← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K212344

# Surgical Mask (K212344)

_Xiantao Dingcheng Non-Woven Product Co., Ltd. · FXX · Nov 16, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K212344

## Device Facts

- **Applicant:** Xiantao Dingcheng Non-Woven Product Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Nov 16, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

## Device Story

Single-use, flat-pleated surgical mask; three-layer construction (spun-bond polypropylene inner/outer layers, melt-blown polypropylene middle layer); includes nose piece (polypropylene/galvanized-iron) and elastic ear straps (polyester/spandex). Device worn over mouth and nose to provide physical barrier against microorganisms, body fluids, and particulates. Used in clinical settings for infection control. No electronic components or software.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Three-layer flat-pleated mask; materials: spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter), polypropylene/galvanized-iron (nose piece), polyester/spandex (ear straps). ASTM F2100 Level 3 performance. Non-sterile, single-use. No software or energy source.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K182515](/device/K182515.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, featuring a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text next to it.

November 16, 2021

Xiantao Dingcheng Non-woven Product Co., Ltd Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai China

Re: K212344

Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: September 30, 2021 Received: October 18, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Liqun Zhao -S

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212344

Device Name Surgical Mask

Indications for Use (Describe)

Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

# 1.0 Submitter's information

Name: Xiantao Dingcheng Non-woven Product Co., Ltd Address: Liukou Industrial Park, Xiantao City, Hubei Province, CHINA Phone Number: +86-18007229722 Contact: Ms. Cheng Qin Date of Preparation: 21/07/2021

#### Designated Submission Correspondent

Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

# 2.0 Device information

Surgical Mask Trade name: Common name: Surqical Mask Classification name: Mask, Surgical Model(s): ear strap, 175×95mm

# 3.0 Classification

Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Panel: Surgical apparel

#### 4.0 Predicate device information

Manufacturer: Wuhan Dymex Healthcare Co., Ltd Surgical Face Mask Device: 510(k) number: K182515 This device has not been subject to a design-related recall.

# 5.0 Device description

{4}------------------------------------------------

The Surgical Mask is single use, three-layer, flat-pleated style with ear straps and nose piece. The mask is manufactured with three layers, the inner and outer layers are made of nonwoven fabrics, and the middle layer is made of melt blown fabrics. The ear straps are held in place over the users' mouth and nose by two elastic ear straps welded to the facemask. The elastic ear straps are not made with natural rubber latex. The nose piece on the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Polypropylene+Galvanize d-iron dual core. The Surgical Mask will be provided in blue. The masks are sold non-sterile and are intended to be single use, disposable devices.

#### 6.0 Indication for Use Statement

Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

#### 7.0 Comparison of Technological Characteristics

| Item           | Proposed device                                                                                                                                                                                                                                                                                                                                                                             | Predicated device                                                                                                                                                                                                                                                                                                                                                                                                   | Remark |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code   | FXX                                                                                                                                                                                                                                                                                                                                                                                         | FXX                                                                                                                                                                                                                                                                                                                                                                                                                 | Same   |
| Regulation No. | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                     | Same   |
| Class          | II                                                                                                                                                                                                                                                                                                                                                                                          | II                                                                                                                                                                                                                                                                                                                                                                                                                  | Same   |
| Product name   | Surgical Mask                                                                                                                                                                                                                                                                                                                                                                               | Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                                                  | -      |
| 510(k) No.     | K212344                                                                                                                                                                                                                                                                                                                                                                                     | K182515                                                                                                                                                                                                                                                                                                                                                                                                             | -      |
| Models         | ear strap, 175×95mm                                                                                                                                                                                                                                                                                                                                                                         | ear strap                                                                                                                                                                                                                                                                                                                                                                                                           | -      |
| Intended Use   | The Surgical Mask is<br>intended to be worn to protect<br>both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. It is intended for use<br>in infection control practices<br>to reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device(s),<br>provided non sterile. | The Surgical Face Masks<br>are intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer ofmicroorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device(s),<br>provided non-sterile. | Same   |
| OTC use        | Yes                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                 | Same   |
| Composite      | Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                      | Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                              | Same   |
| Internal layer | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                     | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                             | Same   |
| Middle layer   | Melt blown polypropylene                                                                                                                                                                                                                                                                                                                                                                    | Melt blown polypropylene                                                                                                                                                                                                                                                                                                                                                                                            | Same   |

