← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K222545

# Disposable Medical Mask (K222545)

_Guangzhou Fuzelong Hygiene Material Co., Ltd. · FXX · May 19, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K222545

## Device Facts

- **Applicant:** Guangzhou Fuzelong Hygiene Material Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** May 19, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

Disposable Medical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.

## Device Story

Disposable medical mask; four-layer construction; polypropylene inner, outer, and middle (melt-blown) layers; galvanized iron wire nose piece; polyester/spandex ear loops or polypropylene ties. Single-use, sterile, flat-pleated design. Used in clinical settings or general infection control to provide physical barrier against blood, body fluids, and particulates. Protects both wearer and patient. Performance meets ASTM F2100 Level 3 standards.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified per ASTM F2100, F1862 (synthetic blood penetration), F2101 (BFE ≥ 99.80%), F2299 (PFE ≥ 99.80%), and EN 14683 (Delta P 4.2 mm H2O/cm²). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Flammability tested per 16 CFR 1610 (Class I).

## Technological Characteristics

Four-layer polypropylene construction; melt-blown middle layer; galvanized iron wire nose piece; polyester/spandex ear loops or polypropylene ties. Sterilized via ethylene oxide (SAL 10^-6). Flat-pleated form factor. ASTM F2100 Level 3 compliant. No software or electronic components.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical mask ([K211454](/device/K211454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

May 19, 2023

Guangzhou Fuzelong Hygiene Material Co., Ltd % Boyle Wang Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K222545

Trade/Device Name: Disposable Medical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: April 1, 2023 Received: April 19, 2023

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Katherine D. Kavlock -S

for

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K222545

Device Name Disposable Medical Mask

Indications for Use (Describe)

Disposable Medical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K222545

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

#### 1.0 submitter's Information

Name: Guangzhou Fuzelong Hygiene Material Co., Ltd Address: #12, Guancun Road, Jiangpu Street, Conghua, Guangzhou, China 510900 Tel:+86-020-87993188 Contact: Haiyan Zeng Date of Preparation: Aug.12, 2022

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong, Shanghai 200120, China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Trade name: Disposable Medical Mask Common name: Surgical Face Mask Classification name: Surgical Face Mask Model: Ear loops, Tie-on

#### 3.0 Classification

FXX - Mask, Surgical. Production code: Classification Name: Surgical Apparel (21 CFR part 878.4040) Classification: Class II Panel: Surgical Apparel

#### 4.0 Predicate Device Information

| Manufacturer:  | Jiangsu Xingtong Biotechnology Group Co., Ltd. |
|----------------|------------------------------------------------|
| Device:        | Surgical mask                                  |
| 510(k) number: | K211454                                        |

{4}------------------------------------------------

## 5.0 Device Description

The Disposable Medical Mask consists of a mask body, a nose piece, and ear loops or ties.

The mask body is divided into four layers, the inner, second and outer layers are made of polypropylene materials; the middle layer is composed of melt-blown cloth (polypropylene); the nose piece is made of galvanized iron wire, the ear loops are made of polyester and spandex, and the ties are made of polypropylene.

The Medical Surgical Mask will be provided in green. The Disposable Medical Mask is sold as sterile and are intended to be single use, disposable devices.

The size specification of the surgical mask:

- Mask body for ear-loop type: 14.5cm × 9cm (S), 17.5cm × 9.5cm (M), 18cm × 9.3cm (L), 22cm×9.5cm (XL);

- Mask body for Tie-on type: 14.5cm X 9cm (S), 17.5cm X 9.5cm (M), 18cm X 9.3cm (L), 22cm×9.5cm (XL).

The smallest sizes mask (14.5cm×9cm) are for adult population only.

#### 6.0 Indication for Use Statement

Disposable Medical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.

