← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K213806

# Disposable Medical Face Mask (K213806)

_Xiantao Zhibo Nonwoven Products Co., Ltd. · FXX · Mar 12, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K213806

## Device Facts

- **Applicant:** Xiantao Zhibo Nonwoven Products Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 12, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

## Device Story

Disposable medical face mask; protects patient and healthcare personnel from microorganisms, body fluids, and particulates. Composed of three non-woven fabric layers (outer, middle melt-blown filter, inner), HDPE nose clip, and nylon/spandex ear loops. Features pleats for coverage from nose to chin. Used in clinical settings for infection control. Single-use, non-sterile, over-the-counter device. Provides barrier protection against blood and body fluids; reduces potential for cross-contamination.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862 (fluid resistance at 160 mmHg), ASTM F2299 (particulate filtration efficiency 98.9-99.5%), ASTM F2101 (bacterial filtration efficiency 99.88-99.97%), and EN 14683 (differential pressure 3.3-3.4 mmH2O/cm2). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

Three-layer non-woven construction: outer/inner spun-bond polypropylene, middle melt-blown polypropylene filter. HDPE nose clip. Nylon/spandex ear loops. Dimensions 175mm x 95mm. Non-sterile. Meets ASTM F2100-19 Level 3 standards. Biocompatible per ISO 10993.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Level 3 Fluid Resistant Procedure/Surgical Mask ([K211827](/device/K211827.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 12, 2022

Xiantao Zhibo Nonwoven Products Co., Ltd % Jarvis Wu Consultant Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Century Ave. Shanghai, Shanghai 200122 China

Re: K213806

Trade/Device Name: Disposable Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 6, 2021 Received: December 13, 2021

Dear Jarvis Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213806

Device Name Disposable Medical Face Mask

#### Indications for Use (Describe)

The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|                                                 |
|                                                 |

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K213806

#### A. Applicant:

Name: XIANTAO ZHIBO NONWOVEN PRODUCTS CO., LTD. Address: NO.8, Hefeng industrial zone, Pengchang Town, Xiantao City, HUBEI, CHINA Contact Person: Fen Peng Tel: +86 18872609993 Mail: 260993463@qq.com Submission Correspondent: Primary contact: Mr. Jarvis Wu Shanghai SUNGO Management Consulting Co., Ltd. 14th floor, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungogloba1.com Secondary contact: Mr. Raymond Luo 14th floor, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com Document Date Prepared: 2022/3/9 B. Device:

Proprietary Name: Disposable Medical Face Mask Common Name: Surgical Face Mask Model: FMZB3

#### Regulatory Information

Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

{4}------------------------------------------------

#### C. Predicate de vice:

| 510K    | Device name                                         | ASTM F2100-19<br>level | Manufacturer                             |
|---------|-----------------------------------------------------|------------------------|------------------------------------------|
| K211827 | Level 3 Fluid  Resistant<br>Procedure/Surgical Mask | Level 3                | Zhejiang Lanhine Medical Products<br>LTD |

#### D. Indications for use/Intended Use:

The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

#### E. Device Description:

Disposable Medical Face Mask is composed of Nylon & Spandex Ear loops, High Density Polyethylene nose clip, Inner layer Non-woven fabric, Middle Melt-blown Non-woven fabric and outer Non-woven fabric. The melt-blown material acts as the filter that stops microbes from entering or exiting the mask. Disposable Medical Face Mask feature pleats. Three pleats are used to expand the mask such that it covers the area from the nose to the chin. The type is ear loop, where a string-like material is attached to the mask and placed behind the ears. The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

