← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K223232

# Procedure mask/Surgical mask/Face mask (K223232)

_Winner Medical Co., Ltd. · FXX · Mar 6, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K223232

## Device Facts

- **Applicant:** Winner Medical Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Procedure mask/Surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

## Device Story

Three-layer, single-use, flat-pleated surgical face mask; inner/outer layers polypropylene nonwoven; middle layer polypropylene melt-blown nonwoven; includes iron/polypropylene nose clip; available in ear loop (polyester/spandex) or tie-on (polypropylene) configurations. Used in clinical settings or general use to provide barrier protection against blood, body fluids, and particulates. User adjusts nose clip for fit; secures via ear loops or ties. Device acts as physical barrier; performance meets ASTM F2100-19 Level 3 standards. Benefits include reduced risk of cross-contamination between patient and personnel.

## Clinical Evidence

Bench testing only. No clinical data included. Performance verified against ASTM F1862 (synthetic blood penetration), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), EN 14683 (differential pressure), 16 CFR 1610 (flammability), and ISO 10993-5/10 (cytotoxicity, sensitization, irritation).

## Technological Characteristics

Three-layer polypropylene nonwoven construction; melt-blown middle layer. Iron/polypropylene nose clip. Ear loops (polyester/spandex) or tie-on (polypropylene). ASTM F2100-19 Level 3 performance. Non-sterile. Dimensions: 17.5x9.5 cm or 14.5x9 cm.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Procedure mask/Surgical mask/Face mask ([K220194](/device/K220194.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 6, 2023

Winner Medical Co., Ltd. Mingni Liu Regulatory Affairs Winner Industrial Park, No. 660 Bulong Road, Longhua District, Shenzhen Guangdong 518109 China

Re: K223232

Trade/Device Name: Procedure mask/Surgical mask/Face mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 8, 2023 Received: February 8, 2023

Dear Mingni Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Katherine D. Kavlock -S

for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K223232

Device Name

Procedure mask/Surgical mask/Face mask

## Indications for Use (Describe)

The Procedure mask/Surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: __K223232

- 1. Date of Preparation: October 18, 2022
- 2. Sponsor Identification

Winner Medical Co., Ltd. Winner Industrial Park, No.660 Bulong Road, Longhua District, Shenzhen Guangdong, China 518109

Establishment Registration Number: 9616433

Contact Person: Mingni Liu Position: Regulatory Affairs Tel: +86-755-28138888-8822 Email: 2346@winnemedical.com

{4}------------------------------------------------

#### 3. Identification of Proposed Device

Trade Name: Procedure mask/Surgical mask/Face mask Common Name: Surgical Face Mask

Regulatory Information Classification Name: Mask, Surgical; Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040; Review Panel: General Hospital;

### Indications for use:

The Procedure mask/Surgical mask/Face mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

### Device Description:

The proposed device is a three-layer, single-use, flat- pleated mask. The inner and outer layers of the mask are made of polypropylene nonwoven, and the middle layer is made of polypropylene melt-blown nonwoven. The proposed devices are available in two types, ear loop and tie-on. The ear loops are made of polyester and spandex, and the ties are made of polypropylene nonwoven. The ear loops/ties are used to secure the mask over the users' mouth and face. The nose clip is made of Iron and polypropylene. Users can adjust the nose piece according to the shape of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The ear loop masks are available in two size, 17.5×9.5 cm and 14.5×9 cm; the tie-on mask is available in one size, 17.5×9.5 cm. And the colors for the ear loop mask are blue and black; the color for the tie-on mask is blue. Both the ear loop and tie-on masks are available in level 3 masks based on ASTM F2100-19 due to the difference in gram weight of the mask body. The proposed device is provided in non-sterile. All specifications of the proposed device are provided in table 1.

{5}------------------------------------------------

| Product name                                 | ASTM<br>F2100-19<br>level | Ear strap type | Size (cm) | Color | Layers | Sterilization |
|----------------------------------------------|---------------------------|----------------|-----------|-------|--------|---------------|
| Procedure<br>mask/Surgical<br>mask/Face mask | Level 3                   | Ear loop       | 17.5x9.5  | Blue  | 3      | Non-sterile   |
|                                              |                           |                | 17.5x9.5  | Black |        |               |
|                                              |                           |                | 14.5x9    | Blue  |        |               |
|                                              |                           |                | 14.5x9    | Black |        |               |
|                                              |                           | Tie-on         | 17.5x9.5  | Blue  |        |               |

