← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K203776

# Surgical Masks, Model SNN200640 & MN112 (K203776)

_Sq Medical Supplies, Inc. · FXX · Mar 25, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203776

## Device Facts

- **Applicant:** Sq Medical Supplies, Inc.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 25, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

## Device Story

Surgical masks (models SNN200640, MN112) are flat-pleated, single-use, disposable face masks. Device features ear loops and a nose piece for facial fit. SNN200640 consists of three layers; MN112 consists of four layers. Masks act as a physical barrier to reduce transfer of microorganisms, body fluids, and particulates between patients and healthcare personnel. Used in clinical or general settings for infection control. Healthcare providers wear the mask to protect themselves and patients from exposure to blood and body fluids. Benefit is reduced risk of cross-contamination and exposure to infectious agents.

## Clinical Evidence

No clinical data. Bench testing only. Compliance with ASTM F2100-19 verified via fluid resistance (ASTM F1862M-17), particulate filtration efficiency (ASTM F2299), bacterial filtration efficiency (ASTM F2101-19), differential pressure (EN 14683:2019), and flammability (16 CFR 1610).

## Technological Characteristics

Flat-pleated design; 3 or 4 layers of spun-bond and melt-blown polypropylene. Ear loops: Spandex/Nylon. Nose piece: Steel wire coated with Polyethylene. Non-sterile. Dimensions: 17.5cm x 9.5cm. Biocompatibility per ISO 10993. Performance meets ASTM F2100-19 Level 1 and Level 3 requirements.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face masks (Ear loops and Tie-on) ([K160269](/device/K160269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. FOOD & DRUG" on top of the word "ADMINISTRATION".

March 25, 2021

SQ Medical Supplies Inc. % Shelly Li Official Correspondent Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 703, 705, Building 1, West Guangzhong Road 555 Shanghai. 200072 China

Re: K203776

Trade/Device Name: Surgical Masks, Model SNN200640 & MN112 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: December 23, 2020 Received: December 28, 2020

Dear Shelly Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K203776

Device Name Surgical Masks

#### Indications for Use (Describe)

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary: K203776

### l. Submitter

SQ Medical Supplies Inc. 8889 West Olympic Blvd. Suite 1000 Beverly Hills, CA 90211. Establishment Registration Number: 3016757871

Contact person: Jonathan Lim Position: Chief Executive Officer Tel.: +1-888-912-8168 E-mail: jonlim@sqmedicalsupplies.com

Preparation date: Feb. 26, 2021

# II. Proposed Device

| Trade Name of Device: | Surgical Masks   |
|-----------------------|------------------|
| Common name:          | Surgical Mask    |
| Regulation Number:    | 21 CFR 878.4040  |
| Regulatory Class:     | Class II         |
| Product code:         | FXX              |
| Review Panel          | General Hospital |

### III. Predicate Devices

| 510(k) Number:  | K160269                                    |
|-----------------|--------------------------------------------|
| Trade name:     | Surgical Face masks (Ear loops and Tie-on) |
| Common name:    | Surgical Mask                              |
| Classification: | Class II                                   |
| Product Code:   | FXX                                        |
| Manufacturer    | San-M Package Co., Ltd.                    |

### IV. Device Description

The Surgical Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Surgical Masks are manufactured with three/four layers. The surgical mask has two models which are SNN200640 and MN12. They are basically the same, the only difference is the SNN200640 has the three layers and MN112 has the fours layers.

{4}------------------------------------------------

The Surgical Masks are single use, disposable device, provided non-sterile.

### V. Indication for Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

| Item               | Proposed Device<br>(K203776)                                                                                                                                                                                                                                                                                                                                             | Predicate Device<br>(K160269)                                                                                                                                                                                                                                                                                                                                        | Decision |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Trade Name         | Surgical Masks                                                                                                                                                                                                                                                                                                                                                           | Surgical Face masks (Ear - loops and Tie-on)                                                                                                                                                                                                                                                                                                                         |          |
| Product Code       | FXX                                                                                                                                                                                                                                                                                                                                                                      | FXX                                                                                                                                                                                                                                                                                                                                                                  | Same     |
| Regulation No.     | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                          | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                      | Same     |
| Class              | Class II                                                                                                                                                                                                                                                                                                                                                                 | Class II                                                                                                                                                                                                                                                                                                                                                             | Same     |
| Mask Style         | Flat-pleated, ear loops, 3/4 layers                                                                                                                                                                                                                                                                                                                                      | Flat-pleated, ear loops or tie-on, 4 layers                                                                                                                                                                                                                                                                                                                          | Similar  |
| Indication for Use | The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | The Surgical facemasks are intended to be worn to protect the patient and healthcare personnel from transfer of microorganisms, blood fluid, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluid. This is a single-use, disposable devices provided non-sterile. | Same     |
| Material           | Inner layer: White double spun-bond polypropylene                                                                                                                                                                                                                                                                                                                        | Polypropylene                                                                                                                                                                                                                                                                                                                                                        |          |
|                    | Middle layer: Layer #1: Polypropylene<br>Layer #2: Melt blown                                                                                                                                                                                                                                                                                                            | Layer #1. Polypropylene spun-bond                                                                                                                                                                                                                                                                                                                                    | Same     |

