← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K203455

# Disposable Medical Mask (K203455)

_Fujian Kang Chen Daily Necessities Co., Ltd. · FXX · Sep 13, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203455

## Device Facts

- **Applicant:** Fujian Kang Chen Daily Necessities Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Sep 13, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Disposable Medical mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. The Disposable Medical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

## Device Story

Disposable Medical Mask (Model k0450) is a 3-layer, flat-pleated, single-use, non-sterile surgical mask. It features spunbond polypropylene outer/inner layers, a melt-blown polypropylene filter middle layer, spandex ear loops, and a polyethylene-coated aluminum nose clip. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against microorganisms, body fluids, and particulates. The device is worn over the nose and mouth, secured by ear loops and adjusted via the nose clip. It functions as a Level 2 barrier per ASTM F2100, providing fluid resistance, bacterial filtration, and particulate filtration to reduce infection risk.

## Clinical Evidence

Bench testing only. Device demonstrated compliance with ASTM F2100-19, ASTM F1862 (fluid resistance: 32/32 pass at 120 mmHg), ASTM F2299 (particulate filtration efficiency: 99.9%), ASTM F2101 (bacterial filtration efficiency: 99.9%), 16 CFR 1610 (flammability: Class 1), and differential pressure (<5.0 mmH2O/cm2). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

3-layer flat-pleated design. Materials: Spunbond polypropylene (outer/inner), melt-blown polypropylene (filter), spandex (ear loops), polyethylene-coated aluminum (nose clip). ASTM F2100 Level 2 barrier. Dimensions: 17.5cm x 9.5cm. Non-sterile, single-use. No energy source or software.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Single-Use Surgical Face Mask with Ear Loop ([K201691](/device/K201691.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". The logo is simple and professional, conveying the organization's role in regulating food and drugs.

September 13, 2021

Fujian Kang Chen Daily Necessities CO., LTD Shuyan Wang Administrative Director No.55 Houdun Road, houmao Industrial Zone, Fengze District Quanzhou, Fujian 362000 China

Re: K203455

Trade/Device Name: Disposable Medical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 5, 2021 Received: August 9, 2021

Dear Shuyan Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203455

Device Name Disposable Medical mask

Indications for Use (Describe)

The Disposable Medical mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. The Disposable Medical mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Model: k0450, blue color, and Level 2 barrier level per ASTM F2100.

| Type of Use (Select one or both, as applicable)                                      |                                                                                                |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510k Summary

Prepared Date:09/13/2021

510k Number: 203455

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

## 1. Submitter Information

Sponsor : FUJIAN KANG CHEN DAILY NECESSITIES CO.,LTD

Address: No.55 Houdun Road, houmao Industrial Zone, Fengze District, Quanzhou

City, Fujian Province, China.

Contact Person: Shuyan Wang (Administrative Director)

Phone: +86-595-28892058

Fax:+86- 595-28892058

E-mail: shuyan@consoncompanies.com

### 2. Subject Device Information

| Type of 510(k):      | Traditional             |
|----------------------|-------------------------|
| Common Name:         | Surgical mask           |
| Trade Name:          | Disposable Medical Mask |
| Model:               | k0450                   |
| Classification Name: | Surgical Apparel        |
| Review Panel:        | General Hospital        |
| Product Code:        | FXX                     |
| Regulation Number:   | 21 CFR 878.4040         |
| Regulation Class:    | II                      |

# 3. Predicate Device Information

| Sponsor:     | Qiqihar Hengxin Medical Supplies,Ltd.       |
|--------------|---------------------------------------------|
| Common Name: | Surgical Mask                               |
| Trade Name:  | Single-Use Surgical Face Mask with Ear Loop |

{4}------------------------------------------------

| 510(k) number: | K201691 |
|----------------|---------|
| Model:         | L       |

| Classification Name: | Surgical Apparel |
|----------------------|------------------|
| Review Panel:        | General Hospital |
| Product Code:        | FXX              |
| Regulation Number:   | 21 CFR 878.4040  |
| Regulation Class:    | II               |

#### 4. Indications for Use

The Disposable Medical mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. The Disposable Medical mask intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Model: k0450, blue color, and Level 2 barrier level as ASTM F2100.

## 5. Device Description

Disposable Medical Mask are non-sterile, single use, 3 layers, flat-pleated style with ear loops and nose clip. The outer layer and inner facing layer of face mask consist of Spunbond Polypropylene, and the middle layer consists of Melt Blown Polypropylene Filter. Each mask contains ear loops to secure the mask over the users' mouth and face and includes a nose clip to provide a firm fit over the nose. The mask is a single use, disposable device, provided non-sterile.

#### 6. Test Summary

Surgical face mask has been evaluated the safety and effectiveness by lab bench testing according to the following standards:

{5}------------------------------------------------

- ASTM F2299 Standard Test Method for Determining the Initial Efficiency of Materials Used in Surgical face masks to Penetration by Particulates Using Latex Spheres.
- . ASTM F1862 Standard test method for resistance of Surgical face masks to penetration by synthetic blood (Horizontal projection of fixed volume at a known velocity).
- . ASTM F 2101-19 Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Surgical face mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus.
- 16 CFR Part 1610 STANDARD FOR THE FLAMMABILITY OF CLOTHING TEXTILES
- ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks

During use, the Non-Woven Face Mask will directly contact with user's skin, so we have it tested to demonstrate conformance to the following standards.

