← Product Code [FYA](/submissions/SU/subpart-e%E2%80%94surgical-devices/FYA) · K201351

# Maxima Surgical Gown (K201351)

_Elevate Textiles · FYA · Mar 11, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FYA/K201351

## Device Facts

- **Applicant:** Elevate Textiles
- **Product Code:** [FYA](/submissions/SU/subpart-e%E2%80%94surgical-devices/FYA.md)
- **Decision Date:** Mar 11, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The reusable Maxima Surgical Gown is intended to be used by operating room personnel during surgical procedures to help protect both the patient and the health care worker from transfer of micro-organisms, body fluids, and particulate material.

## Device Story

Maxima Surgical Gown; reusable woven surgical gown; 100% filament polyester; provides AAMI Level 3 liquid barrier protection in critical zones and back; sleeves feature permanent seam tape. Used in operating rooms by surgical personnel. Provided non-sterile; requires laundering and steam sterilization (135°C/274°F for 3 minutes) by healthcare facilities or contract services before first and subsequent uses. Protects against fluid/particulate transfer; supports sterile surgical environment.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via biocompatibility (ISO 10993-5, ISO 10993-10), water resistance (AATCC 127, AATCC 42), mechanical strength (ASTM D5034, ASTM D5587, ASTM D1683), flammability (16 CFR 1610), and lint generation (EN ISO 9073-10).

## Technological Characteristics

100% woven polyester; AAMI Level 3 liquid barrier performance per ANSI/AAMI PB70:2012. Mechanical specs: ASTM D5034 (breaking strength), ASTM D5587 (tear strength), ASTM D1683 (seam strength), ASTM D3776 (mass/area). Flammability: 16 CFR 1610 Class 1. Sterilization: Prevac steam autoclave. Reusable up to 75 cycles.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Blockade Surgical Gown ([K150598](/device/K150598.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 11, 2022

Elevate Textiles Nathaniel Terry Technical Director 22 American Street Mt. Holly, North Carolina 28210

Re: K201351

Trade/Device Name: Maxima Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FYA Dated: December 8, 2021 Received: December 13, 2021

Dear Nathaniel Terry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201351

Device Name Maxima Surgical Gown

Indications for Use (Describe)

The reusable Maxima Surgical Gown is intended to be used by operating room personnel during surgical procedures to help protect both the patient and the health care worker from transfer of micro-organisms, body fluids, and particulate material.

The reusable Maxima Surgical Gown is provided non-sterile. Sterilization parameters are as follows: * Prevac Steam Sterilization - 135°C/274°F (temperature) / 3 minutes (exposure time) / 20 minutes (dry time)

Elevate Textiles Item Numbers / gown color / liquid barrier protection / gown size MAXAL3S / Royal Blue B103 / AAMI Level 3 / Small MAXAL3M / Royal Blue B103 / AAMI Level 3 / Medium MAXAL3L / Royal Blue B103 / AAMI Level 3 / Large

| Type of Use (Select one or both, as applicable) |                                                                                     |
|-------------------------------------------------|-------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CER 801 Subpart D)    |                                                                                     |
|                                                 | <div> <span> <b>X</b> Over-The-Counter Use (21 CER 801 Subpart C)    </span> </div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K201351)

## [As required by section 807.92(c)]

#### Submitter:

 Elevate Textiles 22 American Street Mt Holly NC, 28120

#### Contact Person:

Nathaniel Terry Technical Director / Regulatory Affairs Telephone: 336-684-1683 Email: Nathaniel.terry@elevatetextiles.com

#### Summary Preparation Date

December 1, 2021

Type of Submission Traditional 510(k)

#### Device Name / Classification

| Trade Name:          | Maxima® Surgical Gown  |
|----------------------|------------------------|
| Common Name:         | Sterile Surgical Gown  |
| Classification Name: | Surgical Gown, Class 2 |

#### Predicate Device - Maxima® Surgical Gown claims substantial equivalence to:

K150598 Blockade Surgical Gown manufactured by Medline Industries, Inc

#### Device Description

The Maxima® Surgical Gown is a reusable woven surgical gown for use in sterile conditions. It is made of 100% filament polyester that provides AAM PB70:2012 Level 3 Liquid Barrier Performance Barrier in the critical zone up to 75 reprocessing cycles. The back of the gown provides AAMI Level 3 Liquid Barrier Performance Barrier up to 75 reprocessing cycles. The seam in the sleeves are sealed with a permanent seam tape which seals the stitch area enabling it to perform as AAMI Level 3 up to 75 reprocessing cycles.

