← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K212097

# Disposable medical surgical mask (K212097)

_Hebei Titans Hongsen Medical Technology Co., Ltd. · FXX · Aug 18, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K212097

## Device Facts

- **Applicant:** Hebei Titans Hongsen Medical Technology Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Aug 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Disposable medical surgical mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The surgical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile.

## Device Story

Disposable, non-sterile, flat-pleated surgical mask; 3-layer construction (polypropylene spunbond outer/inner layers, polypropylene meltblown middle filter layer); includes polypropylene-coated steel nose piece and nylon/spandex ear loops. Used in clinical settings by healthcare personnel and patients to reduce exposure to blood, body fluids, and particulates. Device provides physical barrier; performance meets ASTM F2100 Level 3 requirements for fluid resistance, bacterial/particulate filtration, and breathability.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F2100-19, ASTM F1862-17 (fluid resistance), ASTM F2299-03 (particulate filtration ≥99%), ASTM F2101-19 (BFE ≥99%), EN 14683 (differential pressure <4.9 mmH2O/cm2), and 16 CFR 1610 (flammability Class 1). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

3-layer mask; outer/inner layers: polypropylene spunbond; middle layer: polypropylene meltblown; nose piece: polypropylene-coated steel wire; ear loops: nylon/spandex. Dimensions: 17.5cm x 9.5cm. Non-sterile, single-use. Meets ASTM F2100 Level 3 specifications.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Disposable Medical Surgical Mask ([K202511](/device/K202511.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 18, 2021

Hebei Titans Hongsen Medical Technology Co., LTD % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd Rm. 912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K212097

Trade/Device Name: Disposable medical surgical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 30, 2021 Received: July 6, 2021

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K212097

Device Name Disposable medical surgical mask

## Indications for Use (Describe)

The Disposable medical surgical mask is intended to be worn to protect and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The surgical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary - K212097

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

- 1. Date of Preparation:2021/08/18
- 2. Sponsor Identification

# Hebei Titans Hongsen Medical Technology Co., Ltd

East Industrial Concentration Zone, Nangong City, Xingtai City, 055750, Hebei Province, China.

Contact Person: Chai Wu Position: Production Manager Tel: +86-17769045117 Fax: +86-0319-7295811 Email: 846920354@qq.com

- Designated Submission Correspondent 3.
Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China Tel: +86-18910677558 Fax: +86-10-56335780 Email: Ray.Wang(@believe-med.com

- 4. Identification of Proposed Device
Trade Name: Disposable medical surgical mask Common Name: Mask, Surgical

Regulatory Information Classification Name: Mask, Surgical Classification: II Product Code: FXX Regulation Number: 878.4040 Review Panel: General Hospital

Indication for use Statement:

{4}------------------------------------------------

The Disposable medical surgical mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The surgical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile.

Device Description:

The Disposable medical surgical mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The surgical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile.

The proposed device(s) are Blue color, and Flat Pleated type mask, utilizing Ear Loops' way for wearing, and they all have Nose Piece design for fitting the facemask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of polypropylene spunbond fabric, and the middle layer is made of polypropylene meltblown fabric.

The Disposable medical surgical mask is held in place over the user's mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are made with nylon and spandex.

The nose piece contained in the proposed device(s) is in the layers of the facemask to allow the user to fit the facemask around their nose, which is made of polypropylene coated steel wire.

The proposed device(s) are sold non-sterile and are intended to be single-use, disposable devices.

- ડ. Identification of Predicate Device(s)
510(k) number: K202511 Device Name: Disposable Medical Surgical Mask Manufacturer: Improve Medical (HuNan) Co., Ltd.

#### Technological Characteristics Comparison 6.

## Table 1 General Comparison

| ITEM | Proposed Device K212097 | Predicate Device K202511 | Remark |
|------|-------------------------|--------------------------|--------|
|      | ASTM F2100 Level 3      | ASTM F2100 Level 3       |        |

{5}------------------------------------------------

| Indication for use |                    | The Disposable medical surgical mask is<br>intended to be worn to protect both the patient<br>and healthcare personnel from the transfer of<br>microorganisms, body fluids, and particulate<br>material. The surgical mask is intended for use<br>in infection control practices to reduce the<br>potential exposure to blood and body fluids.<br>This is a single-use, disposable device(s),<br>provided non-sterile. | The Disposable Medical Surgical Masks are<br>intended to be worn to protect both the<br>patient and healthcare personnel from<br>transfer of microorganisms, body fluids and<br>particulate material. These face masks are<br>intended for use in infection control<br>practices to reduce the potential exposure to<br>blood and body fluids. This is a single use,<br>disposable device(s), provided non-sterile. | SAME     |
|--------------------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Basic Design       |                    | Ear Loops, Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                      | Ear loops, Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                   | SAME     |
| Materials          | Outer Facing Layer | polypropylene spunbond fabric                                                                                                                                                                                                                                                                                                                                                                                          | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                             | Analysis |
|                    | Middle Layer       | polypropylene meltblown fabric                                                                                                                                                                                                                                                                                                                                                                                         | Melt-blown polypropylene filter                                                                                                                                                                                                                                                                                                                                                                                     |          |
|                    | Inner Facing Layer | polypropylene spunbond fabric                                                                                                                                                                                                                                                                                                                                                                                          | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                             |          |
|                    | Nose Piece         | polypropylene coated steel wire                                                                                                                                                                                                                                                                                                                                                                                        | Polyethylene coated steel wire                                                                                                                                                                                                                                                                                                                                                                                      |          |
|                    | Ear Loops          | nylon , spandex                                                                                                                                                                                                                                                                                                                                                                                                        | Polyester                                                                                                                                                                                                                                                                                                                                                                                                           |          |
| Color              |                    | Blue                                                                                                                                                                                                                                                                                                                                                                                                                   | Blue                                                                                                                                                                                                                                                                                                                                                                                                                | SAME     |
| Dimension          |                    | $17.5 \pm 5\% \text{ cm}$                                                                                                                                                                                                                                                                                                                                                                                              | $17.5 \text{cm} \pm 0.2 \text{cm}$                                                                                                                                                                                                                                                                                                                                                                                  | Similar  |
|                    | (Length, Width)    | $9.5 \pm 5\% \text{cm}$                                                                                                                                                                                                                                                                                                                                                                                                | $9.5 \pm 0.2 \text{cm}$                                                                                                                                                                                                                                                                                                                                                                                             |          |
| OTC use            |                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                 | SAME     |
| Single Use         |                    | Single use, disposable                                                                                                                                                                                                                                                                                                                                                                                                 | Single use, disposable                                                                                                                                                                                                                                                                                                                                                                                              | SAME     |
| Sterile            |                    | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                            | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                         | SAME     |

