← Product Code [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX) · K202615

# Nordiwell Medical Face Mask (K202615)

_Changzhou Combat Protective Equipment Co., Ltd. · FXX · Dec 31, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K202615

## Device Facts

- **Applicant:** Changzhou Combat Protective Equipment Co., Ltd.
- **Product Code:** [FXX](/submissions/SU/subpart-e%E2%80%94surgical-devices/FXX.md)
- **Decision Date:** Dec 31, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Intended Use

When properly worn, the surgical face masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only.

## Device Story

Nordiwell Medical Face Mask is a 3-layer, flat-pleated, single-use surgical mask. Construction includes an outer layer of blue spunbond polypropylene, a middle melt-blown polypropylene filter layer, and an inner white spunbond polypropylene layer. A malleable, polypropylene-coated galvanized nose wire ensures a secure fit over the nose and mouth; ear-loops are made of spandex and nylon. The device is used in clinical or general settings to protect patients and healthcare workers from microorganisms, body fluids, and particulate matter. It functions as a physical barrier; no electronic or software components are involved. The device is provided non-sterile and is latex-free.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862-17 (fluid resistance), ASTM F2299-17 (particulate filtration), ASTM F2101-19 (bacterial filtration), EN 14683:2019/ASTM F2100-19 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed per ISO 10993-5:2009 and ISO 10993-10:2010.

## Technological Characteristics

3-layer flat-pleated mask. Materials: Spunbond polypropylene (outer/inner), melt-blown polypropylene (filter). Nose wire: Polypropylene-coated galvanized wire. Ear-loops: Spandex/Nylon. Dimensions: 17.5cm x 9.5cm. Non-sterile. Biocompatibility: ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010. Performance standards: ASTM F1862-17, ASTM F2299-17, ASTM F2101-19, EN 14683:2019, 16 CFR 1610.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Avianz® Surgical Face Mask ([K200847](/device/K200847.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 31, 2020

Changzhou Combat Protective Equipment Co., Ltd. Yanmei Song Official Correspondent QingSiTang village, Henglin Town, Wujin District Changzhou, Jiangsu 213101 China

Re: K202615

Trade/Device Name: Nordiwell Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 23, 2020 Received: November 30, 2020

Dear Yanmei Song:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K202615

Device Name Nordiwell Medical Face Mask

Indications for Use (Describe)

When properly worn, the surgical face masks are intended to patient and healthcare workers from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K202615

### 510(k) Summary

# 1. Contact Information

#### 1.1. Applicant

Applicant Name: Changzhou Combat Protective Equipment Co., Ltd. Address: Qingsitang, Henglin, Changzhou, Jiangsu, 213101 China Contact Person: Yanmei Song Title: Sales Manager Telephone: +86-519-85190068; +86-15051998125 E-mail: xinyaoutdoor08@126.com

#### 1.2. Consultant

Company: Sinow Medical AS Address: Vestre Fantoft åsen 44, 5072, Bergen, Norway Contact Person: Huifang Zhao Telephone: +86 13961151430 Email: zhao@bergemed.com

### 2. Device information

Trade Name: Nordiwell Medical Face Mask Common Name: Surgical Face Mask Regulatory information: Classification Name: Mask, Surgical Classification: II Product Code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: General Hospital

# 3. Legally Marketed Primary Predicate Device

Product name: _ Avianz® Surgical Face Mask 510(k) Number: K200847 Product Code: FXX Manufacture: MEXPO INTERNATIONAL INC.

{4}------------------------------------------------

### 4. Indication for use

When properly worn, the medical face masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only.

#### 5. Device Description

The Nordiwell Medical Face Masks are composed of 3-layers and are flat-pleated. The mask materials consist of an outer layer (polypropylene spunbond, blue), inner layer (polypropylene spunbond white), filter (polypropylene melt-blown, white) and ear-loops. The mask contains a malleable nosepiece to provide a firm fit over the nose and then to secure the mask over the users' mouth and face. The mask is a single use, provided non-sterile.

This device is not made from Natural Rubber Latex.

Shelf-life of the device is 2 years.

