← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K230093

# Disposable 4 Layers Surgical Face Mask (K230093)

_Social Medical Supply, LLC · FXX · Mar 2, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K230093

## Device Facts

- **Applicant:** Social Medical Supply, LLC
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 2, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Disposable 4 layers surgical face mask is intended to be worn by adults, and protects both patients and healthcare personnel from microorganisms, body fluids and particulate material transfer within a professional healthcare environment. The mask is designed for use in infection control practices to reduce the potential exposure to blood and body fluids. It is a single-use, disposable device(s) , provided non-sterile.

## Device Story

Disposable 4-layer surgical face mask; adult-sized; earloop design. Acts as physical barrier against fluids and particulates; provides Level 3 barrier protection. Composed of spunbond polypropylene outer/inner layers, meltblown polypropylene filter, and electrostatic ethylene-propylene cotton layer. Used in professional healthcare settings for infection control. Protects patients and healthcare personnel from cross-contamination. Single-use; non-sterile.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified per ASTM F2100-19: Fluid resistance (ASTM F1862, 29/32 pass at 160mmHg), Particulate Filtration Efficiency (ASTM F2299, ≥98%), Bacterial Filtration Efficiency (ASTM F2101, 99.9%), Differential Pressure (EN 14683, <6.0 mmH2O/cm²), and Flammability (16 CFR 1610, Class 1). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

4-layer construction: spunbond polypropylene (outer/inner), meltblown polypropylene (filter), electrostatic ethylene-propylene cotton (middle). Malleable polyethylene nose wire; polyester/spandex earloops. Dimensions: 17.7cm x 9.5cm. Non-sterile, single-use. Performance meets ASTM F2100-19 Level 3 standards.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K221272](/device/K221272.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

March 2, 2023

Social Medical Supply LLC % Mr. Jett Lee Regulation Manager Guangdong Jianda Medical Technology Co Ltd 906 Room, Longxiang Garden, Tianhe district Guangzhou, Guangdong 510000 China

Re: K230093

Trade/Device Name: Disposable 4 Layers Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 12, 2023 Received: January 12, 2023

Dear Mr. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230093

Device Name Disposable 4 layers surgical face mask

#### Indications for Use (Describe)

The Disposable 4 layers surgical face mask is intended to be worn by adults, and protects both patients and healtheare personnel from microorganisms, body fluids and particulate material transfer within a professional healthcare environment. The mask is designed for use in infection control practices to reduce the posure to blood and body fluids. It is a single-use, disposable device(s), and provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <checkbox></checkbox>            |
|----------------------------------------------|----------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <checkbox checked=""></checkbox> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a traditional 510(K) submission, and there were no prior submissions for the subject device.

## 1. Submitter's Information

## Sponsor

- Company Name: SOCIAL MEDICAL SUPPLY LLC �
- � Address: 11528 Harry Hines Blvd, Suite A102 Dallas,TX75229
- � Tel:+1-214-416-9009 Fax:+1-214-416-9009
- Email: support@wellbefore.com �
- � Contact Person (including title): Shahzil Amin (General Manager)

## Manufacturer

- Company Name: Jiangxi Xianmiyan Medical Technology Co., Itd �
- Address: South side of the Ganda line of Xiangtian industrial Park, Jing'an County, Yichun � City, Jiangxi Province, China
- Tel:+86-0754-87712777 Fax:+86-0754-87712777 �
- � Email: finnick@wellbefore.com
- Contact Person (including title): Finnick Lin(General Manager) �

## Application Correspondent:

- � Guangdong Jianda Medical Technology Co Ltd
- Address: 906 Room, Longxiang Garden, Tianhe district, Guangzhou, China �
- � Contact Person: Mr. Jett Lee
- � Title: Regulation Manager
- � Tel: +86 13512755282
- Email: jianda-lee@foxmail.com �

## 2. Subject Device Information

- � Type of 510(k) submission: Traditional
- Classification/Common Name: Surgical Face Mask �
- Trade Name: Disposable 4 layers surgical face mask �
- Model: S1 �
- � Review Panel: Surgical Apparel
- � Product Code: FXX

{4}------------------------------------------------

- � Regulation Number: 878.4040
- � Requlation Class: 2

## 3. Predicate Device Information

- 510(k) number: K221272 �
- Sponsor: Shandong Xingyu Gloves Co., Ltd. �
- � Classification/Common Name: Surgical Face Mask
- � Trade Name: Surgical Face Mask
- � Model: Ear Loop
- � Review Panel: Surgical Apparel
- � Product Code: FXX
- � Regulation Number: 878.4040
- � Regulation Class: 2

