← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K221273

# Medical Face Mask (M001) (K221273)

_Xingyu Medical Tech Co., Ltd. · FXX · Jun 28, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K221273

## Device Facts

- **Applicant:** Xingyu Medical Tech Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Jun 28, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Medical Face Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

## Device Story

Medical Face Mask (M001) is a single-use, disposable, non-sterile, flat-pleated mask; worn over mouth and nose to protect patients and healthcare personnel from microorganisms, body fluids, and particulates. Construction includes three layers: inner/outer spun-bond polypropylene; middle melt-blown polypropylene filter. Features include metal-core plastic nose clip and polyester/spandex elastic ear loops. Used in clinical or general settings for infection control. Provides physical barrier to reduce exposure to blood and body fluids. No electronic components or software.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19, ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), EN 14683 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

Three-layer construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (middle). Nose clip: metal-core plastic (iron wire/polypropylene). Ear loops: polyester/spandex. Dimensions: 175mm x 95mm. Non-sterile, single-use. Meets ASTM F2100-19 Level 2 & 3 requirements.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K211631](/device/K211631.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 28, 2022

Xingyu Medical Tech Co., Ltd % Jarvis Wu Consultant Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai. 200122 China

Re: K221273

Trade/Device Name: Medical Face Mask (M001) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 2, 2022 Received: May 2, 2022

Dear Jarvis Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221273

Device Name Medical Face Mask (M001)

#### Indications for Use (Describe)

The Medical Face Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

| <input type="checkbox"/> Research involving HHS-conducted or -supported Research | <input checked="" type="checkbox"/> Secondary Research involving HHS-conducted or -supported Research |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

# 510(k) Summary K221273

Document Date Prepared:2022/4/21

#### A. Applicant:

Name: Xingyu Medical Tech Co., Ltd.

Address: No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

Contact Person: Carrie Li Tel: 0086-18263618867 Mail: carrie@xingyugloves.com

Submission Correspondent: Primary contact: Mr. Jarvis Wu Shanghai SUNGO Management Consulting Co., Ltd. 14th floor, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo 14th floor, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

#### B. Device:

Regulatory Information

Proprietary Name: Medical Face Mask Common Name: Surgical Face Mask Model(s): M001

Classification Name: Surgical Face Mask Classification: Class II Product code: FXX

{4}------------------------------------------------

Xingyu Medical Tech Co., Ltd. No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

Regulation Number: 878.4040

Review Panel: Surgical Apparel

C. Predicate de vice:

| 510K    | Device name        | ASTM F2100-19<br>level | Manufacturer                                      |
|---------|--------------------|------------------------|---------------------------------------------------|
| K211631 | Surgical Face Mask | Level 2 & 3            | Macheng thimble Technology<br>Investment Co., Ltd |

(Note: Predicate device has NOT been subject to any Medical Device Recalls, including design-related recall.)

### D. Indications use of the device:

The Medical Face Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

### E. Device Description:

The Proposed Device Medical Face Mask (Model: M001) is available in white, blue and black colors, and flat pleated type mask, utilizing ear loops way for wearing, and it has Nose clips design for fitting the face mask around the nose.

The proposed device is manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The model of proposed device, ear loops, is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of Polyester and Spandex. The nose piece contained in the proposed device is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Metal Core Plastic (Iron wire & Polypropylene). The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

## F. Comparison with predicate de vice

Table 1 General Comparison

| Device       | Proposed Device               | Predicate device                                     | Comparison |
|--------------|-------------------------------|------------------------------------------------------|------------|
| Manufacturer | Xingyu Medical Tech Co., Ltd. | Macheng thimble<br>Technology Investment<br>Co., Ltd | -          |

{5}------------------------------------------------

No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

| 510K number            |                                                                                   | K221273                                                                                                                                                                                                                                                                                                                                                                                                  | K211631                                                                                                                                                                                                                                                                                                                                                                                                        | -         |
|------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Device name            |                                                                                   | Medical Face Mask                                                                                                                                                                                                                                                                                                                                                                                        | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                             | Same      |
| Classification         |                                                                                   | Class II Device, FXX (21<br>CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                 | Class II Device, FXX (21<br>CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                       | Same      |
| Indications for use    |                                                                                   | The Medical Face Mask is<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These face<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This a<br>single use, disposable device(s),<br>provided non-sterile. | The Surgical Face Masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material.<br>These face masks are intended<br>for use in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. This a single<br>use, disposable<br>device(s), provided non-sterile. | Same      |
| Material               | Outer<br>layer                                                                    | Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                         | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                        | Same      |
|                        | Middle<br>layer                                                                   | Melt-blown Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                              | Melt blown polypropylene filter                                                                                                                                                                                                                                                                                                                                                                                | Same      |
|                        | Inner layer                                                                       | Non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                         | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                        | Same      |
|                        | Nose clip                                                                         | Metal Core Plastic<br>(Iron wire & Polypropylene)                                                                                                                                                                                                                                                                                                                                                        | Malleable aluminum wire                                                                                                                                                                                                                                                                                                                                                                                        | Different |
|                        | Ear loops                                                                         | Polyester and Spandex                                                                                                                                                                                                                                                                                                                                                                                    | Polyester                                                                                                                                                                                                                                                                                                                                                                                                      | Different |
| Color                  |                                                                                   | White, Blue, Black                                                                                                                                                                                                                                                                                                                                                                                       | Blue                                                                                                                                                                                                                                                                                                                                                                                                           | Different |
| Dimension<br>(length)  |                                                                                   | 175mm+/-5mm                                                                                                                                                                                                                                                                                                                                                                                              | 17.5cm+/-0.5cm                                                                                                                                                                                                                                                                                                                                                                                                 | Same      |
| Dimension<br>(Width)   |                                                                                   | 95mm+/-5mm                                                                                                                                                                                                                                                                                                                                                                                               | 9.5cm+/-0.5cm                                                                                                                                                                                                                                                                                                                                                                                                  | Same      |
| OTC use                |                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                            | Same      |
| Sterility              | Non-Sterile                                                                       | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                           |           |
| Use                    | Single Use, Disposable                                                            | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                           |           |
| ASTM F2100-19<br>level | Level 2 & 3                                                                       | Level 2 & 3                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                           |           |
| Biocompatibility       | Meet ISO10993, proved non-<br>cytotoxicity, non-irritating and<br>non-sensitizing | Meet ISO10993, proved non-<br>cytotoxicity, non-irritating and<br>non-sensitizing                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                           |           |

