← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K220191

# Disposable Face Mask (K220191)

_Jiujiang Taixin Technology Co., Ltd. · FXX · May 24, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K220191

## Device Facts

- **Applicant:** Jiujiang Taixin Technology Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** May 24, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The DISPOSABLE FACE MASK is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

## Device Story

Disposable face mask (Models TX-005, TX-006) designed for infection control; protects patient and healthcare personnel from microorganisms, body fluids, and particulates. Flat pleated construction with nose clip and elastic ear loops. Three-layer design: spunbonded non-woven outer/inner layers; melt-blown non-woven middle layer. Nose clip made of PE-coated tin-plate wire. Device worn over mouth and nose; secured by ear loops. Single-use, non-sterile, disposable. Used in clinical settings to reduce exposure to blood and body fluids. Benefits patient and provider by providing physical barrier against contaminants.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F2101 (BFE >99.8%), ASTM F2299 (PFE >98.5%), ASTM F1862 (fluid resistance 120-160 mmHg), and EN 14683 (differential pressure 4.9-5.0 mmH2O/cm²). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10 (cytotoxicity, irritation, sensitization).

## Technological Characteristics

Three-layer non-woven construction (spunbonded/melt-blown). PE-coated tin-plate wire nose clip. Spandex ear loops. Dimensions 175mm x 95mm. Non-sterile. Complies with ASTM F2100 (Level 2/3), ASTM F1862, ASTM F2101, ISO 10993-5, and ISO 10993-10.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Mask ([K202424](/device/K202424.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 24, 2022

Jiujiang Taixin Technology Co., Ltd. % Cassie Lee Manager Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road Huangpu District Guangzhou, Guangdong 510700 China

Re: K220191

Trade/Device Name: DISPOSABLE FACE MASK (Model: TX-005, TX-006) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: April 19, 2022 Received: April 25, 2022

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220191

Device Name DISPOSABLE FACE MASK (Model: TX-005, TX-006)

Indications for Use (Describe)

The DISPOSABLE FACE MASK is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| <span></span> Residential Use (Act 250/Section 248 Review) | <span></span> Commercial or Industrial Use (Act 250/Section 248 Review) |
|------------------------------------------------------------|-------------------------------------------------------------------------|
|------------------------------------------------------------|-------------------------------------------------------------------------|

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary of K220191

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

Subject Device: DISPOSABLE FACE MASK (Model: TX-005, TX-006)

510(k) Number: K220191

1. Date of the summary prepared: April 19, 2022

### 2. Submitter's Information

510(k) Owner's Name: Jiujiang Taixin Technology Co., Ltd. Establishment Registration Number: 3017207447 Address: Zone A, Ruichang Science and Technology Park, Ruichang City, Jiujiang City, Jiangxi Province, China Contact Person: Xiaojie Li Email: 43590288@qq.com

#### Application Correspondent:

Contact Person: Ms. Cassie Lee Share Info (Guangzhou) Medical Consultant Ltd. Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Tel: +86 20 8266 2446

# 3. Subject Device Information

Email: regulatory@share-info.com

Type of 510(k): Traditional Classification Name: Surgical Face Mask Trade Name: DISPOSABLE FACE MASK Model Name: TX-005, TX-006

{4}------------------------------------------------

Review Panel: General Hospital Product Code: FXX Regulation Number: 21 CFR 878.4040 Regulatory Class: II

#### 4. Primary Predicate Device Information

Sponsor: Foshan Xinbao Technology Co., Ltd. Trade Name: Surgical Mask Common name: Surgical apparel Classification Name: Mask, Surgical 510(K) Number: K202424 Review Panel: General Hospital Product Code: FXX Regulation Number: 21 CFR 878.4040 Regulation Class: II

#### 5. Indications for Use

The DISPOSABLE FACE MASK is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

#### 6. Device Description

The DISPOSABLE FACE MASK is flat pleated style mask, utilizing ear loops way for wearing, and they all has nose clip design for fitting the face mask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer facing layers are made of spunbonded non-woven, and the middle layer is made of melt-blown non-woven the model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural 

{5}------------------------------------------------

rubber latex.

The nose clip contained in the proposed device(s) is in the layers of face mask to allow the user to fit the mask around their nose, which is made of PE coated Tin-plate wire.