Table 1 Procedure mask/Surgical mask/Face mask Description

#### 4. Identification of Predicate Device

510(k) Number: K220194 Product Name: Procedure mask/Surgical mask/Face mask

#### 5. Summary of Non-Clinical Test

Non clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- A ASTM F1862/F1862M: 2017 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity);
- > ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres;
- > ASTM F2101: 2019 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus;
- A ASTM F2100: 2019 Standard Specification for Performance of Materials Used in Medical Face Masks
- A EN 14683:2019+AC: 2019 Annex C Medical face masks - Requirements and test methods
- > 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;
- > ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization

{6}------------------------------------------------

| Test Methodology                                   | Purpose                                                                                                                                                                                                                                                                                                                                                          | Acceptance Criteria                                                                                                                                                            | Result                                                                                                                                                                                                                                     |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Resistance to<br>Penetration by<br>Synthetic blood | The test was performed in<br>accordance with ASTM<br>F1862/F1862M: 2017 Standard Test<br>Method for Resistance of Medical<br>Face Masks to Penetration by<br>Synthetic Blood (Horizontal<br>Projection of Fixed Volume at a<br>Known Velocity) to evaluate the<br>effectiveness of the test article from<br>possible exposure to blood and other<br>body fluids. | Level 3: No penetration at<br>160 mmHg                                                                                                                                         | Level 3: Pass at<br>160mmHg                                                                                                                                                                                                                |
| Particulate<br>Filtration<br>Efficiency            | The test was performed in<br>accordance with ASTM<br>F2299/F2299M-03 (2017) Standard<br>Test Method for Determining the<br>Initial Efficiency of Material Used<br>in medical Face Masks to<br>Penetration by Particulates using<br>Latex Spheres to determine the<br>particle filtration efficiency (PFE) of<br>the test article.                                | Level 3: ≥98%                                                                                                                                                                  | Blue mask:<br>Pass at 99.75 %<br>Black mask:<br>Pass at 99.03 %                                                                                                                                                                            |
| Bacterial Filtration<br>Efficiency                 | The test was performed in<br>accordance with ASTM F2101:<br>2019 Standard Test Method for<br>Evaluating the Bacterial Filtration<br>Efficiency (BFE) of Medical Face<br>Mask Materials, Using a Biological<br>Aerosol of Staphylococcus aureus to<br>determine the bacterial filtration<br>efficiency (BFE) of the test article.                                 | Level 3: ≥98%                                                                                                                                                                  | Blue mask:<br>Pass at 99.87 %<br>Black mask:<br>Pass at 98.90 %                                                                                                                                                                            |
|                                                    |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                |                                                                                                                                                                                                                                            |
| Differential<br>Pressure                           | The test was performed in<br>accordance with EN<br>14683:2019+AC: 2019 Annex C<br>Medical face masks - Requirements<br>and test methods to determine the<br>differential pressure of the test<br>article.                                                                                                                                                        | Level 3:<6.0 mmH2O/cm2                                                                                                                                                         | Blue mask:<br>Pass at 4.8<br>mmH2O/cm2<br>Black mask:<br>Pass at 3.7<br>mmH2O/cm2                                                                                                                                                          |
| Flammability                                       | The test was performed in<br>accordance with 16 CFR Part 1610<br>Standard for the Flammability of<br>Clothing Textiles to evaluate the<br>flammability of the test article.                                                                                                                                                                                      | Class 1                                                                                                                                                                        | Class 1                                                                                                                                                                                                                                    |
| Cytotoxicity                                       | The test was performed in<br>accordance with ISO 10993-5 Third<br>edition 2009-06-01 Biological<br>evaluation of medical devices - Part<br>5: Tests for in vitro cytotoxicity to<br>evaluate the cytotoxicity of the test<br>article.                                                                                                                            | The viability should be<br>≥70% of the blank. And<br>the 50% extract of the test<br>sample should have at least<br>the same or a higher<br>viability than the 100%<br>extract. | The viability was<br>≥70% of the<br>blank. And the<br>50% extract of the<br>test sample had a<br>higher viability<br>than the 100%<br>extract.<br>Under the<br>conditions of the<br>study, the<br>proposed device<br>was<br>non-cytotoxic. |
| Sensitization                                      | The test was performed in<br>accordance with ISO 10993-10<br>Third Edition 2010-08-01<br>Biological evaluation of medical                                                                                                                                                                                                                                        | Non-sensitizing                                                                                                                                                                | Under the<br>conditions of the<br>study, the<br>proposed device                                                                                                                                                                            |
|                                                    | devices - Part 10: Tests for irritation<br>and skin sensitization to evaluate the<br>sensitization of the test article.                                                                                                                                                                                                                                          |                                                                                                                                                                                | was<br>non-sensitizing.                                                                                                                                                                                                                    |
| Irritation                                         | The test was performed in<br>accordance with ISO 10993-10<br>Third Edition 2010-08-01<br>Biological evaluation of medical<br>devices - Part 10: Tests for irritation<br>and skin sensitization to evaluate the<br>irritation of the test article.                                                                                                                | Non-irritating                                                                                                                                                                 | Under the<br>conditions of the<br>study, the<br>proposed device<br>was non-irritating.                                                                                                                                                     |

