← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K210042

# Disposable Surgical Mask (K210042)

_Suzhou Letian Protective Products Co., Ltd. · FXX · Nov 17, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K210042

## Device Facts

- **Applicant:** Suzhou Letian Protective Products Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Nov 17, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Disposable Surgical Masks are single use, disposable device, provided non-sterile.

## Device Story

Disposable surgical mask; three-layer flat-pleated design; outer/inner layers spun-bond polypropylene; middle layer melt-blown polypropylene; includes malleable high-density polyethylene nose piece and ear loops (80% spandex/20% polyester). Device worn over mouth and face to provide physical barrier against microorganisms, body fluids, and particulates. Used in clinical settings for infection control. Protects both patient and healthcare personnel. Non-sterile, single-use device.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19, ASTM F1862 (synthetic blood penetration), ASTM F2299 (PFE ≥99.5%), ASTM F2101 (BFE ≥99.9%), and EN 14683 (Delta P <4.3 mmH2O/cm2). Biocompatibility testing (ISO 10993-5, 10993-10) confirmed non-cytotoxic, non-irritating, and non-sensitizing properties.

## Technological Characteristics

Three-layer flat-pleated mask. Materials: Spun-bond polypropylene (outer/inner), melt-blown polypropylene (middle). Nose piece: High-density polyethylene. Ear loops: 80% Spandex/20% Polyester. Standards: ASTM F2100-19, ASTM F1862, ASTM F2299, ASTM F2101, MIL-M-36954C, 16 CFR 1610, ISO 10993. Non-sterile, single-use.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Masks ([K182514](/device/K182514.md))

## Reference Devices

- Face Mask ([K210007](/device/K210007.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 17, 2021

Suzhou Letian Protective Products Co., Ltd. % Ryan Li RA Manager Shanghai Mind-link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District Shanghai. 200040 China

Re: K210042

Trade/Device Name: Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 22, 2021 Received: September 29, 2021

Dear Ryan Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K210042

Device Name

Disposable Surgical Mask, Blue(Model; LT-0175A, LT-0145A) Disposable Surgical Mask, White (Model: LT-0175B, LT-0145B)

Indications for Use (Describe)

The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Disposable Surgical Masks are single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

### 510(K) Summary-K210042

#### I. SUBMITTER:

Suzhou Letian Protective Products Co., Ltd. No. 12 Zhenfeng Road, Nanfeng Town, Suzhou City, Jiangsu Province, 215600, China

Contact Person: Huang Qizhi Title: Integrated Manager Tel: +86 13701567601

Email: huangqizhi@szltfhypwwgc.onexmail.com

Submission Correspondent: Ryan Li Email: ryan.li(@mind-link.net Tel:+86 13701581791

Summary prepared: September 17th, 2021

#### II. DEVICE

Name of Device: Disposable Surgical Mask Regulation Number: 21 CFR PART 878.4040 Common Name: Surgical Mask Classification Name: Surgical Mask Regulatory Class: II

Product Code: FXX

#### III. PREDICATE DEVICE

{4}------------------------------------------------

Primary predicate device: Surgical Face Masks (K182514)

#### IV. REFERENCE DEVICE

Reference device: Face Mask (K210007)

#### V. DEVICE DESCRIPTION

Disposable Surgical Mask is composed of three layers and is flat-pleated. The mask materials consist of an outer layer (spun-bond polypropylene), a middle layer, between the outer layer and inner layers (melt-blown polypropylene), and an inner layer (spun-bond polypropylene). Each mask contains ear loops to secure the mask over the users' mouth and face and includes a malleable nose piece (High-density Polyethylene) to provide a firm fit over the nose.

There are four models for Disposable Surgical Mask with different colors and sizes. All six models are Ear Loop type in blue or white colors and two sizes including 145mm×90mm and 175×95mm.

#### VI. AVAILABLE MODELS

| REF No.  | Product  | Model Description |          | Mask Color |       |
|----------|----------|-------------------|----------|------------|-------|
|          | Size     | Mask              | Ear Loop | Blue       | White |
| LT-0145A | 145×90mm | X                 | X        | X          |       |
| LT-0175A | 175×95mm | X                 | X        | X          |       |
| LT-0145B | 145×90mm | X                 | X        |            | X     |
| LT-0175B | 175×95mm | X                 | X        |            | X     |

{5}------------------------------------------------

#### VII. INDICATIONS FOR USE

The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Disposable Surgical Face Masks are single use, disposable device, provided non-sterile.