The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

### 7. Summary of Technological Characteristics

The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

| Elements of<br>Comparison               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                              | Verdict           |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Company                                 | Jiujiang Taixin Technology Co.,<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                      | Zhejiang The Purples Protective<br>Products Co.,Ltd                                                                                                                                                                                                                                                                                                                                           | --                |
| 510 (k)                                 | K220191                                                                                                                                                                                                                                                                                                                                                                                                      | K202424                                                                                                                                                                                                                                                                                                                                                                                       | --                |
| Trade Name                              | DISPOSABLE FACE MASK                                                                                                                                                                                                                                                                                                                                                                                         | Surgical Mask                                                                                                                                                                                                                                                                                                                                                                                 | --                |
| Classification                          | Class II Device, FXX (21<br>CFR 878.4040)                                                                                                                                                                                                                                                                                                                                                                    | Class II Device, FXX (21<br>CFR 878.4040)                                                                                                                                                                                                                                                                                                                                                     | Same              |
| Intended use                            | The DISPOSABLE FACE MASK is<br>intended to be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids, and<br>particulate material. These face<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single-use, disposable device,<br>provided non-sterile. | The Surgical Masks are intended<br>to be worn to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material. These masks are<br>intended for use in infection<br>control practices to reduce<br>potential exposure to blood and<br>body fluids.<br>This is a single use, disposable<br>device, provided nonsterile. | Same              |
| Outer facing<br>layer                   | Spunbonded non-woven                                                                                                                                                                                                                                                                                                                                                                                         | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                       | Similar<br>Note 1 |
| Middle layer                            | Melt-Blown non-woven                                                                                                                                                                                                                                                                                                                                                                                         | Melt blown polypropylene                                                                                                                                                                                                                                                                                                                                                                      | Similar<br>Note 1 |
| Inner facing<br>layer                   | Melt-Blown non-woven                                                                                                                                                                                                                                                                                                                                                                                         | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                       | Similar<br>Note 1 |
| Nose clip                               | PE coated Tin-plate wire                                                                                                                                                                                                                                                                                                                                                                                     | Galvanized iron wire                                                                                                                                                                                                                                                                                                                                                                          | Similar<br>Note 1 |
| Ear loops                               | Spandex                                                                                                                                                                                                                                                                                                                                                                                                      | Nylon and Spandex                                                                                                                                                                                                                                                                                                                                                                             | Similar<br>Note 1 |
| Color                                   | white + blue                                                                                                                                                                                                                                                                                                                                                                                                 | Blue                                                                                                                                                                                                                                                                                                                                                                                          | Similar<br>Note 1 |
| Specification<br>and<br>Dimension       | 175mm×95mm                                                                                                                                                                                                                                                                                                                                                                                                   | Width: 17.5cm±1cm<br>Length: 9.5cm±1cm                                                                                                                                                                                                                                                                                                                                                        | Same              |
| Sterility                               | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                  | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                   | Same              |
| Use                                     | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                       | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                        | Same              |
| ASTM F2100<br>Level                     | Level 3<br>(Model: TX-006)                                                                                                                                                                                                                                                                                                                                                                                   | Level 2<br>(Model:<br>TX-005)                                                                                                                                                                                                                                                                                                                                                                 | Same              |
| Fluid<br>Resistance<br>Performance      | Pass at 160 mmHg<br>(Model: TX-006)                                                                                                                                                                                                                                                                                                                                                                          | Pass at 120 mmHg<br>(Model:<br>TX-005)                                                                                                                                                                                                                                                                                                                                                        | Same              |
| Particulate<br>Filtration<br>Efficiency | 98.57%<br>(Model: TX-006)                                                                                                                                                                                                                                                                                                                                                                                    | 98.6%<br>(Model:<br>TX-005)                                                                                                                                                                                                                                                                                                                                                                   | Similar<br>Note 2 |
| Bacterial<br>Filtration                 | 99.83%<br>(Model: TX-006)                                                                                                                                                                                                                                                                                                                                                                                    | 99.84%<br>(Model:<br>TX-005)                                                                                                                                                                                                                                                                                                                                                                  | Similar<br>Note 2 |

{6}------------------------------------------------

{7}------------------------------------------------

| Efficiency               |                                                                                              | TX-005)                                 |                                                                                              |                           |                   |
|--------------------------|----------------------------------------------------------------------------------------------|-----------------------------------------|----------------------------------------------------------------------------------------------|---------------------------|-------------------|
| Differential<br>Pressure | 5.0mm<br>H2 O/cm²<br>(Model: TX-006)                                                         | 4.9mm<br>H2 O/cm²<br>(Model:<br>TX-005) | Pass at <6.0<br>mmH2O/cm2                                                                    | Pass at <6.0<br>mmH2O/cm2 | Similar<br>Note 2 |
| Flammability             | Class 1                                                                                      |                                         | Class 1                                                                                      |                           | Same              |
| Cytotoxicity             | Under the conditions of the study, the device is noncytotoxic.<br>Comply with ISO 10993-5.   |                                         | Under the conditions of the study, the device is noncytotoxic.<br>Comply with ISO 10993-5.   |                           | Same              |
| Irritation               | Under the conditions of the study, the device is nonirritating. Comply with ISO 10993-10.    |                                         | Under the conditions of the study, the device is nonirritating. Comply with ISO 10993-10.    |                           | Same              |
| Sensitization            | Under the conditions of the study, the device is nonsensitizing.<br>Comply with ISO 10993-10 |                                         | Under the conditions of the study, the device is nonsensitizing.<br>Comply with ISO 10993-10 |                           | Same              |

# Comparison in Detail(s):

# Note 1:

Although the "Outer facing Layer", "Middle layer", "Inner facing layer", "Ear loops" and "Color" of subject device are slightly difference with predicate device, it meets the requirement standard ASTM F2100, ASTM F1862, ASTM F2101, ISO 10993-5 and ISO 10993-10.