Table 2 Summary of Performance Testing

{7}------------------------------------------------

{8}------------------------------------------------

- 6. Clinical Test Conclusion
No clinical study is included in this submission.

- 7. Technological Characteristics Comparison

| ITEM                                    | Proposed Device                                                                                                                                                                                                                                  | Predicate Device<br>K220194                                                                                                                                                                                                                                                                                                                                                                              | Remark                                                        |           |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|-----------|
| Product Code                            | FXX                                                                                                                                                                                                                                              | FXX                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                          |           |
| Regulation No.                          | 21 CFR 878.4040                                                                                                                                                                                                                                  | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                          |           |
| Class                                   | II                                                                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                          |           |
| Indications for<br>Use                  | The Procedure<br>mask/Surgical<br>mask is intended<br>to be worn to<br>protect both the<br>patient and<br>healthcare<br>personnel from<br>transfer of<br>microorganisms,<br>body fluids, and<br>particulate<br>material. These<br>face masks are | The Procedure mask/Surgical mask is intended to be<br>worn to protect both the patient and healthcare<br>personnel from transfer of microorganisms, body<br>fluids, and particulate material. These face masks<br>are intended for use in infection control practices to<br>reduce the potential exposure to blood and body<br>fluids. This is a single-use, disposable device,<br>provided non-sterile. | Same                                                          |           |
|                                         | intended for use in<br>infection control<br>practices to reduce<br>the potential<br>exposure to blood<br>and body fluids.<br>This is a<br>single-use,<br>disposable device,<br>provided<br>non-sterile.                                          |                                                                                                                                                                                                                                                                                                                                                                                                          |                                                               |           |
| Mask style                              | Flat pleated                                                                                                                                                                                                                                     | Flat pleated                                                                                                                                                                                                                                                                                                                                                                                             |                                                               | Same      |
| Design feature                          | Ear loop / Tie-on                                                                                                                                                                                                                                | Ear loop / Tie-on                                                                                                                                                                                                                                                                                                                                                                                        |                                                               | Same      |
| Dimension                               | Ear loops:<br>Body: 175 mm×95 mm, nose clip: 110mm,<br>Ear-loop: 178mm<br>Body: 145 mm×90 mm, nose clip: 85mm, Ear-loop: 145mm                                                                                                                   | Ear loops:<br>Body: 175 mm×95 mm, nose clip: 110mm,<br>Ear-loop: 178mm<br>Body: 145 mm×90 mm, nose clip: 85mm, Ear-loop: 145mm<br><br>Tie-on:<br>Body: 175 mm×95 mm, nose clip: 110mm,<br>Ear-loop: 900mm                                                                                                                                                                                                |                                                               | Same      |
|                                         | Tie-on:<br>Body: 175 mm×95 mm, nose clip: 110mm,<br>Ear-loop: 900mm                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                          |                                                               |           |
| ASTM F2100<br>Level                     | Level 3                                                                                                                                                                                                                                          | Level 1                                                                                                                                                                                                                                                                                                                                                                                                  | Level 2                                                       | Same      |
| Fluid<br>Resistance                     | Pass at 160mmHg                                                                                                                                                                                                                                  | Pass at 80mmHg                                                                                                                                                                                                                                                                                                                                                                                           | Pass at 120 mmHg                                              | Same      |
| Particulate<br>filtration<br>efficiency | Blue mask:<br>Pass at 99.87 %<br>Black mask:<br>Pass at 98.90 %                                                                                                                                                                                  | Blue mask:<br>Pass at 96.05%<br>Black mask:<br>Pass at 96.03%                                                                                                                                                                                                                                                                                                                                            | Blue mask:<br>Pass at 98.78%<br>Black mask:<br>Pass at 98.75% | Different |
| Bacterial<br>filtration                 | Blue mask:<br>Pass at 99.75%                                                                                                                                                                                                                     | Blue mask:<br>Pass at 98.25%                                                                                                                                                                                                                                                                                                                                                                             | Blue mask:<br>Pass at 98.72%                                  | Different |

