← Product Code [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO) · K190077

# Biogel® PI UltraTouch S Surgical Glove, Biogel® Pl Ultra Touch S Indicator Underglove (K190077)

_Molnlycke Health Care Us, LLC · KGO · May 24, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K190077

## Device Facts

- **Applicant:** Molnlycke Health Care Us, LLC
- **Product Code:** [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO.md)
- **Decision Date:** May 24, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Biogel® PI UltraTouch S Surgical Glove is a disposable device made of polyisoprene, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants. The Biogel® PI UltraTouch S Indicator Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

## Device Story

Disposable, sterile, powder-free surgical gloves made of synthetic polyisoprene; straw-colored overglove and blue underglove. Used in surgical settings by healthcare professionals to provide a barrier against infectious materials and contaminants. Gloves can be worn independently or as a double-gloving pair. No electronic or software components; purely mechanical barrier device.

## Clinical Evidence

Bench testing only. No clinical data required. Testing included biocompatibility (ISO 10993-10, ISO 10993-11), physical dimensions, tensile strength, elongation at break, freedom from holes (ASTM D5151), and residual powder levels (ASTM D6124).

## Technological Characteristics

Material: Synthetic polyisoprene. Design: Single-use, sterile, powder-free, hand-specific, beaded cuff. Sterilization: Radiation (10^-6 SAL). Compliance: ASTM D3577-09(2015) for physical properties/dimensions, ASTM D5151-06(2015) for hole detection, and ASTM D6124-06(2017) for powder residual.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Biogel® PI UltraTouch Surgical Glove ([K050184](/device/K050184.md))
- Biogel® Indicator Underglove ([K111413](/device/K111413.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

My 24, 2019

Molnlycke Health Care, US LLC Leonard Stewart Regulatory Affairs Specialist 5550 Peachtree Parkway, Suite 500 Norcross, GA 30092

Re: K190077

Trade/Device Name: Biogel® PI UltraTouch S Surgical Glove, Biogel® PI Ultra Touch S Indicator Underglove Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: February 28, 2019 Received: March 1, 2019

Dear Leonard Stewart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For David Krause, PhD Acting Director Division of Infection Control and Plastic Surgery Office of Surgical & Infection Control Devices Office of Product Evaluation & Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K190077

Device Name

Biogel® PI UltraTouch S Surgical Glove and Biogel® PI UltraTouch S Indicator Underglove

Indications for Use (Describe)

The Bioge® PI UltraTouch S Surgical Glove is a disposable device made of polyisoprene, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

The Bioge® PI UltraTouch S Indicator Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY (K190077)

#### for

# Biogel® PI UltraTouch S Surgical Glove and Biogel®Pl UltraTouch S Indicator Underglove

| Date Prepared:            | May 22, 2019                                                                                                                                                                          |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Sponsor:       | Mölnlycke Health Care US, LLC<br>5550 Peachtree Parkway, Suite 500<br>Norcross, GA 30092<br>Registration number: 3004763499<br>Owner/Operator Number: 8030877                         |
| Submission Correspondent: | Leonard Stewart<br>Regulatory Affairs Specialist<br>Tel: 470-375-0178<br>Fax: 678-245-7746<br>email: leonard.stewart@molnlycke.com                                                    |
| Trade/Proprietary Names:  | Biogel® PI UltraTouch S Surgical Glove and Biogel®PI<br>UltraTouch S Indicator Underglove                                                                                             |
| Regulation Name:          | Non-powered surgeon's glove                                                                                                                                                           |
| Common Name:              | Surgeon's Glove                                                                                                                                                                       |
| Classification Name:      | Surgeon's Glove                                                                                                                                                                       |
| Device Class:             | Class I                                                                                                                                                                               |
| Regulation Number:        | 21 CFR 878.4460                                                                                                                                                                       |
| Product Code:             | KGO                                                                                                                                                                                   |
| 510(k) Submission Number: | K190077                                                                                                                                                                               |
| Predicate Device Name(s): | Biogel® PI UltraTouch Surgical Glove (SKINSENSE®POLY-<br>ISOPRENE POWDER-FREE NON-LATEX GLOVE, K050184),<br>Biogel® Indicator Underglove (BIOGEL® INDICATOR ®<br>UNDERGLOVE, K111413) |

### Description of Device:

Subject of this submission are two surgical gloves: a single-use, sterile, straw-colored overglove which is a disposable, powder-free surgical glove made from synthetic polyisoprene, and a single-use, sterile, blue underglove which is a disposable, powder-free surgical glove made from synthetic polyisoprene. The overglove, and underglove may be used independently or worn as a double-gloving pair if desired.

