← Product Code [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO) · K110272

# STERILE POLYISOPRENE POWDER-FREE SURGICAL GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUG (K110272)

_Cardinal Health, Inc. · KGO · Jun 8, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K110272

## Device Facts

- **Applicant:** Cardinal Health, Inc.
- **Product Code:** [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO.md)
- **Decision Date:** Jun 8, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Intended Use

This powder-free surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM and the most go res world of Medical Gloves to Permeation by Chemotherapy Drugs:

## Device Story

Disposable, sterile, powder-free surgical glove; formulated from polyisoprene synthetic rubber latex; features anti-slip finish, independent thumb, and tapered mechanically locking cuffs to reduce roll down. Used by operating room personnel to protect surgical wounds from contamination. Tested for permeation resistance against 25 chemotherapy drugs per ASTM D6978. Provides barrier protection; healthcare providers use gloves during surgical procedures and handling of chemotherapy agents. Benefits include protection of patient from contamination and protection of clinician from drug exposure.

## Clinical Evidence

Bench testing only. No clinical data required. Testing included biocompatibility (ISO 10993-10), physical properties, dimensions, freedom from holes (ASTM D5151), and chemotherapy drug permeation (ASTM D6978).

## Technological Characteristics

Material: Polyisoprene synthetic rubber latex. Design: Sterile, powder-free, hand-specific, independent thumb, beaded cuff, lubricated. Standards: ASTM D3577 (physical properties/dimensions), ASTM D6978 (chemotherapy permeation), 21 CFR 800.20 (freedom from holes). Biocompatibility: ISO 10993-10. Powder residual: ≤2.0 mg/glove.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Esteem SMT Sterile Polyisoprene Powder-Free Surgical Gloves ([K093300](/device/K093300.md))
- Duraprene Sterile Neoprene Powder-Free Surgical Gloves ([K013302](/device/K013302.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic above the company name. The graphic is made up of curved lines that resemble a bird in flight. The text "CardinalHealth" is written in a bold, sans-serif font.

### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## JUN - 8 2511

#### STERILE POLYISOPRENE POWDER-FREE SURGICAL GLOVES

(A summary of safety and effectiveness information in accordance with the requirements of 21 CFR 807.92)

| Applicant:                                                        | Cardinal Health<br>1430 Waukegan Road<br>McGaw Park, IL 60085                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number:                             | 1423537                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulatory Affairs<br>Contact:                                    | Tatyana Bogdan, RAC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Telephone:                                                        | 847-887-2325                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Fax:                                                              | 847-887-2717                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| E-mail:                                                           | tatyana.bogdan-curvin@cardinalhealth.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Summary Prepared:                                                 | May 13, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Trade Name:                                                       | Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with<br>Chemotherapy Drugs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Common Name:                                                      | Surgeon's Gloves                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name:                                              | Surgeon's Gloves                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Panel:                                             | General and Plastic Surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Regulation:                                                       | 21 CFR 878.4460                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code(s):                                                  | KGO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Legally marketed<br>device(s) to which<br>equivalence is claimed: | 1. Esteem SMT Sterile Polyisoprene Powder-Free Surgical Gloves ( 510(k) K093300, product code KGO)<br>2. Duraprene Sterile Neoprene Powder-Free Surgical Gloves (510(k) K013302, product code KGO)                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Reason for 510(k)<br>Submission:                                  | Addition of new indications for use: Tested for Use with Chemotherapy Drugs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description:                                               | The proposed device is a disposable device. It is not made with natural rubber<br>latex. Instead, it is formulated from polyisoprene, which is a synthetic rubber<br>latex. This Sterile Polyisoprene Powder-Free Surgical Glove is manufactured<br>using exact same material used in the currently cleared device, Esteem SMT<br>glove (K093300) that has been legally marketed by Cardinal Health. The glove<br>is manufactured using molds that feature anti-slip finish, independent thumb,<br>and tapered mechanically locking cuffs to help reduce cuff roll down. It is<br>offered powder-free and sterile. |