# VI. Comparison of Technological Characteristics with the Predicate Devices

{5}------------------------------------------------

|                     |                | polypropylene filter<br>(Middle Layer #1 is not applicable<br>SNN200640) | Layer #2: Polypropylene melt<br>blown                                                                                                                                     |           |
|---------------------|----------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
|                     | Outer<br>layer | Blue double spun-bond<br>polypropylene                                   | Polypropylene                                                                                                                                                             | Same      |
|                     | Ear loops      | Spandex+Nylon                                                            | Ear loops: Polyester,<br>polyurethane;<br>Side tapes: Polyester<br>spun-bond (ear loops mask<br>only)<br>Ties tapes: Polypropylene<br>spun-bond or polyester<br>spun-bond | Different |
|                     | Nose<br>piece  | Steel wire coated by<br>Polyethylene                                     | Steel wire coated by<br>Polyethylene                                                                                                                                      | Same      |
| Color               |                | Blue                                                                     | Blue ,white                                                                                                                                                               | Similar   |
| Dimension           |                | 17.5cmx9.5cm                                                             | 17.5cmx9.0cm<br>18.0cmx9.0cm                                                                                                                                              | Same      |
|                     | OTC Use        | Yes                                                                      | Yes                                                                                                                                                                       | Same      |
| Sterility           |                | Non-sterile                                                              | Non-sterile                                                                                                                                                               | Same      |
| For single Use      |                | Yes                                                                      | Yes                                                                                                                                                                       | Same      |
| ASTM F2100<br>Level |                | Level 1: Pass at 80mmHg<br>Level 3: Pass at 160mmHg                      | Level 1: Pass at 80mmHg<br>Level 2: Pass at 120mmHg<br>Level 3: Pass at 160mmHg                                                                                           | Similar   |
| Biocompatibility    |                | Confirm to the requirements of<br>ISO 10993 series standards             | Confirm to the requirements of<br>ISO 10993 series standards                                                                                                              | Same      |

# VII. Non-Clinical Testing

Non clinical performance tests were conducted to verify that the proposed device met all design specifications. The below table shows the test results of test article, which demonstrated that the proposed device complies with the standards of ASTM F2100-19:

| Standard          | Purpose                            | Acceptance Criteria            |                                 | Results |
|-------------------|------------------------------------|--------------------------------|---------------------------------|---------|
|                   |                                    | Level 1                        | Level 3                         |         |
| ASTM<br>F1862M-17 | Fluid<br>Resistance<br>Performance | 29 out of 32 pass<br>at 80mmHg | 29 out of 32 pass<br>at 160mmHg | Pass    |
| ASTM F2299        | Particulate                        | ≥95%                           | ≥98%                            | Pass    |

{6}------------------------------------------------

|                             | Filtration<br>Efficiency              |                       |               |      |
|-----------------------------|---------------------------------------|-----------------------|---------------|------|
| ASTM<br>F2101-19            | Bacterial<br>Filtration<br>Efficiency | ≥95%                  | ≥98%          | Pass |
| EN<br>14683:2019<br>Annex C | Differential<br>Pressure              | <5.0mmH2O/cm²         | <6.0mmH2O/cm² | Pass |
| 16 CFR 1610                 | Flammability                          | Class I non flammable |               | Pass |

# VIII. Clinical Testing

No clinical study is included in this submission.

# IX. Conclusion

The conclusion drawn from the non-clinical performance testing data demonstrates that the subject device is as safe, as effective, and performs as well as or better than the predicate device, Surgical Face masks (K160269).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203776](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203776)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