# Table 3 Comparison of Procedure mask/Surgical mask

{9}------------------------------------------------

{10}------------------------------------------------

| efficiency               |                                               | Black mask:<br>Pass at 99.03 %                                              | Black mask:<br>Pass at 98.25%                                               | Black mask:<br>Pass at 98.73%                                               |           |      |
|--------------------------|-----------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------|------|
| Differential<br>pressure |                                               | Blue mask:<br>Pass at 4.8 mmH2O/cm²<br>Black mask:<br>Pass at 3.7 mmH2O/cm² | Blue mask:<br>Pass at 3.5 mmH2O/cm²<br>Black mask:<br>Pass at 3.5 mmH2O/cm² | Blue mask:<br>Pass at 3.5 mmH2O/cm²<br>Black mask:<br>Pass at 3.6 mmH2O/cm² | Different |      |
|                          | Flammability                                  |                                                                             | Class 1                                                                     | Class 1                                                                     | Class 1   | Same |
| Label/Labeling           |                                               | Complied with 21 CFR part 801                                               | Complied with 21 CFR part 801                                               |                                                                             | Same      |      |
| Materials                |                                               |                                                                             |                                                                             |                                                                             |           |      |
|                          | Ear loop<br>Polyester and spandex             |                                                                             | Polyester and spandex                                                       |                                                                             |           |      |
|                          | Tie strings<br>38g /m² polypropylene nonwoven |                                                                             | 38 /m² polypropylene nonwoven                                               |                                                                             |           |      |
|                          | Nose clip<br>Iron and polypropylene           |                                                                             | Iron and polypropylene                                                      |                                                                             |           |      |
|                          | Outer<br>material                             | 35g/m² polypropylene nonwoven                                               | Outer<br>material<br>25g/m² polypropylene nonwoven                          | 30g/m² polypropylene nonwoven                                               | Different |      |
| Mask<br>body             | Middle<br>material                            | 33g/m² polypropylene melt-blown nonwoven                                    | Middle<br>material<br>25g/m² polypropylene melt-blown nonwoven              | 30g/m² polypropylene melt-blown nonwoven                                    |           |      |
|                          | Inner<br>material                             | 25g/m² polypropylene nonwoven                                               | Inner<br>material<br>25g/m² polypropylene nonwoven                          | 30g/m² polypropylene nonwoven                                               |           |      |
| Colors                   |                                               | Blue; Black                                                                 | Blue; Black                                                                 |                                                                             | Same      |      |
| Biocompatibility         |                                               |                                                                             |                                                                             |                                                                             |           |      |
| Cytotoxicity             |                                               | Under the conditions of the study, the proposed device                      | Under the conditions of the study, the proposed device was non-cytotoxic.   |                                                                             | Same      |      |

{11}------------------------------------------------

|               | was<br>non-cytotoxic.                                                                            |                                                                                |      |
|---------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------|
| Sensitization | Under<br>the<br>conditions of the<br>study,<br>the<br>proposed device<br>was<br>non-sensitizing. | Under the conditions of the study, the proposed<br>device was non-sensitizing. | Same |
| Irritation    | Under<br>the<br>conditions of the<br>study,<br>the<br>proposed device<br>was non-irritating.     | Under the conditions of the study, the proposed<br>device was non-irritating.  | Same |
| Sterilization |                                                                                                  |                                                                                |      |
| Method        | Non-sterile                                                                                      | Non-sterile                                                                    | Same |

## Different - Particulate efficiency

The test result for particulate filtration efficiency for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 3 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of proposed device.

# Different - Bacterial filtration efficiency

The test result for bacterial filtration efficiency for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 3 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of proposed device.

# Different - Differential pressure

The test result for differential pressure for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 3 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of proposed device.

# Different - Materials

The material for the proposed device is the same as the predicate device, both of the proposed device and the predicate device are manufactured by Winner Medical. Level 2 and level 3 masks based on ASTM F2100-19 are made of the same material, except for the difference in the gram weight of the inner, outer and middle layer of the mask. The material of the tie on mask can be covered by the material of the ear loop mask. Therefore, biocompatibility tests were performed on level 2 ear loop blue and black masks, the results does not show any adverse effect. Thus, this difference in the gram weight will 

{12}------------------------------------------------

not affect the safety and effectiveness of the proposed device.

#### 8. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device K220194.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K223232](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K223232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
