← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K211696

# Disposable Surgical Face Mask (K211696)

_Anhui Cleanpro Pharmpack Co., Ltd. · FXX · Nov 30, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K211696

## Device Facts

- **Applicant:** Anhui Cleanpro Pharmpack Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Nov 30, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

## Device Story

Disposable surgical face mask; flat-pleated, three-layer design. Outer/inner layers: spun-bond polypropylene; middle layer: melt-blown polypropylene filter. Features malleable aluminum nose piece for fit; available in tie-on or elastic ear-loop configurations. Used in clinical settings for infection control; protects wearer and patient from fluid/particulate transfer. Single-use, non-sterile device.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F2299 (particulate filtration ≥99%), ASTM F1862 (fluid resistance at 160 mmHg), ASTM F2101 (bacterial filtration ≥99%), differential pressure (breathability <6.0 mmH2O/cm²), and 16 CFR 1610 (flammability Class 1). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).

## Technological Characteristics

Three-layer construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter). Malleable aluminum nose piece. Tie-on or polyester elastic ear loops. Non-sterile. ASTM F2100 Level 3 performance. No electronic components or software.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Mask ([K133070](/device/K133070.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

November 30, 2021

Anhui Cleanpro Pharmpack Co., Ltd. Chong Xiuming , Regulatory Manager Yeshan Industrial Zone Tianchang, Anhui 239300 CHINA

Re: K211696

Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 20, 2021 Received: October 25, 2021

Dear Chong Xiuming:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K211696

Device Name Disposable Surgical Face Mask

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non- sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(K) Summary

#### This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR §807.92

#### Date prepared: October 20.2021

#### 1. Submitter Name and Address:

| Owner Name:                         | Anhui Cleanpro Pharmpack Co., Ltd.                    |
|-------------------------------------|-------------------------------------------------------|
| Establishment Registration Number : | 3016771217                                            |
| Address:                            | Yeshan Industrial Zone, Tianchang City, Anhui, China. |
| Post Code:                          | 239300                                                |
| Contactor Name:                     | Chong Xiuming                                         |
| TEL:                                | +86-550-7981778                                       |
| E-mail:                             | carlchong@126.com                                     |

#### Manufacturer Name: Anhui Cleanpro Pharmpack Co., Ltd.

| Address:   | Yeshan Industrial Zone, Tianchang City, Anhui, China. |
|------------|-------------------------------------------------------|
| Post Code: | 239300                                                |

#### US Agent:

| US Agent: | LI QIAN                                                   |
|-----------|-----------------------------------------------------------|
| Address:  | 4319 Abbotts Bridge Rd STE 3 Duluth Duluth, GA  US  30097 |
| TEL:      | 404 4261248 Ext                                           |
| Email :   | Li@Carelifeusa.Com                                        |

#### 2. Submission Devices Information:

Trade/Proprietary Name: Disposable Surgical Face Mask Common Name: Disposable Surgical Face Mask Classification name: Surgical Apparel. Class: II. Product codes: FXX Submission Type: 510(K) Regulation Number: 21 CFR 878.4040

#### 3. Predicate Devices Information:

Company Name: BH Medical Products Co., Ltd. Address: No.90 Zhangjia Industrial Area, Xiling, Changzhou, Jiangsu, China 213024 Trade Name: Surgical Face Mask Class: II.

{4}------------------------------------------------

Yeshan Industrial Zone, Tianchang City, Anhui Provice, China.

[Disposable Surgical Face Mask]

510(K) Number: K133070 Product codes: FXX Submission Type: 510(K) Regulation Number: 21 CFR 878.4040

### 4.Devices Description:

The proposed device(s) are Blue color, and Flat Pleated type mask, utilizing Tie-On or Ear Loops way for wearing, and they all has Nose Piece design for fitting the face mask around the nose.

The outer layer is blue, and the colorant material is identified as Phthalocyanine Blue , CAS number: 147-14-8.

The proposed device(s) are manufactured with three layers. the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.

The model of proposed device, tie-on, is held in place over the uers's mouth and nose by four ties welded to the face mask. The tie is made of spun-bond polypropylene.

The model of proposed device, ear loops, is held in place over the uers's mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex.

The nose piece contained in the proposed device(s) is in the layers of facemask to allow the user to fit the facemask around their nose, which is made of malleable aluminum wire.

