Fluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound Visor (62114); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops (62115); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116)

K232812 · Owens & Minor (O&M) Halyard, Inc. · FXX · Jun 14, 2024 · General, Plastic Surgery

Device Facts

Record IDK232812
Device NameFluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound Visor (62114); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops (62115); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116)
ApplicantOwens & Minor (O&M) Halyard, Inc.
Product CodeFXX · General, Plastic Surgery
Decision DateJun 14, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Fluidshield* 2 Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Surgical Mask is a single use, disposable device, provided non-sterile. The Fluidshield* 2 Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Procedure Mask is a single use, disposable device, provided non-sterile.

Device Story

Fluidshield* 2 masks are 4-layer, non-woven, single-use, disposable surgical and procedure masks. Surgical versions feature ties; procedure versions feature earloops and a malleable nosepiece. Some models include a wraparound visor. Masks are worn by healthcare personnel and patients to provide a barrier against microorganisms, body fluids, and particulates. They are used in clinical settings for infection control. The device acts as a physical barrier; no electronic or algorithmic components are involved. Benefits include reduced potential exposure to blood and body fluids for the wearer.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100 (Level 2), ASTM F2101 (Bacterial Filtration Efficiency >98%), ASTM F2299 (Particulate Filtration Efficiency >98%), EN 14683 (Differential Pressure <6.0 mmH2O/cm2), ASTM F1862 (Fluid Resistance 120 mmHg), 16 CFR Part 1610 (Flammability Class 1), and ISO 10993 (Biocompatibility: Cytotoxicity, Sensitization, Irritation).

Technological Characteristics

4-layer non-woven construction: polyester cellulose, spunbond polypropylene, and meltblown polypropylene filter media. Features include aluminum nose wire, polyester lycra earloops or polypropylene/spunlace ties, and optional polyester film visors. Single-use, non-sterile. Complies with ASTM F2100 Level 2, ASTM F2101, ASTM F2299, EN 14683, ASTM F1862, 16 CFR 1610, and ISO 10993 standards.