#### 7.0 Comparison to the Predicate Device

| Table 1 General Comparison            |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                              |                                                                                                                                   |                         |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Item                                  | Subject Device                                                                                                                                                                                                                               | Predicate Device<br>K211454                                                                                                                                                                                                                                                                                  | Remark                                                                                                                            |                         |
| Product Name                          | Medical Surgical Mask                                                                                                                                                                                                                        | Surgical mask                                                                                                                                                                                                                                                                                                | --                                                                                                                                |                         |
| Product Code                          | FXX                                                                                                                                                                                                                                          | FXX                                                                                                                                                                                                                                                                                                          | Same                                                                                                                              |                         |
| Regulation No.                        | 21 CFR 878.4040                                                                                                                                                                                                                              | 21 CFR 878.4040                                                                                                                                                                                                                                                                                              | Same                                                                                                                              |                         |
| Class                                 | II                                                                                                                                                                                                                                           | II                                                                                                                                                                                                                                                                                                           | Same                                                                                                                              |                         |
| Intended Use&<br>Indications for use  | Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the | The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to                                                              | Same                                                                                                                              |                         |
|                                       |                                                                                                                                                                                                                                              | potential exposure to blood<br>and body fluids. This is a<br>single use, disposable<br>device(s), provided as<br>sterile.                                                                                                                                                                                    | reduce the potential<br>exposure to blood and body<br>fluids. This is a single-use,<br>disposable device, provided<br>as sterile. |                         |
| Design features                       |                                                                                                                                                                                                                                              | Ear Loops,<br>Tie-on,<br>4 layers                                                                                                                                                                                                                                                                            | Ear loops: XT10A1;<br>Tie-on: XT10B1;<br>3 layers                                                                                 | Different<br>Analysis 1 |
| Mask Styles                           |                                                                                                                                                                                                                                              | Flat pleated                                                                                                                                                                                                                                                                                                 | Flat pleated                                                                                                                      | Same                    |
| Material                              | Outer<br>facing<br>layer                                                                                                                                                                                                                     | Polypropylene                                                                                                                                                                                                                                                                                                | Polypropylene                                                                                                                     | Same                    |
|                                       | Middle<br>layer                                                                                                                                                                                                                              | 1. Polypropylene<br>2. Melt-blown cloth<br>(polypropylene)                                                                                                                                                                                                                                                   | 1. Melt-blown cloth<br>(polypropylene)                                                                                            | Similar                 |
|                                       | Inner<br>Facing<br>layer                                                                                                                                                                                                                     | Polypropylene                                                                                                                                                                                                                                                                                                | Polypropylene                                                                                                                     | Same                    |
|                                       | Nose<br>piece                                                                                                                                                                                                                                | Galvanized iron wire                                                                                                                                                                                                                                                                                         | Polyethylene coated<br>steel wire                                                                                                 | Different<br>Analysis 1 |
|                                       | Ear<br>loops,<br>Ties                                                                                                                                                                                                                        | -Ear loops: Polyester,<br>spandex<br>-Ties: Polypropylene                                                                                                                                                                                                                                                    | -Ear loops: Polyester silk &<br>Polyurethane filament<br>-Ties: Polypropylene                                                     | Similar<br>Analysis 1   |
| Color                                 |                                                                                                                                                                                                                                              | Green                                                                                                                                                                                                                                                                                                        | Blue                                                                                                                              | Different<br>Analysis 2 |
| Dimension                             |                                                                                                                                                                                                                                              | - Mask body for ear-loop<br>type:<br>14.5cm×9cm (S),<br>17.5cm×9.5cm (M),<br>18cm×9.3cm (L),<br>22cm×9.5cm (XL);<br>- Mask body for Tie-on type:<br>14.5cm×9cm (S),<br>17.5cm×9.5cm (M),<br>18cm×9.3cm (L),<br>22cm×9.5cm (XL).<br>The smallest sizes mask<br>(14.5cm×9cm) are for adult<br>population only. | Mask body for ear-loop<br>type:<br>17.5cm×9.5cm &<br>14.5cm×9.5cm<br>Mask body for Tie-on type:<br>17.5cm×9.5cm                   | Different<br>Analysis 3 |
| OTC use                               |                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                               | Same                    |
| Shelf life                            |                                                                                                                                                                                                                                              | 2 years                                                                                                                                                                                                                                                                                                      | 2 years                                                                                                                           | Same                    |
| Single Use                            | Yes                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                          | Same                                                                                                                              |                         |
| Sterility                             | Sterile                                                                                                                                                                                                                                      | Sterile                                                                                                                                                                                                                                                                                                      | Same                                                                                                                              |                         |
| Sterilization<br>method and<br>S.A.L. | Sterilized by ethylene<br>oxide gas, SAL=10-6                                                                                                                                                                                                | Sterilized by ethylene<br>oxide gas, SAL=10-6                                                                                                                                                                                                                                                                | Same                                                                                                                              |                         |
| ASTM F2100 Level                      | Level 3                                                                                                                                                                                                                                      | Level 3                                                                                                                                                                                                                                                                                                      | Same                                                                                                                              |                         |