#### Table 1 Technological Characteristic Comparison Table

{5}------------------------------------------------

| Material           | External layer | Spun-bond polypropylene                     | Spun-bond polypropylene     | Same          |
|--------------------|----------------|---------------------------------------------|-----------------------------|---------------|
|                    | Nose piece     | Polypropylene+Galvanize<br>d-iron dual core | Malleable polyethylene wire | * Different 1 |
|                    | ear strap      | Polyester, spandex                          | spandex                     | * Different 2 |
| Color              |                | Blue                                        | Yellow                      | * Different 3 |
| Dimension (Length) |                | $17.5cm\pm0.5cm$                            | $17.5cm\pm0.2cm$            | * Different 4 |
| Dimension (Width)  |                | $9.5cm\pm0.5cm$                             | $9.5cm\pm0.2cm$             | * Different 5 |
| Sterility          |                | Non-Sterile                                 | Non-Sterile                 | Same          |
| Single Use         |                | Yes                                         | Yes                         | Same          |
| Sterile            |                | No                                          | No                          | Same          |
| ASTM F2100 Level   |                | Level 3                                     | Level 2                     | * Different 6 |

* Different analysis:

Different 1-3: the two devices have some difference in materials and product color, product materials safety is proved by its biocompatibility, and the difference does not raise additional questions for safety and effectivenessof device.

Different 4-5: the two devices share same dimensions otherwise the tolerance is different, the little deviation in tolerance does not raise additional questions for safety and effectiveness of device.

{6}------------------------------------------------

Different 6: the two devices in different levels, the difference does not raise additional questions for safety and effectiveness of device.

# 8.0 Non-Clinical Test Conclusion

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification [510(k)] Submission issued on March 5, 2004.

Following standards are applicable for the device of Surgical Mask:

- ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks
- ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- ASTM F1862-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- ASTM F2299- 03 (Reapproved 2017) Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
- EN 14683-2019 Medical Face Masks – Requirements and Test Methods
- 16 CFR Part 1610(a) Standard for The Flammability of Clothing Textiles
- ISO 10993-1:2018 Biological evaluation of medical devices -- Part 1: Evaluation and testing
- O ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.
- ISO 10993-10 Third Edition 2010-08-01, Biological evaluation of medical devices - Part 10: Tests for irritation and skinsensitization.