#### F. Comparison with predicate de vice

Table 1 General Comparison

| Device                 | Proposed Device                                                            | Predicate device                                                           | Comparison                                                                                                                                                                                                 |                                                                                                                                                                                                       |
|------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | XIANTAO ZHIBO NONWOVEN<br>PRODUCTS CO., LTD.                               | Zhejiang Lanhine Medical<br>Products LTD.                                  | -                                                                                                                                                                                                          |                                                                                                                                                                                                       |
| 510K number            | K213806                                                                    | K211827                                                                    | -                                                                                                                                                                                                          |                                                                                                                                                                                                       |
| Classification         | Class II Device, FXX (21<br>CFR878.4040)                                   | Class II Device, FXX (21<br>CFR878.4040)                                   | Same                                                                                                                                                                                                       |                                                                                                                                                                                                       |
|                        |                                                                            |                                                                            |                                                                                                                                                                                                            |                                                                                                                                                                                                       |
|                        |                                                                            | The Disposable Medical Face                                                | The Level 3 Fluid Resistant<br>Procedure/Surgical Mask<br>(model: 15604F, 15704F)                                                                                                                          |                                                                                                                                                                                                       |
|                        |                                                                            |                                                                            | Mask is intended to be worn to<br>protect both the patient and<br>healthcare personnel from transfer                                                                                                       | are intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of                                                                                                   |
|                        |                                                                            |                                                                            | of microorganisms, body fluids<br>and particulate material. These                                                                                                                                          | microorganisms, body fluids and<br>particulate material. These face                                                                                                                                   |
|                        |                                                                            | Indications for use                                                        | face masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single use, disposable device(s),<br>provided non-sterile. | masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids.<br>This is a single use, disposable<br>device(s), provided non-sterile. |
| Material               | Outer<br>layer                                                             | Spun-bond polypropylene                                                    | Non-woven Fabric<br>(Polypropylene)                                                                                                                                                                        | Same                                                                                                                                                                                                  |
|                        | Middle<br>layer                                                            | Melt blown polypropylene filter                                            | Melton brown Fabric<br>(Polypropylene)                                                                                                                                                                     | Same                                                                                                                                                                                                  |
|                        | Inner<br>layer                                                             | Spun-bond polypropylene                                                    | Non-woven Fabric<br>(Polypropylene)                                                                                                                                                                        | Same                                                                                                                                                                                                  |
|                        | Nose<br>clip                                                               | High Density Polyethylene<br>(HDPE)                                        | Polypropylene coating iron                                                                                                                                                                                 | Different                                                                                                                                                                                             |
|                        | Ear<br>loops                                                               | Nylon and Spandex                                                          | Polyurethane                                                                                                                                                                                               | Different                                                                                                                                                                                             |
| Color                  |                                                                            | White                                                                      | Blue                                                                                                                                                                                                       | Different                                                                                                                                                                                             |
| Dimension<br>(length)  |                                                                            | $175mm \pm 5mm$                                                            | $17.5 cm +/- 0.5 cm$                                                                                                                                                                                       | Same                                                                                                                                                                                                  |
| Dimension<br>(Width)   |                                                                            | $95mm \pm 5mm$                                                             | $9.5 cm +/- 0.5 cm$                                                                                                                                                                                        | Same                                                                                                                                                                                                  |
| OTC use                |                                                                            | Yes                                                                        | Yes                                                                                                                                                                                                        | Same                                                                                                                                                                                                  |
| Sterility              |                                                                            | Non-Sterile                                                                | Non-Sterile                                                                                                                                                                                                | Same                                                                                                                                                                                                  |
| Use                    |                                                                            | Single Use, Disposable                                                     | Single Use, Disposable                                                                                                                                                                                     | Same                                                                                                                                                                                                  |
| ASTM F2100-19<br>level |                                                                            | Level 3                                                                    | Level 3                                                                                                                                                                                                    | Same                                                                                                                                                                                                  |
| Biocompatibility       | Meet ISO10993 ,proved non-cytotoxicity, non-irritating and non-sensitizing | Meet ISO10993 ,proved non-cytotoxicity, non-irritating and non-sensitizing | Same                                                                                                                                                                                                       |                                                                                                                                                                                                       |

{5}------------------------------------------------

{6}------------------------------------------------

From the comparison we found the color, the material of the current nose clip and the ear loop were different from the predicate device. The biocompatibility tests were conducted to both components to ensure their compliance to the ISO10993-10. There is no new risk generated from the difference of the material.

#### G. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

- A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- A ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks
- > ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity);
- A EN 14683, Medical Face Masks-Requirements and Test Methods;
- A ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;
- > ASTM F2299. Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
- A 16 CFR 1610, Standard for the Flammability of clothing textiles;

{7}------------------------------------------------

Table 2 - Performance Testing

| Item                                                   | Purpose                                                                                                                                           | Proposed device                                                                                                                       | Acceptance<br>Criteria                          | Result |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|--------|
| Fluid Resistance<br>Performance<br>ASTM F1862          | Assess the performance of a<br>mask to resistance to a<br>synthetic blood preparation<br>targeted toward the mask at a<br>set pressure            | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>32 out of 32 pass at<br>160 mmHg                            | 29 out of 32 pass at<br>160 mmHg for<br>level 3 | PASS   |
| Particulate<br>Filtration<br>Efficiency ASTM<br>F2299  | Assess the performance of<br>a mask to penetration by<br>sub-micron polystyrene<br>latex particles of 0.1 micron                                  | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>Lot1: 98.9%<br>Lot2: 99.3%<br>Lot3: 99.5 %                  | ≥ 98%                                           | PASS   |
| Bacterial<br>Filtration<br>Efficiency ASTM<br>F2101    | Assess the performance of<br>a mask to penetration by a<br>prepared solution with known<br>concentration of<br>an indicator bacterial<br>organism | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>Lot1: 99.88%<br>Lot2: 99.95%<br>Lot3: 99.97%                | ≥ 98%                                           | PASS   |
| Differential<br>Pressure (Delta P)<br>EN 14683 Annex C | Assess the performance of<br>a mask for resistance to<br>air movement through the<br>materials of the face of the<br>mask                         | 3 non-consecutive<br>lots tested, using a<br>sample size of<br>32/lot.<br>Lot1: 3.3 H2O/cm²<br>Lot2: 3.4 H2O/cm²<br>Lot3: 3.3 H₂O/cm² | < 6.0mmH₂O/cm²                                  | PASS   |

{8}------------------------------------------------

Table 3 – Biocompatibility Testing

| Test Method   | Purpose                                                           | Acceptance Criteria | Result                                                                          |
|---------------|-------------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------|
| Cytotoxicity  | Assess the potential risk<br>of Cytotoxicity of mask<br>material  | Non-Cytotoxic       | PASS<br>Under the conditions of the<br>study, the device is non-<br>cytotoxic.  |
| Irritation    | Assess the potential risk<br>of Irritation of mask<br>material    | Non-Irritating      | PASS<br>Under the conditions of the<br>study, the device is non-<br>irritating. |
| Sensitization | Assess the potential risk<br>of Sensitization of mask<br>material | Non-Sensitizing     | PASS<br>Under the conditions of the<br>study, the device is non-<br>sensitizing |

## H. Clinical Test Conclusion

No clinical study is included in this submission.

## I. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Disposable Medical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K211827.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K213806](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K213806)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
