← Product Code [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO) · K093300

# ESTEEM SMT POLYISOPRENE POWDER-FREE SURGICAL STERILE GLOVES (K093300)

_Cardinal Health, Inc. · KGO · Oct 30, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K093300

## Device Facts

- **Applicant:** Cardinal Health, Inc.
- **Product Code:** [KGO](/submissions/SU/subpart-e%E2%80%94surgical-devices/KGO.md)
- **Decision Date:** Oct 30, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery
- **Attributes:** 3rd-Party Reviewed

## Intended Use

Esteem SMT Polyisoprene Powder-Free Surgical Gloves are a disposable device that is intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities.

## Device Story

Disposable surgical gloves; manufactured from synthetic rubber latex; powder-free; sterile. Worn by operating room personnel to provide a protective barrier between the wearer and the surgical site; prevents contamination of surgical wounds. Used in hospitals and healthcare facilities. Benefits include protection against cross-contamination during surgical procedures.

## Clinical Evidence

No clinical data was required. Bench testing included primary skin irritation, guinea pig maximization for sensitization, dimensions, physical characteristics, freedom from holes (21 CFR 800.20), and residual powder testing (ASTM D6124). All results met ASTM D3577 requirements.

## Technological Characteristics

Synthetic rubber latex; powder-free; sterile. Dimensions and physical properties conform to ASTM D3577. Residual powder levels <2mg per glove (ASTM D6124). Biocompatibility confirmed via primary skin irritation and sensitization testing.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Esteem Polyisoprene Powder-Free Surgical Gloves ([K011721](/device/K011721.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 093300

Page 1 of 2

Image /page/0/Picture/2 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The text "CardinalHealth" is written in a bold, sans-serif font, with the first letter of each word capitalized.

# OCT 8 0 2009

### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## ESTEEM SMT POLYISOPRENE POWDER-FREE SURGICAL GLOVES

| Manufacturer:               | Cardinal Healthcare 222 LTD.                                                                                                                                                                                                                             |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                             | 7/111 Moo 4, Highway 331 Mabyangporn Pluakdaeng,                                                                                                                                                                                                         |
|                             | Rayong, 21140, Thailand                                                                                                                                                                                                                                  |
| Regulatory Affairs Contact: | Tatyana Bogdan, RAC                                                                                                                                                                                                                                      |
|                             | Cardinal Health, Inc.                                                                                                                                                                                                                                    |
|                             | 1430 Waukegan Road                                                                                                                                                                                                                                       |
|                             | McGaw Park, IL 60085                                                                                                                                                                                                                                     |
| Telephone:                  | 847-578-2325                                                                                                                                                                                                                                             |
| Date Summary Prepared:      | October 14, 2009                                                                                                                                                                                                                                         |
| Product Trade Name:         | Esteem SMT Polyisoprene Powder-Free Surgical Gloves                                                                                                                                                                                                      |
| Common Name:                | Surgeon's Gloves                                                                                                                                                                                                                                         |
| Classification Name:        | Surgeon's Gloves                                                                                                                                                                                                                                         |
| Device Description:         | Esteem SMT Polyisoprene Powder-Free Surgical gloves are formulated using synthetic rubber latex. They are offered powder-free and sterile.                                                                                                               |
| Intended Use:               | Esteem SMT Polyisoprene Powder-Free Surgical Gloves are a disposable device that is intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities. |
| Predicate Devices:          | Esteem Polyisoprene Powder-Free Surgical Gloves previously cleared under 510(k) K011721 (product code KGO);                                                                                                                                              |
| Substantial Equivalence:    | Esteem SMT Polyisoprene Powder-Free Surgical gloves are substantially equivalent to Esteem Polyisoprene Powder-Free                                                                                                                                      |

·

5

{1}------------------------------------------------

K093300

page 2 of 2

Surgical gloves. Substantial equivalence can be established in regard to intended use, physical characteristics, design and product features. Both gloves are made of synthetic rubber latex using similar manufacturing process and packaging.

### Performance Testing:

| Test:                    | Result:                                                                                                                                                                                                                                                       |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Skin Irritation  | Gloves are non-irritating.                                                                                                                                                                                                                                    |
| Guinea Pig Maximization  | Gloves do not display any potential for sensitization.                                                                                                                                                                                                        |
| Dimensions               | Gloves meet requirements of ASTM D3577.                                                                                                                                                                                                                       |
| Physical Characteristics | Gloves meet requirements for rubber surgical gloves per ASTM<br>D3577.                                                                                                                                                                                        |
| Freedom from Holes       | Gloves meet requirements of 21 CFR 800.20 and ASTM D3577                                                                                                                                                                                                      |
| Powder Residual          | Gloves meet powder level requirements for "Powder-Free"<br>designation per ASTM D3577 tested using ASTM standard<br>D6124, Standard test method for residual powder on medical<br>gloves. Results generated values below 2mg of residual powder<br>per glove. |
| Clinical Data:           | No clinical data was required.                                                                                                                                                                                                                                |
| Conclusion:              | The Esteem SMT Polyisoprene Powder-Free Surgical Gloves meet<br>the technological characteristics of ASTM D3577 and are<br>substantially equivalent to predicate devices identified in this<br>510(k) summary.                                                |

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is centered in the image. The font is a serif font, and the text is black. The background is white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Cardinal Health, Incorporated C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories Incorporated 333 Pfingsten Road Northbrook, Illinois 60062-2096

# OCT 3 0 2009

Re: K093300

Trade/Device Name: Esteem SMT Polyisoprene Powder-Free Surgical Sterile Gloves Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: October 19, 2009 Received: October 21, 2009

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2- Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthem Orenton

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# FDA CDRH DMC

## OCT 21 2009

### Received

# Indications for Use

# 510(k) Number (if known): _ | 093300

| Device Name: | Esteem SMT Polyisoprene Powder-Free Surgical Sterile Gloves |
|--------------|-------------------------------------------------------------|
|--------------|-------------------------------------------------------------|

Indications for Use:

These powder-free surgeon's gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

|  | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--|--------------------------------------------------------|
|--|--------------------------------------------------------|

Shippy to the paris Wills

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K O 93300

Cardinal Health, Inc. Premarket Notification Submission - Traditional 510(k)

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K093300](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/KGO/K093300)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