# Table 2 Summary of Non-Clinical Performance Testing

| No. | Name of the Test<br>Methodology / Standard                                                                      | Purpose                                                                                                                                   | Acceptance Criteria                                                                                                                                             | Results                                                                                                                                            |
|-----|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 1   | ISO 10993-10:2010<br>Biological Evaluation<br>Of Medical Devices -<br>Part 10: Tests For<br>Irritation And Skin | This part of ISO<br>10993 assesses<br>possible contact<br>hazards from<br>chemicals released<br>from medical<br>devices, which may        | Skin Sensitization<br>Test:<br>provided grades less<br>than 1, otherwise<br>sensitization.                                                                      | All grades are 0.<br><br>All animals were survived and no<br>abnormal signs were observed<br>during the study.                                     |
| 2   | Sensitization.                                                                                                  | produce skin and mucosal irritation, eye irritation or skin sensitization.                                                                | Skin Irritation Test: If the primary irritation index is 0-0,4, the response category is Negligible. 0,5-1,9 means slight 2-4,9 means moderate 5-8 means severe | The primary irritation index is 0. The response of the proposed device was categorized as negligible under the test condition                      |
| 3   | ISO 10993-5:2009<br>Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity          | This part of ISO 10993 describes test methods to assess the in vitro cytotoxicity of medical devices.                                     | The viab.% of the 100% extract of the test article is the final result, and if viability is reduced to <70% of the blank, it has cytotoxic potential.           | Viab.% of 100% test article extract is 90.4%<br>It means the proposed device have no potential toxicity to L-929 in the MTT method                 |
| 4   | Bacterial filtration efficiency (BFE) (%)<br>ASTM F2101-19                                                      | The purpose of the test is to evaluate the Bacterial filtration efficiency (BFE) (%)                                                      | ≥98                                                                                                                                                             | Average Lot 1: 99.3%<br>Average Lot 2: 99.2%<br>Average Lot 3: 99.3%<br>Pass                                                                       |
| 5   | Differential pressure (mmH2O/cm²)<br>EN 14683                                                                   | The purpose of the test is to evaluate the Different pressure (mmH2O/cm²)                                                                 | <6.0 mmH2O/cm²                                                                                                                                                  | Average Lot 1: 4.4 mmH2O/cm²<br>Average Lot 2: 4.6 mmH2O/cm²<br>Average Lot 3: 4.6 mmH2O/cm²<br>Pass                                               |
| 6   | Sub-micron<br>particulate filtration<br>efficiency at 0.1<br>micron, % (PFE)                                    | The purpose of<br>the test is to<br>evaluate the<br>Sub-micron<br>particulate<br>filtration<br>efficiency at<br>0.1<br>micron, %<br>(PFE) | ≥98                                                                                                                                                             | Average Lot 1: 98.44%<br>Average Lot 2: 98.26%<br>Average Lot 3: 98.30%<br>Pass                                                                    |
| 7   | Resistance to<br>penetration by<br>synthetic blood,<br>Minimum<br>pressure in                                   | The purpose of<br>the test is to<br>evaluate the<br>Resistance to<br>penetration by<br>synthetic                                          | 29 of 32 test<br>articles passed at<br>160mmHg                                                                                                                  | Lot 1: 31 of 32 test<br>articles passed at<br>160mmHg;<br>Lot 2: 32 of 32 test<br>articles passed at                                               |
|     | mmHg for pass<br>result                                                                                         | blood,<br>Minimum<br>pressure in<br>mmHg for pass<br>result                                                                               |                                                                                                                                                                 | 160mmHg;<br>Lot 3: 32 of 32 test<br>articles passed at<br>160mmHg Pass                                                                             |
| 8   | Flammability                                                                                                    | The purpose of<br>the test is to<br>evaluate the<br>Flammability                                                                          | Class 1                                                                                                                                                         | Class 1, Non Flammable<br>Pass                                                                                                                     |
| 9   | Tensile strength of<br>Ear Straps                                                                               | to evaluate the<br>tensile strength                                                                                                       | >10 N                                                                                                                                                           | Average Lot 1:<br>Left: 14.3 N<br>Right: 14.4N<br>Average Lot 2:<br>Left: 14.1 N<br>Right: 14.7N<br>Average Lot 3:<br>Left: 14.6 N<br>Right: 13.9N |
| 10  | Product dimensions                                                                                              | to evaluate the                                                                                                                           | Length:                                                                                                                                                         | Pass<br>Average Lot 1:                                                                                                                             |
|     | Product dimensions                                                                                              | $17.5cm\pm0.5cm$<br>Width: $9.5cm\pm0.5cm$                                                                                                | (Length): 17.7 cm<br>(Width): 9.8 cm                                                                                                                            |                                                                                                                                                    |
|     |                                                                                                                 |                                                                                                                                           | Average Lot 2:<br>(Length): 17.8 cm<br>(Width): 9.6 cm                                                                                                          |                                                                                                                                                    |
|     |                                                                                                                 |                                                                                                                                           | Average Lot 3:<br>(Length): 17.4 cm<br>(Width): 9.6 cm                                                                                                          |                                                                                                                                                    |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

# 8.0 Clinical Test Conclusion

No clinical study implemented for the Surgical Mask.

# 9.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better the legally marketed device..

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K212344](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K212344)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
