← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K221976

# Surgical Face Mask (Non-sterile) (K221976)

_Xiantao Junhui Plastic Products Co., Ltd. · FXX · Sep 20, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K221976

## Device Facts

- **Applicant:** Xiantao Junhui Plastic Products Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Sep 20, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Surgical Face Mask (Non-sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

## Device Story

Surgical Face Mask (Non-sterile) is a single-use, disposable, flat-pleated mask; worn over the nose and mouth to provide a physical barrier against microorganisms, body fluids, and particulates. Device consists of three layers: inner and outer layers of spun-bonded polypropylene non-woven fabric; middle layer of polypropylene melt-blown non-woven fabric. Features include a nose piece (galvanized iron wire coated with PE) for facial fit and two ear loops (nylon and spandex) for securement. Used in clinical settings for infection control; operated by healthcare personnel or patients. Provides protection by filtering bacteria and particulates and resisting fluid penetration. Benefits include reduced risk of cross-contamination and exposure to blood/body fluids during medical procedures.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19, ASTM F2101 (BFE ≥98%), ASTM F2299 (sub-micron particulate filtration ≥98%), ASTM F1862 (synthetic blood penetration at 160 mmHg), and EN 14683 (breathability <6.0 mmH2O/cm²). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Flammability tested per 16 CFR 1610 (Class I).

## Technological Characteristics

Three-layer flat-pleated mask. Materials: Spun-bonded polypropylene (inner/outer), melt-blown polypropylene (middle). Nose piece: PE-coated galvanized iron wire. Ear loops: Nylon and spandex. Dimensions: 175mm x 95mm. Non-sterile. Complies with ASTM F2100-19. Biocompatibility per ISO 10993.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Level 3 Fluid Resistant Procedure/Surgical Mask ([K211827](/device/K211827.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

September 20, 2022

Xiantao Junhui Plastic Products Co., Ltd. % Ms. Ivy Wang Technical Manager Shanghai Sungo Management Consulting Co., Ltd. 14th Floor, 1500# Century Avenue Shanghai, 200122 China

Re: K221976

Trade/Device Name: Surgical Face Mask (Non-sterile) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: July 5, 2022 Received: July 5, 2022

Dear Ms. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Submission Number (if known)

K221976

Device Name

Surgical Face Mask (Non-sterile)

#### Indications for Use (Describe)

The Surgical Face Mask (Non-sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

(As requirement by 21 CFR 807.92)

Date prepared: 15th, June, 2022

#### A. Applicant:

Name: XIANTAO JUNHUI PLASTIC PRODUCTS CO., LTD. Address: No. 3, Babu Industrial Park, Pengchang Town, Xiantao City, HUBEI, CHINA Contact: Fiona Tian Title: Sales Manager Tel: +86-15027285720 Email: fiona@xtjhslzp.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com

# B. Device:

Trade Name: Surgical Face Mask (Non-sterile) Model: Flat type/over-the-ear

Regulatory Information Common Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation: 21 CFR 878.4040 - Surgical apparel

# C. Identification of Primary Predicate device:

K211827 Zhejiang Lanhine Medical Products LTD. Trade Name: Level 3 Fluid Resistant Procedure/Surgical Mask Common Name: Surgical Face Mask Model(s): 15604F, 15704F

{4}------------------------------------------------

#### XIANTAO JUNHUI PLASTIC PRODUCTS CO., LTD. No. 3, Babu Industrial Park, Pengchang Town, Xiantao City, HUBEI, CHINA

Regulatory Information: Common Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation: 21 CFR 878.4040 - Surgical apparel

#### D. Indications for use of the device:

The Surgical Face Mask (Non-sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Surgical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

#### E. Device Description:

The Surgical Face Masks (Non-sterile) are blue color, three-layer, flat-pleaded masks with nose piece and ear loops, which are composed of inner layer and outer layer. The colorant is blue polypropylene (PP) master batch.