Maxima® Surgical Gown will be manufactured in small, medium, and larges sizes. Maxima® Surgical Gowns will be sold as non-sterile reusable surgical gowns that are to be laundered, sterilized and processed by health care facilities and or contract laundry and sterliization companies before first use and after each subsequent use.

#### Indications for Use

The Maxima Surgical Gown is a reusable surgical gown intended to be used by operating room personnel during surgical procedures to help protect both the patient and the health care worker of micro-organisms, body fluids, and particulate material.

The Maxima Surgical Gown is provided non-sterilization parameters are as follows:

* Prevac Steam Sterilization - 135°C/274°F (temperature) / 3 minutes (exposure time) / 20 minutes (dry time)

Elevate Textiles Item Numbers / Gown Color / Liquid Barrier Protection / Gown Size

MAXAL3S / Royal B103 / AAMI Level 3 / Small MAXAL3M MAXAL3L

{4}------------------------------------------------

# Summary of Technological Characteristics

| Device<br>Characteristic      | Proposed Device                                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                                                        | Comparison<br>Analysis |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name                  | Maxima® Surgical Gown                                                                                                                                                                                                                 | Blockade Surgical Gown                                                                                                                                                                                                                  | N/A                    |
| 510(k) Reference              | K201351                                                                                                                                                                                                                               | K150598                                                                                                                                                                                                                                 | N/A                    |
| Product Owner                 | Elevate Textiles                                                                                                                                                                                                                      | Medline Industries Inc.                                                                                                                                                                                                                 | N/A                    |
| Product Code                  | FYA                                                                                                                                                                                                                                   | FYA                                                                                                                                                                                                                                     | same                   |
| Intended Use                  | Intended to be used by<br>operating room personnel<br>during surgical procedures to<br>help protect both the patient<br>and the health care worker<br>from transfer of micro-<br>organisms, body fluids, and<br>particulate material. | Intended to be used by<br>operating room personnel<br>during surgical procedures<br>to help protect both the<br>patient and the health care<br>worker from transfer of<br>micro-organisms, body<br>fluids, and particulate<br>material. | same                   |
| Regulation Number             | 21 CFR 878.4040                                                                                                                                                                                                                       | 21 CFR 878.4040                                                                                                                                                                                                                         | same                   |
| Color                         | Royal B103                                                                                                                                                                                                                            | Ceil, Jade                                                                                                                                                                                                                              | different              |
| Design Features               | belt ties<br>knit cuffs<br>neck closure: metal snaps                                                                                                                                                                                  | belt ties<br>knit cuffs<br>neck closure: metal snaps                                                                                                                                                                                    | same                   |
| Sizes                         | Small, Medium, Large                                                                                                                                                                                                                  | Large, X-Large XX-Large                                                                                                                                                                                                                 | similar                |
| Materials                     | Woven polyester                                                                                                                                                                                                                       | Woven polyester                                                                                                                                                                                                                         | same                   |
| Performance<br>Specifications | Level 3 ANSI/AAMI PB70<br>Barrier protection                                                                                                                                                                                          | Level 3 ANSI/AAMI PB70<br>Barrier protection                                                                                                                                                                                            | same                   |
| Prescription vs. OTC          | ОТС                                                                                                                                                                                                                                   | OTC                                                                                                                                                                                                                                     | same                   |
| Contact Durations             | Surface, Breached or<br>Compromised, < 24 hours                                                                                                                                                                                       | Surface, Breached or<br>Compromised, < 24 hours                                                                                                                                                                                         | same                   |
| Sterile vs. non-<br>sterile   | Sterile when used                                                                                                                                                                                                                     | Sterile when used                                                                                                                                                                                                                       | same                   |
| Single Use vs.<br>Reusable    | reusable                                                                                                                                                                                                                              | reusable                                                                                                                                                                                                                                | same                   |
| Biocompatibility              | Under the test conditions, the<br>subject device was shown to<br>be non-cytotoxic, non-irritating<br>and non-sensitizing per ISO<br>10993-5 & ISO 10993-10.                                                                           | Met requirements per: ISO<br>10993-5 Cytotoxicity ISO<br>10993-10 Irritation ISO<br>10993-10 Sensitization                                                                                                                              | same                   |
| Flammability                  | Meets the requirements of<br>CPSC 16 CFR 1610 Class 1                                                                                                                                                                                 | Meets the requirements of<br>CPSC 16 CFR 1610 Class 1                                                                                                                                                                                   | same                   |
| Lint Level                    | Log10 (lint count) 1.89                                                                                                                                                                                                               | Log10 (lint count) 1.89                                                                                                                                                                                                                 | same                   |
| Tensile Strength              | Warp: 138.9 Ibf<br>Fill: 82.01 Ibf                                                                                                                                                                                                    | Warp: 138.9 Ibf<br>Fill: 82.01 Ibf                                                                                                                                                                                                      | same                   |
| Seam Strength                 | Warp: 78 Ibf<br>Fill: 65.4 Ibf                                                                                                                                                                                                        | Warp: 78 Ibf<br>Fill: 65.4 Ibf                                                                                                                                                                                                          | same                   |
| Tear Resistance               | Warp: 25.3 Ibf<br>Fill: 7.4 Ibf                                                                                                                                                                                                       | Warp: 25.3 Ibf<br>Fill: 7.4 Ibf                                                                                                                                                                                                         | same                   |
| Sterilization Method          | Prevac steam autoclave                                                                                                                                                                                                                | Prevac steam autoclave                                                                                                                                                                                                                  | same                   |

