D-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply

K213970 · Danameco Medical Joint Stock Corporation · FXX · Jan 23, 2022 · General, Plastic Surgery

Device Facts

Record IDK213970
Device NameD-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply
ApplicantDanameco Medical Joint Stock Corporation
Product CodeFXX · General, Plastic Surgery
Decision DateJan 23, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The D-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The surgical face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. These are single use, disposable devices, provided non-sterile.

Device Story

Three-layer, flat-pleated surgical face mask; constructed of nonwoven polypropylene (inner/outer layers) and melt-blown polypropylene (middle layer). Features ear loops (polyester/spandex) and polypropylene nose piece for fit. Single-use, disposable, non-sterile. Used in clinical settings for infection control; worn by healthcare personnel and patients to provide barrier protection against blood, body fluids, and particulates. Provides Level 3 fluid resistance.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration), ASTM F2101-19 (bacterial filtration), and 16 CFR 1610 (flammability). Biocompatibility confirmed via ISO 10993-5 and ISO 10993-10.

Technological Characteristics

Three-layer nonwoven polypropylene construction; melt-blown middle layer. Flat-pleated design. Dimensions: 17.5cm x 9.5cm. Ear loops: 80% polyester/20% spandex. Polypropylene nose piece. Non-sterile. Performance: ASTM F2100 Level 3 (Fluid resistance 160 mmHg, PFE ≥99% at 0.1µm, BFE ≥99%, Delta P < 4 mmH2O/cm²).

