Medline Orbis Surgical Gown

K202447 · Medline Industries, Inc. · FYA · Apr 2, 2021 · General, Plastic Surgery

Device Facts

Record IDK202447
Device NameMedline Orbis Surgical Gown
ApplicantMedline Industries, Inc.
Product CodeFYA · General, Plastic Surgery
Decision DateApr 2, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The Medline Orbis Surgical Gown is a single use, Level 4 surgical gown intended to be worn by healthcare professionals during surgical procedures to help protect both the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Orbis Surgical Gown meets the Level 4 requirements in accordance with ANSI/AAMIPB70:2012 "Liquid barrier and performance classification of protective apparel and drapes intended for use in healthcare facilities".

Device Story

Medline Orbis Surgical Gown is a single-use, Level 4 protective garment. Worn by healthcare professionals in surgical settings to prevent cross-contamination of microorganisms, body fluids, and particulates between the wearer and the surgical field. Device functions as a physical barrier; no electronic or mechanical components. Provides protection per ANSI/AAMI PB70:2012 standards.

Clinical Evidence

No clinical data; bench testing only per ANSI/AAMI PB70:2012 requirements for Level 4 liquid barrier performance.

Technological Characteristics

Single-use surgical gown; Level 4 barrier protection; meets ANSI/AAMI PB70:2012 standards for liquid barrier performance and classification of protective apparel.

Indications for Use

Indicated for healthcare professionals during surgical procedures to provide barrier protection against microorganisms, body fluids, and particulate matter.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. The logo is simple and professional, and it is easily recognizable. April 2, 2021 Medline Industries, Inc. Adam Ostrower Regulatory Affairs Sr. Specialist Three Lakes Drive Northfield. Illinois 60093 ## Re: K202447 Trade/Device Name: Medline Orbis Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: March 24, 2021 Received: March 25, 2021 ## Dear Adam Ostrower: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. # Clarence W. Murray III -S For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202447 Device Name Medline Orbis Surgical Gown #### Indications for Use (Describe) The Medline Orbis Surgical Gown is a single use, Level 4 surgical gown intended to be worn by healthcare professionals during surgical procedures to help protect both the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Orbis Surgical Gown meets the Level 4 requirements in accordance with ANSI/AAMIPB70:2012 "Liquid barrier and performance classification of protective apparel and drapes intended for use in healthcare facilities". | Type of Use (Select one or both, as applicable) | <table border="0"><tr><td><table border="0"><tr><td></td></tr></table></td><td><table border="0"><tr><td></td></tr></table></td></tr></table> | <table border="0"><tr><td></td></tr></table> | | <table border="0"><tr><td></td></tr></table> | | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--|----------------------------------------------|--| | <table border="0"><tr><td></td></tr></table> | | <table border="0"><tr><td></td></tr></table> | | | | | | | | | | | | | | | | | | | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...