### Indications for Use:

The Biogel® PI UltraTouch S Surgical Glove is a disposable device made of polvisoprene, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

{4}------------------------------------------------

The Biogel® PI UltraTouch S Indicator Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants

|                                        | Summary of technological characteristics of the subject device<br>compared to the predicate device                                                                                                                             |                                                                                                                                                                                                                |           |                                                                                                                                                                                                                                        |                                                                                                                                                                                                                               |           |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
|                                        | (overglove)                                                                                                                                                                                                                    |                                                                                                                                                                                                                |           | (underglove)                                                                                                                                                                                                                           |                                                                                                                                                                                                                               |           |
|                                        | Biogel® PI<br>UltraTouch S<br>Surgical<br>Glove<br>(Subject<br>Device)                                                                                                                                                         | Biogel® PI<br>UltraTouch<br>Surgical<br>Glove<br>(Predicate<br>Device)                                                                                                                                         | Comment   | Biogel® PI<br>UltraTouch S<br>Indicator<br>Underglove<br>(Subject<br>Device)                                                                                                                                                           | Biogel® PI<br>Indicator<br>Underglove<br>(Predicate<br>Device)                                                                                                                                                                | Comment   |
| 510(k)<br>Number                       | K190077                                                                                                                                                                                                                        | K050184                                                                                                                                                                                                        | -         | K190077                                                                                                                                                                                                                                | K111413                                                                                                                                                                                                                       | -         |
| Manufacturer                           | Mölnlycke                                                                                                                                                                                                                      | Mölnlycke                                                                                                                                                                                                      | Identical | Mölnlycke                                                                                                                                                                                                                              | Mölnlycke                                                                                                                                                                                                                     | Identical |
| Regulation<br>number                   | 21CFR<br>878.4460                                                                                                                                                                                                              | 21CFR<br>878.4460                                                                                                                                                                                              | Identical | 21CFR<br>878.4460                                                                                                                                                                                                                      | 21CFR<br>878.4460                                                                                                                                                                                                             | Identical |
| Regulation<br>name                     | Surgeon's<br>Glove                                                                                                                                                                                                             | Surgeon's<br>Glove                                                                                                                                                                                             | Identical | Surgeon's<br>Glove                                                                                                                                                                                                                     | Surgeon's<br>Glove                                                                                                                                                                                                            | Identical |
| Regulatory<br>class                    | Class 1                                                                                                                                                                                                                        | Class 1                                                                                                                                                                                                        | Identical | Class 1                                                                                                                                                                                                                                | Class 1                                                                                                                                                                                                                       | Identical |
| Product code                           | KGO                                                                                                                                                                                                                            | KGO                                                                                                                                                                                                            | Identical | KGO                                                                                                                                                                                                                                    | KGO                                                                                                                                                                                                                           | Identical |
| Intended use                           | Powder-Free<br>Surgeon's<br>Glove                                                                                                                                                                                              | Powder-Free<br>Surgeon's<br>Glove                                                                                                                                                                              | Identical | Powder-Free<br>Surgeon's<br>Glove                                                                                                                                                                                                      | Powder-Free<br>Surgeon's<br>Glove                                                                                                                                                                                             | Identical |
| Indication for<br>use                  | Biogel® PI<br>UltraTouch S<br>Surgical Glove<br>(Overglove) is<br>a disposable<br>device made<br>of<br>polyisoprene,<br>that is<br>intended to be<br>worn on the<br>hands, usually<br>in surgical<br>settings, to<br>provide a | Biogel® PI<br>UltraTouch<br>Surgical<br>glove(Overgl<br>ove) A<br>powder-free<br>sterile<br>surgeon's<br>glove is a<br>disposable<br>device made<br>of<br>polyisoprene<br>that is<br>intended to<br>be worn on | Similar   | Biogel® PI<br>UltraTouch S<br>Indicator<br>Underglove is a<br>disposable<br>device made of<br>polyisoprene,<br>blue in color<br>that is intended<br>to be worn on<br>the hands,<br>usually in<br>surgical<br>settings, to<br>provide a | Biogel® PI<br>Indicator<br>Underglove is<br>a disposable<br>device made<br>of<br>polyisoprene,<br>blue in color<br>that is<br>intended to<br>be worn on<br>the hands,<br>usually in a<br>surgical<br>setting, to<br>provide a | Similar   |
|                                        | Summary of technological characteristics of the subject device<br>compared to the predicate device                                                                                                                             |                                                                                                                                                                                                                |           |                                                                                                                                                                                                                                        |                                                                                                                                                                                                                               |           |