{6}------------------------------------------------

No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

From the comparison we found the material of proposed device's nose clip and Ear loops, as well as its color, are different from the predicate device. The biocompatibility tests were conducted to both components to ensure their complance to the ISO10993-5 and ISO10993-10. There is no new risk generated from the difference of the material.

### G. Non-Clinical TestConclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004:

- A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- A ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks
- A A.STM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity):
- > EN 14683, Medical Face Masks-Requirements and Test Methods;
- A ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;
- A ASTM F2299, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
- A 16 CFR 1610, Standard for the Flammability of clothing textiles;

Table 2 - Performance Testing

| Item                                                      | Purpose                                                                                                                                                    | Proposed device                                                                                                                                                              | Acceptance<br>Criteria                                                                             | Result |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------|
|                                                           |                                                                                                                                                            |                                                                                                                                                                              |                                                                                                    |        |
| Fluid Resistance<br>Performance<br>ASTM F1862             | Assess the<br>performance of a<br>mask to resistance<br>to a synthetic blood<br>preparation<br>targeted toward the<br>mask at a set<br>pressure            | 32 out of 32 per<br>lot pass at 120<br>mmHg for 3<br>non-consecutive<br>lots tested;<br>32 out of 32 per<br>lot pass at 160<br>mmHg for 3<br>non-consecutive<br>lots tested; | 29 out of 32 pass at<br>120 mmHg for<br>Level 2<br>29 out of 32 pass at<br>160 mmHg for<br>Level 3 | PASS   |
| Particulate<br>Filtration<br>Efficiency ASTM<br>F2299     | Assess the<br>performance of<br>a mask to<br>penetration by<br>sub-micron<br>polystyrene<br>latex particles of<br>0.1 micron                               | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>Average Result-<br>Level 2 : 99.2%<br>Level 3 : 99.4%                                              | ≥ 98%                                                                                              | PASS   |
| Bacterial<br>Filtration<br>Efficiency ASTM<br>F2101       | Assess the<br>performance of<br>a mask to<br>penetration by a<br>prepared solution<br>with known<br>concentration of<br>an indicator<br>bacterial organism | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>Average Result-<br>Level 2 : 99.2%<br>Level 3 : 99.4%                                              | ≥ 98%                                                                                              | PASS   |
| Differential<br>Pressure (Delta P)<br>EN 14683 Annex<br>C | Assess the<br>performance of<br>a mask for<br>resistance to<br>air movement<br>through the                                                                 | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>Average Result-                                                                                    | < 6.0mmH2O/cm²                                                                                     | PASS   |
| Flammability 16<br>CFR 1610                               | materials of the<br>face of the mask                                                                                                                       | Level 2 : 4.4<br>mmH2O/cm²<br>Level 3 : 5.1<br>mmH2O/cm²                                                                                                                     |                                                                                                    |        |
|                                                           | Assess the<br>resistance of a<br>mask to ignition                                                                                                          | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br><br>Average Result-<br>Level 2 : Class 1<br>Level 3 : Class 1                                      | Class 1                                                                                            | PASS   |

{7}------------------------------------------------

No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

{8}------------------------------------------------

No. 2189 Yaoqian Road, Gaomi Economic Development Zone, Weifang City, Shandong Province, China

Table 3 - Biocompatibility Testing

| Test Method   | Purpose                                                           | Acceptance Criteria | Result                                                                          |
|---------------|-------------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------|
| Cytotoxicity  | Assess the potential risk<br>of Cytotoxicity of mask<br>material  | Non-Cytotoxic       | PASS<br>Under the conditions of the<br>study, the device is non-<br>cytotoxic.  |
| Irritation    | Assess the potential risk<br>of Irritation of mask<br>material    | Non-Irritating      | PASS<br>Under the conditions of the<br>study, the device is non-<br>irritating. |
| Sensitization | Assess the potential risk<br>of Sensitization of mask<br>material | Non-Sensitizing     | PASS<br>Under the conditions of the<br>study, the device is non-<br>sensitizing |

## H. Clinical Test Conclusion

No clinical study is included in this submission.

## I. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Medical Face Mask (Model: M001) is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K211631.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K221273](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K221273)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
