COMPLEAT SURGICAL MASK
Device Facts
| Record ID | K962654 |
|---|---|
| Device Name | COMPLEAT SURGICAL MASK |
| Applicant | Carl Parker Associates, Inc. |
| Product Code | FXX · General, Plastic Surgery |
| Decision Date | Jan 27, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4040 |
| Device Class | Class 2 |
Intended Use
to act as physical barrier including fluid-resistant properties, provide only very low impediment to breathing, avoid fogging of surgeons' glasses, effective filtration of particles, droplets, not hinder speech.
Device Story
Disposable surgical mask; three-layer construction; non-woven polypropylene outer/inner layers; 'Web Dynamics' air filtration media middle layer. Designed for nose/mouth coverage; flat-fold arcuate segment design. Used in clinical/surgical settings by healthcare personnel. Functions as physical barrier; provides fluid resistance; minimizes spectacle fogging; facilitates breathing. Benefits patient/clinician by reducing exposure to airborne particles and body fluids during procedures.
Clinical Evidence
Bench testing only. Testing included Synthetic Blood Penetration Resistance (Protocol 9617002-01), Bacterial Filtration Efficiency (BFE), Delta P, Primary Dermal Irritation, Particulate Shedding analysis, flammability, and cytotoxicity.
Technological Characteristics
Three-layer flat-fold design; non-woven polypropylene layers; 'Web Dynamics' air filtration media. Disposable. No energy source. No software.
Indications for Use
Indicated for use as a surgical mask to provide a physical barrier against fluids, particles, and droplets for clinicians and surgical staff.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
- Fluid Resistant Masks, Busse Hosp Disposables (K830890)
- Fluid Shield Procedure Mask, Technol Inc. (K932137)
- Fluid Resistant Mask, Mid America Medical, Inc. (K934974)
Related Devices
- K210267 — Disposable Surgical Mask · Xiamen Eagledon Pharmaceutical Co., Ltd. · Jul 8, 2021
- K971130 — DEFEND FLUID RESISTANT PLEATED MASK · Carl Parker Associates, Inc. · Jun 27, 1997
- K212944 — Disposable Surgical Mask · Heilongjiang Yinjia Medical Equipment Co., Ltd. · Dec 15, 2021
- K211451 — Disposable Surgical Mask (non sterile) · Qinhuangdao Taizhi Medical Technology Co., Ltd. · Aug 25, 2021
- K210222 — Disposable Medical Mask · Guangdong Lide Medical Technology Co., Ltd. · Nov 2, 2021