← Product Code [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO) · K171276

# Sterile Polyisoprene Surgical Powder Free Glove Tested for Use with Chemotherapy Drugs (K171276)

_Tg Medical Sdn. Bhd. · KGO · Aug 21, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K171276

## Device Facts

- **Applicant:** Tg Medical Sdn. Bhd.
- **Product Code:** [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO.md)
- **Decision Date:** Aug 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Intended Use

This powder free surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating, nom personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with Chemotherapy Drugs in accordance ASTM D6978-05 (2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

## Device Story

Sterile polyisoprene surgical powder-free glove; disposable synthetic rubber. Used by operating room personnel to protect surgical wounds from contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05 (2013). Provides barrier protection for clinicians handling hazardous chemotherapy agents. Benefits patient by reducing risk of surgical site contamination and protects healthcare workers from chemotherapy drug exposure.

## Clinical Evidence

Bench testing only. Testing performed in accordance with ASTM D6978-05 (2013) to assess permeation breakthrough times for nine specific chemotherapy drugs (Carmustine, Cyclophosphamide, Doxorubicin Hydrochloride, Etoposide, Fluorouracil, Methotrexate, Paclitaxel, Thiotepa, Vincristine Sulfate).

## Technological Characteristics

Material: Synthetic polyisoprene rubber. Form factor: Sterile, powder-free surgical glove. Standards: ASTM D6978-05 (2013) for chemotherapy drug permeation. Sterilization: Not specified.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 21, 2017

TG Medical SDN. BHD. Fatima Mohamad OA. Assistant Manager Lot 5091. Jalan Teratai Batu 5, Off Jalan Meru Klang, 41050 My

Re: K171276

Trade/Device Name: Sterile Polyisoprene Surgical Powder Free Glove Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460

Regulation Name: Surgeon's Glove Regulatory Class: Class I Product Code: KGO, LZC Dated: July 24, 2017 Received: July 28, 2017

Dear Fatima Mohamad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tara A. Ryan -S 2017.08.21 17:25:27 -04'00

for

Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

#### 510(k) Number (if known)

# K171276

#### Device Name

Sterile Polyiosprene Surgical Powder Free Glove Tested for Use with Chemotherapy Drugs

### Indications for Use (Describe)

This powder free surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating, nom personnel to protect a surgical wound from contamination.

In addition, these gloves were tested for use with Chemotherapy Drugs in accordance ASTM D6978-05 (2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drugs Permeation

(Breakthrough Detection Time) in Minutes

The following chemicals have been tested with these gloves.

| Chemotherapy Drugs         | Concentration | Breakthrough Detection Time in Minutes |
|----------------------------|---------------|----------------------------------------|
| Carmustine (BCNU)          | 3.3mg/ml      | 13.0                                   |
| Cyclophosphamide (Cytoxan) | 20.0mg/ml     | >240                                   |
| Doxorubicin Hydrochloride  | 2.0mg/ml      | >240                                   |
| Etoposide (Toposar)        | 20.0mg/ml     | >240                                   |
| Fluorouracil               | 50.0mg/ml     | >240                                   |
| Methotrexate               | 25.0mg/ml     | >240                                   |
| Paclitaxel                 | 6.0mg/ml      | >240                                   |
| Thiotepa                   | 10.0mg/ml     | 24.9                                   |
| Vincristine Sulfate        | 1.0mg/ml      | >240                                   |

Please note that the following drugs have low permeation times: Carmustine (BCNU): 13.0 minutes and Thiotepa: 24.9

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 804 Subpart D)

[X] Over-The-Counter Use (21 GFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate on any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K171276](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K171276)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).