-

{1}------------------------------------------------

#### Intended Use:

This powder-free surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM m and the most go res world of Medical Gloves to Permeation by Chemotherapy Drugs:

|     | Chemotherapy Drug and Concentration | Min Breakthrough Detection Time, 0.01<br>µg/cm²/minute |
|-----|-------------------------------------|--------------------------------------------------------|
| 1.  | Blenoxane (15 mg/ml)                | >240                                                   |
| 2.  | Busulfan (6 mg/ml)                  | >240                                                   |
| 3.  | Carmustine (BCNU) (3.3 mg/ml)       | 0.37                                                   |
| 4.  | Cisplastin, (1.0 mg/ml)             | >240                                                   |
| 5.  | Cytarabine (100 mg/ml)              | >240                                                   |
| 6.  | Cyclophosphamide (20 mg/ml)         | >240                                                   |
| 7.  | Dacarbazine (10 mg/ml)              | >240                                                   |
| 8.  | Doxorubicin HCl (2.0 mg/ml)         | >240                                                   |
| 9.  | Ellence (25 mg/ml)                  | >240                                                   |
| 10. | Etoposide (Toposar) (20 mg/ml)      | >240                                                   |
| 11. | Fludarabine (25 mg/ml)              | >240                                                   |
| 12. | Fluorouracil (50 mg/ml)             | >240                                                   |
| 13. | Idarubicin (1 mg/ml)                | >240                                                   |
| 14. | Ifosfamide (50 mg/ml)               | >240                                                   |
| 15. | Mechlorethamine HCl (1 mg/ml)       | >240                                                   |
| 16. | Melphalan (5 mg/ml)                 | >240                                                   |
| 17. | Methotrexate (25 mg/ml)             | >240                                                   |
| 18. | Mitoxantrone (2 mg/ml)              | >240                                                   |
| 19. | Mitomycin C (0.5 mg/ml)             | >240                                                   |
| 20. | Paclitaxel (Taxol) (6.0 mg/ml)      | >240                                                   |
| 21. | Paraplatin (10 mg/ml)               | >240                                                   |
| 22. | Rituximab (10 mg/ml)                | >240                                                   |
| 23. | Thiotepa (10 mg/ml)                 | 0.44                                                   |
| 24. | Trisenox (0.1 mg/ml)                | >240                                                   |
| 25. | Vincristine sulfate (1 mg/ml)       | >240                                                   |

Please note that the following drugs have extremely low permeation time of less than 30 minutes: Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 0.37 minute; Thiotepa (10 mg/ml) has a minimum breakthrough time of 0.44 minute.

