← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K202903

# Surgical Face Mask (K202903)

_Rizhao Sanqi Medical & Health Articles Co., Ltd. · FXX · Mar 24, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K202903

## Device Facts

- **Applicant:** Rizhao Sanqi Medical & Health Articles Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Mar 24, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. When worn properly, these face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

## Device Story

Surgical face mask; 3-layer, flat-pleated, nonwoven polypropylene construction; includes malleable nose clip for fit; available with ear loops (spandex) or ties (polypropylene); optional visor. Single-use, disposable, non-sterile. Used in clinical settings or by general public for infection control; worn by healthcare personnel and patients. Acts as physical barrier to reduce transfer of microorganisms, body fluids, and particulates. Protects wearer and patient from cross-contamination; reduces exposure to blood and body fluids.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F2100, ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), EN 14683 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

## Technological Characteristics

3-layer nonwoven polypropylene; melt-blown filter layer; spandex ear loops or polypropylene ties; malleable nose clip. Dimensions 175x95mm. Non-sterile. Performance levels 1, 2, and 3 per ASTM F2100. No software or electronic components.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Surgical Face Masks (Ear loops and Tie-on) ([K160269](/device/K160269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 24, 2022

Rizhao Sanqi Medical & Health Articles Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai, Shanghai 200122 China

Re: K202903

Trade/Device Name: Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 24, 2021 Received: December 29, 2021

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202903

Device Name Surgical Face Mask

Indications for Use (Describe)

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. When worn properly, these face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Level 1 Surgical Face Mask models: SQ-1001 Level 2 Surgical Face Mask models: SQ-2001, SQ-2001H, SQ-2004, SQ-2004H Level 3 Surgical Face Mask models: SQ-3001, SQ-3001H

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

## K202903

Date of Summary prepared: 2021-12-24

## A. Applicant:

Rizhao Sanqi Medical & Health Articles Co., Ltd. Address: Heshan Industry park, Donggang District, Rizhao, Shandong Province, China Contact person: Julia Yu Title: Sales Manager Tel: +86-633-7663737 Fax: +86-633-7653718 Email: 3qmedical@3qcn.cc

Submission Correspondent:

Primary contact: Ms. Ivy Wang

Shanghai SUNGO Management Consulting Co., Ltd.

Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China

Tel: +86-21-58817802

Email: haiyu.wang@sungoglobal.com

Secondary contact: Mr. Raymond Luo

Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China

Tel: +86-21-68828050

Email: fda.sungo@gmail.com

## B. Device:

Trade Name: SURGICAL FACE MASK Common Name: SURGICAL MASK Model:

Table 1 – Surgical face mask model numbers

| Mask Style         | Ear loops | Tie-on   | Color              |
|--------------------|-----------|----------|--------------------|
| Level 1            | SQ-1001   | -        |                    |
| Level 2            | SQ-2001   | SQ-2001H | Blue, white, pink, |
| Level 2 with visor | SQ-2004   | SQ-2004H | green, yellow      |
| Level 3            | SQ-3001   | SQ-3001H |                    |

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

{4}------------------------------------------------

| C. Predicate device: |                                            |                         |
|----------------------|--------------------------------------------|-------------------------|
| K160269              | Surgical Face Masks (Ear loops and Tie-on) | SAN-M PACKAGE CO., LTD. |

## D. Indications for use of the device:

The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. When worn properly, these face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

## E. Device Description:

The Surgical Face Masks are 3-layer, flat-folded masks constructed of nonwoven polypropylene materials. The mask is provided with ear loops (Spandex) or ties (polypropylene). A malleable nose clip is placed in the layers of facemask for comfort and individualized fit. The surgical face mask will be provided in blue, white, pink, green, vellow and with the option for a visor. The surgical face masks are single-use, disposable devices, provided non-sterile.