The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

### 5. Indications for use:

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided nonsterile.

### 6. Comparison of technological characteristics with the predicate:

The primary components of The Disposable Surgical Face Masks are manufactured to identical or similar specifications of the predicated devices listed above. The intended use, basic design, function and materials used are identical or similar to the predicate devices.

{5}------------------------------------------------

### Anhui Cleanpro Pharmpack Co., Ltd. Yeshan Industrial Zone, Tianchang City, Anhui Provice, China. [Disposable Surgical Face Mask]

### Table 1 General Comparison

| Element of Comparison                                 | Submission Device                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>K133070                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                  |            |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|------------|
| Manufacturer                                          | Anhui Cleanpro<br>Pharmpack Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                               | BH Medical Products<br>Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                     | /                                                           |            |
| 510 (k)                                               | K211696                                                                                                                                                                                                                                                                                                                                                                                                                             | K133070                                                                                                                                                                                                                                                                                                                                                                                                              | /                                                           |            |
| Proprietary or Model Name                             | Surgical Face Mask, Ear<br>Loops, Model:CRWKE001<br>Surgical Face Mask, Tie-on,<br>Model:CRWKT001                                                                                                                                                                                                                                                                                                                                   | Surgical Face Mask, Ear<br>Loops, Model 101B, 101G,<br>136B, 136G, 137B, 137G<br>Surgical Face Mask, Tie-on,<br>Model 145B, 145G, 143B,<br>143G,<br>138B, 138G, 142B, 142G,<br>151B, 151G.                                                                                                                                                                                                                           | Similar                                                     |            |
| Classification                                        | Class II Device, FXX<br>(21CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                                             | Class II Device, FXX<br>(21CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                              | Same                                                        |            |
| Intended use/Indications for<br>Use                   | The Disposable Surgical Face<br>Masks are intended to be<br>worn to protect both the<br>patient and healthcare<br>personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This<br>is a single use, disposable<br>device(s), provided non-<br>sterile. | The Surgical Face Masks<br>are intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device(s),<br>provided non-sterile. | Same                                                        |            |
| Ear loop model and<br>Tie-on model                    | Ear Loops, Tie-On                                                                                                                                                                                                                                                                                                                                                                                                                   | Ear Loops, Tie-On                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                        |            |
| Mask styles                                           | Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                                              | Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                        |            |
| Materials                                             | Outer Facing Layer                                                                                                                                                                                                                                                                                                                                                                                                                  | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                              | Spun-bond polypropylene                                     | Same       |
|                                                       | Middle Layer                                                                                                                                                                                                                                                                                                                                                                                                                        | Melt blown polypropylene<br>filter                                                                                                                                                                                                                                                                                                                                                                                   | Melt blown polypropylene<br>filter                          | Same       |
|                                                       | Inner Facing Layer                                                                                                                                                                                                                                                                                                                                                                                                                  | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                              | Spun-bond polypropylene                                     | Same       |
|                                                       | Tie-on                                                                                                                                                                                                                                                                                                                                                                                                                              | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                              | Spun-bond polypropylene                                     | Same       |
|                                                       | Nose Piece                                                                                                                                                                                                                                                                                                                                                                                                                          | Malleable aluminum wire                                                                                                                                                                                                                                                                                                                                                                                              | Malleable aluminum wire                                     | Same       |
|                                                       | Ear Loops                                                                                                                                                                                                                                                                                                                                                                                                                           | Polyester                                                                                                                                                                                                                                                                                                                                                                                                            | Polyester                                                   | Same       |
| Design features                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                     | Color : Blue                                                                                                                                                                                                                                                                                                                                                                                                         | Color : Blue, Green                                         | Similar    |
| Dimension (Length)                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                     | 17.5 cm +/- 0.5cm                                                                                                                                                                                                                                                                                                                                                                                                    | 6.8"+/-0.25"                                                | Similar    |
