← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K203064

# ICU Health Surgical Mask (K203064)

_Plastikon Industries · FXX · Mar 23, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203064

## Device Facts

- **Applicant:** Plastikon Industries
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 23, 2021
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

## Device Story

Surgical mask; three-layer, flat-pleated design; outer/inner layers polypropylene spunbond; middle layer polypropylene melt-blown; includes ear loops and malleable polyethylene-coated steel nosepiece. Single-use, disposable, non-sterile. Used in clinical settings by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Device functions as physical barrier; performance verified via ASTM F2100-19 standards for fluid resistance, filtration efficiency, and breathability. Benefits include reduced risk of cross-contamination and exposure to infectious materials during medical procedures.

## Clinical Evidence

No clinical data. Bench testing only. Device performance validated against ASTM F2100-19, including fluid resistance (ASTM F1862-17), PFE (ASTM F2299), BFE (ASTM F2101-19), differential pressure (EN 14683:2019), and flammability (16 CFR Part 1610). Biocompatibility confirmed per ISO 10993-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

Three-layer construction: polypropylene spunbond outer/inner layers, polypropylene melt-blown filter layer. Malleable polyethylene-coated steel wire nosepiece. Elastic laminate ear loops. Flat-pleated form factor. Non-sterile. Tested to ASTM F2100-19, ASTM F1862-17, ASTM F2299, ASTM F2101-19, EN 14683:2019, 16 CFR 1610, ISO 10993-1, ISO 10993-5, and ISO 10993-10.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- San-M Package Co, [K160269](/device/K160269.md) Surgical Mask

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

March 23, 2021

Plastikon Industries % Mary Gallup RA Consultant Hantel Technologies, Inc 3496 Breakwater Ct Hayward, California 94545

Re: K203064

Trade/Device Name: Surgical Mask, ASTM Level 2 or Level 3 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 12, 2021 Received: February 17, 2021

Dear Mary Gallup:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K203064

Device Name Surgical Mask, ASTM Level 2 or Level 3

#### Indications for Use (Describe)

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary - K203064

| Device Name:           | Surgical Mask, ASTM Level 2 or Level 3                                                                                                                                                                                                                                                                                                                                                 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor Information:   | Plastikon Industries<br>688 Sandoval Way<br>Hayward, CA 94545                                                                                                                                                                                                                                                                                                                          |
| Contact person:        | Paul Gutwald<br>Title: Chief Operating Officer<br>Phone Number: (510) 400-1010<br>Fax Number: (510) 400-1141<br>e-mail: pgutwald@plastikon.com                                                                                                                                                                                                                                         |
| Date of Summary:       | 03/15/2021                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name:           | Surgical Mask                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name:   | Surgical Apparel                                                                                                                                                                                                                                                                                                                                                                       |
| Proprietary Name:      | Surgical Mask - ASTM Level 2 or Level 3                                                                                                                                                                                                                                                                                                                                                |
| Review Panel:          | General and Plastic Surgery                                                                                                                                                                                                                                                                                                                                                            |
| Product Code:          | FXX                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Classification: | Class II per (21 CFR §878.4040)                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device:      | San-M Package Co, K160269 Surgical Mask                                                                                                                                                                                                                                                                                                                                                |
| Intended Use:          | The Surgical Mask is intended to be worn to protect both the<br>patient and healthcare personnel from transfer of<br>microorganisms, body fluids, and particulate material. These<br>face masks are intended for use in infection control practices to<br>reduce potential exposure to blood and body fluids. The face<br>mask is single use, disposable device, provided non-sterile. |

## Available Model Numbers

PLASMA Surgical-1 Surgical Mask

#### Device Description:

The surgical mask is composed of three-layers and are flat-pleated. The mask materials consist of an outer layer (polypropylene spunbond, white), filter middle layer (polypropylene melt-blown, white) and inner layer (spunbond polypropylene, white). Each mask contains ear loops to secure the mask over the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. The mask is a single use, disposable device, provided non-sterile.

The device is not made from Natural Rubber Latex.