# Note 2:

Although the "Particulate Filtration Efficiency", "Bacterial Filtration Efficiency" and "Differential Pressure" of subject device is a little different from the predicate device, and they all meet the requirements of essential performance standard ASTM F2100.

#### 8. Summary of Non-Clinical Performance Testing

# Performance Testing summary:

| Test item | Test method | Pass criteria | Test results |
|-----------|-------------|---------------|--------------|
|-----------|-------------|---------------|--------------|

{8}------------------------------------------------

| (Performance Level 2<br>and Level 3)                                                                |                                                                                                                                                                                                                                |                                                                                                                | /Verdict                                                                                                                |
|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Bacterial filtration<br>efficiency                                                                  | ASTM F2101-14 Standard Test<br>Method for Evaluating the Bacterial<br>Filtration Efficiency (BFE) of Medical<br>Face Mask Materials, Using a<br>Biological Aerosol of Staphylococcus<br>aureus according to ASTM<br>F2100:2019 | Level 2: ≥98%<br><br>Level 3: ≥98%                                                                             | 99.84% / Pass<br>Model: TX-005<br><br>99.83% / Pass<br>Model: TX-006                                                    |
| Differential pressure<br>(Delta-P)                                                                  | EN 14683: 2019, Annex C Medical<br>face masks - Requirements and test<br>methods according to ASTM<br>F2100:2019                                                                                                               | Level 2: <6.0 mm H2O/cm²<br><br>Level 3:<6.0 mm H2O/cm²                                                        | 4.9 mm H2O/cm² /<br>Pass<br>Model: TX-005<br><br>5.0 mm H2O/cm² /<br>Pass<br>ModelTX-006                                |
| Sub-micron particulate<br>filtration<br>efficiency<br>at 0.1 µm of Polystyrene<br>Latex Spheres     | ASTM F2299-03 Standard Test<br>Method for Determining the Initial<br>Efficiency of Materials Used in<br>Medical Face Masks to Penetration<br>by Particulates Using Latex Spheres<br>according to ASTM F2100:2019               | Level 2: ≥ 98%<br><br>Level 3: ≥ 98%                                                                           | 98.60% / Pass<br>Model: TX-005<br><br>98.57% / Pass<br>Model: TX-006                                                    |
| Resistance to<br>penetration by synthetic<br>blood, minimum pressure<br>in mm Hg<br>for pass result | ASTM F1862/F1862M-17<br>Standard Test Method for<br>Resistance of Medical Face Masks<br>to Penetration by Synthetic Blood<br>(Horizontal Projection of Fixed<br>Volume at a Known Velocity)<br>according to ASTM F2100:2019    | Level 2: Fluid<br>resistant claimed<br>at 120 mm Hg<br><br>Level 3: Fluid<br>resistant claimed<br>at 160 mm Hg | Fluid Resistant<br>claimed at 120 mm<br>Hg<br>/ Pass<br>Model: TX-005<br><br>Fluid Resistant<br>claimed at 160 mm<br>Hg |
|                                                                                                     |                                                                                                                                                                                                                                |                                                                                                                | / Pass                                                                                                                  |
|                                                                                                     |                                                                                                                                                                                                                                |                                                                                                                | Model: TX-006                                                                                                           |
| Flame spread                                                                                        | 16 CFR Part 1610 Standard for the<br>Flammability of Clothing according<br>to ASTM F2100:2019                                                                                                                                  | Level 2: Class 1                                                                                               | Class 1 / Pass<br>Model: TX-005                                                                                         |
|                                                                                                     |                                                                                                                                                                                                                                | Level 3: Class 1                                                                                               | Class 1 / Pass<br>Model: TX-006                                                                                         |

{9}------------------------------------------------

# Biocompatibility Testing Summary:

According to ISO 10993-1: 2018, the nature of body contact for the subject device is the Surface Device category, Skin Contact, and duration of the contact is A-Limited (<24 h). The two levels(Model:TX-005,TX-006) of masks are produced using the same material and process. The following tests for the subject device were conducted to demonstrate that the subject device is biocompatible and safe for its intended use:

| Test Item     | Proposed device                                                                                                           | Result |
|---------------|---------------------------------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity  | Under the conditions of the study, the subject device extract was determined to be non-cytotoxic.                         | PASS   |
| Irritation    | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.  | PASS   |
| Sensitization | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing. | PASS   |

# 9. Summary of Clinical Performance Test

No clinical study is included in this submission.

# 10. Final Conclusion

The subject device is a safe, as effective, and perform as well or better than the legally marketed predicated K202424.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K220191](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K220191)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