- ISO 10993-5, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- . ISO 10993-10, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

#### Summary of Comparison and Technological Characteristics 7.

| Elements of<br>Comparison                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Results                       |                     |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|---------------------|
| Product Name                                            | Disposable Medical<br>Mask (Model: k0450)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Single-Use Surgical<br>Mask with Ear Loop<br>(Model:L)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | --                            |                     |
| 510(k) Number                                           | K203455                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K201691                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                               |                     |
| General Comparison                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                               |                     |
| Elements of<br>Comparison                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Results                       |                     |
| Indication for Use                                      | The Disposable<br>Medical mask is<br>intended to be worn<br>to protect both the<br>patient and<br>healthcare personnel<br>from the transfer of<br>microorganisms,<br>body fluids and<br>particulate material.<br>The Disposable<br>Medical mask is<br>intended for use in<br>infection control<br>practices to reduce<br>the potential<br>exposure to blood<br>and body fluids.<br>This is a single use,<br>disposable device,<br>provided non-sterile.<br>Model: k0450, blue<br>color, and Level 2<br>barrier level as ASTM<br>F2100. | The Single-Use surgical<br>mask with Ear Loop is<br>intended to be worn to<br>protect both the patient<br>and healthcare<br>personnel from the<br>transfer of<br>microorganisms, body<br>fluids and particulate<br>material. The Single-Use<br>Mask with Ear Loop<br>intended for use in<br>infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device(s),<br>provided non-sterile.<br>Model:M and L, blue<br>color, and Level 2 barrier<br>level as ASTM F2100 | Same                          |                     |
| Model                                                   | 3 layers, Ear Loops,<br>Flat-Pleated Style                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 3 layers, Ear Loops,<br>Flat-Pleated Style                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                               |                     |
| Material                                                | Outer<br>facing layer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Spun-bond non-woven<br>fabric | Different<br>Note 1 |
| Elements of<br>Comparison                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Results                       |                     |
| Middle<br>layer                                         | Melt blown<br>polypropylene filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Melt blown non-woven<br>fabric                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different<br>Note 2           |                     |
| Inner<br>facing layer                                   | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Spun-bond non-woven<br>fabric                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Different<br>Note 3           |                     |
| Nose clip                                               | aluminum wire coated<br>by polyethylene.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Malleable aluminum<br>wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different<br>Note4            |                     |
| Ear loops                                               | Spandex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Polyester                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different<br>Note 5           |                     |
| Color                                                   | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                          |                     |
| Dimension (Width)                                       | 9.5cm ± 0.5cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 9.0cm ± 1.0cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Different<br>Note 6           |                     |
| Dimension (Length)                                      | 17.5cm ± 0.5cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 18.0cm ± 1.0cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                               |                     |
| Sterility                                               | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                          |                     |
| Use                                                     | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                          |                     |
| ASTM F2100 Level                                        | Level2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Level 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                          |                     |
| Fluid Resistance<br>Performance (mmHg)                  | 32 out of 32 pass at 120<br>mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 31 Out of 32 pass at<br>120 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar                       |                     |
| Particulate Filtration<br>Efficiency<br>Performance (%) | 99.9%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ≥ 99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                          |                     |
| Bacterial Filtration<br>Efficiency<br>Performance (%)   | 99.9%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ≥ 99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                          |                     |
| Flammability class                                      | Class 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Class 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                          |                     |
| Differential Pressure<br>(Delta-P)<br>(mm H2O/cm2)      | <5.0 mmH2O/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | < 5.0 mmH20/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                          |                     |

## Table 1 - General Comparison

{6}------------------------------------------------

{7}------------------------------------------------

Note1: Although their are differences in the material of the outer facing layer, the difference does not raise any issues of safety and effectiveness based on the results obtained in the biocompatibility studies and performance studies.

{8}------------------------------------------------

Note2: Although their are differences in the material of the middle layer, the difference does not raise any issues of safety and effectiveness based on the results obtained in the biocompatibility studies and performance studies.

Note3: Although their are differences in the material of the inner facing layer, the difference does not raise any issues of safety and effectiveness based on the results obtained in the biocompatibility studies and performance studies.

Note4: Although their are differences in the material of the nose clip, the difference does not raise any issues of safety and effectiveness based on the results obtained in the biocompatibility studies and performance studies.

Note5: Although their are differences in the material of the ear loops, the difference does not raise any issues of safety and effectiveness based on the results obtained in the biocompatibility studies and performance studies.

Note 6: Although their are differences in the dimension that do not raise any issues of safety and effectiveness based on the results obtained in the biocompatibility studies and performance studies.

#### 8. Non-clinical Tests Performed on the Proposed Device

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification [510(k)] Submission issued on March 5, 2004.

| Test purpose                                                  | Test Method   | Pass Criteria                    | Results                          |
|---------------------------------------------------------------|---------------|----------------------------------|----------------------------------|
| Fluid<br>Resistance<br>Performance<br>(mmHg)                  | ASTM<br>F1862 | 29 Out of 32 pass at<br>120 mmHg | 32 out of 32 pass at 120<br>mmHg |
| Particulate<br>Filtration<br>Efficiency<br>Performance<br>(%) | ASTM<br>F2299 | ≥ 98%                            | 99.9%                            |
| Bacterial<br>Filtration<br>Efficiency                         | ASTM<br>F2101 | ≥ 98%                            | 99.9%                            |

{9}------------------------------------------------

| Performance<br>(%)                                 |               |                 |                 |
|----------------------------------------------------|---------------|-----------------|-----------------|
| Flammability class                                 | 16 CFR 1610   | Class 1         | Class 1         |
| Differential Pressure<br>(Delta-P)<br>(mm H2O/cm2) | ASTM F2100-19 | < 5.0 mmH2O/cm2 | < 5.0 mmH2O/cm2 |

# 9. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K203455, the Disposable Medical Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201691.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203455](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203455)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