## 4. Device Description

The WellBefore 4 Layers disposable surgical mask is an adult-sized surgical mask equipped with earloops, designed to cower the user's nose and mouth. The surgical face masks are consists of Outer Layer(Non-woven fabric made from spunbond polypropylene), Second Layer(meltblown filtration fabric made from polypropylene), Third Layer(Electrostastic cotton made from ethylene-propylene side by side), Inner Layer(Non-woven fabric made from spunbond polypropylene), Nose Bridge(Double-wirenosepinmade by polypropylene, and Earloops(Polyster Fiber and Spandex). The mask acts as a physical barrier from fuids and particulate materials, and providing a level 3 barrier. It is a single-use, disposable device(s) , provided non-sterile.

#### 5. Intended Use / Indications for Use

The Disposable 4 layers surgical face mask is intended to be worn by adults, and protects both patients and healthcare personnel from microorganisms, body fluids and particulate material transfer within a professional healthcare environment. The mask is designed for use in infection control practices to reduce the potential exposure to blood and body fluids. It is a single-use, disposable device(s) , provided non-sterile.

#### 6. Test Summary

#### 1)Non-clinical Test performed on the proposed device

The Surgical Mask has been evaluated for its safety and performance by lab bench testing as following:

- Three lots tested according to ASTM F2100-19 standard �

| Performance requirement and test method recommended in ASTM F2100 |                                           |             |
|-------------------------------------------------------------------|-------------------------------------------|-------------|
| Item                                                              | Performance requirement                   | Test method |
| Fluid Resistance Performance                                      | 29 out of 32 pass at 160mmHg              | ASTM F1862  |
| Particulate Filtration Efficiency                                 | $                 \u226598%             $ | ASTM F2299  |

{5}------------------------------------------------

| Bacterial Filtration Efficiency | $\u226598%$       | ASTM F2101             |
|---------------------------------|-------------------|------------------------|
| Differential Pressure (Delta P) | $< 6.0 mmH2O/cm²$ | EN 14683:2019, Annex C |
| Flammability                    | Class 1           | 16 CFR 1610            |

- � Biocompatibility test according to ISO 10993-1 standard
	- 1) Cytotoxicity Test with MTT Method according to ISO 10993-5
	- 2) Skin Irritation Test with test method: 0.9% Sodium Chloride Injection Extract and Soybean Oil Extract according to ISO 10993-10
	- 3) Skin Sensitization with test method: 0.9% Sodium Chloride Injection Extract and Soybean Oil Extract according to ISO 10993-10

#### Performance testing result

| Performance<br>Testing Items                                    | Test result                                                                   | Criteria                                                                      | Remark |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|--------|
| Fluid Resistance<br>Performance<br>ASTM F1862                   | 3 lots with at least 29 out of 32<br>passing at 160mmHg (level 3)             | 3 lots with at least 29 out of 32<br>passing at 160mmHg (level 3)             | Pass   |
| Particulate Filtration<br>Efficiency<br>ASTM F2299              | 3 lots at ≥98%                                                                | 3 lots at ≥98%                                                                | Pass   |
| Bacterial Filtration<br>Efficiency<br>ASTM F2101                | 3 lots at 99.9%                                                               | 3 lots at ≥98%                                                                | Pass   |
| Differential<br>Pressure (Delta P)<br>EN 14683:2019,<br>Annex C | 3 lots passing at<br>at <6.0 mmH₂O/cm²                                        | 3 lots passing<br>at <6.0 mmH₂O/cm²                                           | Pass   |
| Flammability<br>16 CFR 1610                                     | Class 1<br>Non Flammable                                                      | Class 1<br>Non Flammable                                                      | Pass   |
| Biocompatibility                                                | Cytotoxicity/Irritation/Sensitization according to ISO10993-5;<br>ISO10993-10 | Cytotoxicity/Irritation/Sensitization according to ISO10993-5;<br>ISO10993-10 | Pass   |

## 2) Clinical Test Conclusion

No clinical study is included in this submission.