# Table 2 Performance Characteristic Comparison

| ITEM                                                                    | Proposed Device<br>K212097       | Predicate Device<br>K202511   | ASTM F2100<br>Requirements | Remark |
|-------------------------------------------------------------------------|----------------------------------|-------------------------------|----------------------------|--------|
| ASTM F2100 Level                                                        | Level 3                          | Level 3                       | Level 3                    | SAME   |
| Fluid Resistance<br>Performance ASTM<br>F1862                           | 32 out of 32 pass at<br>160 mmHg | 32 out of 32 pass at 160 mmHg | 160 mmHg                   | SAME   |
| Particulate Filtration<br>Efficiency ASTM<br>F2299                      | ≥99%                             | Pass at 98.6%                 | ≥ 98%                      |        |
| Bacterial Filtration<br>Efficiency ASTM<br>F2101                        | ≥99%                             | Pass at 99.9%                 | ≥ 98%                      |        |
| Differential<br>Pressure (Delta P)<br>EN 14683:2019+<br>AC:2019 Annex C | < 4.9 mmH2O/cm2                  | Pass at 3.5 mmH2O/cm2         | < 6.0 mmH2O/cm2            |        |
| Flammability<br>16 CFR 1610                                             | Class 1                          | Class 1                       | Class 1                    | SAME   |

{6}------------------------------------------------

| ITEM          | Proposed Device K212097 | Predicate Device K202511 | Remark |
|---------------|-------------------------|--------------------------|--------|
| Cytotoxicity  | Non-cytotoxic           | Non-cytotoxic            | SAME   |
| Irritation    | Non-irritating          | Non-irritating           | SAME   |
| Sensitization | Non-sensitizing         | Non-sensitizing          | SAME   |

Table 3 Biocompatibility Comparison

Analysis:

The Disposable medical surgical mask is substantially equivalent to the Improve Medical (HuNan) Products Disposable Medical Surgical Mask. Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Disposable Medical Surgical Mask cleared under K202511.

- 7. Non-Clinical Test Conclusion
The test results demonstrated that the proposed device complies with the following standards:

- A ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- > ASTM F2100-19, Standard Specification For Performance Of Materials Used In Medical Face Masks.
- ASTM F1862-17, Standard Test Method For Resistance Of Medical Face Masks To > Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity)
- > EN 14683-2019+AC:2019 Annex C, Medical face masks - Requirements and test methods;
- > ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency (Bfe) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus;
- > ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
- > 16 CFR 1610, Standard for the Flammability of clothing textiles;

| Test Method | Purpose                                         | Acceptance Criteria | Results   |
|-------------|-------------------------------------------------|---------------------|-----------|
| ASTM F1862  | Resistance to penetration<br>by synthetic blood | 160 mm Hg           | 160 mm Hg |

{7}------------------------------------------------

| ASTM F2299          | Sub-micron particulate<br>filtration efficiency at 0.1<br>micron | $\geq 98%$                 | $\geq 99%$                                                      |
|---------------------|------------------------------------------------------------------|----------------------------|-----------------------------------------------------------------|
| ASTM F2101          | Bacterial Filtration<br>Efficiency                               | $\geq 98%$                 | $\geq 99%$                                                      |
| EN 14683<br>Annex C | Differential Pressure                                            | $< 6.0 mm H2O/cm²$         | $< 4.9 mm H2O/cm²$                                              |
| 16 CFR 1610         | Flammability                                                     | Class 1                    | Class 1                                                         |
| ISO 10993-10        | Irritation                                                       | No irritation effect       | Under the conditions of<br>the study, no<br>irritation effect   |
|                     | Sensitization                                                    | No sensitization<br>effect | Under conditions of<br>the study, no<br>sensitization effect    |
| ISO 10993-5         | Cytotoxicity                                                     | No cytotoxicity effect     | Under the conditions<br>of the study, no<br>cytotoxicity effect |

### 8. Clinical Test Conclusion

No clinical study is included in this submission.

### 9. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better the legally marketed device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K212097](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K212097)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