# VIII. TECHNOLOGICAL CHARACTERISTICS COMPARISON

The Disposable Surgical Masks are compared with the predicate device (Surgical Face Masks (K182514)):

{6}------------------------------------------------

| Device                                                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                       | Primary Predicate Device                                                                                                                                                                                                                                                                                                                                                                                   |                   |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
|                                                          | Disposable Surgical Mask<br>(K210042)                                                                                                                                                                                                                                                                                                                                                                                                                | Surgical Face Mask<br>(K182514 )                                                                                                                                                                                                                                                                                                                                                                           | Comparison        |
| Intended Use                                             | The Disposable Surgical Face Masks<br>are intended to be worn to protect both<br>the patient and healthcare personnel<br>from transfer of microorganisms, body<br>fluids and particulate material.<br><br>These face masks are intended for use<br>in infection control practices to reduce<br>the potential exposure to blood and<br>body fluids. The Disposable Surgical<br>Face Masks are single use, disposable<br>device, provided non-sterile. | The surgical face masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms,body fluids, and<br>particulate material.These face<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single-use, disposable device,<br>provided non-sterile. | Same              |
| Classification<br>Product Code                           | FXX                                                                                                                                                                                                                                                                                                                                                                                                                                                  | FXX                                                                                                                                                                                                                                                                                                                                                                                                        | Same              |
| Ear Loop<br>Model                                        | Ear Loops                                                                                                                                                                                                                                                                                                                                                                                                                                            | Ear Loops                                                                                                                                                                                                                                                                                                                                                                                                  | Same              |
| Outer Facing Layer                                       | Spun-bond<br>polypropylene<br>non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                                       | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                    | Similar Note 1    |
| Middle Layer                                             | Melt-blown polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                             | Melt blown polypropylene<br>filter                                                                                                                                                                                                                                                                                                                                                                         | Similar<br>Note 1 |
| Inner Facing Layer                                       | Spun-bond<br>polypropylene<br>non-woven fabric                                                                                                                                                                                                                                                                                                                                                                                                       | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                    | Similar Note 1    |
| Nose Piece                                               | High-density polyethylene                                                                                                                                                                                                                                                                                                                                                                                                                            | Malleable aluminum wire                                                                                                                                                                                                                                                                                                                                                                                    | Similar<br>Note 1 |
| Ear Loops                                                | 80% Spandex<br>20% Polyester                                                                                                                                                                                                                                                                                                                                                                                                                         | Polyester                                                                                                                                                                                                                                                                                                                                                                                                  | Similar Note 1    |
| Color                                                    | Blue, white                                                                                                                                                                                                                                                                                                                                                                                                                                          | White                                                                                                                                                                                                                                                                                                                                                                                                      | Different Note 1  |
| Style                                                    | Flat - Pleated                                                                                                                                                                                                                                                                                                                                                                                                                                       | Flat - Pleated                                                                                                                                                                                                                                                                                                                                                                                             | Same              |
| Multiple<br>Layers                                       | 3 Layers                                                                                                                                                                                                                                                                                                                                                                                                                                             | 3 Layers                                                                                                                                                                                                                                                                                                                                                                                                   | Same              |
| Single Use                                               | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                           | Single use                                                                                                                                                                                                                                                                                                                                                                                                 | Same              |
| Sterile                                                  | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                          | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                | Same              |
| Length $\times$ Width                                    | $145 \times 95$ mm ( $\pm$ 10mm)<br>$175 \times 95$ mm ( $\pm$ 10mm)                                                                                                                                                                                                                                                                                                                                                                                 | Length: 95 $\pm$ 10mm<br>Width:175 $\pm$ 10mm                                                                                                                                                                                                                                                                                                                                                              | Different Note 2  |
| Fluid Resistance<br>ASTM F1862                           | 32 out of 32 pass at<br>120mmHg                                                                                                                                                                                                                                                                                                                                                                                                                      | 32 out of 32 pass at<br>120mmHg                                                                                                                                                                                                                                                                                                                                                                            | Same              |
| Particulate Filtration<br>Efficiency (PFE) ASTM<br>F2299 | Pass at $\ge$ 99.5%                                                                                                                                                                                                                                                                                                                                                                                                                                  | Pass at 99.88%                                                                                                                                                                                                                                                                                                                                                                                             | Similar           |
| Bacterial Filtration<br>Efficiency (BFE) ASTM<br>F2101   | Pass at $\ge$ 99.9%                                                                                                                                                                                                                                                                                                                                                                                                                                  | Pass at 99.6%                                                                                                                                                                                                                                                                                                                                                                                              | Similar           |
| Differential Pressure<br>(Delta P)<br>MIL-M-36954<br>C   | Pass at $<4.3$ mmH2O/cm2                                                                                                                                                                                                                                                                                                                                                                                                                             | Pass at 3.0 mmH2O/cm2                                                                                                                                                                                                                                                                                                                                                                                      | Similar           |
| Flammability 16 CFR<br>PART 1610                         | Class 1 Non-Flammable                                                                                                                                                                                                                                                                                                                                                                                                                                | Class 1 Non-Flammable                                                                                                                                                                                                                                                                                                                                                                                      | Same              |
| Cytotoxicity                                             | Under the conditions of the study, the<br>subject device extract was determined to<br>be non-cytotoxic.                                                                                                                                                                                                                                                                                                                                              | Under the conditions of the study, the<br>subject device extract was<br>determined to be non-cytotoxic.                                                                                                                                                                                                                                                                                                    | Same              |
| Irritation                                               | Under the conditions of the study, the<br>subject device non-polar and polar<br>extracts were determined to be non-<br>irritating.                                                                                                                                                                                                                                                                                                                   | Under the conditions of the study, the<br>subject device non-polar and polar<br>extracts were determined to be non-<br>irritating.                                                                                                                                                                                                                                                                         | Same              |
| Sensitization                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                            |                   |
|                                                          | Under the conditions of the study, the<br>subject device non-polar and polar<br>extracts were determined to be non-<br>sensitizing.                                                                                                                                                                                                                                                                                                                  | Under the conditions of the study, the<br>subject device non-polar and polar<br>extracts were determined to be non-<br>sensitizing.                                                                                                                                                                                                                                                                        | Same              |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