#### Table 1 General Comnarison

{5}------------------------------------------------

{6}------------------------------------------------

Analysis 1: the two devices have some difference in design features and materials, product materials safety is proved by its biocompatibility, and the difference does not raise additional questions for safety and effectiveness of device.

Analysis 2: The subject device (Green) has different color to the predicate device (Blue), but all proposed devices are conducted the biocompatibility test. The difference does not raise additional questions for safety and effectiveness of device.

Analysis 3: the two devices have some difference in dimensions, the little deviation in dimensions does not raise additional questions for safety and effectiveness of device.

## 8.0 Non-Clinical Test Conclusion

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Medical face masks

- Premarket Notification [510(k)] Submission issued on March 5, 2004.

ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization

ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks

ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity);

ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

ASTM F2299, stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;

16 CFR 1610, Standard for the Flammability of clothing textiles;

| Item      | Purpose     | Acceptance Criteria  | Results |
|-----------|-------------|----------------------|---------|
| Synthetic | Demonstrate | 29 samples out of 32 | Pass    |

Table 2 - Performance Testing

{7}------------------------------------------------

| Blood<br>Penetration<br>ASTM<br>F1862                        | resistance<br>to liquid<br>penetration   | pass (AQL 4%)<br>Level 3 pass at<br>160mmHg | 32 out of 32 pass at 160<br>mmHg              |
|--------------------------------------------------------------|------------------------------------------|---------------------------------------------|-----------------------------------------------|
| Particulate<br>Filtration<br>Efficiency<br>ASTM<br>F2299     | Demonstrate<br>particulate<br>filtration | Level 3 pass at ≥98%                        | Pass<br>Filtration Efficiency (%)<br>≥ 99.80% |
| Bacterial<br>Filtration<br>Efficiency<br>ASTM<br>F2101       | Demonstrate<br>bacterial<br>filtration   | Level 3 pass at ≥98%                        | Pass<br>Percent BFE (%)<br>≥ 99.80%           |
| Differential<br>Pressure<br>(Delta P)<br>EN 14683<br>Annex C | Demonstrate<br>breathability             | Level 3 pass at ≤ 6.0<br>mmH2O/cm²          | Pass<br>Average 4.2 mm H2O/cm²                |
| Flammability<br>16 CFR<br>1610                               | Demonstrate<br>flame<br>resistance       | Class I                                     | Pass                                          |

#### Table 4 - Biocompatibility Testing

| Item          | Subject Device                                                            | Result |
|---------------|---------------------------------------------------------------------------|--------|
| Cytotoxicity  | Under the conditions of the study, the subject device was non-cytotoxic   | Pass   |
| Irritation    | Under the conditions of the study, the subject device was non-irritating  | Pass   |
| Sensitization | Under the conditions of the study, the subject device was non-sensitizing | Pass   |

#### 9.0 _Clinical Test Conclusion

No clinical study is included in this submission.

#### 10.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K211454.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K222545](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K222545)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