{4}------------------------------------------------

# Technological Characteristics:

The Surgical Mask is compared with the predicate device. The results are shown below in the Technological Characteristics Comparison Table:

| Items(S)                                | Surgical Mask (K203064)<br>Subject Device                                                                                                                                                                                                                                                                                                                                                      |         | Predicate Device (K160269)<br>San-M Package Co.<br>Surgical Face Masks                                                                                                                                                                                                                                                                                                                                      |         |         | Comparison                                                                                                                                                                                                     |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                         | Level 2                                                                                                                                                                                                                                                                                                                                                                                        | Level 3 | Level 1                                                                                                                                                                                                                                                                                                                                                                                                     | Level 2 | Level 3 |                                                                                                                                                                                                                |
| Intended Use/<br>Indications for<br>Use | The surgical mask is intended<br>to be worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids,<br>and particulate material. This<br>face mask is intended for use<br>in infection control practices to<br>reduce potential exposure to<br>blood and body fluids. This is<br>a single-use, disposable<br>device, provided non-sterile. |         | The surgical face masks are<br>intended to be worn to protect<br>both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce potential exposure to<br>blood and body fluids. This is<br>a single-use, disposable<br>device, provided non-sterile. |         |         | Same                                                                                                                                                                                                           |
|                                         | Materials                                                                                                                                                                                                                                                                                                                                                                                      |         |                                                                                                                                                                                                                                                                                                                                                                                                             |         |         |                                                                                                                                                                                                                |
| Outer Material                          | Polypropylene Spunbond,<br>White                                                                                                                                                                                                                                                                                                                                                               |         | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                               |         |         | Same                                                                                                                                                                                                           |
| Filter Web<br>(Middle)                  | Polypropylene Meltblown,<br>white                                                                                                                                                                                                                                                                                                                                                              |         | 1. Polypropylene<br>spunbond<br>2. Polypropylene<br>meltblown                                                                                                                                                                                                                                                                                                                                               |         |         | Same                                                                                                                                                                                                           |
| Inner Material                          | Polypropylene Spunbond,<br>white                                                                                                                                                                                                                                                                                                                                                               |         | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                               |         |         | Same                                                                                                                                                                                                           |
| Nose Clamp                              | Polyethylene Coated Steel<br>Wire                                                                                                                                                                                                                                                                                                                                                              |         | Polyethylene coated steel wire                                                                                                                                                                                                                                                                                                                                                                              |         |         | Same                                                                                                                                                                                                           |
| Ear Loops / Tie<br>Tapes                | Elastic Laminate –<br>No Tie Tapes                                                                                                                                                                                                                                                                                                                                                             |         | Ear loops: Polyester,<br>polyurethane<br>Side tapes:<br>Polyester spunbond (ear loops<br>mask only)<br><br>Tie tapes:<br>Polypropylene spunbond or<br>polyester spunbond                                                                                                                                                                                                                                    |         |         | Same – The<br>subject device<br>is an ear loop<br>type<br>attachment.<br>The ear loop<br>materials has<br>been tested for<br>biocompatibility<br>and does not<br>does not affect<br>the device<br>performance. |
|                                         | Design Features                                                                                                                                                                                                                                                                                                                                                                                |         |                                                                                                                                                                                                                                                                                                                                                                                                             |         |         |                                                                                                                                                                                                                |
| Colors                                  | White (Outer)                                                                                                                                                                                                                                                                                                                                                                                  |         | Colors: white or blueVisor option: polyester                                                                                                                                                                                                                                                                                                                                                                |         |         | Different<br>The subject<br>device does not                                                                                                                                                                    |

{5}------------------------------------------------

| Style           | Flat - Pleated | Flat - Pleated | come in any other colors. The color of the device does not affect the device performance or Biocompatibility as demonstrated by the test results. |
|-----------------|----------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Multiple Layers | Yes            | Yes            | Same                                                                                                                                              |
| Single Use      | Yes            | Yes            | Same                                                                                                                                              |
| Sterility       |                |                |                                                                                                                                                   |
| Sterile         | Non-Sterile    | Non-Sterile    | Same                                                                                                                                              |
| Dimensions      |                |                |                                                                                                                                                   |
| Length          | $95 \pm 19mm$  | $90 \pm 3mm$   | Same                                                                                                                                              |
| Width           | $175 \pm 19mm$ | $175 \pm 5mm$  | Same                                                                                                                                              |

| Performance Testing (ASTM F2100) | Subject Device<br>Level 2 Level 3 | Predicate Device<br>Level 1 Level 2 Level 3 | Comparison |
|----------------------------------|-----------------------------------|---------------------------------------------|------------|
|----------------------------------|-----------------------------------|---------------------------------------------|------------|