## 7. Summary of Technological Characteristic

The technological characteristics, features, specifications, materials, and intended use of Surgical Mask is similar to the predicate device quoted above. The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.

| Device       | Proposed Device           | Predicate Device                 | Comparison |
|--------------|---------------------------|----------------------------------|------------|
| 510(K)       | Apply                     | K221272                          | --         |
| Sponsor      | SOCIAL MEDICAL SUPPLY LLC | Shandong Xingyu Gloves Co., Ltd. | --         |
| Manufacturer | Jiangxi Xianmiyan Medical | Shandong Xingyu Gloves Co., Ltd. | --         |

{6}------------------------------------------------

| Technology Co., ltd                                |                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                          |                             |
|----------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Product Name                                       |                                                                                 | Disposable 4 layers surgical face<br>mask                                                                                                                                                                                                                                                                                                                                                                                                                                       | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                       | --                          |
| Classification                                     |                                                                                 | Class II, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Class II, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                           | Same                        |
| Intended Use                                       |                                                                                 | The Disposable 4 layers surgical<br>face mask is intended to be worn<br>by adults, and protects both<br>patients and healthcare personnel<br>from microorganisms, body fluids<br>and particulate material transfer<br>within a professional healthcare<br>environment. The mask is<br>designed for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. It is a single-use,<br>disposable device(s), provided<br>non-sterile. | The Medical Face Mask is<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These face<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This a<br>single use, disposable device(s),<br>provided non-sterile. | substantially<br>equivalent |
| Material                                           | Outer Layer                                                                     | Non-woven fabric (spunbond PP)                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                         | Same                        |
|                                                    | Middle Layer                                                                    | Second Melt blown polypropylene<br>filter<br>Third Electrostatic cotton made<br>Layer from ethylene-propylene                                                                                                                                                                                                                                                                                                                                                                   | Melt-blown Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                              | Different<br>Note1          |
|                                                    |                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                          |                             |
|                                                    | Inner Facing<br>Layer                                                           | Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                         | Same                        |
|                                                    | Noise Piece                                                                     | Malleable polyethylene wire                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Metal Core Plastic<br>(Iron wire & Polypropylene)                                                                                                                                                                                                                                                                                                                                                        | Same                        |
| Ear Loops                                          |                                                                                 | Polyster and Spandex                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Polyster and Spandex                                                                                                                                                                                                                                                                                                                                                                                     | Same                        |
| Color                                              |                                                                                 | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Blue, Black                                                                                                                                                                                                                                                                                                                                                                                              | Different<br>Note2          |
| Dimension                                          |                                                                                 | 17.7cm*9.5cm, ±1cm tolerance                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 17.5cm*9.5cm, ±0.5cm tolerance                                                                                                                                                                                                                                                                                                                                                                           | Different<br>Note2          |
| OTC Use                                            |                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                      | Same                        |
| Sterility                                          |                                                                                 | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                              | Same                        |
| Use                                                |                                                                                 | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                   | Same                        |
| Performance<br>Testing (ASTM<br>F2100)             |                                                                                 | Level 2 & 3                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Level 2 & 3                                                                                                                                                                                                                                                                                                                                                                                              | Same                        |
| Fluid Resistance<br>Performance<br>ASTM F1862      |                                                                                 | 29 out of 32 pass at 120mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 32 out of 32 pass at 120mmHg                                                                                                                                                                                                                                                                                                                                                                             | Different<br>Note3          |
| Particulate Filtration<br>Efficiency<br>ASTM F2299 |                                                                                 | ≥98%                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | ≥98%                                                                                                                                                                                                                                                                                                                                                                                                     | Same                        |
| Bacterial Filtration<br>Efficiency<br>ASTM F2101   | $\ge$ 98%                                                                       | $\ge$ 98%                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                     |                             |
| Differential<br>Pressure (Delta P)                 | <6.0 mmH2O/cm²                                                                  | <6.0 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                     |                             |
| Flammability<br>16 CFR 1610                        | Class 1<br>Non Flammable                                                        | Class 1<br>Non Flammable                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                     |                             |
| Biocompatibility                                   | Meet ISO10993 ,proved<br>noncytotoxicity, non-irritating and<br>non-sensitizing | Meet ISO10993 ,proved<br>noncytotoxicity, non-irritating and<br>non-sensitizing                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                     |                             |

{7}------------------------------------------------

## Comparison in Detail(s):

#### Note 1:

The difference in the middel layer material and layer numbers does not raise additional questions for safety and effectiveness. The biocompatibility tests were conducted to both components to ensure their compliance to the ISO10993-5 and ISO10993-10. Performance tests were in accordance with ASTM F2100 standard. There is no new risk generated from the difference of the material and layer number .

## Note 2:

The difference in the dimensions and colors does not raise additional questions for safety and effectiveness. The dimensions of proposed device is design for adults. Performance testing including biocompatibility evaluation has been performed on the finished device which includes all construction materials and color additives.

#### Note 3:

The difference in Fluid Resistance Performance ASTM F1862 is does not raise additional questions for safety and effectiveness.

## 8. Summary for clinical test

Clinical performance is not deemed necessary.

#### 9. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the devices are as safe, as effective, and perform as well as or better than the legally marketed devices identified.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K230093](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K230093)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
