |           |
| Aged<br>12MPa minimum                                                                             | 18.53MPa                                                                                                                                                                                          | 12 MPa min                                                                                                                                                                                                                                                                        | Similar   |
| Ultimate Elongation                                                                               |                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                   |           |
| Aged<br>490% minimum                                                                              | 653%                                                                                                                                                                                              | 490% min                                                                                                                                                                                                                                                                          | Similar   |
| Freedom from Holes                                                                                | AQL 1.5                                                                                                                                                                                           | AQL 2.5                                                                                                                                                                                                                                                                           | Similar   |
| Powder residue for<br>powder free glove<br>Powder content<br>< 2 mg/Glove                         | 0.40mg/Glove                                                                                                                                                                                      | < 2mg/Glove                                                                                                                                                                                                                                                                       | Similar   |
| Skin Irritation                                                                                   | Under conditions of the testing, not an irritant                                                                                                                                                  | Under conditions of the testing, not an irritant                                                                                                                                                                                                                                  | Similar   |
| Skin Sensitization                                                                                | Under the conditions of the testing, not a sensitizer                                                                                                                                             | Under the conditions of the testing, not a sensitizer                                                                                                                                                                                                                             | Similar   |
| In vitro cytotoxicity                                                                             | Under the conditions of the testing, not cytotoxic                                                                                                                                                | Under the conditions of the testing, not cytotoxic                                                                                                                                                                                                                                | Similar   |
| Material Mediated<br>pyrogenicity                                                                 | Under the conditions of the testing, non-pyrogenic                                                                                                                                                | No data available                                                                                                                                                                                                                                                                 | Different |
| Bacterial Endotoxin Test                                                                          | < 20 EU/glove                                                                                                                                                                                     | No data available                                                                                                                                                                                                                                                                 | Different |
| Systemic Toxicity                                                                                 | Under the conditions of the testing, no acute systemic<br>toxicity                                                                                                                                | No data available                                                                                                                                                                                                                                                                 | Different |
| Sterilization Modality                                                                            | Radiation                                                                                                                                                                                         | Radiation/EtO                                                                                                                                                                                                                                                                     | Similar   |
| Sterility Assurance Level                                                                         | 10-6                                                                                                                                                                                              | 10-6                                                                                                                                                                                                                                                                              | Same      |
| Shelf Life                                                                                        | 3 years                                                                                                                                                                                           | Unknown                                                                                                                                                                                                                                                                           | Different |