## Comparison in Detail(s): Note 1:

Although the material including outer facing layer, middle layer, inner facing layer, nose piece and ear loops, as well as the color of the subject device and the color is different from the predicate device.

## Note 2:

Although LT-O145A and LT-O145B models of the subject device are smaller than the predicate device, these models match the dimension with reference device (K210007, Model: Type A), the barrier protection performance of the subject device is same with the barrier protection performance of predicate device and the reference device, which is the Level 2 barrier protection.

#### IX. PERFORMANCE DATA

### Non-Clinical Performance Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications found in the test methodology and standard using 3 nonconsecutive lots. The test results demonstrated that the Disposable Surgical Masks complies with the following standards:

- ASTM F2100-19 Standard Specification for Performance of Materials Used ● in Medical Face Masks
- ASTM F1862 Standard Test Method for Resistance of Medical Face Masks to . Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- EN 14683:2019+AC2019(E) Annex C ●

ASTM F2299 Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres

- . ASTM F2101 Standard Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Page 7 of 9

{10}------------------------------------------------

Aerosol of Staphylococcus aureus

- MIL-M- 36954C Military Specification, Mask, Surgical, Disposable .
- 16 CFR Part 1610 Standard for the Flammability of Clothing ●
- ISO10993-1 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process
- . ISO10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity of medical devices
- ISO10993-10 Biological Evaluation of Medical Devices Part 10: Tests for . Irritation and Skin Sensitization

| Test Methodology                     | Purpose                                                                                                                                                                                                                                                                                                                     | Acceptance Criteria  | Results                |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|------------------------|
| Bacterial Filtration<br>Effciency    | The test was performed in<br>accordance with ASTM F2101:<br>2019 Standard Test Method for<br>Evaluating the Bacterial<br>Filtration Efficiency (BFE) of<br>Medical Face<br><br>Mask Materials, Using a<br>Biological Aerosol of<br>Staphylococcus aureus to<br><br>determine the bacterial<br>filtration efficiency(BFE) of | Level 2≥98%          | 32/32 Passed at ≥99.9% |
| Particulate Filtration<br>Efficiency | The test was performed in<br>accordance with ASTM F2299<br>Standard Test Method for<br>Determining the Initial<br>Efficiency of Material Used<br>in medical Face Masks to<br>Penetration by Particulates<br>using Latex Spheres to<br>determine the particle filtration<br>efficiency (PFE) of                              | Level 2≥98%          | 32/32 Pass at ≥99.5%   |
| Differential Pressure                | The test was performed in                                                                                                                                                                                                                                                                                                   | Level 2<6.0mmH2O/cm² | 32/32 Pass at <4.3     |

{11}------------------------------------------------

| (delta-P)                                                                                              | accordance with EN<br>14683:2019+AC;2019(E)<br>Annex C                                                                                                                                                                                                                                                                                                      |                                       | mmH2O/cm²                           |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|-------------------------------------|
| Resistance to<br>penetration by<br>synthetic blood,<br>minimum pressure in<br>mm Hg for pass<br>result | The test was performed in<br>accordance with ASTM<br>F1862/F1862M Standard Test<br>Method for Resistance of<br>Medical Face Masks to<br>Penetration by Synthetic Blood<br>(HorizontalProjection of Fixed<br>Volume at a Known Velocity)<br>to evaluate the effectiveness of<br>the test sample from possible<br>exposure to blood and other<br>body fluids. | Level2: No penetration at<br>120 mmHg | 32/32 Passed at<br>120mmHg          |
| Flammability                                                                                           | The test was performed in<br>accordance with 16 CFR Part<br>1610 Standard for the<br>Flammability of Clothing to<br>measure the flammability of<br>masks                                                                                                                                                                                                    | Level2: Class 1                       | 32/32 Passed Class 1<br>requirement |

Clinical Test Conclusion

No clinical study is included in this submission.

#### X. CONCLUSION

The conclusion drawn from the nonclinical tests demonstrates that the subject device, the Disposable Surgical Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device Surgical Face Masks (K182514).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K210042](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K210042)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).