## F. Technological Comparison with predicate device

| Device                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                             | Result                                            |           |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|-----------|
| 510K #                                  | K202903                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K160269                                                                                                                                                                                                                                                                                                                                                                                                      |                                                   |           |
| Manufacturer                            | Rizhao Sanqi Medical & Health Articles<br>Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                 | SAN-M PACKAGE CO., LTD.                                                                                                                                                                                                                                                                                                                                                                                      |                                                   |           |
| Model Name                              | SURGICAL FACE MASK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Surgical Face Masks (Ear loops<br>and Tie-on)                                                                                                                                                                                                                                                                                                                                                                | Similar                                           |           |
| Classification                          | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                                        | Same                                              |           |
| Intend use                              | The Surgical Face Mask is intended to<br>be worn to protect both the patient<br>and healthcare personnel from<br>transfer of microorganisms, body<br>fluids, and particulate material.<br>When worn properly, these face masks<br>are intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and body<br>fluids. This is a single-use, disposable<br>device, provided non-sterile. This is a<br>single use, disposable device(s),<br>provided non-sterile. | The surgical face masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids, and<br>particulate material. These face<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single-use, disposable device,<br>provided non-sterile. | Same                                              |           |
| Materials                               | Level 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Level 1                                                                                                                                                                                                                                                                                                                                                                                                      | -                                                 |           |
|                                         | Level 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Level 2                                                                                                                                                                                                                                                                                                                                                                                                      |                                                   |           |
|                                         | Level 3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Level 3                                                                                                                                                                                                                                                                                                                                                                                                      |                                                   |           |
| Outer layer                             | Spunbond Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                | Same                                              |           |
| Inner layer                             | Spunbond Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                | Same                                              |           |
| Filter layer                            | Melt-blown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1. Polypropylene spunbond                                                                                                                                                                                                                                                                                                                                                                                    | Similar                                           |           |
|                                         | 1.Melt-blown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                              |                                                   |           |
|                                         | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Polypropylene<br>2.Microporous Film                                                                                                                                                                                                                                                                                                                                                                          | 2. Polypropylene meltblown                        |           |
| Nose wire                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Steel coated by polypropylene                                                                                                                                                                                                                                                                                                                                                                                | Polyethylene coated steel wire                    | Different |
| Ear loops                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Spandex                                                                                                                                                                                                                                                                                                                                                                                                      | Polyester, polyurethane                           | Different |
| Tie-on                                  | Spunbond Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                              | Polypropylene spunbond or<br>polyester spunbond   | Similar   |
| Design Features                         | Ear Loops, Tie-on                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                              | Ear Loops, Tie-on                                 | Same      |
| Offer with visor                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                               | Same      |
| Mask style                              | 3 Flat Pleated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 4 Flat Pleated                                                                                                                                                                                                                                                                                                                                                                                               | 4 Flat Pleated                                    | Similar   |
| Color                                   | Blue, white, pink, green, yellow                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                              | white or blue                                     | Different |
| Dimension<br>(Length)                   | 175±5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                              | 175±5mm                                           | Same      |
| Dimension<br>(Width)                    | 95±5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                              | 95±5mm                                            | Same      |
| OTC use                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                               | Same      |
| Sterility                               | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                              | Non-Sterile                                       | Same      |
| Use                                     | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                              | Single Use, Disposable                            | Same      |
| Biocompatibility                        | Non-cytotoxic, Non-sensitizing,<br>non-irritating                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                              | Non-cytotoxic, Non-sensitizing,<br>non-irritating | Same      |
| Performance<br>Testing (ASTM<br>2100)   | Level 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Level 2 Level 3                                                                                                                                                                                                                                                                                                                                                                                              | Level 1 Level 2 Level 3                           | -         |
| Fluid<br>Resistance                     | Meet ASTM F1862-17                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                              | Meet ASTM F1862-13                                | similar   |
| Particulate<br>Filtration<br>Efficiency | Meet ASTM F2299-17                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                              | Meet ASTM F2299-03                                | Similar   |
| Bacterial<br>Filtration<br>Efficiency   | Meet ASTM F2101-19                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                              | Meet ASTM F2101-14                                | Similar   |
| Differential<br>Pressure                | Meet EN 14683: 2019, Annex C                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                              | Meet MIL-M36945C                                  | Different |
| Flammability                            | Meet 16 CFR 1610                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                              | Meet 16 CFR 1610                                  | Similar   |
| Biocompatibility                        | Non-cytotoxic, non-sensitizing,<br>non-irritating                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                              | Non-cytotoxic, non-sensitizing,<br>non-irritating | Same      |

Table 2 General Comparison

{5}------------------------------------------------

Rizhao Sanqi Medical & Health Articles Co., Ltd. Heshan Industry park, Donggang District, Rizhao, Shandong Province, China