|                                        | potentially<br>infectious<br>material and<br>other<br>contaminants                                                                                                                                                             | the hands,<br>usually in<br>surgical<br>settings; to<br>provide a<br>barrier<br>against<br>potentially<br>infectious<br>materials and<br>other<br>contaminants                                                 |           | potentially<br>infectious<br>material and<br>other<br>contaminants                                                                                                                                                                     | barrier<br>against<br>potentially<br>infectious<br>material and<br>other<br>contaminants                                                                                                                                      |           |
| Material                               | Synthetic<br>Polyisoprene                                                                                                                                                                                                      | Synthetic<br>Polyisoprene                                                                                                                                                                                      | Identical | Synthetic<br>Polyisoprene                                                                                                                                                                                                              | Synthetic<br>Polyisoprene                                                                                                                                                                                                     | Identical |
| Design                                 | Single use                                                                                                                                                                                                                     | Single use                                                                                                                                                                                                     | Identical | Single use                                                                                                                                                                                                                             | Single use                                                                                                                                                                                                                    | Identical |
|                                        | Sterile                                                                                                                                                                                                                        | Sterile                                                                                                                                                                                                        | Identical | Sterile                                                                                                                                                                                                                                | Sterile                                                                                                                                                                                                                       | Identical |
|                                        | Powder-free                                                                                                                                                                                                                    | Powder-free                                                                                                                                                                                                    | Identical | Powder-free                                                                                                                                                                                                                            | Powder-free                                                                                                                                                                                                                   | Identical |
|                                        | Hand specific                                                                                                                                                                                                                  | Hand specific                                                                                                                                                                                                  | Identical | Hand specific                                                                                                                                                                                                                          | Hand specific                                                                                                                                                                                                                 | Identical |
|                                        | Beaded Cuff                                                                                                                                                                                                                    | Beaded cuff                                                                                                                                                                                                    | Identical | Beaded cuff                                                                                                                                                                                                                            | Beaded cuff                                                                                                                                                                                                                   | Identical |
| Coating                                | Yes                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                            | Identical | Yes                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                           | Identical |
| Color                                  | Straw<br>(Natural)                                                                                                                                                                                                             | Straw<br>(Natural)                                                                                                                                                                                             | Identical | Blue                                                                                                                                                                                                                                   | Blue                                                                                                                                                                                                                          | Identical |
| Sterilization<br>method                | Radiation                                                                                                                                                                                                                      | Radiation                                                                                                                                                                                                      | Identical | Radiation                                                                                                                                                                                                                              | Radiation                                                                                                                                                                                                                     | Identical |
| Sterility<br>Assurance<br>Level (SAL)  | 10-6 SAL                                                                                                                                                                                                                       | 10-6 SAL                                                                                                                                                                                                       | Identical | 10-6 SAL                                                                                                                                                                                                                               | 10-6 SAL                                                                                                                                                                                                                      | Identical |
| Dimensions<br>& physical<br>properties | Meets ASTM<br>D3577-<br>09(2015)                                                                                                                                                                                               | Meets ASTM<br>D3577-<br>09(2015)                                                                                                                                                                               | Identical | Meets ASTM<br>D3577-<br>09(2015)                                                                                                                                                                                                       | Meets ASTM<br>D3577-<br>09(2015)                                                                                                                                                                                              | Identical |
| Freedom<br>from holes                  | AQL meets 21<br>CFR 800.20<br>and ASTM<br>D3577-<br>09(2015)<br>requirements                                                                                                                                                   | AQL meets<br>21 CFR<br>800.20 and<br>ASTM<br>D3577-<br>09(2015)<br>requirements                                                                                                                                | Identical | AQL meets 21<br>CFR 800.20<br>and ASTM<br>D3577-<br>09(2015)<br>requirements                                                                                                                                                           | AQL meets<br>21 CFR<br>800.20 and<br>ASTM<br>D3577-<br>09(2015)<br>requirements                                                                                                                                               | Identical |
| Powder<br>residual                     | Meets<br>requirements<br>of ≤ 2.0<br>mg/glove for<br>Powder-free<br>designation<br>per ASTM<br>D3577-<br>09(2015)                                                                                                              | Meets<br>requirements<br>of ≤ 2.0<br>mg/glove for<br>Powder-free<br>designation<br>per ASTM<br>D3577-<br>09(2015)                                                                                              | Identical | Meets<br>requirements of<br>≤ 2.0 mg/glove<br>for Powder-free<br>designation per<br>ASTM D3577-<br>09(2015)                                                                                                                            | Meets<br>requirements<br>of ≤ 2.0<br>mg/glove for<br>Powder-free<br>designation<br>per ASTM<br>D3577-<br>09(2015)                                                                                                             | Identical |

