← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K955557

# FLUID REISISTANT EYESHIELD MASK (K955557)

_American Threshold Healthcare · FXX · Apr 25, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K955557

## Device Facts

- **Applicant:** American Threshold Healthcare
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Apr 25, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

This product is intended for use in infection control practices to minimize contamination caused by exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids.

## Device Story

Fluid Resistant Eyeshield Mask provides barrier protection for healthcare workers. Device consists of three-layer construction: spunbonded polyolefin or cellulosic outer/inner facings with meltblown polyolefin filter media. Layers are sonically welded; includes malleable aluminum nose wire for fit and optical quality eyeshield attached via ultrasonic welding. Worn by clinicians in surgical or clinical settings to reduce exposure to blood, body fluids, and exhaled microorganisms. Device filters particles at >95% efficiency (1.0 and 0.1 micron) and maintains breathability (average pressure drop ≤ 5.0 mm water).

## Clinical Evidence

Bench testing only. Testing included: 1) Water Resistance (Impact Penetration) per Nelson Labs SOP#MBG/037; 2) Differential Pressure (Delta P) per Nelson Labs SOP#ARO/006; 3) Latex Particle Challenge per Nelson Labs SOP#ARO/011. Results: Average Delta P 2.19-2.54 mm water; Filtration efficiency >98% for 0.1 and 1.0 micron particles.

## Technological Characteristics

Materials: Spunbonded polyolefin, cellulosic nonwoven, meltblown polyolefin filter media, aluminum nose wire. Construction: Three-layer laminate, sonically welded edges. Form factor: Mask with integrated optical quality eyeshield. Performance: >95% particle filtration efficiency, ≤ 5.0 mm water pressure drop.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Tecnol Fluid Resistant Surgical Mask ([K874608](/device/K874608.md))
- 3M Fluid Resistant Mask ([K910110](/device/K910110.md))

## Reference Devices

- American Threshold Surgical Mask ([K801036](/device/K801036.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K955557
510 (k) SUMMARY
APR 25 1996

Submitted by: American Threshold Industries
240 Sardis Road
Enka, N.C. 28828

Contact Person: Frances Mayhew, Quality Assurance Manager

Date of Submission: Original - December 4, 1995

Device Name: Fluid Resistant Eyeshield Mask
Classification Name: MASK, SURGICAL

Predicate Device(s):
1. Tecnol Fluid Resistant Surgical Mask, 510 (k) #K874608, dated December 2, 1987.
2. 3M Fluid Resistant Mask, 510 (k) #K910110, dated March 20, 1991

Device Description:

The Fluid Resistant Eyeshield Masks are constructed of a light weight spunbonded polyolefin or cellulosic outer facing, a meltblown polyolefin filter media, and a light weight spunbonded polyolefin or cellulosic inner facing. The three layers of the mask body are collated and sonically welded around the edges to enclose the filter media. The bodies have either light weight polyolefin or cellulosic nonwoven material sonically welded or otherwise attached to both sides of the body (ties on the surgeons tie masks) or polyolefin/elastic earlongs sonically welded or otherwise attached to both sides of the body. The masks have a malleable aluminum wire encapsulated in the facing to form the mask to fit the nose. An optical quality eyeshield is attached to the mask by ultrasonic welding.

All of the materials used in the construction of the masks are being used in other currently marketed devices, such as our regular surgical mask, 510 (k) #K801036.

These masks filter particles at an average efficiency of greater than 95% when tested at both 1.0 micron and 0.1 micron particle sizes. The breathability of the masks (measured by the pressure drop across the mask) averages no more than 5.0 mm of water.

All of the materials used in the construction of the mask are used in other currently marketed devices, these include:

- American Threshold Surgical Mask 510(k) #K801036
- Tecnol Surgical Mask 510(k) #K874608
- 3M Surgical Mask 510(k) #K910110

{1}

K955557

# Statement of Intended Use:

This product is intended for use in infection control practices to minimize contamination caused by exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids.

# Testing Submitted to Determine Substantial Equivalence and Device Claims:

1. Water Resistance, Impact Penetration Test, Nelson Laboratories SOP#MBG/037
- This procedure is designed to measure the resistance of materials to water penetration under contact with sprayed water. This test was performed to demonstrate substantial equivalence.

2. Differential Pressure, Delta P Test, Nelson Laboratories SOP#ARO/006
- The differential pressure test determines the air exchange differential of porous materials and is important for determining the breathability of surgical face masks. This test was performed to demonstrate substantial equivalence.

3. Latex Particle Challenge, Nelson Laboratories SOP#ARO/011
- This procedure provides for evaluation of non-viable particle retention or filtration efficiency of barrier materials. This test was performed to substantiate device claims.

# Conclusions Drawn from the Results of the Above Testing:

1. Water Resistance, Impact Penetration Test Results:

American Threshold Fluid Resistant Mask:
- Average per-cent increase in blotter weight = 2.64% increase in the first study and 9.15% in the second.

Tecnol Fluid Resistant Mask:
- Average per-cent increase in blotter weight = 49.33% increase in the first study and 2.99% in the second.

2. Differential Pressure, Delta P Test Results:

American Threshold Fluid Resistant Mask:
- Differential Pressure average was 2.19 in the first study and 2.54 in the second.

Tecnol Fluid Resistant Mask:
- Differential Pressure average was 1.99 in the first study and 2.68 in the second.

{2}

K95557

3. Latex Particle Challenge Test Results:

American Threshold Fluid Resistant Mask: (with or without eyeshield)

-Average filtration efficiency for Polyolefin Inner Facing:

1.0 micron particle size = 99.3% filtration efficiency
0.1 micron particle size = 98.4% filtration efficiency

-Average filtration efficiency for Cellulosic Inner Facing:

1.0 micron particle size = 99.2% filtration efficiency
0.1 micron particle size = 99.2% filtration efficiency

Based on the above test results American Threshold Industries considers their fluid resistant mask to be substantially equivalent to the other fluid resistant masks currently on the market.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K955557](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K955557)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