Indications for Use

Indicated for healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices. Single-use, disposable, non-sterile.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. June 14, 2024 Owens & Minor (O&M) Halyard, Inc. Angela Bunn Sr. Director, Global Regulatory Affairs 9120 Lockwood Blvd Mechanicsville, Virginia 23116 ### Re: K232812 Trade/Device Name: Fluidshield* 2 Fog-Free Surgical Mask (62113); Fluidshield* 2 Fog-Free Surgical Mask with Wraparound Visor (62114); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops (62115); Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 9, 2024 Received: May 9, 2024 Dear Angela Bunn: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/7 description: The image shows the text "Allan Guan-S" in a large, sans-serif font. The text is black and is set against a white background. There is a faint, light blue watermark in the background that is partially obscured by the text. For Bifeng Qian, M.D., Ph.D. Assistant Director {2}------------------------------------------------ DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232812 #### Device Name Fluidshield * 2 Fog-Free Surgical Mask (62113) Fluidshield * 2 Fog-Free Surgical Mask with Wraparound Visor (62114) Fluidshield * 2 Procedure Mask with SO SOFT* Earloops (62115) Fluidshield * 2 Procedure Mask with SO SOFT* Earloops and WrapAround Visor (62116) Indications for Use (Describe) The Fluidshield* 2 Surgical Mask is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Surgical Mask is a single use, disposable device, provided non-sterile. The Fluidshield* 2 Procedure Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. The Fluidshield* 2 Procedure Mask is a single use, disposable device, provided non-sterile. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) x Dver-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the Owens & Minor logo, which consists of the letters "OM" in a stylized font, with the words "Owens & Minor" to the right of the letters. The letters "OM" are in a dark red color, while the words "Owens & Minor" are in a light gray color. Below the logo, the text "510(k) Summary K232812" is written in a bold font. | Date Summary<br>was Prepared | June 14, 2024 | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Submitter | O & M Halyard, Inc.<br>1 Edison Drive<br>Alpharetta, GA 30005 | | Primary Contact for<br>this 510(k) Submission | Angela L. Bunn, RAC<br>Tel: 470-347-7147<br>Email: angela.bunn@owens-minor.com | | Marketed Common<br>Name | Surgical Mask and Procedure Mask | | Device Submission Trade<br>Name and Description | Fluidshield * 2 Fog-Free Surgical Mask (62113)<br>Fluidshield * 2 Fog-Free Surgical Mask with Wraparound Visor<br>(62114)<br>Fluidshield * 2 Procedure Mask with SO SOFT* Earloops,<br>(62115)<br>Fluidshield* 2 Procedure Mask with SO SOFT* Earloops and<br>WrapAround Visor (62116) | | Device Common Name | Surgical Mask and Procedure Mask | | Device Product Code<br>and Classification Name | FXX Class II, 21 CFR §878.4040 Surgical Apparel | | Predicate Device | KC 200 Fluidshield* Surgical Masks and Procedure Masks and<br>KC 300 Fluidshield* Surgical Masks and Procedure Masks<br>(K111402) | | Subject Device<br>Description | The subject device(s) are ASTM Level 2 Surgical and<br>Procedure Mask(s). Fluidshield* 2 Surgical Mask(s),<br>Fluidshield* 2 Procedure Mask(s) family of products are a four-<br>layer mask, constructed of well-known non-woven | | | materials used in facial protection. The Surgical Masks are<br>provided with ties while the Procedure Masks are provided with<br>earloops and a malleable nosepiece.<br><br>The Fluidshield* 2 Surgical Mask(s), Fluidshield* 2 Procedure<br>Mask(s) family will be provided in designs with and without a<br>protective visor. The Fluidshield* 2 Surgical Mask(s),<br>Fluidshield* 2 Procedure Mask(s) family are single use,<br>disposable devices, provided non-sterile. | | Indications for Use | The Fluidshield* 2 Surgical Mask is intended to be worn to protect<br>both the patient and healthcare personnel from transfer of<br>microorganisms, body fluids, and particulate material. These<br>face masks are intended for use in infection control practices to<br>reduce the potential exposure of the wearer to blood and body<br>fluids. The Fluidshield* 2 Surgical Mask is a single use,<br>disposable device, provided non-sterile.