{2}------------------------------------------------

### KIIO2112

| Summary of the technological characteristics of the device compared to the predicate device |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristic                                                                              | New Device<br>(K110272)                                                                                 | Predicate<br>Sterile Polyisoprene Powder<br>Free Surgical Glove<br>(K093300)                                                                                                                                                                                                                                                                                    | Predicate<br>Sterile Neoprene Powder-<br>Free Surgical Glove<br>(K013302)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Material<br>Composition                                                                     | Polyisoprene Synthetic<br>Rubber Latex                                                                  | Polyisoprene Synthetic<br>Rubber Latex                                                                                                                                                                                                                                                                                                                          | Synthetic Neoprene Polymer<br>coated with Nitrile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
| Design                                                                                      | Single Use<br>Sterile<br>Powder-free<br>Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated | Single Use<br>Sterile<br>Powder-free<br>Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated                                                                                                                                                                                                                                                         | Single Use<br>Sterile<br>Powder-free<br>Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
| Intended Use                                                                                | Powder-Free Surgeon's<br>Glove                                                                          | Powder-Free Surgeon's<br>Glove                                                                                                                                                                                                                                                                                                                                  | Powder-Free Surgeon's<br>Glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |
| Indications for Use                                                                         | Tested for Use with<br>Chemotherapy Drugs                                                               | Not Tested                                                                                                                                                                                                                                                                                                                                                      | Tested for Use with<br>Chemotherapy Drugs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Dimensions &<br>Physical Properties                                                         | Meets ASTM D3577                                                                                        | Meets ASTM D3577                                                                                                                                                                                                                                                                                                                                                | Meets ASTM D3577                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |
| Freedom from<br>Holes                                                                       | AQL meets 21CFR 800.20 &<br>ASTM D3577 requirements                                                     | AQL meets 21CFR 800.20 &<br>ASTM D3577 requirements                                                                                                                                                                                                                                                                                                             | AQL meets 21CFR 800.20 &<br>ASTM D3577 requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |
| Powder Residual                                                                             | Meets requirements of ≤2.0<br>mg/glove for Powder-Free<br>designation per ASTM D3577                    | Meets requirements of ≤2.0<br>mg/glove for Powder-Free<br>designation per ASTM D3577                                                                                                                                                                                                                                                                            | Meets requirements of ≤2.0<br>mg/glove for Powder-Free<br>designation per ASTM D3577                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  |
| PERFORMANCE DATA                                                                            |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE*       |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Performance Test Summary-New Device                                                         |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Characteristic                                                                              |                                                                                                         | Standard/Test/FDA Guidance                                                                                                                                                                                                                                                                                                                                      | Results Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Biocompatibility:                                                                           |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Primary Skin Irritation                                                                     |                                                                                                         | ISO 10993-10                                                                                                                                                                                                                                                                                                                                                    | Gloves are non-irritating.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |
| Guinea Pig Maximization                                                                     |                                                                                                         | ISO 10993-10                                                                                                                                                                                                                                                                                                                                                    | Gloves do not display any potential for<br>sensitization.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Physical Characteristics:                                                                   |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Dimensions                                                                                  |                                                                                                         | ASTM D3577                                                                                                                                                                                                                                                                                                                                                      | Meet requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
| Physical Properties                                                                         |                                                                                                         | ASTM D3577                                                                                                                                                                                                                                                                                                                                                      | Meet requirements for rubber surgical gloves                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |
| Freedom from Holes                                                                          |                                                                                                         | 21 CFR 800.20 & ASTM D3577                                                                                                                                                                                                                                                                                                                                      | Tested in accordance with ASTM D5151 with<br>acceptable results                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Powder Residual                                                                             |                                                                                                         | ASTM D3577 tested using<br>ASTM standard D6124                                                                                                                                                                                                                                                                                                                  | Gloves meet powder level requirements for<br>"Powder-Free" designation per ASTM D3577.<br>Results generated values < 2mg of residual<br>powder per glove.                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Chemotherapy Drug<br>Permeation                                                             |                                                                                                         | ASTM D6978                                                                                                                                                                                                                                                                                                                                                      | Gloves were tested using ASTM D6978. Under<br>the test conditions prescribed by the test, the<br>minimum normalized breakthrough detection<br>times for each of the chemotherapy drugs<br>tested exceeded the maximum testing time of                                                                                                                                                                                                                                                                                                                        |  |
|                                                                                             |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 | 240 minutes except for Carmustine (BCNU)<br>(3.3 mg/ml) and Thiotepa (10 mg/ml), which<br>showed permeation time of less than 30 minutes.                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Comparative Performance Information Summary                                                 |                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Characteristic                                                                              | Requirement                                                                                             | New Device                                                                                                                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |
| Biocompatibility:                                                                           | ISO 10993-1                                                                                             | Meets requirements                                                                                                                                                                                                                                                                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Primary Skin Irritation                                                                     | ISO 10993-10                                                                                            | Pass                                                                                                                                                                                                                                                                                                                                                            | Pass                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  |
| Guinea Pig Maximization                                                                     | ISO 10993-10                                                                                            | Pass                                                                                                                                                                                                                                                                                                                                                            | Pass                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  |
| Dimensions                                                                                  | ASTM D3577                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Physical Properties                                                                         | ASTM D3577                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Freedom from Holes                                                                          | 21CFR 800.20<br>& ASTM D3577                                                                            | Meets requirements                                                                                                                                                                                                                                                                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Powder Residual                                                                             | ASTM D3577                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                              | Meets requirements                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
| Chemotherapy Drug<br>Permeation                                                             | ASTM D6978                                                                                              | Under the test conditions<br>prescribed by the test, the<br>minimum normalized<br>breakthrough detection<br>times for each of the 25<br>chemotherapy drugs tested<br>exceeded the maximum<br>testing time of 240 minutes<br>except for Carmustine<br>(BCNU) (3.3 mg/ml) and<br>Thiotepa (10 mg/ml), which<br>showed permeation time of<br>less than 30 minutes. | Duraprene gloves cleared<br>under K013302 were tested<br>with 10 chemotherapy drugs<br>in accordance with and met<br>requirements of ASTM F739<br>standard. ASTM F739<br>standard was superseded by<br>ASTM D6978-05. The testing<br>method is the same. The<br>difference is in that the<br>thinnest area of the glove<br>(palm or cuff) is tested and<br>the test results are reported<br>as minimum breakthrough<br>times per ASTM D6978, as<br>opposed to testing a random<br>specimen and reporting the<br>average breakthrough times<br>per ASTM F739. |  |

K110272 Traditional 510(k) Premarket Notification Submission – 510(k) Summary Cardinal Health, Inc.