{5}------------------------------------------------

#### F. Summary of Non-Clinical Performance Testing

| Test Method                                                                                                         | Purpose                                                                               | Acceptance Criteria                                                 | Result                                 |          |      |
|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------|----------------------------------------|----------|------|
| ASTM D5151-2019                                                                                                     | Freedom from holes<br>Before aging                                                    | AQL 1.5<br>3 non-consecutive lots                                   | Pass                                   |          |      |
| ASTM D5151-2019                                                                                                     | Freedom from holes<br>After accelerated aging<br>70 $\pm$ 2°C for 166 $\pm$ 2 h       | AQL 1.5<br>3 non-consecutive lots                                   | Pass                                   |          |      |
| ASTM D5151-2019                                                                                                     | Freedom from holes<br>After accelerated aging<br>50 $\pm$ 2°C for 90 $\pm$ 1 days     | AQL 1.5<br>3 non-consecutive lots                                   | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Tensile strength<br>Before aging                                                      | Minimum 17MPa for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots  | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical                                                  | Tensile strength<br>After accelerated aging<br>70 $\pm$ 2°C for 166 $\pm$ 2 h         | Minimum 12MPa for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots  | Pass                                   |          |      |
| Gloves                                                                                                              |                                                                                       |                                                                     |                                        |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Tensile strength<br>After accelerated aging<br>$50\pm2$ °C for $90\pm1$ days          | Minimum 12MPa for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots  | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Ultimate Elongation<br>Before aging                                                   | Minimum 650% for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots   | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Ultimate elongation<br>After accelerated aging<br>$70\pm2$ °C for $166\pm2$ h         | Minimum 490% for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots   | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Ultimate elongation<br>After accelerated aging<br>$50\pm2$ °C for $90\pm1$ days       | Minimum 490% for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots   | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Stress at 500% elongation<br>Before accelerated aging                                 | Maximum 7.0MPa for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Stress at 500% elongation<br>After accelerated aging<br>$70\pm2$ °C for $166\pm2$ h   | Maximum 7.0MPa for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical<br>Gloves                                        | Stress at 500% elongation<br>After accelerated aging<br>$50\pm2$ °C for $90\pm1$ days | Maximum 7.0MPa for all<br>sizes<br>AQL 4%<br>3 non-consecutive lots | Pass                                   |          |      |
| ASTM F1929,<br>Test Method for<br>detecting seal leaks in<br>porous medical<br>packaging by dye<br>penetration      | Package integrity<br>After accelerated aging<br>$70\pm2$ °C for $166\pm2$ h           | No penetration<br>AQL 0.65%<br>3 non-consecutive lots               | Pass                                   |          |      |
| ASTM F1929,<br>Test Method for<br>detecting seal leaks in<br>porous medical<br>packaging by dye<br>penetration      | Package integrity<br>After accelerated aging<br>$50\pm2$ °C for $90\pm1$ days         | No penetration<br>AQL 0.65%<br>3 non-consecutive lots               | Pass                                   |          |      |
| ASTM D3577-19,<br>Standard<br>Specification for<br>Rubber Surgical                                                  | To determine the length of<br>the gloves                                              | Min 265mm for all sizes                                             | Size 6                                 | Pass     |      |
|                                                                                                                     |                                                                                       |                                                                     | Size 6 ½                               | Pass     |      |
|                                                                                                                     |                                                                                       |                                                                     | Size 7                                 | Pass     |      |
| Gloves                                                                                                              |                                                                                       |                                                                     |                                        | Size 7 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8½  | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 9   | Pass |
| ASTM D3577-19, Standard Specification for Rubber Surgical Gloves                                                    | To determine the width of the gloves                                                  | Size 6                                                              | 76±6mm                                 | Size 6   | Pass |
|                                                                                                                     |                                                                                       | Size 6 ½                                                            | 83±6mm                                 | Size 6 ½ | Pass |
|                                                                                                                     |                                                                                       | Size 7                                                              | 89±6mm                                 | Size 7   | Pass |
|                                                                                                                     |                                                                                       | Size 7 ½                                                            | 95±6mm                                 | Size 7 ½ | Pass |
|                                                                                                                     |                                                                                       | Size 8                                                              | 102±6mm                                | Size 8   | Pass |
|                                                                                                                     |                                                                                       | Size 8 ½                                                            | 108±6mm                                | Size 8 ½ | Pass |
|                                                                                                                     |                                                                                       | Size 9                                                              | 114±6mm                                | Size 9   | Pass |
| ASTM D3577-19, Standard Specification for Rubber Surgical Gloves                                                    | To determine finger thickness                                                         |                                                                     | Minimum 0.10mm for all sizes<br>AQL 4% | Size 6   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 6 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 7   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 7 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 9   | Pass |
| ASTM D3577-19, Standard Specification for Rubber Surgical Gloves                                                    | To determine palm thickness                                                           |                                                                     | Minimum 0.10mm for all sizes<br>AQL 4% | Size 6   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 6 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 7   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 7 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 9   | Pass |
| ASTM D3577-19, Standard Specification for Rubber Surgical Gloves                                                    | To determine cuff thickness                                                           |                                                                     | Minimum 0.10mm for all sizes<br>AQL 4% | Size 6   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 6 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 7   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 7 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8   | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 8 ½ | Pass |
|                                                                                                                     |                                                                                       |                                                                     |                                        | Size 9   | Pass |
| ASTM D6124-06R17, Standard Test Method for Residual Powder on Medical Gloves                                        | Less than 2mg/glove                                                                   |                                                                     | Maximum 2mg/glove for all sizes        | Pass     |      |
| ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10- Tests for irritation and skin sensitization | Irritation                                                                            | Under the conditions of the testing, not an irritant                |                                        | Pass     |      |
| ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10- Tests for irritation and skin sensitization | Sensitization                                                                         | Under the conditions of the testing, not a sensitizer               |                                        | Pass     |      |
| ISO 10993-5:2009(E) Cytotoxicity                                                                                    |                                                                                       | Under the conditions of                                             |                                        | Pass     |      |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

| Biological Evaluation<br>of Medical Devices -<br>Part 5-Tests for<br>in-vitro Cytotoxicity                                                      |                                   | the testing, non-cytotoxic                                            |      |
|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|-----------------------------------------------------------------------|------|
| ISO 10993-<br>11:2017(E)<br>Biological Evaluation<br>of Medical Devices -<br>Part 11- Tests for<br>Systemic<br>Toxicity and<br>Biological Tests | Acute systemic toxicity           | Under the conditions of<br>the testing, no acute<br>systemic toxicity | Pass |
| USP 41<br><151>Pyrogen Test                                                                                                                     | Material mediated<br>pyrogenicity | Under the conditions of<br>the testing, non-pyrogenic                 | Pass |
| USP 42 <85><br>Bacterial Endotoxin<br>Test                                                                                                      | Bacterial endotoxin limits        | Less than 20EU/glove                                                  | Pass |

#### G. Summary of Clinical Performance Testing

No clinical performance testing was performed in support of this submission.

### H. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device Sterile Nitrile Surgical Gloves Powder Free is as safe, as effective and performs as well or better than the legally marketed predicate device "PRIMUS NITRILE GLOVES" Sterile Nitrile Surgical Gloves Powder Free (K170515).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K222350](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K222350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