| Device                                          | Subject Device (K202615)<br>Nordiwell Medical Face<br>Mask,<br>Model 952001                                                                                                                                                      | Predicate Device (K200847)<br>Avianz® Surgical Face Mask                                                                                                                                                                         | Comparison |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intend use                                      | When properly worn, the surgical face masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and particulate matter. This device is non-sterile and for single use only. | When properly worn, the surgical face masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and airborne particles. This device is non-sterile and for single use only. | Same       |
| Material                                        |                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                  |            |
| Outer Layer                                     | Spunbond Polypropylene                                                                                                                                                                                                           | Spunbond Polypropylene                                                                                                                                                                                                           | Same       |
| Filter(Middle layer)                            | Melt Blown Polypropylene<br>Filter                                                                                                                                                                                               | Melt Blown Polypropylene Filter                                                                                                                                                                                                  | Same       |
| Inner Layer                                     | Spunbond Polypropylene                                                                                                                                                                                                           | Spunbond Polypropylene                                                                                                                                                                                                           | Same       |
| Nose Wire                                       | Single Galvanize Wire, Coated                                                                                                                                                                                                    | Single Galvanize Wire, Coated                                                                                                                                                                                                    | Similar    |
|                                                 | by Polypropylene                                                                                                                                                                                                                 | by PE                                                                                                                                                                                                                            |            |
| Ear Loops                                       | Spandex and Nylon, natural<br>rubber latex free                                                                                                                                                                                  | Not made with natural rubber<br>latex                                                                                                                                                                                            | Same       |
|                                                 | Design Features                                                                                                                                                                                                                  |                                                                                                                                                                                                                                  |            |
| Colors                                          | Blue                                                                                                                                                                                                                             | White                                                                                                                                                                                                                            | Different  |
| Style                                           | Flat - Pleated                                                                                                                                                                                                                   | Flat - Pleated                                                                                                                                                                                                                   | Same       |
| Multiple<br>Layers                              | Yes                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                              | Same       |
| Single Use                                      | Yes                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                              | Same       |
| OTC Use                                         | Yes                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                              | Same       |
|                                                 | Sterility                                                                                                                                                                                                                        |                                                                                                                                                                                                                                  |            |
| Sterile                                         | Non-Sterile                                                                                                                                                                                                                      | Non-Sterile                                                                                                                                                                                                                      | Same       |
|                                                 | Dimension                                                                                                                                                                                                                        |                                                                                                                                                                                                                                  |            |
| Length                                          | 17.5cm ±0.5cm                                                                                                                                                                                                                    | 17.5cm ±0.5cm                                                                                                                                                                                                                    | Same       |
| Width                                           | 9.5cm ±0.5cm                                                                                                                                                                                                                     | 9.0cm ±0.5cm                                                                                                                                                                                                                     | Similar    |
|                                                 | Performance                                                                                                                                                                                                                      |                                                                                                                                                                                                                                  |            |
| Fluid Resistance Performance<br>ASTM F1862      | 32 of 32 pass at 160 mmHg                                                                                                                                                                                                        | 30 Out of 32 pass at 120 mmHg                                                                                                                                                                                                    | Better     |
| Particulate Filtration Efficiency<br>ASTM F2299 | > 99.9%                                                                                                                                                                                                                          | 99.9%                                                                                                                                                                                                                            | Same       |
| Bacterial Filtration Efficiency<br>ASTM F2101   | > 99.9%                                                                                                                                                                                                                          | ≥98%                                                                                                                                                                                                                             | Better     |
| Differential Pressure<br>(Delta P)              | 5.6 mmH20/cm²<br>( EN 14683:2019, Annex C<br>and ASTM F2100-19)                                                                                                                                                                  | 3.0 mmH20/cm²<br>(MILM-36954C)                                                                                                                                                                                                   | Similar    |
| Flammability 16 CFR 1610                        | Class 1                                                                                                                                                                                                                          | Class 1                                                                                                                                                                                                                          | Same       |
|                                                 | Biocompatibility                                                                                                                                                                                                                 |                                                                                                                                                                                                                                  |            |
| In Vitro Cytotoxicity Test<br>ISO 10993-5: 2009 | Non-Cytotoxic                                                                                                                                                                                                                    | Non-Cytotoxic                                                                                                                                                                                                                    | Same       |
| Skin Irritation Test<br>ISO 10993-10:2010       | Non-Irritating                                                                                                                                                                                                                   | Non-Irritating                                                                                                                                                                                                                   | Same       |
| Skin Sensitization Test<br>ISO 10993-10:2010    | Non-Sensitizing                                                                                                                                                                                                                  | Non-Sensitizing                                                                                                                                                                                                                  | Same       |

# 6. Substantially Equivalent (SE) Comparison

{5}------------------------------------------------

# 7. Non-Clinical Test

Non clinical tests were conducted to verify that the proposed device met all design

{6}------------------------------------------------

specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

(1) Performance test:

Bench test was conducted on Medical Face Mask for proposed device to determine substantially equivalence. The bench tests include the following tests:

| Item                                                                               | Proposed Device           | Acceptance Criteria        | Result |
|------------------------------------------------------------------------------------|---------------------------|----------------------------|--------|
| Fluid Resistance Performance<br>ASTM F1862-17                                      | 32 of 32 pass at 160 mmHg | ≥29 of 32 pass at 160 mmHg | Pass   |
| Particulate Filtration<br>Efficiency ASTM F2299-17                                 | > 99.9%                   | ≥ 98%                      | Pass   |
| Bacterial Filtration Efficiency<br>ASTM F2101-19                                   | > 99.9%                   | ≥ 98%                      | Pass   |
| Differential Pressure<br>(Delta P) EN 14683:2019,<br>Annex C and ASTM<br>F2100-19. | 5.6 mmH20/cm²             | <6.0 mmH20/cm²             | Pass   |
| Flammability 16 CFR 1610                                                           | Class 1                   | Class 1                    | Pass   |

(2) Biocompatibility evaluation and test

Biocompatibility evaluation conducted in accordance with the FDA's 2016 guidance and ISO10993-1:2018 supports that the subject devices are biocompatible.

The biocompatibility test includes the following tests:

In Vitro Cytotoxicity Test(ISO 10993-5: 2009)

Skin Irritation Test(ISO 10993-10:2010)

Skin Sensitization Test(ISO 10993-10:2010)

(3) Shelf-life accelerated aging test

The accelerated aging tests were conducted with 3 non-consecutive lots. The following performances were tested both for before and after aged products:

Fluid Resistance Performance according to ASTM F1862-17.

Particulate Filtration Efficiency according to ASTM F2299-17.

Bacterial Filtration Efficiency according to ASTM F2101-19.

Differential Pressure (Delta P) according to EN 14683:2019, Annex C and ASTM F2100-19.

Flammability according to 16 CFR 1610.

### 8. Clinical Test Conclusion

{7}------------------------------------------------

No clinical study is included in this submission.

# 9. Conclusion

The summary includes the conclusions drawn from the nonclinical and clinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K202615](https://fda.innolitics.com/submissions/HO/subpart-e%E2%80%94surgical-devices/FXX/K202615)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