The inner layer and outer layer of the mask are made of spun-bonded non-woven fabric (polypropylene), the middle layer is made of polypropylene melt-blown non-woven fabric. The ear loop of the subject mask is held in place over the users' mouth and nose by two ear loops welded to the face mask. The ear loop is made with nylon and spandex. The nose piece in the layers of face mask is to allow the user to fit the mask around their nose, which is made of galvanized iron wire coated by PE.

The Surgical Face Mask (Non-sterile) is sold non-sterile and are intended to be single use, disposable devices.

The mask is designed and manufactured in accordance with ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks.

# F. Comparison of technological characteristics with the predicate device

The Surgical Face Masks are essentially the same as or similar to the predicate device in terms of the indications for use, design and construction, performance characteristics. Provided below table 1 is a comparison of the proposed device with the predicate device.

| Device       | Proposed Device                              | Predicate Device                                   | Result  |
|--------------|----------------------------------------------|----------------------------------------------------|---------|
| 510K #       | -                                            | K211827                                            | -       |
| Manufacturer | XIANTAO JUNHUI PLASTIC<br>PRODUCTS CO., LTD. | Zhejiang Lanhine Medical Products LTD.             | -       |
| Product Name | Surgical Face Mask (Non-sterile)             | Level 3 Fluid Resistant<br>Procedure/Surgical Mask | Similar |
| Level        | Level 3                                      | Level 3                                            | Same    |

#### Table 1 Comparison of Proposed and Predicate Devices

{5}------------------------------------------------

#### XIANTAO JUNHUI PLASTIC PRODUCTS CO., LTD. No. 3, Babu Industrial Park, Pengchang Town, Xiantao City, HUBEI, CHINA

| Product Code              | FXX                                                                                                                                                                                                                                                                                                                                                                                                                    | FXX                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same      |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Regulation<br>Number      | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                        | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same      |
| Indications<br>for<br>use | The Surgical Face Mask (Non-sterile) is<br>intended to be worn to protect both the<br>patient and healthcare personnel from<br>transfer of microorganisms, body fluids<br>and particulate material. The Surgical Face<br>Mask is intended for use in infection<br>control practices to reduce the potential<br>exposure to blood and body fluids. This a<br>single use, disposable device(s), provided<br>non-sterile. | The Level 3 Fluid Resistant<br>Procedure/Surgical Masks (model:<br>15604F, 15704F) are intended to be<br>worn to protect both the patient and<br>healthcare personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These face masks<br>are intended for use in infection control<br>practices to reduce the potential<br>exposure to blood and body fluids. This<br>a single use, disposable device(s),<br>provided non-sterile. | Same      |
| Design Feature            | Ear loops, flat pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                      | Ear loops, tie-on, flat pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar   |
| Color                     | Blue                                                                                                                                                                                                                                                                                                                                                                                                                   | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same      |
| Dimension                 | $175 \pm 5mm$<br>$95 \pm 5mm$                                                                                                                                                                                                                                                                                                                                                                                          | $17.5cm \pm 0.5cm$<br>$9.5cm \pm 0.5cm$                                                                                                                                                                                                                                                                                                                                                                                                                          | Same      |
| Sterility                 | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                            | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same      |
| Use                       | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                                 | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same      |
| Material                  |                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |           |
| Outer layer               | Spun-bonded non-woven fabric<br>(polypropylene)                                                                                                                                                                                                                                                                                                                                                                        | Non-woven fabric<br>(Polypropylene)                                                                                                                                                                                                                                                                                                                                                                                                                              | Same      |
| Middle layer              | Polypropylene melt-blown non-woven<br>fabric                                                                                                                                                                                                                                                                                                                                                                           | Melton blown fabric<br>(Polypropylene)                                                                                                                                                                                                                                                                                                                                                                                                                           | Same      |
| Inner layer               | Spun-bonded non-woven fabric<br>(polypropylene)                                                                                                                                                                                                                                                                                                                                                                        | Non-woven fabric<br>(Polypropylene)                                                                                                                                                                                                                                                                                                                                                                                                                              | Same      |
| Nose clip                 | Galvanized iron wire coated by PE                                                                                                                                                                                                                                                                                                                                                                                      | Polypropylene coating iron                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different |
| Ear loops                 | Nylon and spandex                                                                                                                                                                                                                                                                                                                                                                                                      | Polyurethane                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different |
| Biocompatibility          | ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                              | ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same      |

#### Difference Analysis:

The proposed device has same intended use, structure, parameter and performance with the predicate device. The materials of inner layer, middle layer and outer layer of the proposed device are also the same as those of the predicate device.