| Dimension (Width)                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                     | 9.5 cm +/- 0.5cm                                                                                                                                                                                                                                                                                                                                                                                                     | 3.5"+/-0.25"<br>4.2"+/-0.25"                                | Similar    |
| OTC use                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                         | Same       |
| Single Use                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                         | Same       |
| Sterile                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                     | No                                                                                                                                                                                                                                                                                                                                                                                                                   | No                                                          | Same       |
| ASTM F2100 Level                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                     | Level 3                                                                                                                                                                                                                                                                                                                                                                                                              | Level 1、Level 2、 Level 3                                    | Similar    |
| Element of<br>Comparison                              | Submission Device                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>K133070                                                                                                                                                                                                                                                                                                                                                                                          | ASTM F2100<br>Requirements<br>for Level 3<br>Classification | Comparison |
| Fluid Resistance<br>Performance<br>ASTM F1862         | 1、LOT NO.: CR202101<br>1~32,32 out of 32 pass at<br>160 mmHg<br>2、LOT NO.: CR202102<br>33~64,32 out of 32 pass at<br>160 mmHg<br>3、LOT NO.: CR202103<br>65~96,32 out of 32 pass at<br>160 mmHg                                                                                                                                                                                                                                      | Pass                                                                                                                                                                                                                                                                                                                                                                                                                 | 29 out of 32 pass<br>at<br>160 mmHg                         | Same       |
| Particulate<br>Filtration<br>Efficiency ASTM<br>F2299 | 1、LOT NO.: CR202101<br>1~32: ≥ 99%<br>2、LOT NO.: CR202102<br>33~64: ≥ 99%<br>3、LOT NO.: CR202103<br>65~96:  ≥ 99%                                                                                                                                                                                                                                                                                                                   | ≥ 98%                                                                                                                                                                                                                                                                                                                                                                                                                | ≥ 98%                                                       | Same       |
| Bacterial Filtration<br>Efficiency ASTM<br>F2101      | 1、LOT NO.: CR202101<br>1~32: ≥ 99%<br>2、LOT NO.: CR202102<br>33~64: ≥ 99%<br>3、LOT NO.: CR202103<br>65~96:  ≥ 99%                                                                                                                                                                                                                                                                                                                   | ≥98%                                                                                                                                                                                                                                                                                                                                                                                                                 | ≥98%                                                        | Same       |
| Differential<br>Pressure (DeltaP)<br>MIL-M-36954C     | 1、LOT NO.: CR202101<br>1~32: <6.0 mmH2O/cm²<br>2、LOT NO.: CR202102<br>33~64:  <6.0 mmH2O/cm²<br>3、LOT NO.: CR202103<br>65~96: <6.0 mmH2O/cm²                                                                                                                                                                                                                                                                                        | <6.0 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                                                                                       | <6.0 mmH2O/cm²                                              | Same       |
| Flammability<br>16 CFR 1610                           | 1、LOT NO.: CR202101<br>1~32: IBE<br>2、LOT NO.: CR202102<br>33~64: IBE<br>3、LOT NO.: CR202103<br>65~96: IBE                                                                                                                                                                                                                                                                                                                          | Class 1<br>Non Flammable                                                                                                                                                                                                                                                                                                                                                                                             | Class 1<br>(Burn time<br>≥3.5 seconds.<br>IBE, or DNI)      | Same       |
| Element of Comparison                                 | Submission Device                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>K133070                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                                  |            |
| Cytotoxicity                                          | Under the conditions of the<br>study, not cytotoxicity<br>effect (Comply with ISO<br>10993-5)                                                                                                                                                                                                                                                                                                                                       | Under the conditions of<br>the study, not<br>cytotoxicity effect<br>(Comply with ISO<br>10993-5 )                                                                                                                                                                                                                                                                                                                    | Same                                                        |            |
| Irritation                                            | Under the conditions of the<br>study, not an Irritant<br>(Comply with ISO 10993-<br>10)                                                                                                                                                                                                                                                                                                                                             | Under the conditions of<br>the study, not an Irritant<br>(Comply with ISO<br>10993-10 )                                                                                                                                                                                                                                                                                                                              | Same                                                        |            |
| Sensitization                                         | Under conditions of the<br>study, not a sensitizer.<br>(Comply with ISO 10993-<br>10)                                                                                                                                                                                                                                                                                                                                               | Under conditions of the<br>study, not a sensitizer.<br>(Comply with ISO<br>10993-10 )                                                                                                                                                                                                                                                                                                                                | Same                                                        |            |

{6}------------------------------------------------

# Yeshan Industrial Zone, Tianchang City, Anhui Provice, China.

[Disposable Surgical Face Mask]

{7}------------------------------------------------

Yeshan Industrial Zone, Tianchang City, Anhui Provice, China.