## G. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was same to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004:

- A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In VitroCytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin

{6}------------------------------------------------

Sensitization

- > ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks
- > ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity);
- > EN 14683, Medical Face Masks-Requirements and Test Methods;
- > ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;
- > ASTM F2299, Standard test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
- > 16 CFR 1610, Standard for the Flammability of clothing textiles;

| Mask Style         | Ear loops | Tie-on   | Color              |
|--------------------|-----------|----------|--------------------|
| Level 1            | SQ-1001   | -        |                    |
| Level 2            | SQ-2001   | SQ-2001H | Blue, white, pink, |
| Level 2 with visor | SQ-2004   | SQ-2004H | green, yellow      |
| Level 3            | SQ-3001   | SQ-3001H |                    |

# For level 1: SQ-1001

| Item                                                   | Purpose                                                                                           | Acceptance Criteria<br>Level 1 | Result                               |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------|
| Fluid Resistance<br>Performance ASTM<br>F1862          | The purpose of the performance testing is to demonstrate the functionality of the subject device. | 29 out of 32 pass at 80 mmHg   | 3 nonconsecutive lots tested<br>PASS |
| Particulate Filtration<br>Efficiency ASTM F2299        |                                                                                                   | $\ge$ 95%                      | PASS                                 |
| Bacterial Filtration<br>Efficiency ASTM F2101          |                                                                                                   | $\ge$ 95%                      | PASS                                 |
| Differential Pressure<br>(Delta P) EN 14683<br>Annex C |                                                                                                   | <5.0 mmH2O/cm²                 | PASS                                 |
| Flammability 16 CFR<br>1610                            |                                                                                                   | Class 1                        | PASS                                 |

For: Blue, white, pink, green, yellow

| Item          | Purpose                                                                        | Acceptance Criteria | Result                                                           |
|---------------|--------------------------------------------------------------------------------|---------------------|------------------------------------------------------------------|
| Cytotoxicity  | The purpose of the testing is to demonstrate the safety of the subject device. | Non-Cytotoxic       | Under the conditions of the study, the device is non-cytotoxic.  |
| Irritation    | The purpose of the testing is to demonstrate the safety of the subject device. | Non-Irritating      | Under the conditions of the study, the device is non-irritating. |
| Sensitization | The purpose of the testing is to demonstrate the safety of the subject device. | Non-Sensitizing     | Under the conditions of the study, the device is non-sensitizing |

{7}------------------------------------------------

# For level 2: SQ-2001 and SQ-2001H

| Item                                                   | Purpose                                                                               | Acceptance Criteria<br>Level 2   | Result                               |
|--------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------|--------------------------------------|
| Fluid Resistance<br>Performance ASTM<br>F1862          | The purpose of the<br>performance testing is<br>to demonstrate the<br>subject device. | 29 out of 32<br>pass at 120 mmHg | 3 nonconsecutive lots tested<br>PASS |
| Particulate Filtration<br>Efficiency ASTM F2299        | The purpose of the<br>performance testing is<br>to demonstrate the<br>subject device. | $\ge$ 98%                        | PASS                                 |
| Bacterial Filtration<br>Efficiency ASTM F2101          | The purpose of the<br>performance testing is<br>to demonstrate the<br>subject device. | $\ge$ 98%                        | PASS                                 |
| Differential Pressure<br>(Delta P) EN 14683<br>Annex C | The purpose of the<br>performance testing is<br>to demonstrate the<br>subject device. | <6.0 mmH2O/cm²                   | PASS                                 |
| Flammability 16 CFR<br>1610                            | The purpose of the<br>performance testing is<br>to demonstrate the<br>subject device. | Class 1                          | PASS                                 |

For: Blue, white, pink, green, yellow

| Item          | Purpose                                                                                    | Acceptance Criteria | Result                                                              |
|---------------|--------------------------------------------------------------------------------------------|---------------------|---------------------------------------------------------------------|
| Cytotoxicity  | The purpose of the<br>testing is to<br>demonstrate the<br>safety of the subject<br>device. | Non-Cytotoxic       | Under the conditions of the study, the<br>device is non-cytotoxic.  |
| Irritation    | The purpose of the<br>testing is to<br>demonstrate the<br>safety of the subject<br>device. | Non-Irritating      | Under the conditions of the study, the<br>device is non-irritating. |
| Sensitization | The purpose of the<br>testing is to<br>demonstrate the<br>safety of the subject<br>device. | Non-Sensitizing     | Under the conditions of the study, the<br>device is non-sensitizing |