Indications for Use

Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material in infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. January 23, 2022 Danameco Medical Joint Stock Corporation Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K213970 Trade/Device Name: D-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: December 18, 2021 Received: December 20, 2021 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K213970 Device Name D-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply Indications for Use (Describe) The D-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The surgical face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. These are single use, disposable devices, provided non-sterile. Type of Use (Select one or both, as applicable) | <input type="checkbox"/> Registration for Non-STP CMA Sales Only | <input checked="" type="checkbox"/> Registration for Non-STP CMA Sales & STP CMA Sales | |------------------------------------------------------------------|----------------------------------------------------------------------------------------| |------------------------------------------------------------------|----------------------------------------------------------------------------------------| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY K213970 This 510(k) Summary as per 21 CFR 807.92 Date of submission: January 21, 2022 ## (1) Applicant information | 510(k) Owner/ Applicant: | DANAMECO MEDICAL JOINT STOCK CORPORATION | |--------------------------|-------------------------------------------------------------------------| | Address: | Head office: | | | DANAMECO MEDICAL JOINT STOCK CORPORATION | | | No 12, Trinh Cong Son Street, Hoa Cuong Nam, | | | Hai Chau District, 550000, Da Nang City, Viet Nam | | | Factory: | | | Quang Nam Medical Device Factory- Brand of Danameco | | | Medical Joint Stock Corporation | | | Trang Nhat 2 Industrial Zone, Dien Hoa Commune, Dien | | | Ban Town, 560000 Quang Nam Province, Viet Nam. | | Contact: | Huynh Thi Li Li (Ms.) | | Email: | info@danameco.com; | | | qa@danameco.com | | Phone: | (+84)-236-3849833 | | Owner/Operator Number: | 10071934 | | FEI Number: | 3016970619 | | US Agent: | Emergo Global Representation LLC at address: 2500 | | | Bee Cave Road, Building 1, Suite 300, Austin, | | | TX US 78746 | | Contact of US agent: | Michael van der Woude, Phone: 512 3279997, | | | Email: USAgent@UL.com | | Correspondent: | Regulatory Technology Services at address: 1000 | | | Westgate Drive, Suite #510, Saint Paul, Minnesota, 55114 | | | Contact of Correspondent: Ms. Prithul Bom - Accredited Person, Reviewer | {4}------------------------------------------------ Phone 612-963-0379 Email: prithul.bom@rts3pro.com ## (2) Subject device | Trade name: | D-Care Surgical Face Mask 3 Ply, | |-------------------------------|----------------------------------| | | D-Care Medical Face Mask 3 Ply | | Model number: | KTY3PLYKV/KTY3PLYVT | | Common Name: | Surgical Face Mask | | Classification Name: | Masks, Surgical | | Review Panel: | General Hospital | | Regulation Medical Specialty: | General & Plastic Surgery | | Product Code: | FXX | | Device Classification: | Class II per 21 CFR §878.4040 | #### (3) Predicate device | Submitter: | Zhejiang Hongyu Medicali Commodity Co.,Ltd | |------------------------|--------------------------------------------| | Device name: | Surgical Face Mask | | 510(k) number: | K211897 | | Product Code: | FXX | | Device Classification: | Class II per 21 CFR §878.4040 | #### (4) Description of device The subject device is three-layers, flat-pleated masks constructed of nonwoven polypropylene materials, the inner and outer layers are made of polypropylene nonwoven, and the middle layer is made of melt blown polypropylene. The subject device is provided with ear loops. Ear loops is made of 80% polyester 20% spandex, not made with natural rubber latex. A polypropylene wire nose piece is placed within the binding for comfort and individualized fit, allow the user to fit the facemask around their nose. The subject device is provided in blue color. The blue colorant is made of polypropylene master batch. {5}------------------------------------------------ This subject device is provided with size 17.5 cm ± 0.5 cm. The subject device is single-use, disposable devices, provided non-sterile. ## (5) Indications for use The D-Care Surgical Face Mask 3 Ply, D-Care Medical Face Mask 3 Ply are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The surgical face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. These are single use, disposable devices, provided non-sterile. # (6) Comparison of Technological Characteristics and Performance testing withthe Predicate device | Item | Subject Device | Predicate Device | Comparison | | |-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--| | 510 (k)<br>number | - | K211897 | Not<br>applicable | | | Applicant | DANAMECO MEDICAL<br>JOINT STOCK<br>CORPORATION | ZHEJIANG HONGYU<br>MEDICALI COMMODITY<br>CO., LTD | Not<br>applicable | | | Product<br>Name | D-Care Surgical Face Mask<br>3 Ply, D-Care Medical Face<br>Mask 3 Ply | Surgical Face mask | Not<br>applicable | | | Model<br>number | KTY3PLYKV/KTY3PLYVT | Ear loops | Not<br>applicable | | | Device class | Class II Device, FXX<br>(21 CFR 878.4040) | Class II Device, FXX<br>(21 CFR 878.4040) | Same | | | Indications<br>for Use | The D-Care Surgical Face<br>Mask 3 Ply, D-Care Medical<br>Face Mask 3 Ply are<br>intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. The surgical face<br>masks are intended for use<br>in infection control practices<br>to reduce the potential<br>exposure to blood and body<br>fluids. These are single use,<br>disposable devices, provided<br>non-sterile. | The Surgical Face Mask is<br>intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. This face mask is<br>intended for use in infection<br>control practices to reduce<br>the potential