<br><br>The Fluidshield* 2 Procedure Mask is intended to be worn to<br>protect both the patient and healthcare personnel from transfer of<br>microorganisms, body fluids, and particulate material. These<br>face masks are intended for use in infection control practices to<br>reduce the potential exposure of the wearer to blood and body<br>fluids. The Fluidshield* 2 Procedure Mask is a single use,<br>disposable device, provided non-sterile. | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Owens & Minor. The logo features a stylized "OM" in a dark red color, with the words "Owens & Minor" in a gray serif font to the right of the symbol. The logo is clean and professional, suggesting a well-established company. | Technological Characteristics Comparison Table | | | | |------------------------------------------------|-----------------|------------------------------------------------------------------|------------| | | Subject Device | Predicate Device<br>(K111402)<br>KC 300 and KC 200<br>Face Masks | Comparison | | FDA Product Code | FXX | FXX | Same | | FDA Classification | Class II | Class II | Same | | Regulation Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | Common Name | Surgical Mask | Surgical Mask | Same | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Owens & Minor, a healthcare logistics company. The logo consists of a stylized "OM" in a dark red color, with the words "Owens & Minor" in a light gray color to the right of the "OM" symbol. The "OM" symbol is made up of several horizontal lines, creating a modern and abstract design. The overall design is clean and professional, reflecting the company's focus on healthcare and logistics. | Device Trade Name | Fluidshield * 2 Fog-Free<br>Surgical Mask (62113) | KC200 Procedure Mask | Similar | |----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------| | | Fluidshield * 2 Fog-Free<br>Surgical Mask with<br>Wraparound Visor (62114) | KC200 Procedure Mask,<br>fog-free with visor | | | | Fluidshield * 2 Procedure<br>Mask with SO SOFT*<br>Earloops, (62115) | KC200 Surgical Mask,<br>fog-free | | | | Fluidshield* 2 Procedure<br>Mask with SO SOFT*<br>Earloops and WrapAround<br>Visor (62116) | KC200 Surgical Mask,<br>fog-free with visor | | | | | KC300 Procedure Mask,<br>Fluidshield*, fog-free | | | | | KC300 Procedure Mask,<br>Fluidshield*, fog-free with<br>visor | | | | | KC300 Surgical Mask,<br>Fluidshield*, fog-free | | | | | KC300 Surgical Mask,<br>Fluidshield*, fog-free with<br>visor | | | Intended Use | The Fluidshield* 2 Surgical<br>Mask is intended to be worn<br>to protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure of the wearer to<br>blood and body fluids. The<br>Fluidshield* 2 Surgical<br>Mask is a single use, | The Kimberly-Clark,<br>KC200 and KC300 Face<br>Mask(s) are intended to.be<br>worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices<br>to reduce the potential<br>exposure of the wearer to<br>blood and body fluids. The<br>Kimberly-Clark, KC200 | Same<br>This<br>submission<br>covers the KC<br>300 and KC<br>200 Blue<br>Surgical or<br>Procedure<br>Masks. | | | device, provided non-sterile.<br>The Fluidshield* 2<br>Procedure Mask is intended<br>to be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure of the wearer to<br>blood and body fluids. The<br>Fluidshield* 2 Procedure<br>Mask is a single use,<br>disposable device, provided<br>non-sterile. | and KC300 face mask(s) is<br>a single use, disposable<br>device(s), provided non-<br>sterile. | | | Materials of Construction for Surgical Mask Family (Codes 62113 and 62114) | | | | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of the letters "OM" in a stylized font, with the "O" made up of horizontal lines. To the right of the letters is the text "Owens & Minor" in a serif font. The "OM" is in a dark red color, while the "Owens & Minor" text is in gray. | Structure | Material of Layer for<br>K232812 | Material cleared under<br>K111402 | Same or Different | |--------------------------------------|------------------------------------------|----------------------------------------------------|-------------------| | Code 62113 Materials of Construction | | | | | Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | Same | | 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | Same | | Media Filter | Meltblown polypropylene | Meltblown polypropylene | Same | | Inner Layer | Polyester Cellulose | Polyester Cellulose | Same | | Top Binding | Spunbond polypropylene | Spunbond polypropylene,<br>Polyester spunlace | Same | | Lower Binding | Spunbond polypropylene | Spunbond polypropylene,<br>Polyester spunlace | Same | | Ties (Surgical) | Spunbond polypropylene | Polyester spunlace | Different | | Foam | Polyester/Polyester<br>Spunlace laminate | Polyester/Polyester<br>Spunlace laminate | Same | | Foam | Polyester | Polyester | Same | | Nose Wire | Aluminum | Aluminum | Same | | Code 62114 Materials of Construction | | | | | Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | Same | | 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | Same | | Media Filter | Meltblown polypropylene | Meltblown polypropylene | Same | | Inner Layer | Polyester Cellulose | Polyester Cellulose | Same | | Top Binding | Spunbond polypropylene | Spunbond polypropylene<br>(or) Polyester spunlace | Same | | Lower Binding | Polyester spunlace | Spunbond