| Fluid<br>Resistance<br>ASTM F1862-<br>17                            | Pass @<br>120<br>mmHg | Pass @<br>160<br>mmHg | Pass @<br>80<br>mmHg | Pass @<br>120<br>mmHg | Pass @<br>160<br>mmHg | Same                                                                                                                                                                                    |
|---------------------------------------------------------------------|-----------------------|-----------------------|----------------------|-----------------------|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Particulate<br>Filtration<br>Efficiency<br>(PFE)<br>ASTM F2299      | Pass @<br>99.8%       | Pass @<br>99.8%       | Pass @<br>99.6%      | Pass @<br>99.6%       | Pass@<br>99.7%        | Different<br>The subject<br>device has<br>slightly higher<br>or improved<br>particulate<br>efficiency. This<br>does not<br>increase risk<br>associated<br>with the use of<br>the device |
| Bacterial<br>Filtration<br>Efficiency<br>(BFE)<br>ASTM F2101-<br>19 | Pass @<br>>99%        | Pass @<br>>99%        | Pass @<br>>98%       | Pass @<br>>98%        | Pass @<br>>98%        | Different<br>The subject<br>device has<br>slightly higher<br>or improved<br>BFE. This<br>does not<br>increase risk<br>associated<br>with the use of                                     |

{6}------------------------------------------------

| Differential<br>Pressure<br>EN 14683:2019<br>Annex C | Pass @<br>5.1mm<br>H₂0/cm² | Pass @<br>5.1mm<br>H₂0/cm² | Pass @<br>2.0mm<br>H₂0/cm² | Pass @<br>1.0mm<br>H₂0/cm² | Pass @<br>2.5 mm<br>H₂0/cm² | the device<br>Different<br>The subject<br>device has<br>higher<br>differential<br>pressure, and<br>continues to<br>meet the<br>standard for<br>ASTM<br>classification<br>for 2 and 3.<br>This does not<br>increase risk<br>associated<br>with the use of<br>the device |
|------------------------------------------------------|----------------------------|----------------------------|----------------------------|----------------------------|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flammability<br>16 CFR Part<br>1610                  | Class 1                    | Class 1                    | Class 1                    | Class 1                    | Class 1                     | Same                                                                                                                                                                                                                                                                   |
| Biocompatibility                                     |                            |                            |                            |                            |                             |                                                                                                                                                                                                                                                                        |
| Test                                                 | Subject Device             | Predicate Device           |                            |                            | Comparison                  |                                                                                                                                                                                                                                                                        |
| Cytotoxicity                                         | Non-cytotoxic              | Non-cytotoxic              |                            |                            | Same                        |                                                                                                                                                                                                                                                                        |
| Irritation                                           | Non-irritating             | Non-irritating             |                            |                            | Same                        |                                                                                                                                                                                                                                                                        |
| Sensitization                                        | Non-sensitizing            | Non-sensitizing            |                            |                            | Same                        |                                                                                                                                                                                                                                                                        |

# Non-Clinical Performance Testing

The surgical mask has been tested according to ASTM 2100-19 -- Standard Specification for Performance of materials used in Medical Face Masks, and meets the requirements to be designated as ASTM Level 2 or Level 3, and have met the requirements for biocompatibility.

| Standard                 | Performance Characteristics, Level 2, 3                                                                          | Surgical Mask                                                    |
|--------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| ASTM F1862-17            | Resistance to Penetration by Synthetic<br>Blood                                                                  | Passed                                                           |
| ASTM F2299               | Particulate Filtration Efficiency (PFE)                                                                          | Passed                                                           |
| ASTM F2101-19            | Bacterial Filtration Efficiency (BFE)                                                                            | Passed                                                           |
| EN 14683:2019<br>Annex C | Differential Pressure                                                                                            | Passed                                                           |
| 16 CFR Part 1610         | Flammability of Clothing Textiles                                                                                | Passed                                                           |
| ISO10993-1               | Biological Evaluation of Medical Devices -<br>Part 1: Evaluation and Testing Within A<br>Risk Management Process | Met the requirements                                             |
| ISO10993-5               | Biological Evaluation of Medical Devices -<br>Part 5: Tests for In Vitro Cytotoxicity                            | No cyctotoxicity or<br>cell lysis was noted.                     |
| ISO10993-10              | Biological Evaluation of Medical Devices<br>Part 10: Tests for Irritation and<br>Sensitization, 1ed              | No evidence of<br>sensitization or<br>irritation was<br>observed |

{7}------------------------------------------------

## Clinical Performance

No Clinical study is included in this submission. The device as designed does not posed any new or additional risks to user or patient.

## Summary of Test Results:

The conclusions drawn from the nonclinical tests demonstrate that the devices are as safe, as effective, and perform as well as or better than the legally marketed device identified as the predicate device, K160269.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203064](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K203064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