The proposed device has different material of nose clip and ear loops to the predicate device. However, biocompatibility test has been performed on the proposed device according to ISO 10993-10 and the results does not show any adverse effect. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

{6}------------------------------------------------

#### G. Summary of Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specification. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Mask - Premarket Notification [510(K)] Submission issued on March 5, 2004:

- A ISO 10993-05:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization
- > ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks
- > ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- A EN 14683, Medical Face Masks - Requirements and Test Methods
- > ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus Aureus
- > ASTM F2299, Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
- A 16 CFR 1610, Standard for the Flammability of Clothing Textiles

Performance testing and biocompatibility testing are summarized in below table 2.

| Test Methodology                                                    | Purpose                                           | Acceptance Criteria<br>for Level 3 Barrier | Result                                                                                                                 |
|---------------------------------------------------------------------|---------------------------------------------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Bacterial Filtration<br>Efficiency<br>ASTM F2101                    | Measure bacterial<br>filtration efficiency        | ≥98%                                       | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 1: ≥98%<br>Lot 2:≥98%<br>Lot 3:≥98%   |
| Differential<br>Pressure<br>(mmH2O/cm²)<br>EN 14683:2019<br>Annex C | Determine<br>breathability of the<br>mask         | <6.0 mmH2O/cm²                             | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 1: <6.0<br>Lot 2: <6.0<br>Lot 3: <6.0 |
| Sub-micron<br>Particulate Filtration<br>Efficiency ASTM<br>F2299-17 | Measure initial particle<br>filtration efficiency | ≥98%                                       | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 1: ≥98%<br>Lot 2: ≥98%<br>Lot 3: ≥98% |
| Resistance<br>to<br>Penetration<br>by                               | Evaluate the resistance<br>to penetration<br>by   | 29 out of 32 pass at<br>160 mmHg           | Passed<br>3 non-consecutive lots tested, using a                                                                       |

#### Table 2: Summary of Performance Testing & Biocompatibility testing

{7}------------------------------------------------

# XIANTAO JUNHUI PLASTIC PRODUCTS CO., LTD. No. 3, Babu Industrial Park, Pengchang Town, Xiantao City, HUBEI, CHINA

| Synthetic Blood<br>ASTM F1862-17         | impact of small<br>volume of synthetic<br>blood                   |                       | sample size of 32/lot<br>Lot 1: 32 out of 32 pass at 160 mmHg<br>Lot 2: 32 out of 32 pass at 160 mmHg<br>Lot 3: 32 out of 32 pass at 160 mmHg |
|------------------------------------------|-------------------------------------------------------------------|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Flammability<br>16 CFR Part<br>1610-2008 | Response of materials<br>to heat and flame                        | Class I               | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 1: Class 1<br>Lot 2: Class 1<br>Lot 3: Class 1               |
| Cytotoxicity                             | Assess the potential<br>risk of cytotoxicity of<br>mask material  | Non-cytotoxic         | Pass<br>Under the condition of this study, the<br>device has no potential toxicity.                                                           |
| Irritation                               | Assess the potential<br>risk of irritation of<br>mask material    | Negligibly irritating | Under the condition of this study, the<br>device is negligibly irritating.                                                                    |
| Sensitization                            | Assess the potential<br>risk of sensitization of<br>mask material | Non-sensitizing       | Under the conditions of the study, the<br>device is non-sensitizing                                                                           |

#### H. Clinical Test Conclusion

No clinical study is included in this submission.

#### I. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Surgical Face Mask (Non-sterile) is substantially equivalent to the legally marketed predicate device cleared under K211827.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K221976](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K221976)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