# Technological Characteristics:

{5}------------------------------------------------

{6}------------------------------------------------

Summary of Non-Clinical Testing:

| Summary of Non-Clinical Testing                                            |                                                                                                                                                                         |                                                                                                                                                          |                                                                           |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
|                                                                            | Standard/Test/FDA Guidance                                                                                                                                              | Biogel® PI<br>UltraTouch S<br>Surgical Glove<br>(Subject Device)                                                                                         | Biogel® PI<br>UltraTouch S<br>Indicator<br>Underglove<br>(Subject Device) |
| Biocompatibility:                                                          |                                                                                                                                                                         |                                                                                                                                                          |                                                                           |
| Primary Skin<br>Irritation                                                 | ISO 10993-10:2010 Biological<br>evaluation of medical devices —<br>Part 10: Tests for irritation and skin<br>sensitization                                              | Under the conditions of the study, not an<br>irritant.                                                                                                   |                                                                           |
| ISO Closed<br>Patch<br>Sensitization                                       | ISO 10993-10:2010 Biological<br>evaluation of medical devices —<br>Part 10: Tests for irritation and skin<br>sensitization                                              | Under the conditions of the study, not a<br>sensitizer.                                                                                                  |                                                                           |
| Acute Systemic<br>Toxicity Study                                           | ISO 10993-11:2010 Biological<br>evaluation of medical devices —<br>Part 11: Tests for systemic toxicity                                                                 | Under the conditions of the study, no<br>mortality or evidence of systemic toxicity<br>from the extracts.                                                |                                                                           |
| Physical<br>characteristics:                                               |                                                                                                                                                                         |                                                                                                                                                          |                                                                           |
| Dimensions                                                                 | ASTM D3577- 09(2015)<br>Standard Specification for<br>Rubber Surgical Gloves                                                                                            | Meets ASTM D3577-09(2015) requirements<br>for length, width, and thickness. Identical to<br>predicate.                                                   |                                                                           |
| Physical<br>Properties                                                     | ASTM D3577- 09(2015)<br>Standard Specification for<br>Rubber Surgical Gloves                                                                                            | Meets ASTM D3577-09(2015)<br>requirements for tensile strength and<br>elongation at break before and after<br>accelerated aging. Identical to Predicate. |                                                                           |
| Freedom from<br>holes                                                      | ASTM D5151-06(2015) Standard<br>Test Method for Detection of Holes<br>in Medical Gloves<br>ASTM D3577- 09(2015) Standard<br>Specification for Rubber Surgical<br>Gloves | Exceeds 21 CFR 800.20 and ASTM<br>D3577-09(2015) requirements of AQL<br>1.5. Identical to predicate.                                                     |                                                                           |
| Powder residual                                                            | ASTM D6124-06(2017) Standard<br>Test Method for Residual Powder<br>on Medical Gloves<br>ASTM D3577- 09(2015) Standard<br>Specification for Rubber Surgical<br>Gloves    | Meets powder level requirements for<br>"Powder-free" designation per ASTM<br>D3577-09(2015). Identical to predicate.                                     |                                                                           |
| Clinical Data Summary - Subject Device (modified version of the predicate) |                                                                                                                                                                         |                                                                                                                                                          |                                                                           |
| Clinical testing                                                           | Clinical data is not required. Identical to predicate.                                                                                                                  |                                                                                                                                                          |                                                                           |

# Conclusion:

Based on the Indication for Use, technological characteristics, and non-clinical performance data, Bioge®® Pl UltraTouch S Surgical Glogel® Pl UltraTouch S Indicator Underglove (K190077) are as safe, as effective, and perform as well as or better than the legally marketed predicate devices Bioge®® PI UltraTouch Surgical Glove (K050184) and the Bioge®® PI Indicator Underglove (K111413).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K190077](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K190077)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