{3}------------------------------------------------

## KHUJUL

# and/or of clinical information

Clinical data is not required.

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL AND CLINICAL DATA 

Non-clinical data demonstrates that Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with Chemotherapy Drugs meet the technological characteristics of ASTM D3577 standard, and are as safe , as effective, and performed as well as the legally marketed devices identified in this summary.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Tatyana Bogdan Regulatory Affairs Manager Cardinal Health, Incorporated 1430 Waukegan Road McGaw Park, Illinois 60085

JUN - 8 2011

Re: K110272

Trade/Device Name: Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO, LZC Dated: May 12, 2011 Received: May 24, 2011

Dear Ms. Bogdan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

### Page 2 - Ms. Bogdan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

## Indications for Use

# 510(k) Number (if known): KLO27 L

Device Name:

### Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with Chemotherapy Drugs

Indications for Use: This powder-free surgeon's glove is a disposable device mate of synthetic Indications for Osc. This powder need bear.gooms gersonnel to protect a surgical wound from contamination.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM In addition, these gioves were tested for ass Min encation by Chemotherapy Drugs:

|    | Chemotherapy Drug and<br>Concentration | Minimum Breakthrough Detection Time<br>in Minutes, 0.01 μg/cm²/minute |
|----|----------------------------------------|-----------------------------------------------------------------------|
| 1. | Blenoxane (15 mg/ml)                   | >240                                                                  |
| 2. | Busulfan (6 mg/ml)                     | >240                                                                  |
| 3. | Carmustine (BCNU) (3.3 mg/ml)          | 0.37                                                                  |
| 4. | Cisplastin, (1.0 mg/ml)                | >240                                                                  |
| 5. | Cytarabine (100 mg/ml)                 | >240                                                                  |
| 6. | Cyclophosphamide (20 mg/ml)            | >240                                                                  |
| 7. | Dacarbazine (10 mg/ml)                 | >240                                                                  |
| 8. | Doxorubicin HCl (2.0 mg/ml)            | >240                                                                  |
| 9. | Ellence (25 mg/ml)                     | >240                                                                  |

(Continued on next page)

X Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clavine-Williams

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K110272

Page / of 2

5

{7}------------------------------------------------

| Continued from previous page) |                                        |                                                                       |
|-------------------------------|----------------------------------------|-----------------------------------------------------------------------|
|                               | Chemotherapy Drug and<br>Concentration | Minimum Breakthrough Detection Time<br>in Minutes, 0.01 µg/cm²/minute |
| 10.                           | Etoposide (Toposar) (20 mg/ml)         | >240                                                                  |
| 11.                           | Fludarabine (25 mg/ml)                 | >240                                                                  |
| 12.                           | Fluorouracil (50 mg/ml)                | >240                                                                  |
| 13.                           | Idarubicin (1 mg/ml)                   | >240                                                                  |
| 14.                           | Ifosfamide (50 mg/ml)                  | >240                                                                  |
| 15.                           | Mechlorethamine HCl (1 mg/ml)          | >240                                                                  |
| 16.                           | Melphalan (5 mg/ml)                    | >240                                                                  |
| 17.                           | Methotrexate (25 mg/ml)                | >240                                                                  |
| 18.                           | Mitoxantrone (2 mg/ml)                 | >240                                                                  |
| 19.                           | Mitomycin C (0.5 mg/ml)                | >240                                                                  |
| 20.                           | Paclitaxel (Taxol) (6.0 mg/ml)         | >240                                                                  |
| 21.                           | Paraplatin (10 mg/ml)                  | >240                                                                  |
| 22.                           | Rituximab (10 mg/ml)                   | >240                                                                  |
| 23.                           | Thiotepa (10 mg/ml)                    | 0.44                                                                  |
| 24.                           | Trisenox (0.1 mg/ml)                   | >240                                                                  |
| 25.                           | Vincristine sulfate (1 mg/ml)          | >240                                                                  |

Please note that the following drugs have extremely low permeation time of less than 30 minutes:

Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 0.37 minute.

Thiotepa (10 mg/ml) has a minimum breakthrough time of 0.44 minute.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use __X (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clausing-Williams

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices l

510(k) Number: K110272

Page 2 of 2

ర్

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K110272](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K110272)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