[Disposable Surgical Face Mask]

#### Table 2 Performance Characteristic Comparison

{8}------------------------------------------------

# Anhui Cleanpro Pharmpack Co., Ltd. Yeshan Industrial Zone, Tianchang City, Anhui Provice, China.

[Disposable Surgical Face Mask]

# Table 3 Biocompatibility Comparison

# 7.Summary non-Clinical Test:

### Discussion of Non-Clinical Tests Performed:

The performance tests of surgical face masks were conducted.

| Test Methodology                                                                                                                                                                        | Test Methodology<br>Purpose                                                                                                     | Acceptance Criteria                      | Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|---------|
| ASTM F2299-03 Standard<br>Test Method for<br>Determining the Initial<br>Efficiency of Materials Used<br>in Medical Face Masks to<br>Penetration by Particulates<br>Using Latex Spheres  | To evaluate the<br>effectiveness of the<br>test article in protecting<br>the user from possible<br>exposure to body<br>fluids.  | ≥98%                                     | Passed  |
| ASTM F1862-17 Standard<br>Test Method For<br>Resistance Of Medical Face<br>Masks To Penetration By<br>Synthetic Blood (Horizontal<br>Projection Of Fixed Volume<br>At A Known Velocity) | To evaluate the<br>effectiveness of the<br>test article in protecting<br>the user from possible<br>exposure to<br>particulates. | Level 3, No<br>penetration at<br>160mmHg | Passed  |
| ASTM F2101-19 Standard<br>Test Method For Evaluating<br>The Bacterial Filtration<br>Efficiency (Bfe) Of Medical                                                                         | To evaluate the<br>bacterial filtration<br>efficiency (BFE) of<br>mask.                                                         | ≥98%                                     | Passed  |

{9}------------------------------------------------

### Yeshan Industrial Zone, Tianchang City, Anhui Provice, China.

[Disposable Surgical Face Mask]

| Face Mask Materials, Using<br>A Biological Aerosol Of<br>Staphylococcus Aureus.<br>(General Plastic<br>Surgery/General Hospital)                     |                                                                                                         |                                                                                                                                                                                               |        |
|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| MIL-M-36954C Military<br>Specification - Mask,<br>Surgical, Disposable 16<br>CFR Part 1610 STANDARD<br>FOR THE FLAMMABILITY<br>OF CLOTHING TEXTILES  | To evaluate the<br>flammability of mask.                                                                | Class 1, Resist<br>ignition or flame<br>( Burn time ≥3.5<br>seconds, IBE, or<br>DNI )                                                                                                         | Passed |
| Differential Pressure (Delta<br>P) EN14683:2019, Annex C<br>and ASTM F2100-19                                                                        | To measure the<br>differential pressure<br>of mask which is<br>related to<br>breathability.             | <6.0 mmH20/cm²                                                                                                                                                                                | Passed |
| ISO 10993-5: 2009,<br>Biological Evaluation Of<br>Medical Devices -- Part 5:<br>Tests For In Vitro<br>Cytotoxicity.<br>(Biocompatibility)            | To evaluate the<br>biological safety of<br>the product which has<br>direct contact with<br>intact skin. | The test article should<br>not have potential<br>toxicity to L-929 in the<br>MTT method.                                                                                                      | Passed |
| ISO 10993-10: 2010,<br>Biological Evaluation Of<br>Medical Devices-Part 10:<br>Tests For Irritation And Skin<br>Sensitization.<br>(Biocompatibility) | To evaluate the<br>biological safety of<br>the product which has<br>direct contact with<br>intact skin. | The test article should<br>not cause delayed<br>dermal contact<br>sensitization in the<br>guinea pig.<br>The dermal irritation<br>response category in<br>the rabbit should be<br>negligible. | Passed |

### Discussion of Clinical Tests Performed: None

### 8. Conclusion:

Based on the comparison and analysis above, the Anhui Cleanpro Pharmpack Co., Ltd. Disposable Surgical Face Mask is as safe, as effective, and performs as well as the legally marketed predicate device, BH Medical Products Co., Ltd. Surgical Face Mask cleared under K133070.

### END

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K211696](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K211696)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