polypropylene<br>(or) Polyester spunlace | Same | | Ties (Surgical) | Polyester spunlace | Polyester spunlace | Same | | Foam | Polyester | Polyester/Polyester | Same | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of the letters "OM" in a stylized font, with the word "Owens" above the word "Minor" to the right of the letters. The letters "OM" are in a dark red color, while the words "Owens" and "Minor" are in a gray color. | Visor | Polyester Film | Spunlace laminate | | | Same | |--------------------------------------------------------------------------|-----------------------------------------------|----------------------------------------------------|--|--|-------------------| | Nose Wire | Aluminum | Aluminum | | | Same | | Materials of Construction for Procedure Mask Family (Codes 62115, 62116) | | | | | | | Structure | Material of Layer for<br>K232812 | Material cleared under<br>K111402 | | | Same or Different | | | Code 62115 Materials of Construction | | | | | | Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | | | Same | | 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | | | Same | | Media Filter | Meltblown polypropylene | Meltblown polypropylene | | | Same | | Inner Layer | Polyester Cellulose | Polyester Cellulose | | | Same | | Top Binding | Polyester spunlace | Polyester spunlace | | | Same | | Lower Binding | Polyester spunlace | Polyester spunlace | | | Same | | Earloops (Procedure) | Polyester lycra | Polyester lycra | | | Same | | Nose Wire | Aluminum | Aluminum | | | Same | | | Code 62116 Materials of Construction | | | | | | Outer Layer | Polyester Cellulose | Spunbond polypropylene<br>(or) Polyester Cellulose | | | Same | | 2nd Layer | Spunbond polypropylene | Spunbond polypropylene | | | Same | | Media Filter | Meltblown polypropylene | Meltblown polypropylene | | | Same | | Inner Layer | Polyester Cellulose | Polyester Cellulose | | | Same | | Top Binding | Polypropylene spunbond | Polyester spunlace | | | Different | | Lower Binding | Polyester spunlace | Polyester spunlace | | | Same | | Earloops (Procedure) | Polyester lycra | Polyester lycra | | | Same | | Foam | Polyester foam/polyester<br>spunlace laminate | Polyester foam/polyester<br>spunlace laminate | | | Same | | Visor | Polyester Film | Polyester Film | | | Same | | Nose Wire | Aluminum | Aluminum | | | Same | | | Design Attributes for Surgical Masks | | | | | | # of Layers | 4-layer | 4-layer | | | Same | | Color | Blue with Orange Print | Blue with Orange Print | | | Same | | Ties (Surgical) | Yes | Yes | | | Same | | Visor | Some designs contain a<br>visor | Some designs contain a<br>visor | | | Same | | Nose Wire | Enclosed | Enclosed | | | Same | | | Design Attributes for Procedure Masks | | | | | | # of Layers | 4-layer | 4-layer | | | Same | | Color | Blue | Blue | | | Same | | Earloops (Procedure) | Yes | Yes | | | Same | | Visor | Some designs contain a<br>visor | Some designs contain a<br>visor | | | Same | | Performance Data/Product Claims | | | | | | | Fluid Resistance Level | 2 | 2 and 3 | | | Same | | EN 14683 | Yes | No | | | Different | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the Owens & Minor logo above a table. The table has two rows and two columns. The first row indicates "Single Use Device" and "Yes", while the second row indicates "Sterility" and "Non-Sterile". | Single Use Device | Yes | Yes | Same | |-------------------|-------------|-------------|------| | Sterility | Non-Sterile | Non-Sterile | Same | ## Non-Clinical Performance Testing Standards Non-clinical tests were conducted to verify that the proposed device(s) met all design specifications as was same/similar to the predicate device(s). The test results demonstrated that the proposed device complies with the following standards: | Purpose | Test | Acceptance Criteria | Result | |-----------------------------------|------------------|---------------------|--------| | Level 2 - Face Mask Performance | ASTM F2100 | ASTM Level 2 | Pass | | Bacterial Filtration Efficiency | ASTM F2101 | >98% | Pass | | Particulate Filtration Efficiency | ASTM F2299 | >98% | Pass | | Differential Pressure | EN 14683 | <6.0 mmH2O/cm2 | Pass | | Fluid Resistance | ASTM F1862 | 120 mmHg | Pass | | Flammability | 16 CFR Part 1610 | Class 1 | Pass | | Biocompatibility | ISO 10993 | | | | Cytotoxicity | ISO 10993-5 | Non-cytotoxic | Pass | | Sensitization | ISO 10993-10 | Non-sensitizing | Pass | | Irritation | ISO 10993-10 | Non-irritant | Pass | ## Summary of Clinical Performance Testing: Not applicable ## Conclusion: The conclusions drawn from the nonclinical tests demonstrate that the Fluidshield* 2 Surgical Mask(s), and Fluidshield* 2 Procedure Mask(s) are as safe, as effective and perform as well as or better than the legally marketed KC200 and KC300 Face Masks (K111402).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...