# For Level 2 with visor: SQ-2004 and SQ-2004H

| Item                                            | Purpose                                                                                                       | Acceptance Criteria              | Result                       |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------|------------------------------|
|                                                 |                                                                                                               | Level 2                          | 3 nonconsecutive lots tested |
| Fluid Resistance<br>Performance ASTM<br>F1862   | The purpose of the<br>performance testing is<br>to demonstrate the<br>functionality of the<br>subject device. | 29 out of 32<br>pass at 120 mmHg | PASS                         |
| Particulate Filtration<br>Efficiency ASTM F2299 |                                                                                                               | $\ge$ 98%                        | PASS                         |
| Bacterial Filtration<br>Efficiency ASTM F2101   |                                                                                                               | $\ge$ 98%                        | PASS                         |

{8}------------------------------------------------

#### Rizhao Sanqi Medical & Health Articles Co., Ltd. Heshan Industry park, Donggang District, Rizhao, Shandong Province, China

| Differential<br>Pressure      |                |      |
|-------------------------------|----------------|------|
| (Delta P) EN 14683<br>Annex C | <6.0 mmH2O/cm² | PASS |
| Flammability 16 CFR<br>1610   | Class 1        | PASS |

For: Blue, white, pink, green, yellow

| Item          | Purpose                                                                                    | Acceptance Criteria | Result                                                               |
|---------------|--------------------------------------------------------------------------------------------|---------------------|----------------------------------------------------------------------|
| Cytotoxicity  | The purpose of the<br>testing is to<br>demonstrate the<br>safety of the subject<br>device. | Non-Cytotoxic       | Under the conditions of the study, the<br>device is non-cytotoxic.   |
| Irritation    | The purpose of the<br>testing is to<br>demonstrate the<br>safety of the subject<br>device. | Non-Irritating      | Under the conditions of the study, the<br>device is non-irritating.  |
| Sensitization | The purpose of the<br>testing is to<br>demonstrate the<br>safety of the subject<br>device. | Non-Sensitizing     | Under the conditions of the study, the<br>device is non-sensitizing. |

## For level 3: SQ-3001 and SQ-3001H

| Item                                                   | Purpose                           | Acceptance Criteria<br>Level 3 | Result                               |
|--------------------------------------------------------|-----------------------------------|--------------------------------|--------------------------------------|
| Fluid Resistance<br>Performance ASTM<br>F1862          | The purpose of the<br>performance | 29 out of 32 pass at 160 mmHg  | 3 nonconsecutive lots tested<br>PASS |
| Particulate Filtration<br>Efficiency ASTM F2299        | testing is to demonstrate         | $\geq$ 98%                     | PASS                                 |
| Bacterial Filtration<br>Efficiency ASTM F2101          | the functionality                 | $\geq$ 98%                     | PASS                                 |
| Differential Pressure<br>(Delta P) EN 14683<br>Annex C | of the subject device.            | <6.0 mmH2O/cm2                 | PASS                                 |
| Flammability 16 CFR<br>1610                            |                                   | Class 1                        | PASS                                 |

#### For: Blue, white, pink, green, yellow

| Item          | Purpose                                                                        | Acceptance Criteria | Result                                                           |
|---------------|--------------------------------------------------------------------------------|---------------------|------------------------------------------------------------------|
| Cytotoxicity  | The purpose of the testing is to demonstrate the safety of the subject device. | Non-Cytotoxic       | Under the conditions of the study, the device is non-cytotoxic.  |
| Irritation    | The purpose of the testing is to demonstrate the safety of the subject device. | Non-Irritating      | Under the conditions of the study, the device is non-irritating. |
| Sensitization | The purpose of the testing is to demonstrate the safety of the subject device. | Non-Sensitizing     | Under the conditions of the study, the device is non-sensitizing |

## H. Clinical Test Conclusion

No clinical study is included in this submission.

{9}------------------------------------------------

#### l. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device K160269.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K202903](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K202903)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