exposure to<br>blood and body fluids. This<br>is single use, disposable<br>device, provided non-sterile. | Same | | | Material | | | | | | Outer layer | Polypropylene non-woven | Spunbond polypropylene | Same | | | Middle layer | Polypropylene meltblown | Meltblown polypropylene | Same | | | Inner layer | Polypropylene non-woven | Spunbond polypropylene | Same | | | Nose piece | Polypropylene | lron core coated with<br>polypropylene resin | Different | | | Design<br>Features | Ear loops: 80% polyester<br>20% spandex | Ear loops: polyester +<br>spandex | Similar | | | Mask Style | Flat Pleated | Flat Pleated | Same | | | Color | Blue | Blue | Same | | | Specification<br>and<br>Dimension | Length: 17.5cm±0.5cm<br>Width: 9.5cm±0.5cm | Length: 17.5cm±0.2cm<br>Width: 9.5cm±0.2cm | Similar | | | Dimension<br>ear loops | 17cm±0.5cm | 17cm±0.2cm | Similar | | | OTC use | Yes | Yes | Same | | | Sterility | Non-Sterile | Non-Sterile | Same | | | Use | Single Use, Disposable | Single Use, Disposable | Same | | | Performance Characteristics | | | | | | Fluid<br>Resistance<br>Performance<br>ASTM<br>F1862 | Pass at 160 mmHg<br>(Level 3 Fluid Resistance) | Pass at 160 mmHg<br>(Level 3 Fluid Resistance) | Same | | | Particulate<br>Filtration<br>Efficiency<br>ASTM<br>F2299 | ≥99% at 0.1µm | ≥99% at 0.1µm | Same | | | Bacterial<br>Filtration<br>Efficiency<br>ASTM<br>F2101-19 | ≥99% | ≥99% | Same | | | Differential<br>Pressure<br>(Delta P) | < 4 mmH2O/cm² | < 4 mmH2O/cm² | Same | | | Flammability<br>16 CFR<br>1610 | Class 1 | Class 1 | Same | | | Biocompatibility | | | | | | Cytotoxicity<br>ISO 10993-5 | Under the conditions of the<br>studies, the subject device is<br>non-cytotoxic | Under the conditions of the<br>study, the device extract<br>was determined to be non-<br>cytotoxic | Same | | | Irritation<br>ISO 10993-<br>10 | Under the conditions of the<br>studies, the subject device is<br>non-irritating | Under the conditions of the<br>study, the device extract<br>was determined to be non-<br>irritant | Same | | | Sensitization<br>ISO 10993-<br>10 | Under the conditions of the<br>studies, the subject device is<br>non-sensitizing | Under the conditions of the<br>study, the device extract<br>was determined to be non-<br>sensitizer | Same | | {6}------------------------------------------------ {7}------------------------------------------------ The subject device is same indication for use, design feature, design style, component of material (except nose piece), performance characteristics and biocompatibility with predicate device, similar dimension of size in comparison with predicate device. The component of ear loop of subject device is similar with predicate device but clearly claim the percent of composition. The subject device is different material of nose piece to predicate device. The subject device which was included the nose piece had been tested and the result was complied with acceptance criteria. {8}------------------------------------------------ ## (7) Summary of Non-clinical test | Performance characteristics | | | | | | | |-----------------------------------------------------|--------------------------------------------------|-------------------------------------------------|------------|--|--|--| | Test Method | Subject Device | Acceptance Criteria | Conclusion | | | | | Fluid Resistance<br>Performance<br>ASTM F1862 | ≥29 out of 32 passed in<br>160 mmHg<br>(Level 3) | 29 out of 32 passed in<br>160 mmHg<br>(Level 3) | Pass | | | | | Particulate Filtration<br>Efficiency<br>ASTM F2299 | ≥99% at 0.1µm | ≥98% at 0.1µm | Pass | | | | | Bacterial Filtration<br>Efficiency<br>ASTM F2101-19 | ≥99% | ≥98% | Pass | | | | | Differential<br>Pressure (Delta P) | < 4 mmH2O/cm² | < 6.0 mm H2O/cm² | Pass | | | | | Flammability<br>16 CFR 1610 | Class 1(≥ 3.5 seconds) | Class 1 (≥ 3.5 seconds) | Pass | | | | Performance testing were performed on three non-consecutive lots to support that the performance specification are maintained cross production lots and the lot-to-lot variability in performance is acceptable. Sample size of each lot is complied with ISO 2859-1, general inspection level II as FDA recommendation and acceptance quality limit (AQL) of 4%. The results of performance testing of subject device demonstrate that the subject device met all design specification as was same to predicate device, complied with requirements in guidance: "Surgical Masks – Premarket Notification [510(k)] Submissions", complied with standard ASTM F2100 at level 3. | Biocompatibility | | | | | |-----------------------------|----------------------------------------------------------------------|--------------------------------|------------|--| | Standard | Subject Device | Acceptance<br>Criteria | Conclusion | | | Cytotoxicity<br>ISO 10993-5 | Under the conditions of the studies,<br>the device is non-cytotoxic. | The device is<br>non-cytotoxic | Pass | | {9}------------------------------------------------ | Skin<br>Sensitization<br>test ISO<br>10993-10 | Under the conditions of the studies,<br>the subject device is non-<br>sensitizing | The device is<br>non-sensitizing | Pass | |-----------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------|------| | Skin Irritation<br>test ISO<br>10993-10 | Under the conditions of the studies,<br>the device is non-irritating. | The device is<br>non-irritating | Pass | The biocompatibility evaluation of this subject device was conducted according to ISO 10993-1: nature of body contact of the subject device is belonged to category Surface device, intact skin, with contact duration A-limited (≤24h) The subject device was evaluated in its final finished form, in blue color. The results of biocompatibility testing of subject device demonstrate that the subject device is biocompatible and safe for its intended use. ## (8) Summary of Clinical test No clinical study is included in this submission. ## (9) Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...