{5}------------------------------------------------

#### Summary of Non-Clinical Testing

Performance testing was performed to verify that the acceptance criteria. The testing done on the Maxima® Surgical Gown was conducted to demonstrate the safety and effectiveness of the subject device with the relevant test methods cited below, including the appropriate biocompatibility test.

| Test Conducted                                                        | Test Standard                                                                                                                  | Acceptance Criteria | Test Result                                               |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------|-----------------------------------------------------------|
| Cytotoxicity ISO MEM<br>Elution Using L-929<br>Mouse Fibroblast Cells | ISO 10993-5 Cytotoxicity:  Biological Evaluation Of Medical<br>Devices - Part 5: Tests For In Vitro Cytotoxicity               | non-cytotoxic       | non-cytotoxic                                             |
| Skin Irritation ISO<br>Intracutaneous Irritation<br>Test              | ISO 10993-10 Irritation: Biological Evaluation of Medical<br>Devices - Part 10: Tests For Skin Sensitization                   | non-irritating      | non-irritating                                            |
| Skin Sensitization ISO<br>Guinea Pig Buehler<br>Sensitization Test    | ISO 10993-10 Sensitization: Biological Evaluation of Medical<br>Devices - Part 10: Tests For Skin Sensitization                | non-sensitizing     | non-sensitizing                                           |
| Water Resistance                                                      | AATCC 127: Test Method for Water Resistance: Hydrostatic<br>Pressure                                                           | >50 cm AQL 4.0      | Avg. 70.3 cm                                              |
| Water Resistance                                                      | AATCC 42: Test Method for Water Resistance: Impact<br>Penetration                                                              | <1.0 g AQL 4.0      | Mean 0.01 g                                               |
| Breaking Strength                                                     | ASTM D5034: Standard Test Method For Breaking Strength<br>And Elongation Of Textile Fabrics (Grab Test)                        | >30 lbf             | Warp: 138.9 lbf<br>Fill: 82.01 lbf                        |
| Tear Strength                                                         | ASTM D 5587: Standard Test Method for Tearing Strength of<br>Fabrics by Trapezoid Procedure                                    | >1.5 lbf            | Warp: 25.3 lbf<br>Fill: 7.4 lbf                           |
| Seam Strength                                                         | ASTM D1683: Standard Test Method For Failure In Sewn<br>Seams Of Woven Fabrics                                                 | >30 lbf             | Warp: 78 lbf<br>Fill: 65.4 lbf                            |
| Weight                                                                | ASTM D3776/D3776M-09a: Standard Test Methods For Mass<br>Per Unit Area (Weight) Of Fabric                                      | Documentation Only  | 88 gsm                                                    |
| Flammability                                                          | 16 CFR 1610: Standard for the Flammability of Clothing<br>Textiles                                                             | Class 1             | Class 1                                                   |
| Lint Generation                                                       | EN ISO 9073-10: Textiles - Test methods for nonwovens -<br>Part 10: Lint and other particles generation in the dry state       | Documentation Only  | log10 (lint count) 1.89                                   |
| Evaporative Resistance                                                | ASTM F1868: Standard Test Method for Thermal and<br>Evaporative Resistance of Clothing Materials Using a Sweating<br>Hot Plate | Documentation Only  | Ret Critical Zone 0.01193<br>Ret Noncritical Zone 0.01157 |

#### Summary of Clinical Testing

Not applicable

#### Conclusion

Based on the information provided in this premarket notification, Elevate Textiles concludes that the Maxima® Surgical Gown is as safe, as effective, and performs as well as or better than the legally marketed predicate device K150598 Medline Blockade Surgical gown.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FYA/K201351](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FYA/K201351)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
